# CALTAG CAL-LYSE (K965232)

_Caltag Laboratories, Inc. · GKZ · May 14, 1997 · Hematology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K965232

## Device Facts

- **Applicant:** Caltag Laboratories, Inc.
- **Product Code:** [GKZ](/productcode/GKZ.md)
- **Decision Date:** May 14, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5220
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

Caltag Cal-Lyse is a lysis solution to enable the lysis of erythrocytes in samples of anticoagulated human peripheral blood. Cal-Lyse lysing solution is intended as an aid in the enumeration of leukocytes that have been stained with Caltag monoclonal antibodies for analysis by flow cytometric methods.

## Device Story

Caltag Cal-Lyse is a chemical lysing solution used in flow cytometric procedures. It is added to anticoagulated human peripheral blood samples previously incubated with fluorochrome-conjugated monoclonal antibodies. The solution lyses red blood cells and fixes leukocytes simultaneously, eliminating red cell debris and unbound antibodies. The resulting sample is analyzed via flow cytometry. The device is used in clinical laboratory settings by trained technicians or laboratory personnel. By removing red blood cells, the solution enables clear identification and enumeration of leukocyte subsets (e.g., B lymphocytes) stained with monoclonal antibodies, facilitating clinical immunophenotyping and diagnostic assessment of blood cell populations.

## Clinical Evidence

Clinical performance evaluated using 175 samples (155 normal, 20 abnormal) across three independent laboratories. Comparison of Caltag CD19 antibodies (R-PE, TRI-COLOR) against Coulter predicates (RD1, FITC) showed high correlation (r² range 0.963–0.977; slopes 0.92–0.94). Leukocyte recovery and red blood cell lysis efficiency were validated in 5 normal donors, showing mean 90.8% leukocyte recovery and 91.8% red cell lysis. Specificity data across diverse ethnic groups confirmed minimal non-specific binding.

## Technological Characteristics

Chemical lysing solution containing paraformaldehyde as a fixative. Designed for use with fluorochrome-conjugated monoclonal antibodies (R-PE, TRI-COLOR) in whole blood flow cytometry. Operates via chemical lysis of erythrocytes and simultaneous fixation of leukocytes. No electronic or software components; standalone chemical reagent.

## Regulatory Identification

An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”

## Predicate Devices

- Coulter CD19 RD1 monoclonal antibody
- Coulter CD19 FITC monoclonal antibody

## Reference Devices

- Ortho-mune Lysing Reagent

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

CALTAG LABORATORIES
K965232 MAY 14 1997

510(K) SUMMARY
SUMMARY OF SAFETY AND EFFECTIVENESS DATA

Caltag Cal-Lyse™ Lysing Solution For Use With Caltag Monoclonal Antibodies
In Flow Cytometric Procedures

NAME AND LOCATION OF MANUFACTURER:

Caltag Laboratories, Inc.
1849 Old Bayshore Highway
Suite 200
Burlingame, CA 94010
(800) 874-4007

NAME OF CONTACT PERSON:

Robert C. Johnson
Executive Vice President
Caltag Laboratories, Inc.

DATE OF PREPARATION OF SUMMARY:

December 24, 1996

Caltag Laboratories Inc. • 1849 Bayshore Blvd., Suite #200 • Burlingame, CA 94010
Telephone (415) 652-0468 • (800) 874-4007 • FAX (415) 652-9030

{1}

-2-

## TRADE NAME OF THE DEVICE:

Caltag Cal-Lyse Lysing Solution For Use With Caltag Monoclonal Antibodies In Flow Cytometric Procedures

## COMMON NAME:

Caltag Cal-Lyse Lysing Solution

## CLASSIFICATION NAME:

Automated Differential Cell Coulter (21 CFR 864.5220)

## LEGALLY MARKETED DEVICE (PREDICATE DEVICE) TO WHICH THE MANUFACTURER IS CLAIMING SUBSTANTIAL EQUIVALENCE:

The Caltag CD19 R-PE monoclonal antibody is substantially equivalent to the Coulter CD19 RD1 monoclonal antibody when red blood are lysed with the Cal-Lyse lysing solution in flow cytometric procedures.

The Caltag CD19 TRI-COLOR monoclonal antibody is substantially equivalent to the Coulter CD19 FITC monoclonal antibody when red blood are lysed with the Cal-Lyse lysing solution in flow cytometric procedures.

The Caltag CD19 R-PE monoclonal antibody is substantially equivalent to the Coulter CD19 FITC monoclonal antibody when red blood are lysed with the Cal-Lyse lysing solution in flow cytometric procedures.

The Caltag CD19 TRI-COLOR monoclonal antibody is substantially equivalent to the Coulter CD19 RD1 monoclonal antibody when red blood are lysed with the Cal-Lyse lysing solution in flow cytometric procedures.

