EXPERT-XL TOTAL BODY ACQUISITION AND ANALYSIS SOFTWARE

K965226 · Lunar Corp. · KPS · Oct 3, 1997 · Radiology

Device Facts

Record IDK965226
Device NameEXPERT-XL TOTAL BODY ACQUISITION AND ANALYSIS SOFTWARE
ApplicantLunar Corp.
Product CodeKPS · Radiology
Decision DateOct 3, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1200
Device ClassClass 2

Indications for Use

EXPERT-XL® Total Body Acquisition and Analysis Software is used with the EXPERT-XL® bone densitometer system. This software acquires and estimates BMD, Lean Tissue Mass, and Fat Tissue Mass for the Total Body and sub-regions. The use of the EXPERT-XL® Total Body Acquisition and Analysis Software is restricted to prescription use only. The operator’s manual for the EXPERT-XL system contains the following statement: “United States Federal Law restricts this device to the sale, distribution, and use by or on the order of a physician.” The EXPERT-XL Total Body Acquisition and Analysis software requires a 3 minute exposure, with an exposure dose of 3 mrem. This software option is substantially equivalent to the DPX Total Body software and poses no new safety or efficacy concerns.

Device Story

EXPERT-XL Total Body Acquisition and Analysis Software is an accessory for the EXPERT-XL bone densitometer. It processes scan data to estimate bone mineral density (BMD), lean tissue mass, and fat tissue mass for the total body and sub-regions. The system requires a 3-minute scan time. Operated by clinical staff under physician order, the software provides quantitative outputs used by clinicians to assess body composition and bone density. The device is used in clinical settings. It benefits patients by providing standardized measurements of tissue mass and BMD, aiding in clinical evaluation.

Clinical Evidence

Bench testing only. The device demonstrated high correlation with predicate DPX series products (r=0.97 for BMD; r=0.95 for lean/fat tissue mass). Precision (%CV) was reported as 0.4% for BMD, 3.1% for lean tissue mass, and 4.8% for fat tissue mass, which is comparable to the predicate series.

Technological Characteristics

Software accessory for bone densitometer. Operates via X-ray acquisition (3-minute scan time, 3 mrem exposure). Provides quantitative analysis of BMD (g/cm2), lean tissue mass (g), and fat tissue mass (g).

Indications for Use

Indicated for use with the EXPERT-XL bone densitometer system to acquire and estimate bone mineral density (BMD), lean tissue mass, and fat tissue mass for the total body and sub-regions. Restricted to prescription use by or on the order of a physician.

Regulatory Classification

Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0} OCT - 3 1997 LUNAR K965226 313 W. BELTLINE HIGHWAY MADISON, WI 53713 (608) 274-2663 ## 9.0 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS This summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR 807.92(c). | Contact Person: | Kenneth D. Buroker LUNAR Corporation 313 West Beltline Highway Madison, WI 53713 | | --- | --- | | Phone: | (608) 288-6460 | | Fax: | (608) 274-0853 | | Date: | December 20, 1996 | | Device/Trade Name: | EXPERT-XL Total Body Acquisition and Analysis Software | | Common Name: | Bone Densitometer | | Classification Name: | Bone Densitometer 21CFR 892.1170 | | Predicate Device: | K88 4625B Total Body Software for Lunar DPX Series K93 5454 Tissue Quantitation for DPX Series | ## 9.1 DESCRIPTION OF THE DEVICE: The EXPERT-XL Total Body Acquisition and Analysis Software is an accessory software option for estimation of Bone Mineral Density (BMD in $\mathrm{g/cm^2}$), Lean Tissue Mass (g), and Fat Tissue Mass (g) for the Total Body and sub-regions. ## 9.2 SUMMARY OF TECHNICAL CHARACTERISTICS Scans of the Total Body on the EXPERT-XL densitometer take approximately 3 minutes. The Total Body BMD estimations correlate highly ($r=0.97$) with corresponding regions estimated with the DPX series of products which already has $510(k)$ clearance. The Lean and Fat Tissue Mass estimations also correlate highly ($r=0.95$). The average short term precision (%CV) is $0.4\%$ for BMD, $3.1\%$ for lean tissue mass, and $4.8\%$ for fat tissue mass which is also comparable to that shown on the DPX series. The radiation exposure of 5 mrem is higher than that for the DPX series densitometers but remains low compared to the maximum permissible dose for the total body. {1} Summary of Safety, Effectiveness Page 2 of 2 ## 9.3 CONCLUSION The results from the EXPERT-XL Total Body Acquisition and Analysis Software are comparable to the DPX results and demonstrate similar precision. No new safety and effectiveness questions are raised with the EXPERT-XL Total Body software accessory. ![img-0.jpeg](img-0.jpeg) Kenneth D. Buroker Name Director, Regulatory Affairs Title {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT - 3 1997 Richard B. Mazess, Ph.D. President Lunar Corporation 313 W. Beltine Highway Madison, WI 53713 Re: K965226 Expert-XL Total Body Acquisition and Analysis Software Dated: July 16, 1997 Received: July 17, 1997 Regulatory class: II 21 CFR 892.1170/Procode: 90 KPS Dear Dr. Mazess: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Lillian Yin, Ph.D. Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} Page 1 of 1 # 3.0 INDICATION FOR USE FORM - 510(k) Number (if known) **K 965 226** - Device Name: EXPERT-XL Total Body Acquisition and Analysis Software - Indications for use: EXPERT-XL® Total Body Acquisition and Analysis Software is used with the EXPERT-XL® bone densitometer system. This software acquires and estimates BMD, Lean Tissue Mass, and Fat Tissue Mass for the Total Body and sub-regions. The use of the EXPERT-XL® Total Body Acquisition and Analysis Software is restricted to prescription use only. The operator’s manual for the EXPERT-XL system contains the following statement: > “United States Federal Law restricts this device to the sale, distribution, and use by or on the order of a physician.” The EXPERT-XL Total Body Acquisition and Analysis software requires a 3 minute exposure, with an exposure dose of 3 mrem. This software option is substantially equivalent to the DPX Total Body software and poses no new safety or efficacy concerns. PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED. Concurrence of CDRH, Office of Device Evaluation (ODE) David G. Lyman (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number **K965226** Prescription Use ☑ (Per 21 CRF 801.109) OR Over-the-Counter Use ☐ (Optional Format 1-2-96) CONFIDENTIAL
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...