K965108 · Beckman Instruments, Inc. · DEW · Mar 6, 1997 · Immunology
Device Facts
Record ID
K965108
Device Name
SYNCHRON LX SYSTEMS IMMUNOGLOBULIN G/A/M REAGENT
Applicant
Beckman Instruments, Inc.
Product Code
DEW · Immunology
Decision Date
Mar 6, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5510
Device Class
Class 2
Indications for Use
The SYNCHRON LX Systems Immunoglobulin G (Ig-G) reagent, when used in conjunction with SYNCHRON LX Calibrator 1, is intended for the quantitative determination of human immunoglobulin G in serum or plasma. This assay is designed for use with clinical chemistry analyzers from Beckman Instruments, such as the SYNCHRON LX™20 Clinical System. The SYNCHRON LX Systems Immunoglobulin A (Ig-A) reagent, when used in conjunction with SYNCHRON LX Calibrator 1, is intended for the quantitative determination of human immunoglobulin A in serum or plasma. This assay is designed for use with clinical chemistry analyzers from Beckman Instruments, such as the SYNCHRON LX™20 Clinical System. The SYNCHRON LX Systems Immunoglobulin M (Ig-M) reagent, when used in conjunction with SYNCHRON LX Calibrator 1, is intended for the quantitative determination of human immunoglobulin M in serum or plasma. This assay is designed for use with clinical chemistry analyzers from Beckman Instruments, such as the SYNCHRON LX™20 Clinical System.
Device Story
Reagents for quantitative determination of human Ig-G, Ig-A, and Ig-M in serum or plasma; used on SYNCHRON LX clinical chemistry analyzers. Principle: antigen-antibody complex formation measured via turbidimetry at 37°C and 340 nm. High antigen concentration detection prevents false-in-range reporting. Reagents stored in polystyrene cartridges onboard instrument. Healthcare providers use results for clinical assessment of immunoglobulin levels. Benefits: automated, standardized quantification of immunoglobulins for diagnostic support.
Clinical Evidence
Bench testing only. Method comparison study (n=72-82) against predicate methods showed high correlation (r=0.9914 to 0.9970). Within-run imprecision (%CV) ranged from 1.3% to 3.9% across two levels for all three analytes. Stability data supports 24-month shelf-life, 14-day calibration stability, and 60-day on-board stability.
Technological Characteristics
Turbidimetric assay; 37°C reaction temperature; 340 nm endpoint detection. Reagents packaged in polystyrene cartridges. Onboard refrigerated storage. Uses tris buffer for sample dilution. Single-point calibration update.
Indications for Use
Indicated for quantitative determination of human immunoglobulin G, A, and M in serum or plasma using clinical chemistry analyzers.
Regulatory Classification
Identification
An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.
{0}
Beckman Instruments, Inc., Section 510(k) Notification
SYNCHRON LX™ Systems Immunoglobulin G, A, & M Reagents
Summary of Safety & Effectiveness
K965108
MAR - 6 1997
Summary of Safety & Effectiveness
SYNCHRON LX™ Systems Immunoglobulin G, A, and M (Ig-G, Ig-A, & Ig-M) Reagents
## 1.0 Submitted By:
Kathleen M. Jaker
Senior Applications Chemist
Beckman Instruments, Inc.
200 S. Kraemer Blvd., W-117
Brea, California 92822-8000
Telephone: (714) 961-3970
FAX: (714) 961-4457
## 2.0 Date Submitted:
18 December 1996
## 3.0 Device Name(s):
### 3.1 Proprietary Names
SYNCHRON LX™ Systems Immunoglobulin G (Ig-G) Reagent
SYNCHRON LX™ Systems Immunoglobulin A (Ig-A) Reagent
SYNCHRON LX™ Systems Immunoglobulin M (Ig-M) Reagent
### 3.2 Classification Names
Immunoglobulin G Test System (21 CFR §866.5510)
Immunoglobulin A Test System (21 CFR §866.5510)
Immunoglobulin M Test System (21 CFR §866.5510)
## 4.0 Predicate Device(s):
| SYNCHRON LX Reagents | Predicate | Predicate Company | Docket Number |
| --- | --- | --- | --- |
| Immunoglobulin G (Ig-G) Reagent | Beckman Immunoglobulin G (IGG) Reagent | Beckman Instruments, Inc. | K771603 |
| Immunoglobulin A (Ig-A) Reagent | Beckman Immunoglobulin A (IGA) Reagent | Beckman Instruments, Inc. | K771603 |
| Immunoglobulin M (Ig-M) Reagent | Beckman Immunoglobulin M (IGM) Reagent | Beckman Instruments, Inc. | K771603 |
file: ig510k.sse
{1}
Beckman Instruments, Inc., Section 510(k) Notification
SYNCHRON LX™ Systems Immunoglobulin G, A, & M Reagents
Summary of Safety & Effectiveness
# 5.0 Description:
The SYNCHRON LX™ Systems Immunoglobulin G, A, and M (Ig-G, Ig-A, Ig-M) Reagents in conjunction with SYNCHRON LX Calibrator 1, are intended for use on Beckman's SYNCHRON LX Clinical Systems.
# 6.0 Intended Use:
The SYNCHRON LX Systems Immunoglobulin G (Ig-G) reagent, when used in conjunction with SYNCHRON LX Calibrator 1, is intended for the quantitative determination of human immunoglobulin G in serum or plasma. This assay is designed for use with clinical chemistry analyzers from Beckman Instruments, such as the SYNCHRON LX™20 Clinical System.
The SYNCHRON LX Systems Immunoglobulin A (Ig-A) reagent, when used in conjunction with SYNCHRON LX Calibrator 1, is intended for the quantitative determination of human immunoglobulin A in serum or plasma. This assay is designed for use with clinical chemistry analyzers from Beckman Instruments, such as the SYNCHRON LX™20 Clinical System.
