ROTABLATOR ROTATIONAL ANGIOPLASTY SYSTEM WITH THE ROTALINK EXCHANGEABLE CATHETER
K970296 · Boston Scientific Corp · MCW · Aug 29, 1997 · Cardiovascular
Device Facts
Record ID
K970296
Device Name
ROTABLATOR ROTATIONAL ANGIOPLASTY SYSTEM WITH THE ROTALINK EXCHANGEABLE CATHETER
Applicant
Boston Scientific Corp
Product Code
MCW · Cardiovascular
Decision Date
Aug 29, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4875
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Rotablator system is intended for percutaneous use in peripheral vessels in patients with occlusive atherosclerotic disease who are acceptable candidates for bypass graft surgery or percutaneous transluminal angioplasty.
Device Story
Rotablator system uses high-speed, diamond-coated burr (140,000-190,000 RPM) to ablate occlusive atherosclerotic material in peripheral vessels; particles removed via reticuloendothelial system. System comprises advancer (air turbine), catheter with burr/drive shaft, console, foot pedal, and guide wire. Physician operates system in clinical setting; console monitors rotational speed; foot pedal controls turbine and guide wire brake. Catheter tracks over guide wire to lesion site; Teflon sheath protects arterial tissue and allows saline infusion for cooling/lubrication. Dynaglide mode (50,000-90,000 RPM) allows catheter exchange while keeping guide wire in place. Docking port holds wireClip torquer to prevent guide wire spinning during exchange. Device restores luminal patency, potentially benefiting patients by treating peripheral arterial occlusions.
Clinical Evidence
No clinical data. Bench testing only. Testing confirmed performance specifications (dynamic stall torque, steady state stall torque, distal flow rate) with 95% confidence and 99% probability. Functional life of catheter verified at 20 minutes. Lesion crossing time differences between A19 and A21 were not statistically significant.
Indicated for patients with occlusive atherosclerotic disease in peripheral vessels who are candidates for bypass graft surgery or percutaneous transluminal angioplasty.
Regulatory Classification
Identification
An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)
Predicate Devices
Rotablator system (A19)
Submission Summary (Full Text)
{0}
2970296
510(k) for Peripheral Rotablator® Rotational Angioplasty System with the
RotaLink™ Exchangeable Catheter
Page v
## TABLE OF CONTENTS
### 510(k) SUMMARY
AUG 29 1997
A. Introduction ... 6
B. Device Description ... 6
C. Intended Use ... 8
D. Comparison to Predicate Device ... 8
1. Functional Changes ... 9
2. Design and Material Modifications to Functional Components ... 10
3. Performance ... 10
E. *In Vitro* Tests ... 10
F. Biocompatibility ... 11
G. Sterilization ... 11
H. Clinical Tests ... 11
{1}
510(k) for Peripheral Rotablator® Rotational Angioplasty System with the
RotaLink™ Exchangeable Catheter
Page 6
# 510(k) SUMMARY
AUG 29 1997
## A. Introduction
This 510(k) is for the new Rotablator® System with the RotaLink™ Exchangeable Catheter (referred to as the A21 Rotablator system throughout this summary). The A21 Rotablator system is designed to allow for the exchange of the catheter by dividing the advancer into two components: an advancer with a short drive shaft and connector, and a catheter with a long drive shaft and connector. The indications for use of this device remain the same.
Submitter: Boston Scientific Corporation Northwest Technology Center, Inc.
17425 N.E. Union Hill Road
Redmond, WA 98052
Contact: Diane Johnson
Phone: (206) 556-1541
Fax: (206) 558-1400
Preparation Date: January 23, 1997
Device Common Name: Rotational Angioplasty System
Device Proprietary Name: Rotablator® Rotational Angioplasty System with the RotaLink™ Exchangeable Catheter
Classification Name: Catheter, Peripheral, Atherectomy (per 21 CFR 870.4875)
Classification Panel: Cardiovascular
Manufacturing Facilities: Boston Scientific Corporation Northwest Manufacturing Center, Inc.
17425 N.E. Union Hill Road
Redmond, WA 98052
## B. Device Description
The Rotablator® Rotational Angioplasty system uses a high speed, rotating, diamond-coated burr to ablate occlusive material and restore luminal patency. The burr spins at 140,000-190,000 RPM and ablates material into very fine particles that are carried distally and removed via the reticuloendothelial system. The A21 Rotablator system comprises four main components: advancer, catheter with diamond-coated burr, console and foot pedal, and guide wire.
