K970241 · Depuy, Inc. · LPH · Aug 14, 1997 · Orthopedic
Device Facts
Record ID
K970241
Device Name
DEPUY PROXIMAL FEMORAL REPLACEMENT PROSTHESIS
Applicant
Depuy, Inc.
Product Code
LPH · Orthopedic
Decision Date
Aug 14, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DePuy Proximal Femoral Replacement Prosthesis is indicated for uncemented or cemented use as the femoral component in total hip arthroplasty for replacing the hip joint in the following indications, as appropriate: 1) a severely painful and/or a severely disabled joint resulting from osteoarthritis, traumatic arthritis, or rheumatoid arthritis, 2) avascular necrosis of the femoral head, 3) acute traumatic fracture of the femoral head or neck, 4) failed previous surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or other total hip replacement, and 5) certain cases of ankylosis.
Device Story
The DePuy Proximal Femoral Replacement Prosthesis is a femoral component for total hip arthroplasty. It consists of a straight, cast Co-Cr-Mo alloy stem with a porous coating (excluding the distal tip) and a 12/14 taper for modular femoral head attachment. The device is implanted by an orthopedic surgeon in a clinical/surgical setting to replace a diseased or damaged hip joint. By providing a stable, modular femoral component, the device restores joint function, reduces pain, and improves mobility for patients suffering from degenerative or traumatic hip conditions. It is designed for both cemented and uncemented fixation.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Material: Cast Co-Cr-Mo alloy. Design: Straight, collarless, porous-coated stem with 12/14 taper for modular femoral heads. Distal tip: Tapered, non-porous coated 'bullet' shape. Fixation: Cemented or uncemented. Dimensions: Available in three femoral sizes.
Indications for Use
Indicated for patients requiring total hip arthroplasty due to severe pain/disability from osteoarthritis, traumatic/rheumatoid arthritis, avascular necrosis, acute traumatic fracture of femoral head/neck, failed previous hip surgery, or ankylosis. Suitable for cemented or uncemented use.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
Porocoat Proximal Femoral Third Prosthesis
Proximal Femur Replacement Stem
Solution Hip Prosthesis
5/8 Porocoat AML Hip Prosthesis
Submission Summary (Full Text)
{0}
08/06/97 WED 07:48 FAX
004
K970241
1 of 2
AUG 14 1997
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
**NAME OF FIRM:**
DePuy Orthopaedics, Inc.
P.O. Box 988
700 Orthopaedic Drive
Warsaw, Indiana 46581-0988
**510(K) CONTACT:**
Sally Foust
Clinical Affairs Associate
(219) 372-7455
FAX (219) 267-7098
e mail: Sally_Foust@ccgate.depuy.com
**TRADE NAME:**
DePuy Proximal Femoral Replacement Prosthesis
**COMMON NAME:**
Hip Prosthesis
**CLASSIFICATION:**
888.3358 Hip joint metal/polymer semi-constrained porous coated uncemented prosthesis
**DEVICE PRODUCT CODE:**
87LPH
**SUBSTANTIALLY EQUIVALENT DEVICES:**
- Porocoat Proximal Femoral Third Prosthesis
- Proximal Femur Replacement Stem
- Solution Hip Prosthesis
- 5/8 Porocoat AML Hip Prosthesis
**DEVICE DESCRIPTION AND INTENDED USE:** The DePuy Proximal Femoral Replacement Prosthesis is indicated for uncemented and cemented use as the femoral component in total hip arthroplasty for replacing the hip joint in the following indications, as appropriate: 1) a severely painful and/or a severely disabled joint resulting from osteoarthritis, traumatic arthritis, or rheumatoid arthritis, 2) avascular necrosis of the femoral head, 3) acute traumatic fracture of the femoral head or neck, 4) failed previous surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or other total hip replacement, and 5) certain cases of ankylosis. It is a porous coated (except for the tapered distal tip), straight, cast Co-Cr-Mo stem available in three femoral sizes. The stem has a driving platform, a self-locking 12/14, 0540" taper for attachment of interchangeable modular femoral heads.
