K970269 · Sims Deltec, Inc. · LJS · Jun 26, 1997 · General Hospital
Device Facts
Record ID
K970269
Device Name
CLINICATH IR PERIPHERALLY INSERTED CATHETER
Applicant
Sims Deltec, Inc.
Product Code
LJS · General Hospital
Decision Date
Jun 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The CliniCath™ IR Peripherally Inserted Central Catheter is used when therapy requires repeated venous access for intravenous injection or infusion and/or venous blood sampling.
Device Story
CliniCath™ IR is a peripherally inserted central catheter (PICC) designed for repeated venous access; facilitates delivery of medications, fluids, nutritional solutions, blood therapy, and blood sampling. Device consists of a single-lumen, radiopaque silicone catheter with a preinserted stiffening wire and flushing assembly. Available with accessories including introducer needle, guidewire, dilator/sheath assembly, syringe, tape measure, mini-scalpel, and injection cap. Used in clinical settings by healthcare professionals to provide central venous access via peripheral vein insertion. Benefits include reliable, repeated access for therapy, reducing need for multiple venipunctures.
Clinical Evidence
No clinical data. Bench testing only. In-vitro functional testing conducted per FDA 'Guidance on Premarket Notification [510(k)] Submissions for Short- and Long-term Intravascular Catheters' (March 1995). Biocompatibility testing performed.
Technological Characteristics
Radiopaque silicone catheter; single-lumen; 4 French (18G) and 5 French (16G) sizes; 65 cm length. Includes preinserted stiffening wire and flushing assembly. Manual insertion via peripheral vein for central tip placement. Non-powered, passive device.
Indications for Use
Indicated for patients requiring repeated venous access for intravenous injection, infusion therapy, or venous blood sampling.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Peripherally Inserted Central Venous Catheter Sets (Cook, Inc.)
Submission Summary (Full Text)
{0}
510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
CliniCath™ IR Peripherally Inserted Central Catheters - K970269
January 22, 1997
JUN 26 1997
I. GENERAL INFORMATION
Applicant's Name and Address: SIMS Deltec, Inc.
1265 Grey Fox Road
St. Paul, MN 55112
Contact Person: Lisa Stone
Manager, Regulatory Affairs
Tel. (612) 628-7224
Common/Usual Name: Peripherally Inserted Venous Catheters
Proprietary Name: CliniCath™ IR Peripherally Inserted Central Catheters
Equivalence Device Comparison: CliniCath™ Peripherally Inserted Venous Catheters
(manufactured by SIMS Deltec, Inc.)
Peripherally Inserted Central Venous Catheter Sets
(manufactured by Cook, Inc.)
II. DEVICE DESCRIPTION
The modified CliniCath™ IR Peripherally Inserted Central Catheters are similar in design and function to the current commercially available CliniCath™ Peripherally Inserted Venous Catheters.
The catheters are designed to permit repeated access to the venous system for the delivery of medications, fluids, and nutritional solutions, and for blood therapy and/or sampling.
The product consists of a single-lumen, radiopaque silicone catheter with preinserted catheter stiffening wire and attached flushing assembly. Catheters will be made available with and without the following accessories: introducer needle, guidewire, dilator/sheath assembly, syringe, tape measure, mini-scalpel and injection cap. Catheters will be provided in single- and multi-unit package configurations.
{1}
510(k) Summary of Safety and Effectiveness
Page 2 of 3
## III. INTENDED USE OF DEVICE
The CliniCath™ IR Peripherally Inserted Central Catheter is used when patient therapy requires repeated venous access for intravenous injection or infusion therapy and/or venous blood sampling.