The advancer houses the air turbine and functions as a guide for the sliding elements that control burr advancement. The turbine is designed with low torque so that it will stall if it
{2}
510(k) for Peripheral Rotablator® Rotational Angioplasty System with the RotaLink™ Exchangeable Catheter Page 7
encounters resistance. This feature prevents inadvertent damage to the artery. The advancer has an air-actuated guide wire brake which is automatically actuated when the operator steps on the foot pedal to turn on the advancer. This feature prevents movement of the guide wire when the drive shaft is turning.
The catheter consists of the burr and drive shaft, the sheath, and the catheter body. The burr is attached to the end of the flexible drive shaft which is connected to an air turbine and powered by compressed air or nitrogen. The drive shaft, which has a central lumen for a guide wire to pass through, is flexible. This feature facilitates passage through tortuous arteries and around multiple sharp bends, and permits the burr to reach lesions in small distal arteries. The catheter is covered by a Teflon™ sheath which protects arterial tissue from the spinning drive shaft and permits the passage of saline to lubricate and cool the drive shaft. A contrast agent can be delivered to the lesion site by flowing between the sheath and the guide catheter. The A21 peripheral catheter will be available in burr sizes ranging from 1.5 mm to 4.0 mm. The various sizes allow for maximum debulking when treating lesions in small or distal vessels as well as those in larger arteries.
The Rotablator catheter tracks over a guide wire which is passed through the lesion to be treated. The guide wire directs the catheter to the lesion and keeps it aligned within the artery. This allows the catheter tip to be redirected as it progresses around bend points, thus decreasing the possibility of the tip contacting the outer wall and potentially causing tissue damage. The guide wire has a radiopaque spring tip which facilitates its passage through the vasculature, minimizes trauma to the vessel, and makes progress visible on fluoroscopy.
The console monitors and controls the rotational speed of the burr and continuously provides the operator with performance information during the procedure. The console has two modes of operation: a high speed for ablation and a lower speed for catheter exchange (i.e., Dynaglide™ mode). In the Dynaglide mode, the drive shaft rotates at a reduced speed of 50,000-90,000 RPM, thus allowing the physician to defeat the guide wire brake and allow the guide wire to remain in place while the catheter is withdrawn. This feature facilitates removal of the catheter from the coronary arteries. The Rotablator console is powered by house current, however, no electricity is transmitted to the advancer or to the patient.
The foot pedal is the on/off control for the advancer air turbine and is mounted in a protective shroud to inhibit accidental actuation. The pedal is fitted with a valve that vents any compressed gas left in the foot pedal hose when the pedal is released, permitting the burr to stop rapidly. The foot pedal also has a toggle switch for activating and deactivating the lower speed used during catheter exchange.
The compressed gas system consists of a regulator mounted on a compressed gas cylinder and a supply hose leading to the control console inlet.
{3}
510(k) for Peripheral Rotablator® Rotational Angioplasty System with the
RotaLink™ Exchangeable Catheter
Page 8
## C. Intended Use
The Rotablator system is intended for percutaneous use in peripheral vessels in patients with occlusive atherosclerotic disease who are acceptable candidates for bypass graft surgery or percutaneous transluminal angioplasty.
## D. Comparison to Predicate Device
The A21 Rotablator system is substantially equivalent to the currently marketed Rotablator system (referred to as A19 throughout this summary). The difference between the two devices is two functional changes and several minor modifications in material and design of the A21, none of which affects function. In addition, the largest burr used in the A19 system, 4.5 mm, will not be manufactured for the A21 system.
The following table summarizes the functional and non-functional changes.
| | A19 | A21 |
| --- | --- | --- |
| Functional | Advancer and catheter one unit.
No docking port. | Advancer and catheter two separate units.
Docking port to hold wireClip. |
| Design and Material Changes (no functional impact):
Drive Mechanism | Adhesive bonds between bearings, turbine shaft, and bearing housing.
Retro-reflector press fit onto turbine shaft.
Cyanoacrylate adhesive bond between turbine shaft and pump shaft.
Epoxy bond between pump shaft and drive coil. | Press fit of turbine components into turbine housing.
Retro-reflector slip fit onto turbine shaft.
Ultraviolet cure adhesive bond between turbine shaft and pump shaft.
