RELIANCE PF FEMORAL STEMS

K970200 · Howmedica Corp. · LZO · Apr 9, 1997 · Orthopedic

Device Facts

Record IDK970200
Device NameRELIANCE PF FEMORAL STEMS
ApplicantHowmedica Corp.
Product CodeLZO · Orthopedic
Decision DateApr 9, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3353
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Reliance® PF stem is a femoral component of a Total Hip System. It is intended to be used for primary or secondary reconstruction of the proximal femur damaged as a result of inflammatory, non-inflammatory, or traumatic joint disease. The Reliance® PF stem is intended to be used in conjunction with Howmedica's V40 femoral heads, Acetabular components, and unipolar and bipolar components. The stem can be press-fit into the proximal femur for fixation by mechanical interlock, or affixed with bone cement.

Device Story

Reliance® PF Femoral Stem; metallic hip prosthesis component. Used for primary/secondary proximal femur reconstruction. Operates via press-fit mechanical interlock or bone cement fixation. Used in conjunction with V40 femoral heads, acetabular components, and unipolar/bipolar components. Physician-operated in surgical setting. Provides structural replacement for damaged hip joint; restores joint function.

Clinical Evidence

Bench testing only. Endurance load testing demonstrated performance exceeding the Semlitsch minimum of 630 lbs.

Technological Characteristics

Material: Forged cobalt-chromium-molybdenum (Vitallium®) alloy. Design: Femoral stem for total hip arthroplasty. Fixation: Press-fit (mechanical interlock) or cemented. Energy source: None (mechanical).

Indications for Use

Indicated for primary or secondary reconstruction of proximal femur damaged by inflammatory, non-inflammatory, or traumatic joint disease.

Regulatory Classification

Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

Predicate Devices

Submission Summary (Full Text)

{0} APR - 9 1997 510(k) Summary K970200 Proprietary Name: Reliance® PF Femoral Stems Common Name: Metallic Hip Stem Classification Name and Reference: 21 CFR 888.3353 Hip Joint Metal/Ceramic/Polymer semi-constrained cemented or nonporous uncemented prosthesis. Proposed Regulatory Class: Class II Device Product Code: LZO For information contact: Frank Maas Manager, Regulatory Affairs Howmedica Inc. 359 Veterans Boulevard Rutherford, NJ 07070 Telephone: (201) 507-7875 Fax: (201) 507-6870 Date Summary Prepared: 1-17-97 The Reliance® PF stem is a femoral component of a Total Hip System. It is intended to be used for primary or secondary reconstruction of the proximal femur damaged as a result of inflammatory, non-inflammatory, or traumatic joint disease. The Reliance® PF stem is intended to be used in conjunction with Howmedica's V40 femoral heads, Acetabular components, and unipolar and bipolar components. The stem can be press-fit into the proximal femur for fixation by mechanical interlock, or affixed with bone cement. The Reliance® PF stem is manufactured from forged cobalt-chromium-molybdenum (Vitallium®) alloy. The substantial equivalence of the Reliance® PF stem is based on an equivalence in intended use, materials, design, operational principles, and relative indications and contraindications to Howmedica's HNR Hip System (K903563), DRG Femoral component (K936126), Premium Femoral Component (K936127), and Dow Corning Wright's McCutchen Hip System. Testing has demonstrated that the endurance load exceeds the Semlitsch minimum of 630 lbs.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...