K970358 · Mennen Medical, Inc. · DSI · Feb 11, 1998 · Cardiovascular
Device Facts
Record ID
K970358
Device Name
ENSEMBLE CENTRAL STATION MONITOR
Applicant
Mennen Medical, Inc.
Product Code
DSI · Cardiovascular
Decision Date
Feb 11, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
ENSEMBLE includes the essential features of a central station monitor for up to 12 patients. It can display vital signs information collected from bedside patient monitoring devices such as: ECG/Heart Rate, Arrhythmia (basic and extended alarms), ST Segment, Respiration, Temperature, Invasive Blood Pressure, Noninvasive Blood Pressure, Pulse Oximetry, and End Tidal carbon-dioxide. The ENSEMBLE may be used in system application in many different clinical specialties within the hospital.
Device Story
ENSEMBLE is a hospital-based, multi-patient central station display terminal. It functions as a component of a physiological monitoring system, receiving vital signs, waveforms, and alarm data from bedside devices (e.g., HORIZON XL, 219 TELEMETRY) via an Ethernet network. It displays data for up to 12 patients simultaneously, including patient name, location, vital sign values, alarm limits, waveforms, and alarm status. The device does not perform primary alarm determination; this occurs at the bedside. However, it provides remote interaction capabilities, allowing clinicians to adjust bedside alarm/parameter settings and initiate recordings or printouts. Operated via mouse, keyboard, touchscreen, trackball, or light pen by hospital staff. It centralizes patient data to facilitate clinical monitoring and decision-making, improving patient oversight.
Clinical Evidence
Bench testing only. The device complies with AAMI/ANSI ES11 (Safe Current Limits), AAMI/ANSI EC13 (Cardiac Monitors), and UL-544 standards. No clinical data provided.
Technological Characteristics
Software-driven central station display terminal. Ethernet connectivity for data acquisition from bedside monitors. Supports mouse, keyboard, touchscreen, trackball, and light pen inputs. Complies with UL 544, ANSI Safe Current Limits, and FDA MDS-201-0004. Non-implantable, non-life-sustaining. No drug/biological components.
Indications for Use
Indicated for hospital-based physiological monitoring of up to 12 patients simultaneously. Displays vital signs including ECG/Heart Rate, Arrhythmia, ST Segment, Respiration, Temperature, Invasive/Noninvasive Blood Pressure, Pulse Oximetry, and End Tidal CO2. For prescription use only.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
MENNEN MEDICAL COMPREHENSIVE CENTRAL STATION (K892066)
HEWLETT-PACKARD M2360A WAVEVUE CENTRAL MONITOR (K921014)
Submission Summary (Full Text)
{0}
K970358
FEB 11 1998
# MENNEN MEDICAL LTD.
Kiryat Weizmann Science Park
P.O.B. 102
Rehovot 76100 Israel
Tel: 972-8-938-3030
Fax: 972-8-940-6519
**Date:** 10 November 1997
**To:** Food and Drug Administration
Center for Devices and Radiological Health Document Control Center (HFZ-401)
1390 Piccard Drive
Rockville MD 20850
**Attn.:** Document Control Clerk
**From:** Kenneth Raichman
Director of Regulatory Affairs
**Topic:** 510(k) Summary
**ENSEMBLE Central Station Monitor**
**Safety and Effectiveness**
## Product Name
**Proprietary:** ENSEMBLE
**Common:** Physiological Monitoring System, Central Station Display Terminal
## Establishment Registration Number and Address
**Registration Number** 9611022
**Mennen Medical LTD.**
Alan Schwebel, Ph.D.
**President**
Kiryat Weizmann Science Park
Rehovot 76100 Israel
Tel: 972-8-938-3030
## Classification (Section 513)
**Classification:** Class III
**Panel:** Anesthesiology and Respiratory/Cardiology
## Performance Standards
None Established.
## Voluntary Standards
AAMI/ANSI ES11 Safe Current Limits for Electromedical Devices
AAMI/ANSI EC13, Cardiac Monitors, Heart Rate Meters and Alarms
UL-544
## Predicate Devices
MENNEN MEDICAL COMPREHENSIVE CENTRAL STATION (K892066).