Laser weld between drive coil and pump shaft and connectors. |
| Design and Material Changes (no functional impact):
Fluid Path | Stainless steel pump rotor.
Polyetherimide front plug.
FEP small hypo tube liner.
Cyanoacrylate adhesive bond between pump rotor and pump shaft.
NA
Diamond grit size: .002-.0031" | Polysulfone pump rotor.
Polysulfone front plug.
PTFE small hypo tube liner (same material as large hypo tube liner).
Pump rotor press fit onto pump shaft or pump rotor insert molded to pump shaft.
Catheter connection made of ABS; new material to fluid path.
Diamond grit size: .0008-.0012" |
{4}
510(k) for Peripheral Rotablator® Rotational Angioplasty System with the
RotaLink™ Exchangeable Catheter
Page 9
# 1. Functional Changes
## a. Catheter and Drive Shaft Connections
In the A21, the drive shaft is made up of two components: the short drive shaft, which is part of the advancer, and the long drive shaft, which is part of the catheter. In the A19, the drive shaft is a single component, and the catheter and advancer are a single unit. By allowing for the separation of the catheter from the advancer, several catheters with various burr sizes can be attached to a single advancer during a procedure. This change is incorporated to facilitate the multiple-burr treatment regime used by many physicians. The connection of the catheter to the advancer involves two systems: drive shaft connection plus slide tube, and the catheter connection. Once the drive shaft and catheter are connected to the advancer, the A21 advancer/catheter is functionally equivalent to the A19 advancer/catheter.
## b. Docking Port
The A21 has a new feature, the docking port. The docking port facilitates conducting the clinical procedure in several ways. First, the wireClip® torquer can be inserted into the docking port, where it is held during the procedure. Second, during an exchange procedure, when the brake defeat button has been depressed, the torquer can be inserted further into the docking port to serve two functions: the docking port holds the torquer in place which prevents spinning of the guide wire, and the torquer prevents the brake from engaging, even when the brake defeat button is not being depressed. This configuration enhances safety and efficiency because neither the operator nor additional personnel are required to hold the torquer.
{5}
510(k) for Peripheral Rotablator® Rotational Angioplasty System with the
RotaLink™ Exchangeable Catheter
Page 10
## 2. Design and Material Modifications to Functional Components
The advancer body has been changed from a solid machined plastic tube to a two-piece snap together shell, i.e., a top and a bottom. The catheter body also is a two piece snap together assembly.
## 3. Performance
All performance specifications for the A21 Rotablator system such as speed, infusate flow, torque, etc., are identical or more stringent than those for the A19 Rotablator system, therefore, the A21 is equivalent in performance to the A19 Rotablator system.
## E. In Vitro Tests
Bench testing was done to show that the performance of the peripheral A21 system conforms to product specifications and that it is as safe and effective as the current peripheral A19 system. Testing was conducted on performance characteristics that are influenced by the catheter’s configuration and are clinically relevant.
Using statistically valid sample sizes, the testing showed with 95% confidence that there is a 99% probability that the following performance specifications will be met: dynamic stall torque, steady state stall torque, and distal flow rate.
The test also indicated the following:
- With 95% confidence, there is a 99% probability that the functional life of the catheter will be 20 minutes.
- The difference in lesion crossing time between the A19 and the A21 were not statistically significant.
Additional testing was done for further comparison of clinically relevant performance characteristics: speed versus console pressure, and stall torque versus console pressure. Five A19 1.5 mm burrs and five A19 3.0 mm burrs were tested - the two burr sizes allow for testing the two types of turbine nozzles in the A19. For the A21 system, one advancer and five catheters were tested. Only one burr size was used for the A21 since they are used with the same type of turbine nozzle.
The results of the speed versus pressure test indicate that the A21 runs slower than the A19 at a given console pressure. The results of the stall torque versus pressure test indicate that at any given console pressure lower than 50 PSI, the A21 delivers less torque than does the A19. However, at console pressures of 50 PSI or greater, this difference decreases. These performance differences between the A19 and the A21 are not significant from the user’s point of view and does not affect the A21’s effectiveness because the maximum suggested operational speed of 180 KRPM can still be attained. Also, the higher operational pressures
{6}
510(k) for Peripheral Rotablator® Rotational Angioplasty System with the RotaLink™ Exchangeable Catheter Page 11
required to attain a given speed for the A21 will serve to offset the lower torques of the A21 since torque increases with pressure.