## IV. DEVICE COMPARISON
| | CliniCath™ IR Peripherally Inserted Central Catheters | CliniCath™ Peripherally Inserted Catheters | Cook Peripherally Inserted Central Venous Catheter Sets |
| --- | --- | --- | --- |
| MANUFACTURER | SIMS Deltec, Inc. | SIMS Deltec, Inc. | Cook, Inc. |
| INDICATION FOR USE | The CliniCath™ IR Peripherally Inserted Central Catheter is used when therapy requires repeated venous access for intravenous injection or infusion and/or venous blood sampling. | The CliniCath™ Peripherally Inserted Central Catheter is used when therapy requires repeated venous access for intravenous injection or infusion and/or venous blood sampling. | Peripherally Inserted Central Venous Catheter is suggested for: (1) Delivery of whole blood or blood products; (2) Drug administration; and (3) Blood sampling. |
| CATHETER INSERTION | Peripheral vein | Peripheral vein | Peripheral vein |
| CATHETER TIP PLACEMENT | Central | Peripheral/Central | Central |
| CATHETER MATERIAL | Radiopaque Silicone | Radiopaque Silicone | Radiopaque Silicone |
| AVAILABLE CATHETER SIZES | 4 French (18 Gauge)
5 French (16 Gauge) | 3 French (20 Gauge)
4 French (18 Gauge)
5 French (16 Gauge) | 3 French
4 French
5 French |
| CATHETER O.D. AND I.D. (Nominal) | 1.4 mm O.D. x 0.7 mm I.D.
1.7 mm O.D. x 1.0 mm I.D. | 1.3 mm O.D. x 0.8 mm I.D.
1.7 mm O.D. x 1.0 mm I.D. | 1.4 mm O.D. x 0.6 mm I.D.
1.7 mm O.D. x 0.9 mm I.D. |
| 4 French
5 French | | | |
| CATHETER LENGTH (Nominal) | 65 cm | 20 cm and 65 cm | 50 cm and 60 cm |
| FLOW RATE (Nominal)
(1 METER GRAVITY FLOW) | 4 French (65 cm)
7 ml/min. ± 0.7 ml/min.
5 French (65 cm)
20 ml/min. ± 1.2 ml/min. | 4 French (65 cm)
479 ml/hr
5 French (65 cm)
1420 ml/hr | 4 French (60 cm)
270 ml/hr
5 French (60 cm)
Sample not available for testing. |
{2}
510(k) Summary of Safety and Effectiveness
Page 3 of 3
# V. SUMMARY OF STUDIES
## A. Functional Testing
In-vitro testing was conducted in accordance with the FDA "Guidance on Premarket Notification [510(k)] Submissions for Short- and Long-term Intravascular Catheters," dated March 1995.
Biocompatibility testing was conducted on the device.
## B. Clinical Studies
Clinical studies were not deemed necessary regarding the CliniCath™ IR Peripherally Inserted Central Catheter due to its similarity in materials; design and function to current commercially available CliniCath™ Peripherally Inserted Venous Catheters.
## C. Conclusion Drawn from the Studies
The results of the testing indicated that the CliniCath™ IR Peripherally Inserted Central Catheter functions according to specification and the materials used in the device are biocompatible. Therefore, the device is considered acceptable for human use.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Ms. Lisa Stone
Manager, Regulatory Affairs
SIMS Deltec, Incorporated
1265 Grey Fox Road
St. Paul, Minnesota 55112
Re: K970269
Trade Name: Clini-Cath™ IR Peripherally Inserted Catheter
Regulatory Class: Unclassified
Product Code: LJS
Dated: June 17, 1997
Received: June 18, 1997
Dear Ms. Stone:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General
{4}
Page 2 - Ms. Stone
(GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the *Federal Register*. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4618. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Timothy A. Uratowski
Director
Division of Dental, Infection Control, and General Hospital Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
510(k) Number (if known): R970269
Device Name: **CliniCath™ IR Peripherally Inserted Central Catheters**
Indications for Use:
*The CliniCath™ IR Peripherally Inserted Central Catheter is used when therapy requires repeated venous access for intravenous injection or infusion and/or venous blood sampling.*
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Patricia Cucinita
(Division Sign-Off) R970269
Division of Dental, Infection Control,
and General Hospital Devices
510(k) Number R970269
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The Counter Use ☐
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.