MARDX ENA IGG IMMUNOSTRIPE TEST SYSTEM / MARDX ENA IGG MARSTRIPE TEST SYSTEM
Applicant
Mardx Diagnostics, Inc.
Product Code
LLL · Immunology
Decision Date
Apr 8, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.5100
Device Class
Class 2
Indications for Use
The MarDx ENA IgG ImmunoStripe Test System is intended for use in testing human serum for the presence of human IgG to extractable nuclear antigens, as an aid in the diagnosis of autoimmune disease, such as systemic lupus erythematosus, Sjogren syndrome, scleroderma, and myositis.
Device Story
Enzyme-linked immunosorbent assay (EIA) for detection of IgG antibodies to extractable nuclear antigens (SSA, SSB, Sm, RNP, Jo-1, Scl-70) in human serum; purified antigens immobilized on nitrocellulose strips; patient serum added to strip; if specific antibodies present, they bind to antigens; enzyme-labeled anti-human IgG conjugate added; substrate solution added; enzyme/antibody complex causes color precipitation at specific locations on strip; visual interpretation of color change indicates presence of specific antibodies; used in clinical laboratory settings; aids healthcare providers in diagnosing autoimmune conditions; provides qualitative diagnostic information.
Clinical Evidence
Clinical study evaluated 264 human serum samples (139 normal, 125 autoimmune disease patients). Compared MarDx ENA IgG Immunostripe against INOVA EIA and MarDx 6-in-1 EIA. Combined sensitivity 95%, specificity 100%. Individual antigen sensitivity ranged from 89% (RNP) to 100% (Sm, Jo-1, Scl-70). Precision study (n=7 samples, 8 replicates across 3 assays) showed 100% precision. Cross-reactivity study (n=7 dsDNA positive samples) showed 0% cross-reactivity.
Indicated for testing human serum for IgG antibodies to extractable nuclear antigens (SSA, SSB, Sm, RNP, Jo-1, Scl-70) as an aid in diagnosing autoimmune diseases including systemic lupus erythematosus, Sjogren syndrome, scleroderma, and myositis.
Regulatory Classification
Identification
An antinuclear antibody immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoimmune antibodies in serum, other body fluids, and tissues that react with cellular nuclear constituents (molecules present in the nucleus of a cell, such as ribonucleic acid, deoxyribonucleic acid, or nuclear proteins). The measurements aid in the diagnosis of systemic lupus erythematosus (a multisystem autoimmune disease in which antibodies attack the victim's own tissues), hepatitis (a liver disease), rheumatoid arthritis, Sjögren's syndrome (arthritis with inflammation of the eye, eyelid, and salivary glands), and systemic sclerosis (chronic hardening and shrinking of many body tissues).
Predicate Devices
INOVA ENA EIA (SSA, SSB, Sm, RNP, Jo-1, Scl-70)
MarDx ENA EIA 6 in 1 Test System
Submission Summary (Full Text)
{0}
ATTACHMENT 1 - 510K Summary
APR - 8 1997
# 510(k) Summary of Safety and Effectiveness Information
## MarDx ENA IgG Immunostripe Test System
K964787
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and proposed 21 CFR Part 807.92.
## Identification of Predicate Devices:
The MarDx ENA IgG Immunostripe Test System is substantially equivalent to the INOVA ENA EIA's for SSA, SSB, Sm, RNP, Jo-1, Scl-70 and the MarDx ENA EIA 6 in 1 Test System for the detection of immunoglobulin to SSA, SSB, Sm, RNP, Jo-1, Scl-70 in human serum.
## Description of New Device:
The MarDx ENA IgG Immunostripe Test System is an enzyme-linked immunosorbent assay (EIA) for the detection of IgG to extractable nuclear antigens in human serum.
## Statement of Intended Use:
The MarDx ENA IgG ImmunoStripe Test System is intended for use in testing human serum for the presence of human IgG to extractable nuclear antigens, as an aid in the diagnosis of autoimmune disease, such as systemic lupus erythematosus, Sjogren syndrome, scleroderma, and myositis.