The Caltag CD19 R-PE monoclonal antibody is substantially equivalent to the Caltag CD19 TRI-COLOR monoclonal antibody when red blood are lysed with the Cal-Lyse lysing solution in flow cytometric procedures.

## DESCRIPTION OF THE DEVICE:

Caltag monoclonal antibodies bind to the surfaces of viable blood cells that express the corresponding antigens. To identify cells bearing these antigenic determinants, peripheral blood samples are incubated with fluorochrome-conjugated monoclonal antibodies. Cells are subsequently washed to remove unbound antibody. Prior to the removal of unbound antibody, Cal-Lyse lysing solution is added to lyse red blood cells. Cells may subsequently be washed, resulting in the elimination of red cell debris as well as unbound antibody.

{2}

-3-

Cal-Lyse lysing solution contains paraformaldehyde as fixative, and no additional fixation is required. Antibody-stained and fixed leukocytes are subsequently analyzed by flow cytometric methods.

INTENDED USE OF THE DEVICE:

Caltag Cal-Lyse is a lysis solution to enable the lysis of erythrocytes in samples of anticoagulated human peripheral blood. Cal-Lyse lysing solution is intended as an aid in the enumeration of leukocytes that have been stained with Caltag monoclonal antibodies for analysis by flow cytometric methods.

SUMMARY OF THE TECHNICAL CHARACTERISTICS OF THE MANUFACTURER'S DEVICE COMPARED TO THE PREDICATE DEVICE:

Comparisons of Caltag CD19 and Coulter CD19 Monoclonal Antibodies In Samples In Which Red Blood Cells Are Lysed With Caltag Cal-Lyse Lysing Solution

|  No. | Item | Caltag Antibodies | Coulter Antibodies | Comparison  |
| --- | --- | --- | --- | --- |
|  1. | Intended Use | Flow Cytometry | Flow Cytometry Immunofluorescence | Substantially equivalent  |
|  2. | Specificity | CD19 | CD19 | Substantially equivalent  |
|  3. | Target cell | B lymphocyte | B lymphocyte | Substantially equivalent  |
|  4. | Chemical form | Monoclonal antibody | Monoclonal antibody | Substantially equivalent  |
|  5. | Fluorochromes | R-PE, TRI-COLOR | FITC, RD1 | Substantially equivalent  |
|  6. | Available forms |  |  |   |
|   | FITC | liquid, PBS | lyophilized | Substantially  |
|   | PE | liquid, PBS | liquid, PBS | equivalent  |
|   | TRI-COLOR | liquid, PBS | not available |   |
|  7. | Sample prep. methods | whole blood | whole blood | Substantially equivalent  |
|  8. | Expected values from this study (n=155) |  |  |   |
|   | R-PE | 5-21% | 4-21% (RD1) | Substantially  |
|   | TRI-COLOR | 4-24% | 3-23% (FITC) | equivalent  |

{3}

-4-

# NON CLINICAL TESTS SUPPORTING A DETERMINATION OF SUBSTANTIAL EQUIVALENCE:

## EXPECTED VALUE DATA

Blood samples were collected from a total of 155 apparently healthy adult normal donors in an age range of 16 to 72, and a mean age of 41.

Samples were stained with Caltag monoclonal antibodies and red blood cells were lysed with Caltag Cal-Lyse lysing solution. Samples were collected and analyzed in each of three independent laboratories. An approximately equal number of males and females were collected and analyzed in each laboratory.

The normal donor population included members of differing ethnic origins, including adult Caucasian, Black, Oriental and Hispanic. Donors in geographically diverse areas of the United States, including the Western, Eastern and South Central regions, participated in this study.

Summary of expected values for the CALTAG T and B cell monoclonal antibodies, CD3 FITC and CD19 R-PE, for all normal donors:

|  procedure | mean % positive | S.D. | Range ±2 S.D. | n  |
| --- | --- | --- | --- | --- |
|  CD3 FITC | 71.0 | 7.4 | 58-86 | 155  |
|  CD19 R-PE | 13.0 | 4.2 | 5-21 | 155  |

Expected values for pediatrics and adolescents have not been established.

The values obtained from normal individuals may vary from laboratory to laboratory; therefore, it is recommended that each laboratory establish its own normal range.

## SPECIFICITY DATA

Blood samples were obtained from healthy normal donors of Caucasian, Black, Hispanic and Oriental ethnic origins. Samples of each donor were stained with Caltag monoclonal antibodies and red blood cells were lysed with Caltag Cal-Lyse lysing solution. Cells contained in the lymphocyte, monocyte and granulocyte regions were selected for analysis. Separate samples from the same donors were prepared for analysis of red blood cells and platelets and stained with each of the Caltag monoclonal antibodies.

