# FUJI MEDICAL LASER IMAGER MF PRINT SERVER MF-PS667 (K970349)

_Fujifilm Medical System U.S.A., Inc. · LLZ · Mar 31, 1997 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K970349

## Device Facts

- **Applicant:** Fujifilm Medical System U.S.A., Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Mar 31, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The indication for use of the Fuji Medical Laser Imager MF Print Server MF-PS667 is to connect imaging modalities on a network to a hard copy system for hard copy output.

## Device Story

Device acts as network print server connecting DICOM or Toshiba medical image data networks to Fuji hard copy printing systems. Input: image data from modalities (CT, MR, NM, US) via network protocols. Processing: protocol conversion; simple image processing including rotation, reversal, and decompression of reversibly compressed (1/2 ratio) image data. Output: formatted image data sent to Fuji hard copy print system. Used in clinical environments; operated by medical personnel. Device does not display images; images are interpreted by physicians. Benefits: enables hard copy output of digital medical images for clinical review.

## Clinical Evidence

Bench testing only. Device complies with UL 1950 Standard for Safety of Information Technology Equipment. No clinical data provided.

## Technological Characteristics

Hardware: Sun SPARC Station, 64-bit micro-SPARC II microprocessor, 64MB RAM, 1.5 GB fixed magnetic disk. Connectivity: Ethernet (IEEE802.3 standard). Software: Protocol conversion, image rotation, reversal, and decompression. No patient contact. Password-protected operation.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Fuji Medical Laser Imager Multi-Formatters MF-300 and MF-300L ([K950721](/device/K950721.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

MAR 31 1997
K970349

# 510(k) Summary
[as required by 21 CFR 807.92]

## Date Prepared [21 CFR 807.92(a)(1)]
January 13, 1997

## Submitter's Information [21 CFR 807.92(a)(1)]
Fuji Medical Systems U.S.A., Inc.
Post Office Box 120035
Stamford, Connecticut 06912-0035
Telephone: +1 203 973-2664
facsimile: +1 203 973-2777
Contact: Robert A. Uzenoff

## Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
The device trade name is Fuji Medical Laser Imager MF Print Server MF-PS667.
The device common name is Print Server, Network Print Server.

## Predicate Device [21 CFR 807.92(a)(3)]
Fuji identifies the subject device as an accessory to the predicate device, the Fuji Medical Laser Imager Multi-Formatters MF-300 and MF-300L. FDA assigned the predicate to regulatory class II citing 21 CFR § 892.1000 (magnetic resonance diagnostic device).
FDA’s accession number for the premarket notification for the predicate device is K950721. FDA cleared the marketing of the predicate device in a letter dated May 24, 1995.

## Description of the Device [21 CFR 807.92(a)(4)]
The device consists of a computer (console, display, keyboard, and mouse) and software. The device connects a DICOM or Toshiba medical image data network to Fuji hard copy printing systems.
The network typically carries images from modalities (e.g. CT, MR, NM, or US) or other image sources being transported under DICOM or Toshiba network protocol. The print server accepts the image data, performs protocol conversion to Fuji’s protocol, and passes it to Fuji’s hard copy print system.

## Intended Use [21 CFR 807.92(a)(5)]
The indication for use of the Fuji Medical Laser Imager MF Print Server MF-PS667 is to connect imaging modalities on a network to a hard copy system for hard copy output.

RA Uzenoff h:usertech_grpdoc510k_667.doc 9/10 97/01/29 13:26
510(k)
Fuji MF Print Server
MF-PS667 510(k)
1997 January 27

{1}

FDA 510(k) Premarket Notification
Fuji Medical Laser Imager MF Print Server MF-PS667
January 27, 1997

## Technological Characteristics [21 CFR 807.92(a)(6)]

The device does not contact the patient, nor does it control any life sustaining devices. Images crossing the print serve are interpreted by a physician, providing ample opportunity for competent human intervention.

The subject and predicate devices are both based on an industry-standard Sun SPARC Station with 64-bit micro-SPACR II microprocessor, standard 64MB RAM, and 1.5 GB fixed magnetic disk storage. Both use a CRT for character display as part of the user interface. No image data is displayed.

Unlike the predicate, the subject device may perform simple image processing: image rotation, reversal, and the decompression of reversibly compressed (ratio: ½) image data. No compressed image data is sent to the DICOM network.

Data loss may occur because of hardware failure (e.g. hard disk crash) or operator error (e.g. turning off power during hard disk access). However, unlike the predicate, the device is a data source or image archive, the gateway transmission can be completed after correcting the cause of the failure. Operation is password protected to prevent unauthorized use.

## Performance Data [21 CFR 807.92(b)(1)]

The IEEE802.3 standard transceiver for Ethernet connection is employed. The subject and predicate devices both use standard data communications controls to detect and correct errors.

The device complies with the UL 1950 Standard for Safety of Information Technology Equipment, Including Electrical Business Equipment.

## Conclusion [21 CFR 807.92(b)(3)]

As is the case with the predicate, the subject device have no patient contact. Nor do the subject device control, monitor, or effect any devices directly connected to or effecting a patient. The images relayed by the subject devices are observed by medical personnel, offering ample opportunity for competent human intervention in the event of a failure.

The device functions as a DICOM standard communications protocol converter, and not as a permanent image store. Standard communications error detection and correction methods are employed. Standard network interface connections are employed. Device failures which might result in a failed transmission may be recovered from by retransmission after correcting the problem. Passwords are required for operation to protect against unauthorized use.

The subject and predicate share the same certification of conformance to the UL 1950 Standard for Safety of Information Technology Equipment, Including Electrical Business Equipment.

We conclude that the subject devices are as safe and effective as the predicate device.

RA Uzenoff h:\user\tech_grp\doc\510k_667.doc 10/10 97/01/29 13:26
510(k)
Fuji MF Print Server
MF-PS667 510(k)
1997 January 27
10

---

**Source:** [https://fda-staging.innolitics.com/device/K970349](https://fda-staging.innolitics.com/device/K970349)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).