GSI CANNULA WITH INTRODUCER

K970201 · General Surgical Innovations · DYB · Apr 21, 1997 · Cardiovascular

Device Facts

Record IDK970201
Device NameGSI CANNULA WITH INTRODUCER
ApplicantGeneral Surgical Innovations
Product CodeDYB · Cardiovascular
Decision DateApr 21, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2

Indications for Use

The intended use of the GSI Cannula with Introducer is the same as the predicate devices (Hasson Stable Access Cannula K903078). The intended use of the predicate device and the GSI Cannula with Introducer is to provide access for surgical and diagnostic instruments for the duration of a surgical procedure.

Device Story

Hand-held cannula with introducer; used in minimally invasive surgery. Device consists of cannula body, inflatable valve/seal, and stopcock. Introducer (sharp or blunt tip) facilitates placement into surgical space. Inflatable valve/seal provides access port for surgical/diagnostic instruments while maintaining pneumoperitoneum via gas insufflation. Operated by surgeons in clinical/OR settings. Provides stable access for instruments; enables minimally invasive procedures; benefits patient by facilitating endoscopic access.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Materials: molded rigid plastic, elastomeric polymer, stainless steel. Components: cannula body, inflatable valve/seal, stopcock, sharp/blunt tip introducer. Disposable, sterile. No specific ASTM standards cited.

Indications for Use

Indicated for patients undergoing minimally invasive surgical or diagnostic procedures requiring instrument access to an endoscopic or laparoscopic surgical space.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Submission Summary (Full Text)

{0} K970201 # Section 10.0: 510(K) Summary # APR 21 1997 This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92(c). ## 10.1 Contact Person Cathleen Mantor General Surgical Innovations, Inc. 3172A Porter Drive Palo Alto, CA 94304 (415) 812-9740 extension 292 (415) 812-9731 fax ## 10.2 Date Summary Prepared January 14, 1997 ## 10.3 Trade Name To Be Determined ## 10.4 Common Name Cannula with Introducer ## 10.5 Predicate Device K903078 Hasson Stable Access Cannula. Manufacturer: Marlow Surgical Technology. ## 10.6 Device Description GSI’s Cannula with Introducer is a hand held device comprised of a Cannula with an Introducer. The Cannula is comprised of a cannula body, an inflatable valve/seal and a stopcock. The inflatable valve/seal is inflated and deflated through the stopcock. The inflatable valve/seal provides a seal for surgical tools and diagnostic instruments during minimally invasive procedures using gas insufflation. The Cannula may be supplied with either a Sharp Tip or Blunt Tip Introducer. The Introducer is used for placement of the Cannula into an endoscopic or laparoscopic surgical space. January 17, 1997 {1} # 10.7 Intended Use The intended use of the GSI Cannula with Introducer is the same as the predicate devices (Hasson Stable Access Cannula K903078). The intended use of the predicate device and the GSI Cannula with Introducer is to provide access for surgical and diagnostic instruments for the duration of a surgical procedure. # 10.8 Comparison to Predicate Devices | | Predicate Device Stable Access Cannula (K903078) | GSI’s Cannula with Introducer | | --- | --- | --- | | Intended Use | Access for minimally invasive procedures | Access for minimally invasive procedures | | Principles of Operation | Insertion through incision, under direct visualization (when possible) when using sharp tip | Insertion through incision, under direct visualization (when possible) when using sharp tip | | Design: -cannula -valve/seal -anchoring -method -insufflation port | rigid flexible valve and seal inflatable balloon type included | flexible inflatable, flexible valve/seal screw type not included | | Materials | molded rigid plastic, elastomeric polymer and stainless steel | molded rigid plastic, elastomeric polymer and stainless steel | | Supplied | Sterile, 5 devices/box w/5mm reducer, disposable | Sterile, 5 devices/box, inflatable valve/seal provides reducer function, disposable | | Introducer Type | blunt & sharp | blunt & sharp | | Cannula Lumen | 10/12 mm | 10 mm | | Compliance w/Standards | No known standards | No known standards | # 10.9 Testing in Support of Substantial Equivalence Determination Results of performance bench tests verify that the differences between the Cannula with Introducer and the predicate device do not affect the safety or efficacy of the GSI Cannula with Introducer. January 17, 1997
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...