## F. Biocompatibility
A biological evaluation was performed on the A21 Rotablator system to give a high degree of assurance that the final product will be safe for human use. These tests included material characterization, cytotoxicity, pyrogenicity and hemocompatibility assays conducted under Good Laboratory Practices (GLP). The infusion pathway portion of the device was evaluated as a whole composite following two EtO sterilization cycles. The A21 Rotablator system was found to be compatible with biological systems.
## G. Sterilization
Following final packaging and labeling, finished devices are 100% ethylene oxide (EtO) sterilized to achieve a sterility assurance level (SAL) of 10⁻⁶, or one nonsterile unit out of one million. The Rotablator device is nonpyrogenic based on the Limulus Amebocyte Lysate (LAL) assay.
## H. Clinical Tests
Clinical studies were not conducted on the A21 design. As part of the design criteria for the A21 Rotablator system, it was specified that the therapeutic benefit of the current A19 Rotablator system not be changed. All parts that come in contact with the patient are, therefore, the same as the corresponding parts in the current system, specifically the burr, the drive shaft, and the sheath material. Any changes made to the system to facilitate the separable catheter design, improve manufacturability and reliability, and reduce the unit cost of the device, have no significant impact on the therapeutic benefit of the device, therefore, no clinical study was required.
The aspects of the device that relate to patient safety are 1) biocompatibility of the materials in the fluid contact path, 2) device sterility, 3) interaction between the burr and the vessel wall, 4) infusion of saline at the treatment site, and 5) functionality of the brake system. All of these aspects of the A21 have been tested and show the device to be safe and the clinically relevant parameters unchanged from those of the A19.
{7}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
Ms. Diane Johnson
Manager, Regulatory Affairs
Boston Scientific Corporation
Northwest Technologies Center, Inc.
17425 N.W. Union Hill Road
Redmond, Washington 98052-3376
AUG 29 1997
Re: K970296
Peripheral Rotablator® Rotational Angioplasty System with the
RotaLink™ Exchangeable Catheter
Regulatory Class: II (Two)
Product Code: MCW
Dated: May 29, 1997
Received: June 2, 1997
Dear Ms. Johnson:
We have reviewed your Section 510(k) notification of intent to market
the device referenced above and we have determined the device is
substantially equivalent (for the indications for use stated in the
enclosure) to devices marketed in interstate commerce prior to
May 28, 1976, the enactment date of the Medical Device Amendments, or
to devices that have been reclassified in accordance with the
provisions of the Federal Food, Drug, and Cosmetic Act (Act). You
may, therefore, market the device, subject to the general controls
provisions of the Act. The general controls provisions of the Act
include requirements for annual registration, listing of devices, good
manufacturing practice, labeling, and prohibitions against misbranding
and adulteration.
If your device is classified (see above) into either class II (Special
Controls) or class III (Premarket Approval), it may be subject to such
additional controls. Existing major regulations affecting your device
can be found in the Code of Federal Regulations, Title 21, Parts 800
to 895. A substantially equivalent determination assumes compliance
with the Current Good Manufacturing Practice requirements, as set
forth in the Quality System Regulation (QS) for Medical Devices:
General regulation (21 CFR Part 820) and that, through periodic QS
inspections, the Food and Drug Administration (FDA) will verify such
assumptions. Failure to comply with the GMP regulation may result in
regulatory action. In addition, FDA may publish further announcements
concerning your device in the Federal Register. Please note: this
response to your premarket notification submission does not affect any
obligation you might have under sections 531 through 542 of the Act
for devices under the Electronic Product Radiation Control provisions,
or other Federal laws or regulations.
{8}
Page 2 - Ms. Diane Johnson
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{9}
510(k) for Peripheral Rotablator® Rotational Angioplasty System with the
RotaLink™ Exchangeable Catheter
Page 12
## SECTION III INDICATIONS FOR USE
Device: Rotablator® Rotational Angioplasty System with RotaLink™ Exchangeable Catheter
The Rotablator system is intended for percutaneous use in peripheral vessels in patients with occlusive atherosclerotic disease who are acceptable candidates for bypass graft surgery or percutaneous transluminal angioplasty.
Dave Johnson
For Boston Scientific Corporation Northwest Technology Center
January 23, 1997
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K970296
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.