K970012 · Medical Data Electronics · DSI · Jun 26, 1997 · Cardiovascular
Device Facts
Record ID
K970012
Device Name
ESCORT LINK CENTRAL STATION MONITOR
Applicant
Medical Data Electronics
Product Code
DSI · Cardiovascular
Decision Date
Jun 26, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1025
Device Class
Class 2
Indications for Use
The ESCORT-LINK™ Central Station Monitor is intended to be used to provide, using a wireless LAN for communication, centralized surveillance and documentation of patient vital sign data and arrhythmia monitoring and alarms for a variable number of ESCORT II bedside monitors and a variable number of UHF telemetry transmitters in the hospital environment. It is intended for use by healthcare practitioners trained in the use of the equipment only.
Device Story
Central station monitor for hospital use; aggregates vital sign and waveform data from ESCORT II bedside monitors and UHF telemetry transmitters. Inputs: patient vital signs/waveforms via proprietary Spread Spectrum LAN or UHF telemetry. Processing: centralized display, storage (up to 24 hours), and alarm management. Output: real-time monitoring, arrhythmia detection, and data retrieval via display or printer. Operated by trained healthcare practitioners. Enhances clinical decision-making through centralized surveillance and configurable cardiac abnormality detection; enables remote monitoring of up to 16 patients.
Clinical Evidence
Bench testing only. Performance testing conducted to verify alarm detection, arrhythmia reporting, and accuracy of vital sign/waveform data. No clinical trials or human subject data presented.
Technological Characteristics
Pentium PC-based system; 16MB RAM; 2GB HDD; MS Windows OS. Connectivity: proprietary Spread Spectrum LAN (902-928 MHz) and UHF telemetry (450-470 MHz). Supports up to 16 patients. Arrhythmia algorithm provided by Zymed Inc. Output via VGA display, thermal array recorder, or LaserJet printer.
Indications for Use
Indicated for centralized surveillance, documentation, and arrhythmia monitoring of patients in a hospital environment. For use by trained healthcare practitioners only.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
ESCORT LINK Central Station Monitor Model 3200
Submission Summary (Full Text)
{0}
K970012
Page 1
JUN 26 1997
# ESCORT LINK Central Station Model 20500
1. Submitter:
Medical Data Electronics
12720 Wentworth Street
Arleta, California 91331
Telephone: 818-768-6411
Telefax: 818-768-4197
Contact: David M. Trueblood
Regulatory Affairs Manager
2. Date of Preparation: December 31, 1996
3. Device Name:
Trade Name: ESCORT LINK Central Station Monitor Model 20500
Common Name: Central Station Monitor
Classification Name: Monitor, Electrocardiograph Detector and Alarm, Arrhythmia
4. Substantial Equivalence:
The ESCORT LINK Central Station Monitor Model 20500 is substantially equivalent to the ESCORT LINK Central Station Monitor Model 3200.
5. Description of the Device:
The ESCORT LINK Central Station Monitor Model 20500 is a Central Station monitor comprised of a standard VGA Display, a standard Personal Computer Base, and an auxiliary base used to mount the Network communications hardware (i.e. Spread Spectrum transponders and UHF telemetry receivers).
The ESCORT LINK Central Station Monitor Model 20500 can provide the centralized display, storage, and recording (or printing) of patient vital sign and waveform data that is being monitored at ESCORT II 100, 300, or 400 bedside monitors or UHF Telemetry Receivers.
Medical Data Electronics, Inc.
12720 Wentworth St. ▲ Arleta, CA 91331-4329 ▲ (818) 768-6411 ▲ Fax (818) 768-7602
{1}
Page 2
Data accumulated at ESCORT II bedsides is sent via a proprietary Spread Spectrum Local Area Network to the Central Station for display and storage. Data accumulated from any of Medical Data Electronics' analog or digital telemetry transmitters is sent directly to the Central Station on standard UHF telemetry frequencies. The Central Station oversees all communications activity, allowing each system component to pass information without interruption to patient monitoring.
The Central Station can provide alarm detecting and reporting for all vital sign parameters that are available to the Central Station. This alarm response is in addition to alarms that are available at the ESCORT 100, 300, or 400 bedsides. Also, arrhythmia monitoring is available for up to 16 patients at the Central Station to provide the configurable ability to detect and report certain cardiac abnormalities.
The Central Station can provide storage of patient data for up to 24 hours. This includes waveform and vital sign information for all available waveforms and parameter values. The waveform and vital sign data that has been stored in a patient file at the Central Station can be retrieved and viewed on the display or printed out on a Thermal Array recorder or a LaserJet® Printer.
