K965198 · Johnson & Johnson International · KWL · Jul 23, 1997 · Orthopedic
Device Facts
Record ID
K965198
Device Name
ULTIMA TPS CEMENTED STEM
Applicant
Johnson & Johnson International
Product Code
KWL · Orthopedic
Decision Date
Jul 23, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3360
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ULTIMA* TPS Cemented Stem is indicated for use in total or partial hip replacement procedures for patients suffering severe pain and disability due to: 1. structural damage in the hip joint from rheumatoid arthritis, 2. osteoarthritis, 3. post-traumatic arthritis, 4. collagen disorders, 5. avascular necrosis, 6. nonunion of femoral fractures, 7. congenital hip dysplasia, 8. protrusio acetabuli, 9. slipped capital femoral epiphysis, and 10. disability due to previous fusion where bone stock is inadequate for other reconstruction techniques. The ULTIMA* TPS Cemented Stem is indicated for use only with PMMA bone cement.
Device Story
ULTIMA* TPS Cemented Stem is a hip joint metal/polymer semi-constrained cemented prosthesis; used in total or partial hip replacement procedures. Device forged from Cobalt-Chromium alloy; features straight, collarless, double-taper geometry; polished finish; 10/12 Morse taper. Available in six sizes with varying offsets and neck options; used with PMMA bone cement and distal centralizer to ensure orientation in medullary canal. Implanted by orthopedic surgeons in clinical settings. Provides structural support for hip joint; alleviates pain and disability; restores function. Substantial equivalence established via mechanical testing (2,300N force, 5,000,000 cycles).
Clinical Evidence
Bench testing only. Mechanical testing demonstrated the device withstands a maximum load of 2,300N force for 5,000,000 cycles, meeting established criteria for substantial equivalence.
Technological Characteristics
Cobalt-Chromium alloy (ASTM F799); PMMA distal centralizer. 10/12 Morse taper; collarless; straight, double-tapered stem. Neck angle 125°. Available in six sizes with multiple offset and neck length options. Non-sterile; intended for use with PMMA bone cement.
Indications for Use
Indicated for patients suffering severe pain and disability requiring total or partial hip replacement due to rheumatoid arthritis, osteoarthritis, post-traumatic arthritis, collagen disorders, avascular necrosis, femoral fracture nonunion, congenital hip dysplasia, protrusio acetabuli, slipped capital femoral epiphysis, or previous fusion with inadequate bone stock. Contraindicated for use without PMMA bone cement.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.
{0}
JUL-22-37 15:45 FROM:J & J/REG AFFAIRS
ID:5088283212
PAGE 5/7
# K965198
## Summary of Safety & Effectiveness Data for the ULTIMA* TPS Cemented Stem
JUL 23 1997
Johnson & Johnson Professional, Inc.
325 Paramount Drive
Raynham, Massachusetts 02767
1. Contact Person
Janet G. Johnson, Assoc. Regulatory Affairs Specialist, (508) 828-3466.
2. Device Name
Proprietary Name: ULTIMA* TPS Cemented Stem
Common Name: Hip joint metal/polymer semi-constrained cemented prosthesis
Classification: Prosthesis, Hip
Name:
Regulatory Class: Class II by 21 CFR § 888.3350
Product Code: 87JDI
Owner/Operator #: 9001269
3. Device Classification
Classification for ULTIMA* TPS Cemented Stem has been placed in Class II by 21 CFR § 888.3350
4. Statement of Substantial Equivalence
The safety and effectiveness of the ULTIMA* TPS Cemented Stem is substantially equivalent in terms of function to the Howmedica Exeter hip stem (#K891454) and to the Zimmer CPT hip stem. Furthermore, testing results demonstrate that the ULTIMA* TPS Cemented Stem is substantially equivalent by withstanding a maximum load of 2,300N force, 5,000,000 cycles, and, as such meets the set criteria for the establishment of "substantial equivalence".
{1}
ID:5088283212
PAGE 6/7
# 5. Indications for Use
The ULTIMA* TPS Cemented Stem is indicated for use in total or partial hip replacement procedures for patients suffering severe pain and disability due to:
1. structural damage in the hip joint from rheumatoid arthritis,
2. osteoarthritis,
3. post-traumatic arthritis,
4. collagen disorders,
5. avascular necrosis,
6. nonunion of femoral fractures,
7. congenital hip dysplasia,
8. protrusio acetabuli,
9. slipped capital femoral epiphysis, and
10. disability due to previous fusion where bone stock is inadequate for other reconstruction techniques.
