K964881 · Vas-Cath, Inc. · LIT · May 28, 1997 · Cardiovascular
Device Facts
Record ID
K964881
Device Name
ULTRAVERSE PTA CATHETER
Applicant
Vas-Cath, Inc.
Product Code
LIT · Cardiovascular
Decision Date
May 28, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Ultraverse™ Small Vessel PTA Balloon Dilatation Catheter is recommended for use in angioplasty of the peripheral vascular system. The catheter is recommended for use in the following vessels: renal, tibial, popliteal, femoral, and peroneal. The catheter is not for use in coronary arteries. The device is for single use only and must not be resterilized or reused.
Device Story
Dual-lumen PTA balloon dilatation catheter; designed for peripheral vascular angioplasty. Input: 0.016-inch guidewire. Operation: catheter inserted over guidewire; balloon inflated with contrast media to dilate stenotic vessels. Output: mechanical vessel dilation. Used in clinical settings by physicians. Radiopaque marker bands facilitate positioning under fluoroscopy. Benefits: restores blood flow in peripheral vessels. Single-use device.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included balloon burst strength, distensibility, inflation/deflation time, fatigue, joint strength, profile, introducer sheath compatibility, flow/pressure, trackability, and tip torque.
Indicated for percutaneous transluminal angioplasty (PTA) of the peripheral vascular system, specifically renal, tibial, popliteal, femoral, and peroneal vessels. Contraindicated for use in coronary arteries.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}
K964881
vas-cath.
INCORPORATED
MAY 28 1997
# 510(k) SUMMARY of SAFETY AND EFFECTIVENESS
## FOR THE
### ULTRAVERSE™ PTA BALLOON DILATATION CATHETERS
2380 Tedlo Street
Mississauga, Ontario
Canada, L5A 3V3
Tel. (905) 848-5800
Fax (905) 848-6638
A. Submitter Information: Vas-Cath Incorporated
Address: 2380 Tedlo Street
Mississauga, Ontario L5Z 3V3
Canada
Telephone Number: (905) 848-5800
Fax Number: (905) 848-6638
Contact Person: Dolores McGirr
Date of Preparation: August 23, 1996
B. Device Name: Ultraverse™ Angioplasty Catheter
3.5F Opti-Plast Catheter
Common/Usual Name: Peripheral Angioplasty Catheter
Classification Name: Surgical Vessel Dilator
C. Predicate Device Name: Ultraverse™ Angioplasty Catheter (K925485)
Trade Name: Ultraverse
D. Device Description:
The Vas-Cath Ultraverse™ PTA Catheter is a dual lumen catheter with a balloon mounted on its distal tip. One lumen accommodates the insertion guidewire and the second provides a channel for inflation on the balloon with contrast media. The catheters were designed to be used in conjunction with a 0.016 inch diameter guidewire. There are two radiopaque marker bands placed beneath the balloon to indicate its position within the vasculature. The balloon inflates to the stated diameter and length at an 8 atmosphere operating pressure.
E. Intended Use:
The Ultraverse™ Small Vessel Peripheral Balloon Dilatation Catheters are recommended for use in the peripheral vascular system, in the following vessels: renal, tibial, popliteal, femoral, and peroneal. This catheter is not for use in coronary arteries. The device is for single use only and must not be resterilized or reused.
F. Technological Characteristics Summary: Summarized in Table 1.
Ultraverse™ Catheter 510(k) Summary Page 1 of 2
{1}
vas-cath. INCORPORATED
TABLE 1 : COMPARISON OF DEVICES
| Characteristics | Predicate K925485 Vas-Cath Ultraverse™ PTA Catheter | Proposed New Sizes for Vas-Cath Ultraverse™ PTA Catheter |
| --- | --- | --- |
| Similarities | | |
| Intended Use | PTA of renal, tibial, popliteal, femoral, and peroneal vessels | PTA of renal, tibial, popliteal, femoral, and peroneal vessels |
| Catheter Shaft Material | Nylon | Nylon |
| Catheter Shaft Diameter | 3.5 French | 3.5 French |
| Inflated Balloon Diameters | 2, 3, 4, 5 mm | 2, 3, 4, 5 mm |
| Balloon Material | Nylon | Nylon |
| Maximum Recommended Inflation Pressure | 8 atmospheres | 8 atmospheres |
| Guidewire Capability | 0.016 inch diameter | 0.016 inch diameter |
| Shaft configuration | Double lumen | Double lumen |
| Radiopaque | Yes | Yes |
| Marker Bands | Yes | Yes |
| Balloon Leg Length | 7 cm | 7 cm |
| Distal Leg Length | 5.5 cm | 5.5 cm |
| Rated Burst Pressures | 2mm by 2cm is 15atm
3mm by 2cm is 15atm
4mm by 2cm is 12 atm
5mm by 2cm is 12 atm | 2mm by 6cm is 15 atm
3mm by 6cm is 15 atm
4mm by 6cm is 12 atm
5mm by 3,4,5 or 6cm is 10 atm |
| Differences | | |
| Inflated Balloon Length | 2 cm | 3, 4, 5, 6 cm |
| Catheter Shaft Length | 75, 100 cm | 75 - 150 cm |
## G. Performance Data
The design, materials and manufacturing processes for the predicate and proposed devices are the same. The additional balloon and shaft sizes proposed in this submission were tested for the following characteristics: balloon minimum burst strength, balloon distensibility, balloon inflation/deflation time, balloon fatigue, joint strength, catheter shaft and balloon profile, introducer sheath compatibility, catheter flow/maximum pressure, trackability/ deflatability, and tip torque. The proposed product is substantially equivalent to the current Ultraverse™ Catheters marketed under 510(k) concurrence K925485.
Ultraverse™ Catheter 510(k) Summary Page 2 of 2
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
MAY 28 1997
Ms. Dolores McGirr
Vas-Cath, Inc.
2380 Tedlo Street
Canada, L5A 3V3
Re: K964881
Ultraverse™ 3.5F PTA Catheters
Regulatory Class: II (two)
Product Code: 74 LIT
Dated: April 2, 1997
Received: April 15, 1997
Dear Ms. McGirr:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}
Page 2 - Ms. Dolores McGirr
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}
# SECTION 1. D INDICATION(S) STATEMENT*
I state in my capacity as Manager of Quality Assurance and Regulatory Affairs of Vas-Cath Inc. that this premarket notification 510(k) for the Ultraverse™ 3.5 French Angioplasty Catheter is indicated for the following.
The Ultraverse™ Small Vessel PTA Balloon Dilatation Catheter is recommended for use in angioplasty of the peripheral vascular system. The catheter is recommended for use in the following vessels: renal, tibial, popliteal, femoral, and peroneal. The catheter is not for use in coronary arteries.

Janet Shaw
Printed Name of Submitter
Date
96/11/15
*Suggested language and format to meet the requirements of sections 513(i) of the Federal Food, Drug, and Cosmetic Act, as amended, and sections 807.92(a)(5) and 801.4 of the code of Federal Regulations, Title 21.
# CONCURRENCE OF OFFICE OF DEVICE EVALUATION
K964881
510(k) Number

Division Sign-Off
Office of Device Evaluation
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.