ABBOTT ADVISOR ONE-STEP PREGNANCY TEST

K970082 · Abbott Laboratories · LCX · Jan 24, 1997 · Clinical Chemistry

Device Facts

Record IDK970082
Device NameABBOTT ADVISOR ONE-STEP PREGNANCY TEST
ApplicantAbbott Laboratories
Product CodeLCX · Clinical Chemistry
Decision DateJan 24, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Indications for Use

The ABBOTT Advisor One-Step Pregnancy Test is a self-performing immunoassay designed for the qualitative determination of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy.

Device Story

The device is a self-performing, lateral-flow immunoassay for qualitative hCG detection in urine. It consists of a molded plastic housing containing an internal test strip. The user applies a direct urine stream to the device; the sample migrates through an absorbent pad to a conjugate pad, mobilizing mouse monoclonal anti-alpha hCG antibody-coated colloidal particles. If hCG is present, it binds to these particles. The complex migrates to a result window containing immobilized polyclonal anti-beta hCG (patient bar) and anti-mouse/anti-beta hCG (control bar). A positive result is indicated by a plus (+) sign; a negative result by a minus (-) sign. A built-in 'Test Timer' changes color to pink/red after approximately 5 minutes to signal the assay is complete. The device is intended for home use by consumers to confirm pregnancy. The visual output allows the user to determine pregnancy status without additional equipment or professional interpretation.

Clinical Evidence

Bench testing confirmed 100% accuracy, sensitivity, and specificity compared to Hybritech ICON and FactPLUS Cup/Dropper tests using 50 positive and 50 negative clinical urine specimens. Consumer clinical study (n=149, females aged 18-45) demonstrated 98% accuracy in result interpretation, with 100% of subjects reporting instructions were easy to understand.

Technological Characteristics

Lateral-flow immunoassay; molded plastic housing; colloidal particle-based detection; monoclonal anti-alpha hCG antibody conjugate; polyclonal anti-beta hCG and anti-mouse antibody capture bars. Qualitative visual readout. No electronic components or external power source.

