MONARCH IOL DELIVERY SYSTEM

K970033 · Alcon Laboratories · KYB · Apr 23, 1997 · Ophthalmic

Device Facts

Record IDK970033
Device NameMONARCH IOL DELIVERY SYSTEM
ApplicantAlcon Laboratories
Product CodeKYB · Ophthalmic
Decision DateApr 23, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4300
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The intended use of this device is to fold and deliver Alcon ACRYSOF® intraocular lenses into the eye for replacement of the human crystalline lens.

Device Story

System consists of reusable titanium handpiece and sterile, single-use cartridge; used for folding and delivering ACRYSOF® intraocular lenses into capsular bag during cataract surgery. Surgeon loads lens into cartridge; handpiece plunger, advanced via screw mechanism, expresses lens through cartridge into eye. Controlled delivery ensures lens unfolds completely without damaging lens geometry, haptic angle, power, or cosmetic attributes. Benefits include reliable, smooth lens placement during lens replacement procedures.

Clinical Evidence

Bench testing only. Performance data demonstrated successful delivery of ACRYSOF® IOLs without affecting lens length, haptic angle, power, resolution, or cosmetic attributes. Lenses unfolded completely within 30 seconds in a controlled manner.

Technological Characteristics

Materials: Reusable titanium handpiece, sterile single-use cartridge. Mechanism: Screw-driven plunger for controlled lens expression. Form factor: Handheld surgical instrument. Sterilization: Autoclavable handpiece, sterile cartridge.

Indications for Use

Indicated for patients undergoing cataract surgery requiring replacement of the human crystalline lens with an Alcon ACRYSOF® intraocular lens.

Regulatory Classification

Identification

An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.

Predicate Devices

Submission Summary (Full Text)

{0} K970033 Alcon LABORATORIES APR 23 1997 ALCON LABORATORIES, INC. 6201 SOUTH FREEWAY FORT WORTH, TEXAS 76134-2099 (817) 293-0450 January 2, 1997 510(K) SUMMARY Submitted by: Martin A. Kaufman Manager, Regulatory Affairs, Surgical Devices Alcon Laboratories, Inc. 6201 South Freeway Fort Worth, TX 76132 (817) 551-8388 (Phone) (817) 551-4630 (Fax) Trade Name: MONARCH™ IOL Delivery System Common Name: IOL Delivery System Classification Name: Intraocular lens Guide, 21 CFR 886.4300 76 {1} K970633 ## 1. Predicate Device The legally marketed device(s) to which we are claiming equivalence to are: a. K903574, ChiroFlex Lens Microsert™, (Chiron Ophthalmics Inc.) b. K913626, Softrans™ Injector, (STAAR Surgical Company) ## 2. Device Description The MONARCH™ IOL Delivery System consists of two parts: an autoclavable, reusable titanium handpiece and a sterile, single-use cartridge. It is a device used for folding and delivering ACRYSOF® intraocular lenses into the eye for replacement of the human crystalline lens. The system provides a controlled means to reliably place the ACRYSOF® intraocular lens into the capsular bag. ## 3. Intended Use of the Device The intended use of this device is to fold and deliver Alcon ACRYSOF® intraocular lenses into the eye for replacement of the human crystalline lens. ## 4. Summary of the Technological Characteristics of the Device The MONARCH™ IOL Delivery System utilizes a sterile, single use cartridge and a reusable handpiece to deliver ACRYSOF® lenses. The cartridge is designed for easy loading and reliable folding of ACRYSOF® lenses. The handpiece accepts the cartridge and delivers the lens by using a plunger to express the lens. The plunger head is contoured to provide a good contact to the lens as well as an adequate clearance for the trailing haptic. The plunger is advanced by a screw mechanism to ensure a smooth and well controlled lens delivery. ## 5. Summary of the Performance Data The performance test demonstrated that the MONARCH™ IOL Delivery System can be used to deliver ACRYSOF® IOLs without affecting the overall length, haptic angle, power, resolution, or cosmetic attributes of the lenses. The test lenses were delivered smoothly and unfolded completely within 30 seconds in a controlled fashion. ## 6. Conclusions The MONARCH™ IOL Delivery System is easy to use, properly delivers the Alcon ACRYSOF® intraocular lens, and is substantially equivalent to devices such as the Softrans™ Injector (Model ET) or the ChiroFlex Lens Microsert™ system. 77
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...