The following specificity data were obtained with the Caltag T and B cell monoclonal antibodies, CD3 FITC and CD19 R-PE, following the lysis of red blood cells with Caltag Cal-Lyse lysing solution:

|  CD3 FITC
Ethnic
Origin | Percent of Stained Cells  |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|   |  Lymph. | Mono. | Gran. | Plt. | RBC  |
|  Caucasian | 65.2 | 1.7 | 1.5 | 0.3 | 0.6  |
|  Caucasian | 81.4 | 1.4 | 0.5 | 0.4 | 0.3  |
|  Hispanic | 79.2 | 1.9 | 0.6 | 0.3 | 0.4  |
|  Oriental | 81.2 | 1.3 | 0.9 | 0.2 | 0.4  |
|  Black | 84.9 | 0.9 | 0.6 | 0.4 | 0.4  |
|  Mean | 78.4 | 1.4 | 0.8 | 0.3 | 0.4  |
|  ±1 S.D. | 7.6 | 0.4 | 0.4 | 0.1 | 0.1  |

{4}

-5-

|  CD19 R-PE  |   |   |   |   |   |
| --- | --- | --- | --- | --- | --- |
|  Ethnic Origin | Percent of Stained Cells  |   |   |   |   |
|   |  Lymph. | Mono. | Gran. | Plt. | RBC  |
|  Caucasian | 18.0 | 0.6 | 0.9 | 0.5 | 0.5  |
|  Caucasian | 13.3 | 1.1 | 0.8 | 0.3 | 0.7  |
|  Hispanic | 12.2 | 0.7 | 0.8 | 0.4 | 1.0  |
|  Oriental | 11.2 | 1.6 | 1.3 | 0.4 | 0.5  |
|  Black | 14.6 | 0.0 | 0.5 | 0.6 | 0.9  |
|  Mean | 13.9 | 0.8 | 0.9 | 0.4 | 0.7  |
|  ±1 S.D. | 2.6 | 0.6 | 0.3 | 0.1 | 0.2  |

Specific and/or nonspecific antibody Fc binding to monocytes in a patient sample can be excluded by proper gating on lymphocytes on the flow cytometer.

## LEUKOCYTE RECOVERY DATA

This study was conducted on blood samples obtained from 5 normal donors consisting of representatives from Caucasian, Black, Hispanic and Oriental ethnic origins. An appropriate hematology analyzer was used to enumerate leukocytes prior to, and immediately following the lysis of red blood cells with Caltag Cal-Lyse lysing solution and with the Ortho-mune™ Lysing Reagent. Leukocyte recovery from members of differing ethnic origins was comparable. Leukocyte recovery following lysis with Cal-Lyse and Ortho-mune lysing reagents was comparable.

It should be noted that washing of cells alone, in the absence of a lysis procedure may result in a modest loss of cells.

In the following table describing leukocyte recovery, leukocyte counts are expressed as cells per cu. mm.

|  Donor No. | Race | Leukocyte Count  |   |   |
| --- | --- | --- | --- | --- |
|   |   |  Prior to Lysis | Following Lysis | Percent Recovered  |
|  1 | Caucasian | 8300 | 7200 | 86.7  |
|  2 | Caucasian | 8100 | 7200 | 88.9  |
|  3 | Black | 5700 | 4800 | 84.2  |
|  4 | Hispanic | 6200 | 6000 | 96.8  |
|  5 | Oriental | 7400 | 7200 | 97.3  |
|  Mean |  | 7140 | 6480 | 90.8  |
|  ± 1 SD |  | 1150 | 1073 | 6.0  |

## RED BLOOD CELL LYSIS DATA

This study was conducted on blood samples obtained from 5 normal donors, to determine whether essentially all red cells were lysed by the lysing solution. Red cells were lysed in a sample of blood from each donor with Caltag Cal-Lyse lysis solution. An appropriate hematology analyzer was used to enumerate red blood cells prior to and immediately following lysis. The differing orders of magnitude of red cells counted prior to and following lysis should be noted in the following table.

{5}

-6-

# Red Blood Cell Count

|  Donor No. | Prior to Lysis cells/cu.mm. x 10^{6} | Following Lysis cells/cu.mm x 10^{5} | Percent Lysed  |
| --- | --- | --- | --- |
|  1 | 5.1 | 3.0 | 94.1  |
|  2 | 5.1 | 5.0 | 90.2  |
|  3 | 4.3 | 3.0 | 93.0  |
|  4 | 3.7 | 3.0 | 91.8  |
|  5 | 4.9 | 5.0 | 89.7  |
|  Mean | 4.6 | 3.8 | 91.8  |
|  ±1 S.D. | 0.6 | 1.1 | 1.9  |

## CLINICAL TESTS SUPPORTING A DETERMINATION OF SUBSTANTIAL EQUIVALENCE:

### CORRELATION DATA

The correlation for the Caltag Cal-Lyse lysing solution was based on the performance of the Caltag B cell monoclonal antibodies CD19 R-PE and CD19 TRI-COLOR. The percent, as well as the mean fluorescence, of B lymphocytes is substantially lower than is observed for T lymphocytes in normal peripheral blood. The resulting analysis of B lymphocytes by flow cytometric methods may be more susceptible to uncontrolled variations in the staining and lysis methods employed.

