Monoclonal mouse anti-human Ki-1 antigen, CD30, clone Ber-H2, is intended for laboratory use to qualitatively identify by light microscopy the lymphocyte activation-associated antigen designated CD30 in acetone fixed, frozen and formalin or B5 fixed, paraffin-embedded tissues. Ber-H2 specifically binds to antigens located on the cell membrane, as well as in the Golgi region of activated lymphoid cells. Positive results aid in the classification of normal and abnormal cells and tissues and serve as an adjunct to conventional histopathology.
Device Story
Immunohistochemical staining reagent; utilizes monoclonal mouse anti-human Ki-1 antigen (CD30, Clone Ber-H2) antibody. Input: acetone-fixed frozen or formalin/B5-fixed paraffin-embedded tissue sections. Principle: antibody binds specifically to CD30 antigen on cell membranes and Golgi regions of activated lymphoid cells. Output: visual staining pattern observable via light microscopy. Used in clinical pathology laboratories by pathologists/technicians. Healthcare providers use staining patterns (membranous/cytoplasmic) as an adjunct to conventional histopathology to classify normal/abnormal cells and aid differential diagnosis of lymphomas (ALCL, Hodgkin's). Benefits: assists in identifying neoplastic cells in samples with low antigen density or heavy macrophage infiltration.
Clinical Evidence
Bench testing only. Evaluated using a panel of normal tissues (formalin-fixed, paraffin-embedded) and three spinal cord specimens. Reproducibility testing performed on eight serial sections from three Hodgkin's Lymphoma blocks; intra- and inter-run testing showed consistent results. Literature review provided for clinical performance in ALCL (100% positivity in 139 cases), Hodgkin's disease (89% positivity in paraffin sections), and cutaneous lymphoid infiltrates.
Technological Characteristics
Monoclonal mouse anti-human antibody (Clone Ber-H2). Supplied in Tris-HCl buffer with sodium azide. Formats: concentrated (M0751) or ready-to-use (N1558) with negative control reagent. Detection via standard immunohistochemical systems (e.g., LSAB2 Peroxidase/Alkaline Phosphatase). Staining principle: antigen-antibody binding visualized by light microscopy. Storage: 2-8°C.
Indications for Use
Indicated for use as part of an antibody panel to aid in the differential diagnosis of large anaplastic cells of undetermined origin. Stains cell membranes of most anaplastic large cell lymphomas (ALCL) and cell membranes/cytoplasm of most Hodgkin's lymphomas. Aids in assessment of cutaneous lymphoid infiltrates with large atypical cells. Not diagnostic for ALCL alone due to limited specificity/sensitivity.
Regulatory Classification
Identification
An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.
{0}
K965022
NOV - 6 1997
# 510(k) Summary
Submitter: DAKO Corporation
6392 Via Real
Carpinteria, CA 93013
(805) 566-6655
Contact: Gretchen M. Murray, Ph.D., RAC Regulatory Affairs Asst. Manager
Date Summary Prepared: July 28, 1997
Device name: 1) Monoclonal Mouse Anti-Human Ki-1 antigen, CD30, Clone Ber-H2 (Product Code No. M0751)
Primary Antibody Device Classification: 2) Ready-To-Use Monoclonal Mouse Anti-Human Ki-1 antigen, CD30, Clone Ber-H2 and Negative Control Reagent (Product Code No. N1558)
Class I or Class II has been proposed for immunohistochemical staining reagents.
Panel: The proposed device classification is under the Hematology and Pathology devices panel, Division of Clinical Laboratory Devices.
Predicate Device: DAKO Monoclonal Mouse Anti-Human LCA (Product Code No. M0701, FDA K896918/B)
Device Description:
1) Monoclonal Mouse Anti-Human Ki-1 antigen, CD30, Clone Ber-H2 (Product Code No. M0751) is a mouse anti-human monoclonal antibody produced as a tissue culture supernatant. The antibody is supplied in 0.05M Tris-HCl buffer, pH 7.2, containing 0.015M sodium azide. (1 ml total volume)
2) Ready To Use Monoclonal Mouse Anti-Human Ki-1 antigen, CD30, Clone Ber-H2 Primary Antibody and Negative Control Reagent (Product Code No. N1558) consists of a mouse anti-human monoclonal antibody produced as a tissue culture supernatant and pre-diluted in 0.05M Tris-HCl buffer, pH 7.6, containing carrier protein and 0.015M sodium azide (7mL total volume). The primary antibody is packaged with a negative control reagent consisting of fetal calf serum in 0.05M Tris-HCl buffer, pH 7.6, containing carrier protein and 0.015M sodium azide (5 ml total volume). The primary antibody and the negative control reagent contain equivalent total protein concentrations.