Indications for Use

Indicated for over-the-counter use by consumers to detect pregnancy as early as the first day of a missed period via direct urine stream sampling.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} ABBOTT Advisor® One-Step Pregnancy Test JAN 24 1997 K970082 # [510(k)] Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination ABBOTT Advisor® One-Step Pregnancy Test Abbott Laboratories 510(k) Notification Page 6 {1} ABBOTT Advisor® One-Step Pregnancy Test # [510(k)] Summary of Safety and Effectiveness Information Supporting a Substantially Equivalent Determination | Trade Name | ABBOTT Advisor One-Step Pregnancy Test | | --- | --- | | Common Name | Consumer Use Home Pregnancy Test | | Classification Name | Human Chorionic Gonadotropin (hCG) Test System | | Device Classification | Class II | | Predicate Device Name | FactPLUS® One Step Pregnancy Test (K962521) | The following information as presented in the 510(k) Notification for the ABBOTT Advisor One-Step Pregnancy Test constitutes data supporting a substantially equivalent determination. ## Intended Use The ABBOTT Advisor One-Step Pregnancy Test is a self-performing immunoassay designed for the qualitative determination of human chorionic gonadotropin (hCG) in urine for early detection of pregnancy. ## Indications The ABBOTT Advisor One-Step Pregnancy Test is an over-the-counter *in vitro* test which can be used by the consumer to detect pregnancy as early as the first day of her missed period, using a direct urine stream sampling method. Some positive results appear as soon as three minutes, however, all results are confirmed when the Test Timer turns pink/red. The Test Timer will turn pink/red approximately 5 minutes after the urine has been added to the test device. ## Device Description The ABBOTT Advisor One-Step Pregnancy Test is an elongated device composed of two pieces of molded plastic which contain the internal test strip. The three openings on the device are the Urine Well, the Result Window and the Test Timer. The Result Window and the Test Timer are protected with a clear seal to prevent potential contamination caused by splashing urine. Abbott Laboratories 510(k) Notification {2} ABBOTT Advisor® One-Step Pregnancy Test The ABBOTT self-performing immunoassay for hCG in urine is based on the principle of a two site immunometric assay. Urine is added to the Urine Well, where it migrates through the Absorbent Pad to the Conjugate Pad portion of the test strip. As the urine moves through the Conjugate Pad, it mobilizes the mouse monoclonal anti-alpha hCG antibody coated colloidal particles. Any hCG present in the urine sample combines with the anti-alpha hCG antibody on the colloidal particle. The urine, now containing an hCG-antibody-colloid complex and unbound colloid, continues to move up the test strip to the Result Window which contains immobilized polyclonal anti-βhCG on a vertical (Patient) bar and anti-mouse/anti-βhCG on a horizontal (Control) bar. The color generated is proportional to the concentration of hCG in the urine sample. Anti-hCG (mouse monoclonal) coated colloidal particles bind to the anti-mouse antibody immobilized on the Control Bar which indicates the assay is performing properly. A positive sign (+) indicates the presence of hCG. Low hCG concentration samples may display a "T" result, which is also interpreted as positive. The absence of detectable hCG will result in a negative sign (-). The assay is complete when the "Test Timer" indicator changes to pink or red, approximately three minutes after the start of the assay. ## Statement of Substantial Equivalence The ABBOTT Advisor One-Step Pregnancy Test is substantially equivalent to, and is the same device as the currently marketed FactPLUS® One Step Pregnancy Test (K962521) which is manufactured by Abbott Laboratories and distributed for over-the-counter sale by Direct Access Diagnostics, Bridgewater, New Jersey. The intended use, technology, methodology, test chemistry, test analyte, component materials, sensitivity, specificity, and accuracy are the same for both the Abbott and the Direct Access Diagnostics devices. The devices can be used the first day of a missed period or menses, and at any time of the day. The only differences between the Abbott and Direct Access Diagnostics devices are the name, the packaging and labeling format, and a modification to the labeled time to result requirement. Although some positive results may be seen as soon as 3 minutes from the start of the assay, customers are instructed to read the results when the Test Timer turns pink/red (approximately 5 minutes from beginning the test). A comparison of Similarities and Differences can be found in Table 1, page 9. Abbott Laboratories 510(k) Notification Page 8 {3} ABBOTT Advisor® One-Step Pregnancy Test # Table 1 ## Substantial Equivalence Similarities and Differences | COMPARATORS | ADVISOR ONE-STEP PREGNANCY TEST (SUBJECT) | FACT PLUS ONE STEP PREGNANCY TEST (PREDICATE) | | --- | --- | --- | | Visual Result Presentation Positive Signal Negative Signal | Plus (+) Sign · Minus (-) Sign | Plus (+) Sign Minus (-) Sign | | Urine Collection & Mode of Addition | Direct Urine Stream | Direct Urine Stream | | Analyte hCG | Same | Same | | Specimen Type Human Urine | Same | Same | | Intended Use Home Pregnancy Test | Same | Same | | Technology Qualitative - sandwich colloidal particle immunoassay | Same | Same | | Time to Assay Result | 5 minutes (some positives seen in 3 minutes) | 3 minutes (some positives seen in 1 minute) | | Components Single unit test device (all reagents contained on a test strip within) | Same | Same | | End of Assay Color Change | Yes - Test Timer Shows adequate sample Wait to read results | Yes * Do not wait for signal before reading results | | Strip Chemistry Conjugate - Monoclonal Anti-αhCG Antibody | Same | Same | | Patient Bar Polyclonal Anti-βhCG Antibody | Same | Same | | Control Bar Polyclonal Anti-βhCG Antibody and Polyclonal Anti-Mouse | Same | Same | | Consumer Use Accuracy | 98% | 98% | | Laboratory Use Accuracy | 100% | 100% | * Do not need to wait for signal before reading results Abbott Laboratories 510(k) Notification Page 9 {4} ABBOTT Advisor® One-Step Pregnancy Test ## Performance Data Accuracy testing was performed by Advanced Care Products using hCG-positive and hCG-negative female clinical urine specimens across three different lots of the Abbott device, which is currently marketed under the brand name of FactPLUS® One Step Pregnancy Test. The device correctly identified 50/50 hCG-positive and 50/50 hCG-negative specimens. Tests were read immediately after the Test Timer turned pink/red, approximately 5 minutes after the start of the test. The hCG-positive samples were confirmed to be positive with the Hybritech ICON pregnancy test. The hCG-negative specimens were confirmed to be negative with the FactPLUS® Cup and Dropper Pregnancy Test. At three minutes agreement was 100% with the ICON and FactPLUS devices. ## Consumer Testing A clinical study was conducted to verify that consumers could obtain accurate test results when using the Abbott manufactured device according to the provided instructions. The consumer population consisted of 149 female volunteers recruited from a central New Jersey location. The volunteers were between the ages of 18 to 45, from various income, educational and employment backgrounds. Correct results were obtained by one hundred forty six (146) of one hundred forty nine (149) subjects for an overall accuracy of 98%. Subjects reported little difficulty with understanding the test instructions. Ninety-two percent reported that the instructions were "very easy to understand" and all subjects (100%) indicated that the instructions were "somewhat" or "very easy to understand." ## Conclusion The ABBOTT Advisor One-Step Pregnancy Test is substantially equivalent, and is in fact the same device as the FactPLUS® One Step Pregnancy Test, distributed by Direct Access Diagnostics. Laboratory testing indicates accuracy of 100%, sensitivity and specificity of 100% each. Results of consumer clinical testing demonstrated that the Abbott device can be performed and correctly interpreted by consumers with an overall accuracy rate of 98%. Prepared and Submitted January 8, 1997 by: Abbott Laboratories Karen L. Gates ADD Regulatory Affairs Abbott Park, IL. 60064 Abbott Laboratories 510(k) Notification
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...