LIFE SUPPORT FOR TRAUMA AND TRANSPORT

K965117 · Northrop Grumman Corp. · MKJ · Jun 26, 1998 · Cardiovascular

Device Facts

Record IDK965117
Device NameLIFE SUPPORT FOR TRAUMA AND TRANSPORT
ApplicantNorthrop Grumman Corp.
Product CodeMKJ · Cardiovascular
Decision DateJun 26, 1998
DecisionST
Submission TypeTraditional
Regulation21 CFR 870.5310
Device ClassClass 3
AttributesTherapeutic

Indications for Use

The Life Support for Trauma and Transport (LSTAT) is a portable system providing Intensive Care Unit (ICU) functionality for adult patients. Its use as a device is to provide in a single platform: physiological monitoring, upper airway or gastrointestinal drainage suction, automated external defibrillation, IV and drug administration pumping, ventilation and oxygen delivery for use in hospitals, aircraft, ambulances, field hospitals and extended care facilities. Defibrillation is only applied to unconscious pulseless and apneic patients weighing over 90 pounds. Operating power is furnished via external AC, external DC or internal rechargeable batteries

Device Story

LSTAT Model 9601 is a portable, integrated ICU platform designed for military/civilian transport. It houses six medical subsystems: physiological monitoring (ECG, NIBP, invasive BP, temperature), blood chemistry assessment, suction, AED, IV/drug administration, and a ventilator. Inputs include patient physiological signals and environmental data (tilt, acceleration, humidity, altitude). The device transforms these into real-time monitoring, life support, and data logs. Operated by clinicians in pre-operative, intra-operative, and post-operative settings, including aircraft and ambulances. The system features a 'cockpit-like' interface for anesthesiologists and a secondary display for remote monitoring. Data logging supports post-event analysis. Benefits include continuous ICU-level care during transport, eliminating the need to move patients between platforms. The structure is EMI-protected, shock-mounted, and compatible with NATO litters and military evacuation vehicles.

Clinical Evidence

Bench testing only. The device underwent comprehensive stress characterization (vibration, shock, thermal, altitude), EMI/EMS testing, and airworthiness qualification. Validation was conducted in collaboration with U.S. Government laboratories (WRAIR, USAARL, Brooks AFB). No clinical trial data provided.

Technological Characteristics

Composite structure assembly; shock-mounted components; EMI-protected. Dimensions: 86.8"L x 23.0"W x 13.2"H. Power: 115/230 VAC, 400 Hz, 25 VDC, and internal rechargeable batteries. Subsystems: ECG, NIBP, invasive BP, temperature, blood chemistry, suction, AED, IV pump, ventilator (oxygen-air blending). Connectivity: Ethernet for secondary display. Environmental monitoring: tilt, acceleration, humidity, ambient temperature, altitude. Sterilization: N/A (integrated system).

Indications for Use

Indicated for adult trauma victims or patients requiring ICU-level physiological monitoring and treatment (resuscitation, stabilization, transport). Applicable for combat/civilian trauma, respiratory distress, pneumothorax, fluid loss, cardiac emergency, or unstable clinical presentation. Defibrillation contraindication: patients weighing under 90 pounds or those who are conscious/have a pulse.

Regulatory Classification

Identification

An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.

Submission Summary (Full Text)