Intended Use: For In Vitro Diagnostic Use
Monoclonal mouse anti-human Ki-1 antigen, CD30, clone Ber-H2, is intended for laboratory use to qualitatively identify by light microscopy the lymphocyte activation-associated antigen designated CD30 in acetone fixed, frozen and formalin or B5 fixed, paraffin-embedded tissues. Ber-H2 specifically binds to antigens located on the cell membrane, as well as in the Golgi region of activated lymphoid cells. Positive results aid in the classification of normal and abnormal cells and tissues and serve as an adjunct to conventional histopathology.
Indicated Use: Monoclonal mouse anti-human Ki-1 antigen, CD30, Clone Ber-H2, may be used as one member of a panel of antibodies to aid in the differential diagnosis of large anaplastic cells of undetermined origin. This antibody stains cell membranes of most cases of anaplastic large cell lymphomas (ALCL), often called Ki-1 lymphomas. It also stains cell membranes and/or cytoplasm of most Hodgkin's lymphomas. Ber-H2 is a valuable aid in the assessment of cutaneous lymphoid infiltrates, particularly when large atypical cells are
2
{1}
present. Most nonhematolymphoid neoplasms are negative with Ber-H2, although there are several significant exceptions.
The staining pattern is most often described as strong membranous and weaker cytoplasmic, specifically paranuclear dot-positivity of the Golgi region. Membrane positivity is seen with embryonal carcinomas, and weak, diffuse cytoplasmic positivity is seen in pancreatic and salivary gland carcinomas.³ The weak, diffuse cytoplasmic staining is considered to be unexpected, non-specific labeling. It may be reduced or removed by changing the protease pretreatment of the paraffin sections and/or further dilution of the Ber-H2 antibody.³
The clinical interpretation of any positive staining or its absence should be complemented by morphological and histological studies with proper controls. Evaluations should be made within the context of the patient's clinical history and other diagnostic tests by a qualified individual having knowledge of all the potential antibody reactivities.
## Experimental Data
### Normal Tissue Testing:
The required panel of normal tissues was tested with this antibody as specified in the 3/28/95 draft of Guidance for Submissions of Immunohistochemistry Applications to the FDA. Additionally tested were three specimens of spinal cord. All tissues were formalin fixed and paraffin embedded. Staining was performed using the DAKO LSAB®2 Peroxidase (Code No. K0677) and LSAB®2 Alkaline Phosphatase (Code No. K0676) kit systems.
Normal tissues exhibiting positive staining with Ber-H2 included the following: a subpopulation of myelocytes in bone marrow, brain (Purkinje cells in cerebellum and ganglion cells in cerebrum), motor neurons and axons in spinal cord, and peri/interfollicular lymphocytes in tonsil. The other tissues were negative for specific staining.
### Reproducibility Testing:
Eight serial sections from each of three different paraffin embedded blocks of Hodgkin’s Lymphoma (pre-screened for low antigen density) were collected for testing. Testing was performed as follows:
- **Intra-run reproducibility**: Following the standard DAKO LSAB®2 Peroxidase Kit protocol (Code No. K0677), three slides from each tissue block were stained with Ready-To-Use DAKO® Monoclonal mouse anti-human Ki-1 antigen, CD30, Clone Ber-H2 (Code No. N1558). Concurrently, one slide from each block was stained with the supplied negative control reagent.
- **Inter-run reproducibility**: Staining one slide from each tissue block, the above procedure was repeated on two additional days. Concurrently, one slide from each block was stained with the supplied negative control reagent.
Reproducibility experiments with Ber-H2 yielded consistent results with intra- and inter-run testing. Consistent test conditions were maintained throughout the study and reagents were stored at 2-8°C between test runs.
