← Product Code [CDR](/productcode/CDR) · K970093

# TYROSINE 50-HOUR REAGENT KIT (K970093)

_Astoria-Pacific, Inc. · CDR · Mar 19, 1998 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K970093

## Device Facts

- **Applicant:** Astoria-Pacific, Inc.
- **Product Code:** [CDR](/productcode/CDR.md)
- **Decision Date:** Mar 19, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1730
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry
- **Attributes:** Pediatric

## Indications for Use

This method is for the quantitative determination of the amino acid tyrosine in whole blood saturated filter paper disks using the API™ 300 SPOTCHECK® Analyzer or the RFA 300 System. Measurements obtained by this device are used in the diagnosis and treatment of diseases such as congenital tyrosinemia, a disease that can cause liver/kidney disorders.

## Device Story

Reagent kit for quantitative determination of tyrosine in whole blood saturated filter paper disks; used with API 300 SPOTCHECK Analyzer or RFA 300 System. Principle: fluorometric reaction of tyrosine with 1-nitroso-2-naphthol, catalyzed by nitrous acid at 80°C; excess reagent removed via sodium m-bisulfite reduction. Excitation 450 nm, emission 550 nm. Designed for mass screening in clinical laboratory settings; operated by laboratory technicians. Provides quantitative tyrosine levels to assist clinicians in diagnosing and managing congenital tyrosinemia. Benefits include standardized, pre-packaged reagents for efficient high-throughput screening (90 samples/hour).

## Clinical Evidence

Clinical testing performed on 1,096 whole blood saturated filter paper specimens. Study included 532 normal neonatal blood samples and 521 spiked samples. Results showed good correlation with expected values for spiked samples; all normal neonatal samples yielded tyrosine values within the normal range.

## Technological Characteristics

Reagent kit containing succinate buffer, saline diluent (with Triton X-100), 1-nitroso-2-naphthol, sodium nitrite, nitric acid, and sodium metabisulfite. Fluorometric sensing principle. Designed for use with automated analyzers (API 300 SPOTCHECK or RFA 300).

## Regulatory Identification

A free tyrosine test system is a device intended to measure free tyrosine (an amono acid) in serum and urine. Measurements obtained by this device are used in the diagnosis and treatment of diseases such as congenital tyrosinemia (a disease that can cause liver/kidney disorders) and as an adjunct to the measurement of phenylalanine in detecting congenital phenylketonuria (a disease that can cause brain damage).

## Predicate Devices

- Technicon Tyrosine test

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

Astoria-Pacific Inc.
Tyrosine 50-Hour Reagent Kit
510(k) K970093

MAR 19 1998

510(k) Summary

IDENTIFICATION

Submitter's Name: Raymond L. Pavitt, President Astoria-Pacific, Inc.

Address: 14600 SE 82nd Drive Clackamas, OR 97015-0830, USA

Telephone: 503-657-3010

FAX: 503-655-7367

Contact Person: Lester B. Garrison, Diagnostics Manager

Date Summary Prepared: February 3, 1998

NAME OF DEVICE

Proprietary Name: Tyrosine 50-Hour Reagent Kit

Common Name: Free Tyrosine Test System

Classification Name: 1-Nitroso-2-naphthol (Fluorometric), Free Tyrosine

LEGALLY MARKETED DEVICE EQUIVALENT TO PROPOSED DEVICE

The Tyrosine 50-Hour Reagent Kit is substantially equivalent to a legally marketed predicate device, Technicon Tyrosine test.

DESCRIPTION OF PROPOSED DEVICE

The proposed device, Tyrosine 50-Hour Reagent Kit, is a set of reagents to be used with the API™ 300 SPOTCHECK® Analyzer or the RFA 300 System for the quantitative determination of the amino acid tyrosine (Tyr) in whole blood saturated filter paper disks. The amount of Tyr is determined by measuring the fluorescent compound produced in the reaction of tyrosine with 1-nitroso-2-naphthol, catalyzed by nitrous acid at 80°C. The excitation wavelength of the product is 450 nm and emission is measured at 550 nm. The method is specific for para-substituted phenolic compounds. Excess 1-nitroso-2-naphthol is removed from the reaction mixture by reduction with sodium m-bisulfite.

The method is designed for mass screening, with enough reagents in each 50-Hour Reagent Kit for 1 week plus start-up (50 hours total) of run time. It is packaged to reduce storage space and to require a minimum of time to prepare. Each component is packaged with the correct weight to prepare the required volume of reagent. The standard is in a concentrated form, to permit easy dilution to prepare a standard curve.

{1}

Astoria-Pacific Inc.
Tyrosine 50-Hour Reagent Kit
510(k) K970093

The materials provided and their components are listed below.

|  Reagent | Active Component(s) | Concentration | Volume  |
| --- | --- | --- | --- |
|  Extraction Buffer | Succinate Buffer | 6.0 mM | 1000 mL  |
|  Saline Diluent, 0.9% | Sodium Chloride | 154. mM | 600 mL  |
|   |  Triton X-100 | 0.05% |   |
|  Complex Reagent | 1-Nitroso-2-Naphthol | 1.5 mM | 500 mL  |
|   |  DMSO | 3.5 M |   |
|  Sodium Nitrite, 2.5% | Sodium Nitrite | 0.36 M | 10 mL  |
|  Nitrous Acid | Nitric Acid, 10% | 50 mL | 50 mL  |
|   |  Sodium Nitrite, 2.5% | 0.25 mL —Added before use  |   |
|  Sodium Metabisulfite | Sodium Metabisulfite | 0.26 M | 500 mL  |
|  Tyrosine Stock Standard, 10 mg/dL | L-Tyrosine | 0.55 mM | 50 mL  |
|   |  Hydrochloric Acid | 6.0 mM |   |

## INTENDED USE

This method is for the quantitative determination of the amino acid tyrosine in whole blood saturated filter paper disks using the API™ 300 SPOTCHECK® Analyzer or the RFA 300 System. Measurements obtained by this device are used in the diagnosis and treatment of diseases such as congenital tyrosinemia, a disease that can cause liver/kidney disorders.

## SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF DEVICE COMPARED TO PREDICATE DEVICE

### Similarities with Predicate Device

The Astoria-Pacific SPOTCHECK® Tyrosine 50-Hour Reagent Kit is similar to the predicate device, Technicon Tyrosine test, having the same technological characteristics and intended use. Both are diagnostic reagent kits to be used for the quantitative, automated, fluorometric determination of the amino acid tyrosine in blood; both have the same chemical principles and reaction mechanism.

The reagents are identical, and there are no new reagents. The proposed device uses approximately the same temperatures, time and ratios of reagents as the predicate device.

### Differences from Predicate Device

While the method and reagents are identical for the proposed and predicate devices, there are minor variations in the surfactants. The Technicon device uses Brij-35, while the proposed device uses Triton X-100, as it is available in a purified grade. In addition, the sample for the proposed device is whole blood spotted on filter paper and extracted with a

{2}

Astoria-Pacific Inc.
Tyrosine 50-Hour Reagent Kit
510(k) K970093

buffer; the sample for the predicate device is not specified. The proposed device is designed for a sampling rate of 90 per hour vs. 50 per hour for the predicate device.

There are also slight differences in the temperature of the assay and the wavelengths of the excitation and emission filters. The temperature of the assay, which affects the level of fluorescence produced, is 80°C for the 50-Hour Reagent Kit and 90°C for the Technicon Tyrosine test; temperatures from 55°C to 90°C are cited in the literature for this reaction. The Proposed device designates 450 nm and 550 nm for the excitation and emission wavelengths, respectively; the predicate device uses 440 nm and 540 nm, respectively. Because there is a broad bandpass of about 40 nm for the filters, and the emitted fluorescent light is broad spectrum, this difference is not consequential.

Other differences include concentration range (0-200 mg/L for Astoria-Pacific vs. 0-2 mg/L for Technicon), detection limit and upper limit of the assay. Also, the Technicon Tyrosine test required the user to prepare all reagents; in the 50-Hour Reagent Kit, the reagents are either pre-made or pre-weighed.

The minor differences between the Astoria-Pacific SPOTCHECK® Tyrosine 50-Hour Reagent Kit and the Technicon Tyrosine test raise no questions of safety or efficacy.

## BRIEF DISCUSSION OF NONCLINICAL TESTS

Because there is such a strong similarity between the Astoria-Pacific SPOTCHECK® Tyrosine 50-Hour Reagent Kit and the Technicon Tyrosine test, no nonclinical testing has been performed to demonstrate substantial equivalence to the predicate device.

## BRIEF DISCUSSION OF CLINICAL TESTS

Clinical testing has been performed on whole blood saturated filter paper disks. These test include 1,096 specimens analyzed by the State of California's Genetic Disease Laboratory, of which 532 were normal neonatal blood samples and 521 were spiked samples. Results correlated well with the expected values for the spiked samples; all normal neonatal blood samples yielded tyrosine values within the normal range. Other testing is described in the product labeling.

## CONCLUSIONS DRAWN FROM THE NONCLINICAL AND CLINICAL TESTS

The clinical tests demonstrate that the proposed device, the Astoria-Pacific SPOTCHECK® Tyrosine 50-Hour Reagent Kit, is safe and performs as well as or better than the predicate device.

## OTHER INFORMATION

No other information has been requested by the FDA at this time.

End of 510(k) SUMMARY.

{3}

DEPARTMENT OF HEALTH &amp; HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850

MAR 19 1998

Raymond L. Pavitt
President
Astoria-Pacific, Inc.
14600 SE 82nd Drive
Clackamas, Oregon 97015-0830

Re: K970093
Tyrosine 50-Hour Reagent Kit
Regulatory Class: II
Product Code: CDR
Dated: February 4, 1998
Received: February 5, 1998

Dear Mr. Pavitt:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health

Enclosure

{5}

MAR-13-1998 12:19 API: ASTORIA-PACIFIC, INC +1 503 655 7367 P.06 Page 1 of 1

510(k) Number (if known): K970093
Device Name: Tyrosine 50-Hour Reagent Kit.
Indications For Use:

This method is for the quantitative determination of the amino acid tyrosine in whole blood saturated filter paper disks using the API™ 300 SPOTCHECK® Analyzer or the RFA 300 System. Measurements obtained by this device are used primarily to screen newborns for congenital tyrosinemia, a disease that can cause liver/kidney disorders.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ☑ OR Over-The-Counter Use ☐
(Per 21 CFR 801.109)
(Division Sign Off)
Division of (illegible)
510(k) Number K970093
(Optional Format 1-2-96)

MAR-13-1998 06:10 301 5945940 93% P.02

---

**Source:** [https://fda-staging.innolitics.com/device/K970093](https://fda-staging.innolitics.com/device/K970093)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
