K970143 · Clinicon Corp. · GEX · Oct 29, 1997 · General, Plastic Surgery

Device Facts

Record IDK970143
Device NameC4
ApplicantClinicon Corp.
Product CodeGEX · General, Plastic Surgery
Decision DateOct 29, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

For use in vaporization, incision and excision of soft tissue in dermatology, general surgery, neurosurgery, oral surgery, oto-rhino-laryngology, and podiatry.

Device Story

The Clinicon C⁴ is a DC-excited, sealed, liquid and convection-cooled CO₂ surgical laser. It generates an invisible infrared beam at 10.6 μm wavelength with an output power of 0-25 watts. The laser energy is delivered to the surgical site via an articulated arm and a surgical handpiece, which produces a collimated beam with a 200-micron spot size. The device is intended for use by clinicians in open surgery for soft tissue vaporization, incision, and excision. The handpieces, constructed from 304 surgical stainless steel, are available in various focal lengths and configurations, including options for use with colposcopes and microscopes via micromanipulators. The output allows the surgeon to perform precise tissue ablation or cutting, providing a therapeutic benefit through controlled thermal interaction with soft tissue.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Sealed, DC-excited CO₂ laser; 10.6 μm wavelength; 0-25W output power; articulated arm delivery; 304 surgical stainless steel handpieces; liquid and convection cooling; 200-micron spot size.

Indications for Use

Indicated for vaporization, incision, and excision of soft tissue in dermatology, general surgery, neurosurgery, oral surgery, oto-rhino-laryngology, and podiatry. Intended for open surgery procedures in non-contact mode.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0} K970143 CLINICON CORPORATION CARLSBAD CALIFORNIA OCT 29 1997 # 510(k) Summary Submitter: Clinicon Coporation Address: 2260 Rutherford Road Carlsbad, CA 92008 Phone number: 619-930-0010 Fax number: 619-930-0074 Contact person: Alan Bunting Date prepared: January 9, 1997 Trade name: C⁴ Common name: Laser Classification name: Class II Substantial equivalence claimed to: 1. Sharplan 20C - K963229 2. JMED CHRYS - K913606 3. Surgical Laser Specialties Ultra MD Laser - 510k Number Unknown # Description: The Clinicon C⁴ CO₂ laser is a sealed, liquid and convection cooled carbon dioxide surgical laser that is DC excited. The output of the laser is a concentrated beam of invisible infrared light at 10.6 micrometer wavelength. The laser has an output power range of 0-25 watts delivered to tissue. The laser system is fitted with a surgical hand piece as a standard accessory, producing a collimated beam with a spot size of 200 microns. The hand piece is manufactured from 304 surgical stainless steel. 000003 2260 Rutherford Road, Carlsbad, CA 92008 • Tel: 619-930-0010 • Fax: 619-930-0074 {1} 000004 # Intended use: For use in vaporization, incision and excision of soft tissue in dermatology, general surgery, neurosurgery, oral surgery, oto-rhino-laryngology, and podiatry. # Summary of technological characteristics: The Clinicon C⁴ CO₂ Surgical Laser energy is delivered to the surgical site via an articulated arm. The substantially equivalent devices are delivered either by an articulated arm or a hollow flexible waveguide. The Clinicon C⁴ CO₂ Surgical Laser's surgical beam wavelength is in the (invisible) infrared region of the spectrum, at 10.6 micrometers (microns). The substantial equivalent devices are 10.6 micron wavelength. # Handpieces HANDPIECE K 50R Surgical Handpiece straight. Focal length 2" HANDPIECE K 100R Surgical handpiece straight. Focal length 4" HANDPIECE K 100RO Surgical handpiece straight. Focal length 4" (OPEN CANNULA) HANDPIECE K 100A Surgical Handpiece Angled. Focal length 4" HANDPIECE K 100B Surgical Handpiece with backstop. Focal length 4" HANDPIECE K 125B Surgical handpiece with backstop. Focal length 5" # Manipulator MICROMANIPULATOR KM Ring adapters available for most popular colposcopes and microscopes. MICROMANIPULATOR KMC Ring adapters available for most popular colposcopes and microscopes (Central mirror) {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. Fritz A. Brauer President Clinicon Corporation 2260 Rutherford Road Carlsbad, California 92008 OCT 29 1997 Re: K970143 Trade Name: C⁴CO₂ Laser Regulatory Class: II Product Code: GEX Dated: August 13, 1997 Received: August 18, 1997 Dear Mr. Brauer: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3} Page 2 - Mr. Fritz A. Brauer This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, ![img-0.jpeg](img-0.jpeg) Enclosure {4} 510(k) Number (if known): K970143 Device Name: C⁴ CO₂ laser Aug 18 12 43 PM '97 ## Indications for Use: For use in vaporization, incision and excision of soft tissue in dermatology, general surgery, neurosurgery, oral surgery, oto-rhino-laryngology, and podiatry. The Clinicon C⁴ CO₂ laser is a sealed, liquid and convection cooled carbon dioxide surgical laser that is DC excited. The output of the laser is a concentrated beam of invisible infrared light at 10.6 micrometer wavelength. The laser has an output power range of 0-25 watts delivered to tissue. The laser system is fitted with a surgical hand piece as a standard accessory, producing a collimated beam with a spot size of 200 microns. The hand-pieces sold with this device will be of the same design and configuration but will have various tip sizes and focal lengths. These hand pieces are only intended for open surgery procedures in the non-contact mode. The hand piece is manufactured from 304 surgical stainless steel. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General Restorative Devices 510(k) Number K970143 Prescription Use ☑ OR Over-the-Counter Use ☐
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...