Published Immunoreactivity
{2}
# Clone Ber-H2 Characteristics
Monoclonal mouse anti-human Ki-1 antigen, CD30, Clone Ber-H2 (Ber-H2) has been shown to react with a cell membrane associated glycoprotein with an apparent molecular mass of 105 to 120 kD.¹⁻⁵ The extracellular domain of the CD30 antigen has significant homology to that of members of the tumor necrosis factor(TNF)/nerve growth factor (NGF) receptor superfamily.⁵,⁶ This evidence suggest that CD30 antigen acts as a receptor whose ligand is a cytokine.⁵ It is hypothesized that the CD30 antigen is associated with lymphoid activation because its presence is induced on B and T lymphocytes in vitro by a number of stimuli, including viruses and lectins.⁵ Positive staining with Ber-H2 is typically membranous, often with cytoplasmic, and/or dot-like staining in the Golgi region.³⁻⁵ Ber-H2 reacts with an epitope that is retained after routine fixation for paraffin embedded tissue sections unlike Ki-1, the first CD30 antibody recognized. Ber-H2 stains both paraffin-embedded and frozen tissue sections, and demonstrates greater intensity than Ki-1 on frozen tissues. The Fourth International Workshop on Leukocyte Differentiation Antigens (Vienna, 1989) recognized Ber-H2 as an antibody directed against the lymphocyte activation-associated antigen designated CD30, also known as the Ki-1 antigen.¹,³,⁷
# Clone Ber-H2 Positivity in Normal Tissues
In normal tissues, Ber-H2 has been characterized as reacting with scattered large activated B and T lymphoid cells localized around and between lymph follicles and at the margin of germinal centers. Some plasma cells also exhibit diffuse cytoplasmic staining in paraffin tissues only. Positive staining of few scattered blasts in the lymphoid tissue of the intestine and in the medullary part of the thymus has also been reported.¹,³,⁵,⁸ Unexpected cytoplasmic staining of normal tissues is described as follows: In bone marrow, cytoplasmic staining has been observed in late stage erythroid and myeloid cells.⁹ Diffuse cytoplasmic staining has also been observed in exocrine pancreas cells. In paraffin embedded sections only, diffuse cytoplasmic staining with Ber-H2 has been observed in a proportion of cerebral cortical neurons, and Purkinje cells of the cerebellum. Fibroblasts have been reported to stain positively in paraffin sections also.¹⁰ The diffuse weak cytoplasmic staining is controversial for specificity with Ber-H2. Some have reported that this staining is of no diagnostic relevance.³,⁵,⁶
Resting peripheral lymphocytes, monocytes, interdigitating cells, follicular dendritic cells, and macrophages were initially reported to be non-reactive with Ber-H2,³,⁵ but subsequent rare Ber-H2 staining of follicular dendritic cells has been observed.⁸ Activated macrophages, including those observed in frozen tissues from granulomatous inflammation, sarcoidosis, toxoplasmosis, and cat-scratch fever as well as in paraffin embedded tissues from one case of miliary tuberculosis also demonstrated Ber-H2 positivity.⁸,¹¹⁻¹²
# Anaplastic Large Cell Lymphomas
Ber-H2 stains most cases of anaplastic large cell lymphomas (ALCL). In a summary of five series of morphologically and immunologically characterized non-Hodgkin’s lymphoma (NHLs), Chang, et al,⁵ reported 100% Ber-H2 positivity in 139 cases of ALCL. ALCLs may account for 3 to 8% of all NHLs.⁵,¹³ Virtually all of the neoplastic cells display membrane staining, often accompanied by paranuclear staining. Because cell membrane staining is seen in other diseases, including all major categories of NHL except lymphoblastic lymphoma, Ber-H2 positivity is not diagnostic of ALCL.³,⁵,⁸
# Hodgkin’s Lymphoma
The CD30 epitope is expressed in Hodgkin (H) and Reed-Sternberg (RS) cells in all types of Hodgkin’s Disease (HD). Ber-H2 staining is particularly useful for demonstrating Hodgkin’s cells in samples with few neoplastic cells or those with heavy infiltration of
{3}
macrophages.⁸ Staining is usually stronger in frozen sections than in paraffin sections.⁵ Ree et al.¹⁰ reported the following staining patterns in paraffin embedded specimens from 89 undisputed Hodgkin's disease patients: 9/89 (10%) were negative, 20/89 (22%) showed membrane staining alone or in combination with paranuclear or cytoplasmic staining and 60/89 (67%) had cytoplasmic and/or paranuclear staining without membrane staining.