## Technological Characteristics of the Device:
The MarDx ENA IgG Immunostripe Test System is an enzyme linked immunosorbent assay to detect IgG, to SSA, SSB, Sm, RNP, Jo-1, Scl-70. Purified antigens are individually attached to specific locations on a solid phase support, nitrocellulose. Diluted test sera is added to a well containing the Immunostripe strip. If the antibodies are present that recognize an antigen, they will bind to the antigen present on the strip. After incubation the nitrocellulose strip is washed to remove unbound antibody. An enzyme labeled antihuman IgG is added to each strip. If antibody is present the conjugate will bind to the antibody attached to the specific antigens. The strip is washed to remove unbound conjugate and a substrate solution is then added. If the enzyme/antibody complex is present, the substrate will undergo a precipitation and color change. After an incubation period, the reaction is stopped and the presence of precipitated substrate is visualized at specific locations on the strip. The presence of a colored precipitation at pre-determined locations on the Immunostripe strip is an indirect measurement of specific antibody for SSA, SSB, Sm, RNP, Jo-1, Scl-70 in the patient specimen.
## Description and Conclusions of the Clinical Studies.
The MarDx ENA IgG Immunostripe Test System is substantially equivalent to the INOVA ENA EIA's for SSA, SSB, Sm, RNP, Jo-1, Scl-70 and the MarDx ENA EIA 6 in 1 Test System for the detection of immunoglobulin to SSA, SSB, Sm, RNP, Jo-1, Scl-70 in human serum. Equivalence is demonstrated by the following comparative results:
{1}
Two hundred and sixty four sera were tested by EIA (INOVA ELISA for ENA antibodies), and the MarDx 6 in 1 ENA EIA, to determine the relative sensitivity and specificity of the MarDx ENA IgG Immunostripe test kit. One hundred and thirty-nine of the sera were from normal individuals which should not have antibodies to ENA antigens. One hundred and twenty five sera were from patients thought to have autoimmune disease and submitted as blind specimens for ENA antigen detection. The MarDx and INOVA assays were performed by the R&D laboratory at MarDx following manufacturer's package insert. Assay sensitivity ranged from 89 to 100%, specificity was 100%. The raw data from the study is prepared in Attachment 5. The data is summarized in Tables 1-11.
Table 1
Specificity of the MarDx ENA IgG Immunostripe Test System relative to the INOVA EIA Test Systems
| | | INOVA EIA | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| MarDx | Positive | 0 | 0 | 0 |
| Immunostripe | | | | |
| | Negative | 0 | 40 | 40 |
| Total | | 0 | 40 | 40 |
Specificity (40/40) = 100%
Table 2
Specificity of the MarDx ENA IgG Immunostripe Test System relative to the MarDx 6 in 1 ENA EIA Test System
| | | MarDx 6 in 1 ENA EIA | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| MarDx | Positive | 0 | 0 | 0 |
| Immunostripe | | | | |
| | Negative | 0 | 139 | 139 |
| Total | | 0 | 139 | 139 |
Specificity (139/139) = 100%
Table 3
Sensitivity of the MarDx ENA IgG Immunostripe Test System relative to the INOVA SSA EIA Test System
| Antigen: SSA | | INOVA SSA EIA | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| MarDx | Positive | 35 | 0 | 35 |
| Immunostripe | | | | |
| | Negative | 2 | 0 | 2 |
| Total | | 37 | 0 | 37 |
SSA Sensitivity (35/37) = 95%
{2}
Table 4
Sensitivity of the MarDx ENA IgG Immunostripe Test System
relative to the INOVA SSB EIA Test System
| Antigen: SSB | | INOVA SSB EIA | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| MarDx | Positive | 16 | 0 | 16 |
| Immunostripe | | | | |
| | Negative | 1 | 0 | 1 |
| Total | | 17 | 0 | 17 |
SSB Sensitivity (16/17) = 94%
Table 5
Sensitivity of the MarDx ENA IgG Immunostripe Test System
relative to the INOVA Sm EIA Test System
| Antigen: Sm | | INOVA Sm EIA | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| MarDx | Positive | 20 | 1 | 21 |