HEWLETT-PACKARD M2360A WAVEVUE CENTRAL MONITOR (K921014).
## Date Prepared
10 November 1997
127
{1}
Food and Drug Administration
10 November 1997
510(k) Summary, ENSEMBLE Safety & Effectiveness
Page 2 of 7.
# Device Description
ENSEMBLE is hospital based, multi-patient, multi-parameter display terminal. As such, it is a component in a physiological patient vital signs monitoring system.
A system consists of any of the several possible combinations of the following components:
## Bedside Devices
HORIZON XL patient monitor, and/or
219 TELEMETRY RECEIVER/TRANSMITTER with NS-7,
(K911616, and K895395, respectively)
## Ethernet™ Communication Network
Documentation Peripherals (Chart Recorders and Laser Printers), and
ENSEMBLE Central Station Monitor.
ENSEMBLE acquires patient vital signs, waveforms and alarm indications from Mennen Medical HORIZON XL and 219 TELEMETRY systems. All processing and alarm determination is made in these “bedside devices”, using proprietary algorithms and software contained therein. Processed information is communicated from the bedside devices to one or more ENSEMBLE Central Station Monitors by the Mennen Medical ENSEMBLE system Ethernet™ network. If documentation is requested, information is transmitted to the appropriate Documentation Peripheral, using the same network.
Although ENSEMBLE does not make alarm determinations, it does provide users with a remote interaction capability, allowing users to make changes to both parameter and alarm settings in the remote bedside devices. Similarly, ENSEMBLE users have a remote interactive capability to initiate recordings or printing of patient information and waveforms.
Vital signs and waveform information from up to 12 patients are displayed simultaneously on an ENSEMBLE Central Station Monitor. Information from each individual patient is presented in a separate portion of the display. Each patient display area includes: Patient Name, Assigned Location (room), Primary Vital Sign Name, Value & Alarm Limits (nominally Heart Rate), Primary Waveform (nominally ECG), Secondary Vital Sign Numeric Data, Alarm Status Messages, and a dedicated “soft key” for recording waveforms. Software configurations allow more than one waveform or parameter to be displayed for each patient.
Primary operation of the ENSEMBLE is accomplished by a desktop mouse. A standard computer keyboard is used for manual entry of alphanumeric information. An optional touchscreen is available for users who prefer to use a touchscreen. Alternative “desktop devices”, such as trackball and light pen, are also available as options. For large installations, with several ENSEMBLEs intended for a single user, the system may be configured to enable one keyboard and one mouse to control up to eight individual ENSEMBLE computers.
ENSEMBLE is a reusable, software driven, central station display terminal, intended for use as part of a physiological monitoring system in a hospital environment. As such it is not a life supporting, or life sustaining device; nor is it implantable and therefore sterility is not a consideration. ENSEMBLE complies with UL 544, ANSI Safe Current Limits for Electromedical Apparatus, and FDA MDS-201-0004 voluntary standards. ENSEMBLE is not a kit, does not contain any drug or biological products and is for prescription use.
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
FEB 11 1998
Allen Schwebel, Ph.D.
Mennen Medical Ltd.
Kiryat Weizmann Science Park
P.O.B. 102
Rehovot 76100
ISRAEL
Re: K970358
Ensemble Central Station Monitor
Regulatory Class: III (three)
Product Code: 74 DSI
Dated: November 10, 1997
Received: November 13, 1997
Dear Dr. Schwebel:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}
970358
10 November 1997
Page 3 of 7.
Food and Drug Administration
510(k) Summary, ENSEMBLE Safety & Effectiveness
## Indications for Use:
ENSEMBLE includes the essential features of a central station monitor for up to 12 patients. It can display vital signs information collected from bedside patient monitoring devices such as: ECG/Heart Rate, Arrhythmia (basic and extended alarms), ST Segment, Respiration, Temperature, Invasive Blood Pressure, Noninvasive Blood Pressure, Pulse Oximetry, and End Tidal carbon-dioxide. The ENSEMBLE may be used in system application in many different clinical specialties within the hospital.
_________________________
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number **K970358**
*Prescription: ☑*
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.