The incidence of positivity depends on the type of tissue processing. In frozen and plastic sections, virtually all cases have been reported to be Ber-H2 positive.⁵ In a summary of a series of ten studies with paraffin sections, only 89% of cases were positive. In addition, the percentage of Ber-H2 positive results with paraffin embedded tissue also varied among types of HD as follows: 25% (14/56) of lymphocyte predominance, 87% (297/340) of nodular sclerosis, 91% (250/274) of mixed cellularity, 86% (24/28) of lymphocyte depletion, and 75% (15/20) of unclassified cases.⁵
## Cutaneous Lymphoid Infiltrates
Ber-H2 positivity is also a valuable aid in the assessment of cutaneous lymphoid infiltrates, when used with a panel of antibodies. Ralfkiaeer, et al.¹⁴ stained specimens from 115 patients with benign dermatoses, pre- or pseudo-malignant disorders, and malignant cutaneous lymphomas. They reported that in 100% of 6 cases of lymphomatoid papulosis, a peripheral T-cell lymphoma, varying numbers of scattered large polymorphic lymphoid cells consistently expressed Ber-H2. The reactivity may be seen as membranous and cytoplasmic.⁶ However, other peripheral T-cell cutaneous lymphomas and 3/14 (21.4%) of cases of mycosis fungoides Sezary syndrome (labeling confined to large cells present) were positive. The lymphoid cells in all other specimens were Ber-H2 negative. While these data suggest that Ber-H2 cannot distinguish cutaneous NHL from lymphomatoid papulosis, they may aid in differentiating the latter from other types of Ber-H2 negative pre- or pseudo-malignant cutaneous infiltrates.¹⁴
## Non-Hematolymphoid Neoplasms
Although generally not found in non-hematolymphoid neoplasms, Ber-H2 positivity has been reported in both frozen and paraffin embedded tissue in salivary gland carcinomas 2/3 (67%), embryonal carcinomas 8/10 (80%), in the embryonal component of mixed germ cell tumors 4/4 (100%) and in 4/21 (19%) cases of testicular germ cell tumors.³,¹⁵,¹⁶ Ber-H2 positivity was also reported in one case of interdigitating cell sarcoma and in 3/4 (75%) cases of follicular dendritic cell sarcoma.⁸
In summary, because of limited specificity and sensitivity, Ber-H2 positivity alone is not diagnostic for ALCL.¹⁷ In cases of HD, the incidence of positivity depends on the type of tissue processing. In frozen and plastic sections, virtually all cases have been reported to be Ber-H2 positive. In paraffin sections only 89% of cases were positive. Similarly, only 25% of cases of lymphocyte predominance HD stained positively in paraffin sections and staining was generally weaker and confined to the cell membrane when compared to that seen in classical HD.⁵
## References
1. Schwarting R, Stein H. Cluster report: CD30. In: Knapp W, Dorken B, Gilks WR, et al., eds. Leukocyte typing IV. White Cell Differentiation Antigens. New York: Oxford University Press, 1989: 419
2. Froese P, Lemke H, Gerdes J, Havsteen B, Schwarting R, Hansen H, Stein H. Biochemical characterization and biosynthesis of the Ki-1 antigen in Hodgkin's derived and virus-transformed human and T lymphoid cell lines. J Immunol 1987; 139: 208 i
{4}
6
3. Schwarting R, Gerdes J, Durkop H, Falini B, Pileri S, Stein H. Ber-H2: a new anti-Ki-1 (CD30) monoclonal antibody directed at a formol-resistant epitope. Blood 1989; 74: 1678
4. Rohde D, Hansen H, Hafner M, Lange H, Mielke V, Hannsman M-L, Lemke H. Cellular localizations and processing of the two molecular forms of the Hodgkin-associated Ki-1 (CD30) antigen. The protein kinase Ki-1/57 occurs in the nucleus. Am J Pathol 1992; 140: 473
5. Chang KL, Arber DA, Weiss LM. CD30: A Review. Applied Immunohistochemistry 1993; 1(4): 244
6. Barclay NA, Birekland ML, Brown MH, Beyers AD, Davis SJ, Somoza C, Williams AF. The Leukocyte Antigen Facts Book. Academic Press, San Diego, 1993.
