The Duracon®, P.C.A.® and Kinemax® Plus Tibial Inserts, All Plastic Tibial Components and Patellar Components (previously cleared in K915512, K923573, K910235, K922048, K932070, K913188, K872735, K871772 and K921640) are intended to be used with Duracon®, P.C.A.® or Kinemax® femoral components, tibial baseplates and patellar components in primary or revision cemented total knee arthroplasty.
Device Story
Duration™ Stabilized UHMWPE knee components are orthopedic implants designed for primary or revision cemented total knee arthroplasty. The device consists of tibial inserts, all-plastic tibial components, and patellar components. These components are intended for use with existing Duracon®, P.C.A.®, or Kinemax® femoral components, tibial baseplates, and patellar components. The primary innovation is a stabilized UHMWPE material processing method intended to improve wear resistance. The device is used by orthopedic surgeons in a clinical/surgical setting. The surgeon selects the components based on patient anatomy and surgical requirements for knee replacement. The stabilized material is intended to reduce volumetric wear over the life of the implant, potentially benefiting the patient by extending the functional lifespan of the knee arthroplasty and reducing the risk of osteolysis associated with wear debris.
Clinical Evidence
Bench testing only. No clinical data provided. Wear testing performed using reciprocating ring-on-block and pin-on-disk evaluations over 2.5 to 5 million cycles. Results showed 30% to 91% reduction in volumetric wear for Duration™ Stabilized UHMWPE compared to conventional gamma-sterilized UHMWPE. Note: In vitro test results have not been shown to correlate with clinical wear mechanisms.
Indicated for patients undergoing primary or revision cemented total knee arthroplasty requiring tibial inserts, all-plastic tibial components, or patellar components.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}
K965173
June 11, 1997
# 510(K) SUMMARY
**Trade Name:** Duration™ Stabilized UHMWPE Knee Components
**Common Name:** UHMWPE Knee Components
This submission includes wear testing data on UHMWPE Knee Components sterilized by the method previously cleared in 510(k) submission K936292. This data is provided to substantiate claims of improved wear due to the processing of the devices. The submission includes Duracon®, P.C.A.® and Kinemax® Plus Tibial Inserts, All Plastic Tibial Components and Patellar Components which are intended to be used with Duracon®, P.C.A.® or Kinemax® Plus femoral components, tibial baseplates and patellar components in primary or revision cemented total knee arthroplasty.
The following are the wear claims that will be made for these devices:
A block of Howmedica’s Duration™ Stabilized UHMWPE showed a 30% reduction in volumetric wear versus the same block of Howmedica’s conventionally gamma sterilized UHMWPE. Testing was performed in a reciprocating ring-on-block wear test for over 5 million cycles, using a circular disk, 2.83” in diameter, 1” wide, CoCr articulating counterface and bovine calf serum as a lubricant. The results of *in vitro* tests have not been shown to correlate with clinical wear mechanisms.
In an independent laboratory test, a 9mm thick circular disk of Howmedica’s Duration™ Stabilized UHMWPE showed a 68% reduction in volumetric wear versus the same circular disk of Howmedica’s conventionally gamma sterilized UHMWPE. Testing was performed in a reciprocating pin-on-disk wear evaluation over 4 million cycles, using a CoCr cylindrical pin with a 1” spherical end as the articulating counterface and bovine calf serum as a lubricant. The results of *in vitro* tests have not been shown to correlate with clinical wear mechanisms.
In an independent laboratory test, a 9 mm thick circular disk of Howmedica’s Duration™ Stabilized UHMWPE, having undergone 23 days of heating in air to simulate 7-9 years of “aging”, showed a 91% reduction in volumetric wear versus the same circular disk of Howmedica’s conventionally gamma sterilized UHMWPE, having undergone 23 days of heating in air to simulate 7-9 years of “aging”. Testing was performed in a
{1}
REGULATORY AFFAIRS ID:2701 JUN 05'97 14:49 No.006
reciprocating pin-on-disk wear evaluations over 2.5 million cycles, using a CoCr
cylindrical pin with a 1" spherical end as the articulating counterface and bovine calf
serum as a lubricant. The results of in vitro tests have not been shown to correlate with
clinical wear mechanisms.
For information, contact: John Dichiara
Manager, Regulatory Affairs
Howmedica Inc.
359 Veterans Boulevard
Rutherford, NJ 07070
(201) 507-7386 - Phone
(201) 507-6870 - Fax
{2}
REGULATORY AFFAIRS ID:2701 JUN 05'97 14:50 No.006 P.15
# DURATION™ WEAR CLAIMS - KNEE COMPONENTS K965173
A block of Howmedica's Duration™ Stabilized UHMWPE showed a 30% reduction in volumetric wear versus the same block of Howmedica's conventionally gamma sterilized UHMWPE. Testing was performed in a reciprocating ring-on-block wear test for over 5 million cycles, using a circular disk, 2.83" in diameter, 1" wide, CoCr articulating counterface and bovine calf serum as a lubricant. The results of in vitro tests have not been shown to correlate with clinical wear mechanisms.
In an independent laboratory test, a 9 mm thick circular disk of Howmedica's Duration™ Stabilized UHMWPE showed a 68% reduction in volumetric wear versus the same circular disk of Howmedica's conventionally gamma sterilized UHMWPE. Testing was performed in a reciprocating pin-on-disk wear evaluation over 4 million cycles, using a CoCr cylindrical pin with a 1" spherical end as the articulating counterface and bovine calf serum as a lubricant. The results of in vitro tests have not been shown to correlate with clinical wear mechanisms.
In an independent laboratory test, a 9 mm thick circular disk of Howmedica's Duration™ Stabilized UHMWPE, having undergone 23 days of heating in air to simulate 7-9 years of "aging", showed a 91% reduction in volumetric wear versus the same circular disk of Howmedica's conventionally gamma sterilized UHMWPE, having undergone 23 days of heating in air to simulate 7-9 years of "aging". Testing was performed in a reciprocating pin-on-disk wear evaluations over 2.5 million cycles, using a CoCr cylindrical pin with a 1" spherical end as the articulating counterface and bovine calf serum as a lubricant. The results of in vitro tests have not been shown to correlate with clinical wear mechanisms.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN 11 1997
Mr. John Dichiara
Regulatory Affairs Manager
Howmedica Inc.
Division of Pfizer Hospital Products Group
359 Veterans Boulevard
Rutherford, New Jersey 07070
Re: K965173
Trade Name: Duration Stabilized UHMWPE - Knee Components
Regulatory Class: II
Product Codes: JWH and HRY
Dated: March 21, 1997
Received: March 25, 1997
Dear Mr. Dichiara:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic
{4}
Page 2 - Mr. John Dichiara
GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the *Federal Register*. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Celis M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}
KEGULHUKY HFFHIKS ID:2701 JUN 05'97 14:50 No.006 P.14
# Indications for Use
510(k) Number (if known): K965173
Device Name: Duration™ Stabilized UHMWPE Knee Components
Indications for Use:
The Duracon®, P.C.A.® and Kinemax® Plus Tibial Inserts, All Plastic Tibial Components and Patellar Components (previously cleared in K915512, K923573, K910235, K922048, K932070, K913188, K872735, K871772 and K921640) are intended to be used with Duracon®, P.C.A.® or Kinemax® femoral components, tibial baseplates and patellar components in primary or revision cemented total knee arthroplasty.
SEE ADDITIONAL PAGE LABELED DURATION™ WEAR CLAIMS
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K965173
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.