Samples obtained from normal and abnormal donors were stained with Caltag and comparable Coulter B cell monoclonal antibodies. Red blood cells were lysed with Caltag Cal-Lyse lysing solution in samples that had been stained with both Caltag and Coulter monoclonal antibodies.

A total of 155 normal donor samples were collected and analyzed in each of three independent laboratories, and analyzed on either the FACscan or Profile flow cytometers.

The normal donor population included members of differing ethnic origins, including adult Caucasian, Black, Oriental and Hispanic. Donors in geographically diverse areas of the United States, including the Western, Eastern and South Central regions, participated in this study. Males and females were represented in approximately equal numbers.

A total of 20 abnormal donors were analyzed on both the FACscan and Profile flow cytometers at a single site. Correlations were obtained for the Caltag and Coulter B cell monoclonal antibodies for all normal and abnormal donors (n=175). Red blood cells from all samples were lysed with Cal-Lyse lysing solution.

Comparison of the Caltag CD19 R-PE conjugated monoclonal antibody with the Coulter CD19 RD1 conjugated monoclonal antibody:

|  procedure | mean % positive | r^{2} value | slope | Y intercept | n  |
| --- | --- | --- | --- | --- | --- |
|  CD19 R-PE | 16.4 | 97.7 | 0.92 | 1.30 | 175  |
|  CD19 RD1 | 16.2 |  |  |  |   |

{6}

-7-

# CD19 R-PE

Linear regression  $y = 1.30 + 0.92x$

Comparison of the CALTAG CD19 R-PE conjugated monoclonal antibody with the Coulter CD19 FITC conjugated monoclonal antibody:

|  procedure | mean % positive | r² value | slope | Y intercept | n  |
| --- | --- | --- | --- | --- | --- |
|  CD19 R-PE | 16.4 | 96.3 | 0.93 | 0.69 | 175  |
|  CD19 FITC | 16.7 |  |  |  |   |

# CD19 R-PE

Linear regression  $y = 0.69 + 0.93x$

Comparison of the CALTAG CD19 TRI-COLOR conjugated monoclonal antibody with the Coulter CD19 RD1 conjugated monoclonal antibody:

|  procedure | mean % positive | r² value | slope | Y intercept | n  |
| --- | --- | --- | --- | --- | --- |
|  CD19 TRI-COLOR | 17.1 | 96.7 | 0.92 | 2.14 | 175  |
|  CD19 RD1 | 16.2 |  |  |  |   |

# CD19 TRI-COLOR

Linear regression  $y = 2.14 + 0.92x$

Comparison of the CALTAG CD19 TRI-COLOR conjugated monoclonal antibody with the Coulter CD19 FITC conjugated monoclonal antibody:

|  procedure | mean % positive | r² value | slope | Y intercept | n  |
| --- | --- | --- | --- | --- | --- |
|  CD19 TRI-COLOR | 17.1 | 97.4 | 0.94 | 1.36 | 175  |
|  CD19 FITC | 16.7 |  |  |  |   |

# CD19 TRI-COLOR

Linear regression  $y = 1.36 + 0.94x$

Comparison of the CALTAG CD19 TRI-COLOR conjugated monoclonal antibody with the CALTAG CD19 R-PE conjugated monoclonal antibody:

|  procedure | mean % positive | r² value | slope | Y intercept | n  |
| --- | --- | --- | --- | --- | --- |
|  CD19 TRI-COLOR | 17.1 | 97.6 | 0.99 | -0.56 | 175  |
|  CD19 R-PE | 16.4 |  |  |  |   |

# CD19 TRI-COLOR

Linear regression  $y = -0.56 + 0.99x$

# BIBLIOGRAPHY

1. Mishell B.B., Shilgi A.M., Selected methods in cellular immunology, W.H. Freeman and Company, 1980.
2. Transport and diffusion of red blood cells, Whittam R. editor, Williams and Wilkins, Baltimore, 1984.
3. Gorgi, J.V., Cheng H., Margolick J. et al, Quality control in the flow cytometric measurement of T-lymphocyte subsets: the multicenter AIDS cohort study experience, Clin. Immunol. Immunopathol. 55:173, 1990.
4. NCCLS Document H42-T, Clinical applications of flow cytometry: Quality Assurance and immunophenotyping of peripheral blood lymphocytes, Tentative Guideline, May, 1992.

---

**Source:** [https://fda-staging.innolitics.com/device/K965232](https://fda-staging.innolitics.com/device/K965232)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).