7. Sakthivel R, Christensson B, Ehlin-Henriksson B, Klein G. Activation antigen expression in EBV-negative Burkitt lymphoma (BL), EBV-converted BL (EBV-BL), and lymphoblastoid cell lines (LCL). In: Knapp W, Dorken B, Gilks WR, et al., eds. Leukocyte typing IV. White Cell Differentiation Antigens. New York: Oxford University Press, 1989: 479
8. Pallesen G. The diagnostic significance of the CD30(Ki-1) antigen. Histopathology 1990;16: 409
9. van der Valk P, Mullink H, Huijgens PC, Tadema TM, Vos W, Meijer CJLM. Immunohistochemistry in bone marrow diagnosis. Value of a panel of monoclonal antibodies on routinely processed bone marrow biopsies. Am J Surg Pathol 1989; 13:97
10. Ree HJ, Neiman RS, Martin AW, Dallenbach F, Stein H. Paraffin section markers for Reed-Sternberg cells. A comparative study of peanut agglutinin, Leu-M1, LN-2, and Ber-H2. Cancer 1989; 63: 2030
11. Epstein ML, Windebank KP, Burt AD, Thomas L, Cant AJ. CD30 expression by peripheral blood monocytes and hepatic macrophages in a child with miliary tuberculosis. J Clin Pathol 1992; 45: 638
12. O'Laughlin S, Braverman M, Smith-Jefferies M, Buckley P. Macrophages (histiocytes) in various reactive and inflammatory conditions express different antigenic phenotypes. Hum Pathol 1992; 23: 1410
13. Stein H, Mason DY, Gerdes J, O'Connor N, Wainscoat J, Pallesen G, Gatter K, Falini B, Delsol G, Lemke H, Schwarting R, Lennert K. The expression of the Hodgkin's disease associated antigen Ki-1 in reactive and neoplastic lymphoid tissues. Evidence that Reed-Sternberg cells and histiocytic malignancies are derived from activated lymphoid cells. Blood 1985; 66: 848
14. Ralfkaier E, Bosq J, Gatter KC, Schwarting R, Gerdes J, Stein H, Mason DY. Expression of a Hodgkin and Reed-Sternberg cell associated antigen(Ki-1) in cutaneous lymphoid infiltrates. Arch Dermatol Res 1987; 279: 285
15. Pallesen G, Hamilton-Dutoit SJ. Ki-1 (CD30) antigen is regularly expressed by tumor cells of embryonal carcinoma. Am J Pathol 1988; 133: 446
16. Hittmair A, Rogatsch H, Hobisch A, Mikuz G, and Feichtinger H. CD30 Expression in Seminoma. Human Pathology 1996; 27(11): 1166
17. Falini B, Pileri S, Stein H, Dieneman D, Dallenbach F, Delsol G, Minelli O, Poggi S, Martelli M, Pallesen G, Palestro G. Variable expression of leukocyte common antigen (CD45) antigen in CD30 (Ki-1)-positive anaplastic large-cell lymphoma: Implications for the differential diagnosis between lymphoid and nonlymphoid malignancies. Hum Pathol 1990; 21: 624
{5}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
Gretchen M. Murray, Ph.D., RAC
Regulatory Affairs Assistant Manager
DAKO Corporation
6392 Via Real
Carpinteria, California 93013
NoV - 6 1997
Re: K965022
Trade Name: DAKO Corporation's Monoclonal Mouse Anti-Human Ki-1
Antigen, CD30, Clone Ber-H2 (Product Code No.
MO751)
Regulatory Class: II
Product Code: DEH
Dated: August 8, 1997
Received: August 11, 1997
Dear Dr. Murray:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
{6}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{7}
Page 1 of 1
510(k) Number (if known): K965022
Device Name: Monoclonal Mouse Anti-Human Ki-1 antigen, CD30, Clone Ber-H2 for Immunoenzymatic Staining (DAKO Code No. M0751)
Monoclonal Mouse Anti-Human Ki-1 antigen, CD30, Clone Ber-H2 Ready-to-Use Antibody and Negative Control for Immunoenzymatic Staining DAKO Code No. N1558)
## Indications For Use:
Monoclonal mouse anti-human Ki-1 antigen, CD30, Clone Ber-H2 (Ber-H2), may be used as one member of a panel of antibodies to aid in the differential diagnosis of large anaplastic cells of undetermined origin. This antibody stains cell membranes of most cases of anaplastic large cell lymphomas (ALCL), often called Ki-1 lymphomas. It also stains cell membranes and/or cytoplasm of most Hodgkin’s lymphomas. Ber-H2 is a valuable aid in the assessment of cutaneous lymphoid infiltrates, particularly when large atypical cells are present. Most nonhematolymphoid neoplasms are negative with Ber-H2, although there are several significant exceptions. Membrane positivity is seen with embryonal carcinomas, and weak, diffuse cytoplasmic positivity is seen in pancreatic and salivary gland carcinomas.
The staining pattern is most often described as strong membranous and weaker cytoplasmic, specifically paranuclear dot-positivity of the Golgi region. The weak, diffuse cytoplasmic staining is considered to be unexpected, non-specific labeling. It may be reduced or removed by changing the protease pretreatment of the paraffin sections and/or further dilution of the Ber-H2 antibody.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use ☑
Use ☐ (Per 21 CFR 801.109)
IVD Use ☐ (Per 21 CFR 801.119)

(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number
Over-The-Counter
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.