The ULTIMA* TPS Cemented stem is indicated for use only with PMMA bone cement.
# 6. Physical Description
The ULTIMA* TPS Cemented Stem is forged from a Cobalt-Chromium alloy which conforms to ASTM F799 (EXHIBIT IV); this material has high strength and good hardness properties and is also highly immune to fatigue. The ULTIMA* TPS Cemented Stem is polished to a bright finish on all surfaces below the spigot and can be used interchangeably in either the left or the right hip. The stem is straight, tapering towards the distal tip in both the mediolateral and anteriorposterior sections. It does not have a proximal collar.
The ULTIMA* TPS Cemented Stem has a 10/12 Morse taper; it will be available in six sizes (0, 1 - 5), each of the 1 - 5 sizes have a choice of either a 37.5 mm (standard) offset, or a 44.0 mm (medialized "M") offset. In addition to this central range, there shall be two long stem options. The stem can be used with various neck heads (22.225, 26, 28 and 32 mm) as well as neck options (-5 (32), -3 (28), 0, +5 and +10). All of the stems in the range have a neck angle of 125°.
| Stem Size | Trunnion Designs | Offset | Neck Heads | Neck Options | Neck Angle |
| --- | --- | --- | --- | --- | --- |
| 0,1,2,3,4,5 | 10/12 taper | 37.5 / 44.0mm | 22.225, 26, 28, 32 mm | -5 (32), -3 (28), 0, +5, +10 | 125° |
{2}
JUL-22-97 15:46 FROM: J & J/REG AFFAIRS
ID:5088283212
PAGE 7/7
## Distal Centralizer
The distal centralizer that is used with the ULTIMA® TPS stem comes in a range of sizes and is manufactured from PMMA. There are three wings on the centralizer which ensure the orientation of the stem distally. These increase in size through the centralizer range to deal with the variation in anatomy of the medullary canal. In revision surgery, a "wingless" centralizer is required, thus a 7mm diameter version is available which simply acts to provide the void, and does not centralize the stem.
Table 1. Similarities and Differences Matrix
| | ULTIMA® TPS Cemented Stem | Howmedica Exeter hip stem | Zimmer CPT hip stem |
| --- | --- | --- | --- |
| DESIGN | | | |
| Polished | Yes | Yes | Yes |
| Collarless | Yes | Yes | Yes |
| Double-taper geometry | Yes | Yes | Yes |
| INTENDED USE | | | |
| Total / partial hip replacement | Yes | Yes | Yes |
| Used with distal centralizer | Yes | Yes | Yes |
| MATERIALS | | | |
| Orthinox™ (stainless steel alloy) | No | Yes | No |
| Zimtron™ (stainless steel alloy) | No | No | Yes |
| Cobalt-Chromium-Molybdenum | Yes | No | No |
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Ms. Janet G. Johnson
*Associate Regulatory Affairs Specialist
Johnson & Johnson Professional, Inc.
325 Paramount Drive
Raynham, Massachusetts 02767-0350
JUL 23 1997
Re: K965198
Ultima* TPS Cemented Stem
Regulatory Class: II
Product Codes: JDI and KWL
Dated: April 22, 1997
Received: April 25, 1997
Dear Ms. Johnson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
Page 2 - Ms. Janet G. Johnson
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}
# Indications for Use for the ULTIMA* TPS Cemented Stem
Johnson & Johnson Professional, Inc.
325 Paramount Drive
Raynham, MA 02767-0350
*Indications for Use* K965198
The ULTIMA* TPS Cemented Stem is indicated for use in total or partial hip replacement procedures for patients suffering severe pain and disability due to:
1. structural damage in the hip joint from rheumatoid arthritis,
2. osteoarthritis,
3. post-traumatic arthritis,
4. collagen disorders,
5. avascular necrosis,
6. nonunion of femoral fractures,
7. congenital hip dysplasia,
8. protrusio acetabuli,
9. slipped capital femoral epiphysis, and
10. disability due to previous fusion where bone stock is inadequate for other reconstruction techniques.
The ULTIMA* TPS Cemented Stem is indicated for use only with PMMA bone cement.
Prescription Use ☑
(Per 21 CFR 801.109)

Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.