NITRA-TOUCH POWDER-FREE NITRILE MEDICAL EXAMINATION GLOVE

K965095 · Ansell Edmont Industrial, Inc. · LZA · Apr 22, 1997 · General Hospital

Device Facts

Record IDK965095
Device NameNITRA-TOUCH POWDER-FREE NITRILE MEDICAL EXAMINATION GLOVE
ApplicantAnsell Edmont Industrial, Inc.
Product CodeLZA · General Hospital
Decision DateApr 22, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6250
Device ClassClass 1

Indications for Use

Nitra-Touch™ examination gloves are disposable device intended for medical purposes that is worn on the examiners hand to prevent contamination between patient and examiner.

Device Story

Nitra-Touch™ is a disposable nitrile examination glove; worn by healthcare examiners to prevent cross-contamination between patient and examiner. Device provides barrier protection; features high puncture resistance (8x greater than vinyl, 2x greater than competitive nitrile) and chemical resistance to glutaraldehyde. Used in clinical settings; operated by healthcare professionals. Output is physical barrier protection; affects clinical decision-making by ensuring sterile/clean environment during patient examinations.

Clinical Evidence

No clinical data. Bench testing only. Performance verified via ASTM D 3578 (dimensions, physical properties), ASTM D 5151 (freedom from holes), ASTM F739-91 (glutaraldehyde resistance), and biocompatibility studies (primary skin irritation in rabbits, guinea pig sensitization).

Technological Characteristics

Nitrile film material. Dimensions and physical properties meet ASTM D 3578 (except ultimate elongation: 500% original, 400% aged). Puncture resistance 1625 lbf/in (ASTM D 120-87). Tensile strength 17.1 MPa. Powder-free (<2 mg residue). Biocompatible.

Indications for Use

Indicated for use as a patient examination glove to prevent contamination between patient and examiner.

Regulatory Classification

Identification

A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0} Ansell Edmont Industrial Ansell Edmont Industrial Inc. 400 Northridge Rd. Suite 800 Atlanta, Georgia 30350 Telephone: (770) 641-1900 FAX: (770) 641-1988 K965055 Nitra-Touch™ Checklist Section 21.0 [1] 510 (k) Summary [2] Ansell Edmont Industrial, Inc. 400 Northridge Road, Suite 800 Atlanta, GA 30350 APR 22 1997 Telephone: 770-641-1900 Fax: 770-641-1988 Contact: Mike W. Hagans Telephone: 614-623-3595 Fax: 614-623-3515 December 12, 1996 [3] Trade Name: Nitra-Touch™ Common Name: Exam Gloves Classification Name: Patient Examination Glove [4] Nitra-Touch™ examination gloves, meet all of the requirements of ASTM D 3578 with the following variation. [5] Nitra-Touch™ examination gloves meet all the current specifications for ASTM D 3578 Rubber Examination Gloves except for the ultimate elongation percentage before and after aging. [6] Nitra-Touch™ examination gloves are disposable device intended for medical purposes that is worn on the examiners hand to prevent contamination between patient and examiner. [7] Nitra-Touch™ examination gloves are summarized with the following technological characteristics compared to ASTM or equivalent standards. | Characteristics | Standard | | --- | --- | | Dimensions | Meets ASTM D 3578 | | Physical Properties | Meets ASTM D 3578 | | Except Ultimate Elongation meets specifications described under Attachment II | Original Ultimate Elongation 500% minimum Aged Ultimate Elongation 400% minimum | Revised 2/21/97 {1} Ansell Edmont Industrial Ansell Edmont Industrial Inc. 400 Northridge Rd. Suite 800 Atlanta, Georgia 30350 Telephone: (770) 641-1900 FAX: (770) 641-1988 Nitra-Touch™ Freedom from holes Meets ASTM D 3578 Meets ASTM D 5151 Powder-Free Meets described test in Attachment V Not more than 2 mg residue by mass. Tensile Strength 60% higher than vinyl (synthetic) examination gloves Puncture resistance of Nitra-Touch™ is eight (8) times greater than vinyl and nearly twice that of competitive nitrile gloves Nitrile film provides excellent chemical resistance to glutaraldehyde. Tensile Strength MPa ASTM D 5250 ASTM D 3578 Vinyl = 10.5 Nitra-Touch™ = 17.1 Puncture lbf/in ASTM D 120-87 Nitra-Touch™ = 1625 Blue Nitrile = 929 Vinyl (Synthetic) = 184 Glutaraldehyde Resistance ASTM T F739-91 Source: TRI/Environmental, Inc. Breakthrough Time = &gt;480 Permeation Rate = &lt;0.013 µg/cm²* min Biocompatibility Primary Skin Irritation in Rabbits Guinea Pig Sensitization Passes Passes [8] The performance test data of the non clinical tests are the same as mentioned immediately above. [9] Clinical data is not needed for medical gloves or for most devices cleared by the 510(k) process. [10] It is concluded that Nitra-Touch™ examination gloves are as safe, as effective, and perform as well as the glove performance standards referenced in Section 7 above and therefore meet: ASTM listed standards, FDA hole requirements, and labeling claims for the product. [11] This summary will include any other information reasonably deemed necessary by The FDA. Revised 2/21/97
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...