{0} K965117 JUN 26 1998 **NORTHROP GRUMMAN** Advanced Technology and Development Center Northrop Grumman Corporation 8900 East Washington Boulevard Pico Rivera, California 90660-0155 Telephone 310 948-9161 ## SECTION 16 ### FDA 510(k) SUMMARY #### Submitted By **Company:** Northrop Grumman Corporation (NGC) Advanced Technology and Development Center (ATDC) N560/XA 8900 E. Washington Blvd. Pico Rivera, CA 90660 **Contact Person:** Louis S. Toth LSTAT Project Engineer (310) 948-6482 (voice) (310) 948-9485 (FAX) **Date:** December 17, 1996 #### Name of the Device **Trade Name:** Life Support for Trauma and Transport (LSTAT) Model 9601 **Common Name:** Intensive Care Unit (ICU) Platform Portable, Individualized ICU Trauma Pod Medical Evacuation Platform **Classification Name:** (This Integrated System Category is Not Listed in 21 CFR Parts 862-892.) #### Device Description The LSTAT Model 9601 is a portable individualization intensive care unit that incorporates resuscitative and life-sustaining medical devices within a single composite structure assembly. The LSTAT functionality is shown in Figure 16-1 as a well equipped ICU bed. {1} ![img-0.jpeg](img-0.jpeg) Figure 16-1. LSTAT Model 9601: General ICU Functionality Physically, the LSTAT Model 9601 is a highly instrumented patient transport platform (interfaced to a NATO litter) with medical devices incorporated within its composite structure assembly. Figure 16-2 shows the appearance of the LSTAT assembly whose unique outer moldline and low profile are defined for military transport compatibility (height = 13.2 inches, length = 86.8 inches, and width = 23.0 inches without patient). Device interfaces are colocated to favor awareness and controls in the "anestheologist position" during intra-operative procedures. The LSTAT incorporates six portable patient monitoring/treatment subsystems/capabilities and five utilities/support subsystems as identified below. The LSTAT Medical Subsystems include: 1. Physiological Monitoring a. ECG b. NIBP c. Invasive Blood Pressure (BP) (two channels) d. Temperature {2} ![img-1.jpeg](img-1.jpeg) Figure 16-2. LSTAT ICU Platform Model 9601 2. Blood Chemistry Assessment (in vitro) 3. Suction 4. Automated External Defibrillator (AED) 5. IV Fluid Administration (and IV drugs) 6. Ventilator a. On-board compressor (for medical air only) b. Oxygen-air blending c. Patient ventilation (controllable: mandatory or assist) d. Ventilation assessment: airway pressure, airway pCO₂ and EₜCO₂, airway flow and tidal volume, and oxygen saturation (pulse oximeter) The LSTAT Utilities/Support Subsystems include: 7. Structure (including NATO litter) 8. Oxygen 9. Electrical Power 10. Environmental Control Subsystem (ECS) 11. Data Logging and Display Subsystem (DDLS) including a secondary display Figure 16-3 illustrates the anesthesiologist's intra-operative position. Figure 16-4 illustrates the operator-patient interface panel where most of the patient connections to the LSTAT are made. {3} ![img-2.jpeg](img-2.jpeg) 96-091-010 ![img-3.jpeg](img-3.jpeg) Figure 16-3. Anesthesiologists Intra-Operative Position (Control and Display Cockpit) Figure 16-4. Operator-Patient Interface Panel (Patient Connections) ## Indications for Use: Medical Applications The LSTAT Model 9601 is used to support resuscitation, stabilization, and transport of adult trauma victims or patients needing physiological monitoring and treatment capabilities of a hospital intensive care unit or trauma care unit for pre-operative, intra-operative, and post operative support during pre, intra, and post-transport phases. The indications for use of selected subsystems/capabilities are dictated by medical protocol, but in general LSTAT use may be anticipated for victims of: combat/civilian casualty, trauma, respiratory distress, pneumothorax, severe fluid loss, cardiac emergency, or unstable clinical presentation. {4} Of course, the LSTAT may be used for transport of less severely injured or traumatized individuals which may not require the use of all the ICU subsystems. ## Indications for Use: Transportation Applications The LSTAT serves as a transportation platform which integrates existing medical devices and the NATO (STANAG) litter into a form-factor small enough to be loaded/carried within the constraints of military evacuation/transport vehicles like: - UH-60 Blackhawk helicopter - UH-1 Huey helicopter - CH-47 Chinook helicopter - C-9 aircraft - C-17 aircraft - C-130 aircraft - C-141 aircraft - MK-II Humvee ground ambulance - Carrier, litter wheeled, (US Army). Of course, other evacuation/transport vehicles (military or civilian) which do not impose the small (and customized) form factor or which are less densely loaded may also use the LSTAT. ## Safety and Efficacy Discussions The LSTAT provides for the safe transport, physiological monitoring, and treatment of trauma victims/patients without compromising efficacy by integrating FDA cleared medical devices and aerospace developed technologies for operation in severe environments. The LSTAT is an integrated medical system whose design includes providing ICU functional capabilities, packaging devices for transport, data logging integration, and military evacuation vehicle/environment compatibility. Its safety and efficacy are based on: 1. Integration of existing legally marketed medical devices which have FDA clearance without change to their functionality, capability, controls, operator interfaces, or