**BASIS OF SUBSTANTIAL EQUIVALENCE:**
The Proximal Femoral Replacement Prosthesis is similar to the Porocoat Proximal Femoral Third Prosthesis, Proximal Femur Replacement Stem, Solution Hip Prosthesis and 5/8 Porocoat AML Hip Prosthesis which have been previously cleared by FDA. The intended use of the Proximal Femoral Replacement Prosthesis is for uncemented and cemented applications just like the Solution Hip Prosthesis, whereas the Porocoat Proximal Femoral Third Prosthesis and Proximal Femur Replacement Stem is intended for cemented use only. The stem and porous coating material, Co-Cr-Mo alloy, for the Proximal Femoral Replacement Prosthesis is identical to that of the Porocoat Proximal Femoral Third Prosthesis and the Solution Hip Prosthesis, whereas the stem and porous coating material for the
{1}
08/06/97 WED 07:48 FAX
005
k970241
2 of 2
Proximal Femur Replacement Stem is Ti-6Al-4V. All of the above stems are porous coated. The placement of the porous coating on the total stem, except for the distal tapered tip, for the Proximal Femoral Replacement Prosthesis is identical to that of the Solution Hip Prosthesis, whereas the porous coating for the Porocoat Proximal Femoral Third Prosthesis and the Proximal Femur Replacement Stem is on the distal stem and the entire stem respectively. The Proximal Femoral Replacement Prosthesis, Porocoat Proximal Femoral Third Prosthesis, and the Solution Hip Prosthesis each has a "bullet" tapered distal tip, whereas the Proximal Femur Replacement Stem has a "rounded" non-tapered distal tip. The Proximal Femoral Replacement Prosthesis and the Proximal Femur Replacement Stem are collarless, whereas the Porocoat Proximal Femoral Third Prosthesis has a mid stem "step" and the Solution Hip Prosthesis has a collar. The Proximal Femoral Replacement Prosthesis, the Proximal Femur Replacement Stem and the Solution Hip Prosthesis are designed to accept modular femoral heads, whereas the Porocoat Proximal Femoral Third Prosthesis is a non-modular design with the femoral head attached.
The subject Proximal Femoral Replacement Prosthesis is similar to the Proximal Femur Replacement Stem previously cleared by FDA with the exception of the change in taper/neck design, the tapered distal stem without porous coating, the material change to Co-Cr-Mo, and the addition of an uncemented use indication. The manufacturing process, intended use (with the exception of the additional indication of uncemented use) and the basic design of the Proximal Femoral Replacement Prosthesis with a 12/14 taper have not changed from those of the Proximal Femur Replacement Stem cleared by FDA in 1990.
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Ms. Sally Foust
Clinical Affairs Associate
DePuy Inc.
P.O. Box 988
700 Orthopaedic Drive
Warsaw, Indiana 46581-0988
Re: K970241
Trade Name: DePuy Proximal Femoral Replacement
(PFR) Prosthesis
Regulatory Class: II
Product Codes: LPH, JDI, and LZO
Dated: June 19, 1997
Received: June 20, 1997
Dear Ms. Foust:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that,
{3}
Page 2 - Ms. Sally Foust
through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and Restorative Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{4}
08/06/97 WED 07:48 FAX
003
510(k) Number (if known) K970241
Device Name DePuy Proximal Femoral Replacement Prosthesis
Indications for Use:
The DePuy Proximal Femoral Replacement Prosthesis is indicated for uncemented or cemented use as the femoral component in total hip arthroplasty for replacing the hip joint in the following indications, as appropriate: 1) a severely painful and/or a severely disabled joint resulting from osteoarthritis, traumatic arthritis, or rheumatoid arthritis, 2) avascular necrosis of the femoral head, 3) acute traumatic fracture of the femoral head or neck, 4) failed previous surgery including joint reconstruction, internal fixation, arthrodesis, hemiarthroplasty, surface replacement arthroplasty, or other total hip replacement, and 5) certain cases of ankylosis.
Concurrence of CDRH, Office of Device Evaluation

Prescription Use ☑ OR Over-The Counter Use ☐ (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.