The SYNCHRON LX Systems Immunoglobulin M (Ig-M) reagent, when used in conjunction with SYNCHRON LX Calibrator 1, is intended for the quantitative determination of human immunoglobulin M in serum or plasma. This assay is designed for use with clinical chemistry analyzers from Beckman Instruments, such as the SYNCHRON LX™20 Clinical System.
# 7.0 Comparison to Predicate(s):
The following tables show similarities and differences between the predicates identified in Section 4.0 of this summary.
SIMILARITIES to the PREDICATE
| Reagent | Aspect/Characteristic | Comments |
| --- | --- | --- |
| SYNCHRON LX Systems Immunoglobulin (Ig-G, Ig-A, & Ig-M) Reagents | Intended Use | Same as Beckman Immunoglobulin reagents; quantitative determination of human immunoglobulins |
| | Chemical Reaction | Same principle as the Immunoglobulin Beckman Reagents; formation of antigen-antibody complex |
| | Antibody | Same source, antibody, processing, and buffer as Beckman Immunoglobulin reagents |
| | Range Expansion | Same as the Beckman Immunochemistry (ARRAY) System which expands usable range with alternate sample dilution ratios |
| | Calibration | Same as the Beckman Immunoglobulin reagents; single point update of manufacturer-determined calibration curve. |
file: ig510k.sse
{2}
Beckman Instruments, Inc., Section 510(k) Notification
SYNCHRON LX™ Systems Immunoglobulin G, A, & M Reagents
Summary of Safety & Effectiveness
DIFFERENCES from the PREDICATE
| Reagent | Aspect/Characteristic | Comments |
| --- | --- | --- |
| SYNCHRON LX Systems Immunoglobulin (Ig-G, Ig-A, & Ig-M) Reagents | Methodology | The SYNCHRON LX reads turbidimetrically and the ARRAY Systems read nephelometrically |
| | Measurement Method | The SYNCHRON LX runs the reaction at 37°C and reads an endpoint at 340 nm; the ARRAY Systems run at 26.5°C and read at a peak rate at 450-550 nm |
| | Antigen Excess Checking | The SYNCHRON LX Reagents are designed so that high antigen concentration will not report in range; the ARRAY Systems add extra antibody to observe for additional activity. |
| | Sample Diluent | The SYNCHRON LX uses a tris buffer for sample dilution; the ARRAY Systems use a phosphate buffered diluent. |
| | Reagent Storage | The SYNCHRON LX reagents are stored refrigerated onboard the instrument; the ARRAY Systems reagents are stored at room temperature onboard the instrument/ refrigerated when not in use. |
| | Packaging | The SYNCHRON LX reagents are packaged in polystyrene cartridges; the ARRAY Systems reagents are in glass bottles. |
file: ig510k.sse
{3}
Beckman Instruments, Inc., Section 510(k) Notification
SYNCHRON LX™ Systems Immunoglobulin G, A, & M Reagents
Summary of Safety & Effectiveness
## 8.0 Summary of Performance Data:
The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method comparison, stability, and imprecision experiments that relate results obtained from the Beckman Immunoglobulin Reagents (IGG, IGA, IGM) to the SYNCHRON LX Systems Immunoglobulin Reagents (Ig-G, Ig-A, Ig-M).
### Method Comparison Study Results
#### SYNCHRON LX Systems Ig-G, Ig-A, & Ig-M Reagents
| Reagent | Slope | Intercept | r | n | Predicate Method |
| --- | --- | --- | --- | --- | --- |
| SYNCHRON LX Immunoglobulin G Reagent (Ig-G) | 0.928 | 72.7 | 0.9970 | 82 | Beckman's Immunoglobulin IGG Reagent |
| SYNCHRON LX Immunoglobulin A Reagent (Ig-A) | 0.924 | 0.68 | 0.9962 | 78 | Beckman's Immunoglobulin IGA Reagent |
| SYNCHRON LX Immunoglobulin M Reagent (Ig-M) | 1.051 | -17.19 | 0.9914 | 72 | Beckman's Immunoglobulin IGM Reagent |
### Estimated Within-Run Imprecision
| Sample | Mean (mg/dL) | S.D. (mg/dL) | %C.V. | N |
| --- | --- | --- | --- | --- |
| Immunoglobulin G (Ig-G) | | | | |
| Level 1 | 527 | 6.9 | 1.3 | 80 |
| Level 2 | 1103 | 16.1 | 1.5 | 80 |
| Sample | Mean (mg/dL) | S.D. (mg/dL) | %C.V. | N |
| --- | --- | --- | --- | --- |
| Immunoglobulin A (Ig-A) | | | | |
| Level 1 | 109 | 1.8 | 1.7 | 80 |
| Level 2 | 242 | 4.0 | 1.7 | 80 |
| Sample | Mean (mg/dL) | S.D. (mg/dL) | %C.V. | N |
| --- | --- | --- | --- | --- |
| Immunoglobulin M (Ig-M) Reagent | | | | |
| Level 1 | 46.6 | 1.8 | 3.9 | 80 |
| Level 2 | 103.5 | 2.3 | 2.2 | 80 |
file: ig510k.sse
{4}
Beckman Instruments, Inc., Section 510(k) Notification
SYNCHRON LX™ Systems Immunoglobulin G, A, & M Reagents
Summary of Safety & Effectiveness
## Stability Study Results
| Reagent | Product Claim |
| --- | --- |
| SYNCHRON LX Systems
Immunoglobulin G, A, & M Reagents | 24 month shelf-life
14 day calibration stability
60 days on-board stability |
This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.
file: ig510k.sse
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.