patient interfaces {5} 2. Provision of utilities and support which meet the requirements of each device’s operation 3. Continuous logging of both patient data and system/device data and alarms for redundant messaging of alarms and for support of post-treatment analysis of any event or observation (24 hours operations before down-load required with continued monitoring during down-load) 4. Aerospace designs for structure, severe environment operations, and airworthiness qualifications. ## Device Equivalents The LSTAT incorporates six patient monitoring/treatment subsystems which contain medical devices which are summarized in Figure 16-5 with their FDA Class type. All devices except the ventilator are currently marketed by their respective suppliers. The ventilator is a customized, integrated subsystem developed for LSTAT. That subsystem incorporates multiple components of devices already legally marketed into its integrated design. | Subsystem | Class | | --- | --- | | Physiological Monitoring | II | | Blood Chemistry | II | | Suction | II | | Defibrillator | III | | IV and Drug Administration | II | | Ventilator | II | Figure 16-5. Patient Monitoring/Treatment Subsystem The listed devices are not just equivalent by predicate reference; they are equivalent by identity. The LSTAT integration design has not changed any of the supplier’s circuitry or patient/operator interfaces. The modifications (discussed below) made to the devices have been to accommodate their integration into the low profile structure and to ensure that the devices will be operational in the severe military/transport environments. {6} # Device Modification Enhancements Minor repackaging of medical devices for severe operational environments and provisions for on-board utilities are the basic modifications and enhancements for medical devices contained within the LSTAT Model 9601. The primary modifications made to the marketed devices have been to: 1. Disassemble them 2. Assemble multiple components onto shock mounts 3. Cluster patient connections and operator controls for “cockpit-like” interface 4. Locate components to maintain low profile within the composite EMI-protected structure 5. Provide environmental control 6. Provide centralized power sources, distribution, and support for all components (without preferential supply or denial of resources) 7. Provide communications with devices for data logging, analysis, and alarms. # System Integration Enhancements The integration approach serves to offer several enhancements as a system: 1. **Power**: Centralization of power utilities permits all devices to be operated from multiple auxiliary (external) power sources and permits single cabling to their respective interfaces: | 115 VAC +/- 10% | 60 Hz +/- 5 Hz 1 phase | | --- | --- | | 230 VAC +/- 10% | 50 Hz +/- 3 Hz 1 phase | | 108-118/200 VAC | 400 Hz +/- 7 Hz any phase | | 25 +/- 5 VDC | | and to provide un-interruptible backup power capabilities to all devices in the event of loss of auxiliary power. The rechargeable battery source also supports stand-alone operations (at least 30 minutes) during movement between locations while auxiliary power is unavailable. With the centralized subsystem, the need to manage recharge/replacement of each subsystem’s batteries individually is eliminated and readiness of all devices concurrently is ensured. No software controls are used in the EPS which could provide power or operations to one {7} device preferentially over another. All power switching is under operator control and hardwired/circuit logic. 2. **Data Logging and Display**: Continuous data logging of patient and system data/status (including all real-time patient parameter waveforms) supports post event analyses of trauma physiology and system operation (flight data recorder equivalency). A secondary display, which is a portable computer, tethered, and capable of displaying any of the subsystems-reported data/alarms via real-time, ethernet communications is provided. This display permits the patient (or system) status to be presented at any location around the LSTAT structure thereby permitting the operator (or observer) to receive information even if the primary device display is not accessible during transport (or operative) scenarios. This secondary display could also include capability for wireless transmission of the data. 3. **Environmental Control**: Accommodations for equipment cooling and temperature are provided. Electromagnetic interference (EMI) protection is provided for control of emanations and electromagnetic susceptibility (EMS). 4. **Structural Support**: The structure provides the physical support for the patient and components within a lightweight composite housing which interfaces with the Blackhawk’s carousel pan and other military evacuation vehicles identified above. The structure locks the NATO litter (which is assumed to be the way the patient is brought to the LSTAT) in place thereby eliminating the need to move the patient between platforms. The physical mounting of the internal components are designed to accommodate the vibration, shock, and rugged environments anticipated. The structure provides monitoring of tilt, acceleration, humidity, ambient temperature, and altitude and includes this information in the data logging records being generated. 