| Immunostripe | | | | |
| | Negative | 0 | 0 | 0 |
| Total | | 20 | 1 | 21 |
Sm Sensitivity (20/20) = 100%
Table 6
Sensitivity of the MarDx ENA IgG Immunostripe Test System
relative to the INOVA RNP EIA Test System
| Antigen: RNP | | INOVA RNP EIA | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| MarDx | Positive | 33 | 0 | 33 |
| Immunostripe | | | | |
| | Negative | 4 | 0 | 4 |
| Total | | 37 | 0 | 37 |
RNP Sensitivity (33/37) = 89%
{3}
Table 7
Sensitivity of the MarDx ENA IgG Immunostripe Test System
relative to the INOVA Jo-1 EIA Test System
| Antigen: Jo-1 | | INOVA Jo-1 EIA | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| MarDx | Positive | 2 | 0 | 2 |
| Immunostripe | | | | |
| | Negative | 0 | 0 | 0 |
| Total | | 2 | 0 | 2 |
Jo-1 Sensitivity (2/2)= 100%
Table 8
Sensitivity of the MarDx ENA IgG Immunostripe Test System
relative to the INOVA Scl-70 EIA Test System
| Antigen: Scl-70 | | INOVA Scl-70 | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| MarDx | Positive | 7 | 0 | 7 |
| Immunostripe | | | | |
| | Negative | 0 | 0 | 0 |
| Total | | 7 | 0 | 7 |
Scl-70 Sensitivity (7/7)= 100%
Table 9
INOVA SUMMARY
Sensitivity of the MarDx ENA IgG Immunostripe Test System
relative to the Six Combined Results from the INOVA ENA EIA Test Systems
| Combined | | INOVA ENA EIA | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| MarDx | Positive | 113 | 0 | 113 |
| Immunostripe | | | | |
| | Negative | 7 | 0 | 7 |
| Total | | 120 | 0 | 120 |
Sensitivity (113/120) = 94%
{4}
Table 10
Sensitivity of the MarDx ENA IgG Immunostripe Test System relative to the MarDx 6 in 1 ENA EIA Test System
| | | MarDx ENA EIA | | |
| --- | --- | --- | --- | --- |
| | | Positive | Negative | Total |
| MarDx | Positive | 67 | 0 | 67 |
| Immunostripe | | | | |
| | Negative | 2 | 0 | 2 |
| Total | | 69 | 0 | 69 |
Sensitivity (67/69) = 97%
Table 11
Summary Table For All Data
| MarDx | | Reference Manufacturer | Number of Specimens | Reference | Sensitivity | Specificity |
| --- | --- | --- | --- | --- | --- | --- |
| Immunostripe | vs | INOVA | 35/37 | SSA | 95% | NA |
| Immunostripe | vs | INOVA | 16/17 | SSB | 94% | NA |
| Immunostripe | vs | INOVA | 20/20 | Sm | 100% | NA |
| Immunostripe | vs | INOVA | 33/37 | RNP | 89% | NA |
| Immunostripe | vs | INOVA | 2/2 | Jo-1 | 100% | NA |
| Immunostripe | vs | INOVA | 7/7 | Scl-70 | 100% | NA |
| Immunostripe | vs | INOVA | 40/40 | Normals | NA | 100% |
| Immunostripe | vs | MarDx | 67/69 | 6 in 1 | 97% | NA |
| Immunostripe | vs | MarDx | 139/139 | Normals | NA | 100% |
| Immunostripe | Combined Sensitivity and Specificity | | | | 95% | 100% |
{5}
# Precision Study
The precision data is presented in Table 10 for 7 sera (Jo-1, 3 sera) assayed 8 times each on three different assays. The data indicates that the MarDx ENA Immunostripe has a high degree of precision both within and between assays.
Table 12
| Sample | | Assay 1 | | Assay 2 | | Assay 3 | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| | | Ratio | Percent | | | | |
| 1 | SSA | 7/7 | 100% | 7/7 | 100% | 7/7 | 100% |
| 2 | SSB | 7/7 | 100% | 7/7 | 100% | 7/7 | 100% |
| 3 | Sm | 7/7 | 100% | 7/7 | 100% | 7/7 | 100% |
| 4 | RNP | 7/7 | 100% | 7/7 | 100% | 7/7 | 100% |
| 5 | Jo-1 | 3/3 | 100% | 3/3 | 100% | 3/3 | 100% |
| 6 | Scl-70 | 7/7 | 100% | 7/7 | 100% | 7/7 | 100% |
| 7 | Normal | 7/7 | 100% | 7/7 | 100% | 7/7 | 100% |
| TOTALS | | 45/45 | 100% | 45/45 | 100% | 45/45 | 100% |
Precision = 100%
# Cross-Reactivity Study
The cross-reactivity data is presented in Table 11 for 7 sera containing dsDNA antibodies. The data indicates that the MarDx ENA Immunostripe does not cross-react with dsDNA.
Table 13
| Sample | Result |
| --- | --- |
| 1 | Neg |
| 2 | Neg |
| 3 | Neg |
| 4 | Neg |
| 5 | Neg |
| 6 | Neg |
| 7 | Neg |
Cross Reactivity to dsDNA = 0%
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.