## 6. Intended Use of the Device:
The ESCORT LINK Central Station Monitor Model 20500 is intended to be used to provide centralized surveillance and documentation of patient vital sign data for a variable number of ESCORT Series 100, 300, and 400 bedside monitors and a variable number of UHF telemetry transmitters in the hospital environment.
## 7. Summary of the Technological Characteristics of the New Device Compared to the Predicate Device:
The MDE ESCORT LINK Central Station Monitor Model 20500 is a new Central Station monitor, however, its intended use and method of operation are substantially equivalent to the MDE ESCORT LINK Central Station Monitor Model 3200.
The primary differences between the two devices are summarized in the following table:
Medical Data Electronics, Inc.
12720 Wentworth St. ▲ Arleta, CA 91331-4329 ▲ (818) 768-6411 ▲ Fax (818) 768-7602
{2}
Page 3
| SPECIFICATION | MODEL 20500 | MODEL 3200 |
| --- | --- | --- |
| PC Specifications | Pentium PC, 16MB RAM, 2GB HDD | 386 PC, 8MB RAM, 200MB HDD |
| Operating System | MS Windows | MS DOS |
| Configuration | External to the PC base | Internal to the PC base |
| Patient Monitoring Capacity | Up to 16 | Up to 32 |
| Viewable Waveforms per Patient | Up to 8 per patient | 1 per patient |
| Viewable Vital Sign Values per Patient | Up to 20 per patient | Up to 6 per patient |
| Telemetry Communication | 902 - 928 Mhz Spread Spectrum communication between the bedside and the Central Station; 450 - 470 Mhz UHF communication between telemetry receivers and the Central Station | 450 - 470 Mhz UHF communication between the bedside and the Central Station |
| Alarm Handling | Alarms are interactively enabled and disabled from either the bedside or the Central Station | Alarms are enabled and disabled from the bedside only |
| Full Disclosure | Up to 3 waveforms per patient can be reviewed at a time. | 1 Waveform per patient can be reviewed at a time. |
| Arrhythmia Algorithm | Provided by Zymed Inc. | Provided by PCI Inc. |
## 8. Device Testing:
The ESCORT LINK Central Station Monitor Model 20500 is designed to functional standards developed by independent and regulatory agencies.
The criteria for these standards are identified in the following FDA documents:
- Reviewer Guidance for Computer Controlled Medical Devices Undergoing 510(k) Review; Office of Device Evaluation, August 1991.
- Reviewer Guidance for Premarket Notification Submissions, Anesthesiology and Respiratory Devices Branch; Division of Cardiovascular, Respiratory and Neurological Devices; November 1993.
Medical Data Electronics, Inc.
12720 Wentworth St. ▲ Arleta, CA 91331-4329 ▲ (818) 768-6411 ▲ Fax (818) 768-7602
{3}
Page 4
Tests demonstrating consideration of and mitigation of hazards identified to have potentially arisen as the result of the modifications in the new device were developed. Conformance to product development procedures and plans is assured by application of system tests, reviews, and product verification and validation studies.
9. Test Conclusions:
The MDE ESCORT LINK Central Station Monitor Model 20500 is shown by performance testing, stressing the areas of alarm detecting and reporting, arrhythmia detecting and reporting, and accuracy of patient vital sign and waveform data, to be a safe, effective Central Station Monitor. The MDE ESCORT LINK Central Station Monitor Model 20500 is substantially equivalent to the ESCORT LINK Central Station Monitor Model 3200.
Medical Data Electronics, Inc.
12720 Wentworth St. ▲ Arteta, CA 91331-4329 ▲ (818) 768-6411 ▲ Fax (818) 768-7602
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN 26 1997
Mr. David M. Trueblood
Medical Data Electronics, Inc.
12720 Wentworth Street
Arleta, California 91331-4329
Re: K970012
ESCORT Link Central Station Monitor Model 20500
Regulatory Class: III (three)
Product Code: 74 DSI
Dated: April 11, 1997
Received: April 14, 1997
Dear Mr. Trueblood:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}
Page 2 - Mr. David M. Trueblood
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{6}
Page 1 of 1
510(k) Number (if known): K970012 1106
Device Name: ESCORT-LINK® CENTRAL STATION MONITOR
FDA/CORH/CDE/CHC
Indications for Use:
Model 20500
The ESCORT-LINK™ Central Station Monitor is intended to be used to provide, using a wireless LAN for communication, centralized surveillance and documentation of patient vital sign data and arrhythmia monitoring and alarms for a variable number of ESCORT II bedside monitors and a variable number of UHF telemetry transmitters in the hospital environment. It is intended for use by healthcare practitioners trained in the use of the equipment only.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
E. Oan, M.D.
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number ________________
Prescription Use ☑ (Per 2.1 CFR 801.109)
OR
Over-the-Counter Use _____
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.