5. **Oxygen**: Compressed and regulated oxygen (480 liters) is provided for controlled blending with ambient air to administer oxygen enriched air through the ventilator or through a mask/cannula. ## Safe Design Analysis Aerospace qualified safety engineers provide analysis and acceptance validation of the LSTAT modification designs, implementations, and hazard analyses (see Hazard Analysis in Appendix A). Supplier hazard analyses were included to ensure no modification created or impacted existing device hazard mitigation, and NGC hazard {8} analysis and mitigation assured that no NGC design or implementation produced unacceptable patient or operator hazard. ## Test Program A comprehensive test program is in place to evaluate devices and total LSTAT assembly for integrated functionality (in vitro subsystem acceptance testing per suppliers directions), stress characterization (vibration, shock, thermal, and altitude), component and system EMI/EMS, manufacturing acceptance tests, usability tests, and airworthiness qualifications. Figure 16-6 illustrates the series and sequence of testing addressed in the Test Plan. ![img-4.jpeg](img-4.jpeg) 96-091-039A Figure 16-6. System Level Validation The design to operating temperature ranges for many pieces of equipment and parts are $-26^{\circ}\mathrm{C}$ to $+49^{\circ}\mathrm{C}$. Some medical displays and fluid handling capabilities may be limited at $0^{\circ}\mathrm{C}$. U.S. Government Laboratories (WRAIR, USAARL, and Brooks AFB) are participating with Northrop Grumman engineers/technicians in the validation/verification of safety and efficacy. Several of the incorporated devices have been through airworthiness qualification testing at USAARL as separate devices: physiological monitor, suction, defibrillator, and IV infusion pump. {9} # Conclusion The LSTAT integration of legally marketed medical devices into a portable, individualized, intensive care unit has maintained the safety and efficacy of the existing devices and provided utilities and support for their operations which have passed the rigors of aerospace design and development. The safety and efficacy of the LSTAT system in military and airborne and environments will be asserted by the U.S. Government and tracked with analysis of the extensive data logging capability of each device. The above 510(k) Summary accurately reflects the information as submitted in other portions of this application. If you have any questions, please direct them to the undersigned at (310) 948-8628 or Fax at (310) 942-3726. Sincerely, Nat F. Piscitelli, Director Business Management and Operations Northrop Grumman Corporation Advanced Technology and Development Center {10} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUN 26 1998 Louis S. Toth, Ph.D. Northrop Grumman Corporation Military Aircraft Systems Division Advanced Systems and Technology Department N800/XA 8900 East Washington Boulevard Pico Rivera, CA 90660 Re: K965117 Life Support for Trauma and Transport Regulatory Class: III (three) Product Code: 74 MKJ Dated: April 24, 1998 Received: April 28, 1998 Dear Dr. Toth: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. You may, therefore, market the device, subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act). The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under section 531 through 542 of the Act for {11} Page 2 - Louis S. Toth, Ph.D. devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. On August 16, 1993, the Final Rule for Device Tracking was published in the *Federal Register*, pages 43442-43455 (copy enclosed). Be advised that under Section 519(e) of the Act as amended by the Safe Medical Devices Act of 1990, FDA has identified the above device as a device which requires tracking. Because the device is subject to tracking, you are required to adopt a method of tracking that follows the devices through the distribution chain and then identifies and follows the patients who receive them. The specific requirements of the regulation are found in 21 CFR 821 as described in the August 16, 1993, *Federal Register* beginning on page 43447. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please not the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act, may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, ![img-5.jpeg](img-5.jpeg) Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosures {12} USE OR DUPLICATION OF NORTHROP GRUMMAN PROPRIETARY INFORMATION IS SUBJECT TO THE RESTRICTIONS ON THE TITLE PAGE OF THIS DOCUMENT NORTHROP GRUMMAN Advanced Technology and Development Center 510(k) Number (If known): K965117 Device Name: Life Support for Trauma and Transport (LSTAT) Indications for Use: The Life Support for Trauma and Transport (LSTAT) is a portable system providing Intensive Care Unit (ICU) functionality for adult patients. Its use as a device is to provide in a single platform: physiological monitoring, upper airway or gastrointestinal drainage suction, automated external defibrillation, IV and drug administration pumping, ventilation and oxygen delivery for use in hospitals, aircraft, ambulances, field hospitals and extended care facilities. Defibrillation is only applied to unconscious pulseless and apneic patients weighing over 90 pounds. Operating power is furnished via external AC, external DC or internal rechargeable batteries (PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-6.jpeg](img-6.jpeg) Prescription Use ☑ OR Over-The-Counter Use ☐ (Per 21 CFR 801.109) Figure 2-2. Indications for Use (Optional Format)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...