# THERMAPREP PLUS OVEN (K970134)

_Tulsa Dental Products, Ltd. · EKM · Mar 21, 1997 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K970134

## Device Facts

- **Applicant:** Tulsa Dental Products, Ltd.
- **Product Code:** [EKM](/productcode/EKM.md)
- **Decision Date:** Mar 21, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3850
- **Device Class:** Class 1
- **Review Panel:** Dental

## Indications for Use

The THERMAPREP® PLUS OVEN is specially developed for heating ThermaFil® endodontic obturators. Do not use the oven for any other purpose.

## Device Story

Heating unit for ThermaFil® endodontic obturators; utilizes microwave technology to achieve uniform, predictable heating; replaces radiant heat method of predicate; operated in clinical setting by dental professionals; reduces heating times for various obturator sizes; facilitates endodontic obturation procedures.

## Clinical Evidence

Bench testing only. Comparison of heating times for various sizes of ThermaFil® endodontic obturators between subject device and predicate (K910224) demonstrated reduced heating times for the subject device.

## Technological Characteristics

Heating unit for endodontic obturators; microwave-based heating principle; replaces radiant heat technology; standalone device.

## Regulatory Identification

Gutta percha is a device made from coagulated sap of certain tropical trees intended to fill the root canal of a tooth. The gutta percha is softened by heat and inserted into the root canal, where it hardens as it cools.

## Predicate Devices

- ThermaPrep® Oven ([K910224](/device/K910224.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

TULSA DENTAL PRODUCTS

Shaping The Future of Endodontics™

K970134

# 510(k) SUMMARY

Page 1 of 1

MAR 21 1997

ADDRESS: 5001 East 68th Street
Tulsa, OK 74136
(918) 493-6598
Telefax (918) 493-6599

CONTACT: Russell Vanderslice
DATE PREPARED: January 9, 1997

TRADE NAME: THERMAPREP® PLUS OVEN

CLASSIFICATION NAME: Accessory to Gutta Percha 21 CFR, 872.3850

PREDICATE DEVICE: ThermaPrep® Oven K910224

DEVICE DESCRIPTION: The THERMAPREP® PLUS OVEN is a heating unit for obturators. The system allows for individual heating of ThermaFil® gutta percha obturators. Microwave technology is used to produce uniform predictable heating results.

INTENDED USE: The THERMAPREP® PLUS OVEN is specially developed for heating ThermaFil® endodontic obturators. Do not use the oven for any other purpose.

TECHNOLOGICAL CHARACTERISTICS: The predicate device, ThermaPrep® Oven (K910224), heats ThermaFil® endodontic obturators by radiant heat. The new THERMAPREP® PLUS OVEN performs the same function by using microwave technology.

Various sizes of ThermaFil® endodontic obturators were heated using the THERMAPREP® PLUS OVEN and using the ThermaPrep® Oven (K910224). Results showed that heating times for all sizes of obturators were greatly reduced using the THERMAPREP® PLUS OVEN.

We believe that the performance data demonstrates the safety and effectiveness of the THERMAPREP® PLUS OVEN for its intended use.

000012

---

**Source:** [https://fda-staging.innolitics.com/device/K970134](https://fda-staging.innolitics.com/device/K970134)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).