K964785 · Victoreen, Inc. · IYE · Jun 13, 1997 · Radiology
Device Facts
Record ID
K964785
Device Name
VERIDOSE DIODES
Applicant
Victoreen, Inc.
Product Code
IYE · Radiology
Decision Date
Jun 13, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The VeriDose solid state detectors are used to make relative measurements of radiation in therapeutic oncology. Silicon diode detectors have been in use for many years for the measurement of both photons and electrons used in radiation. The VeriDose is intended for the detection of ionizing radiation, it should only be used by persons who have the proper interpretation of its readings and the appropriate safety procedures to be followed in the presence of radiation. The diodes are taped or otherwise attached to a patient's skin during a therapeutic oncology session where the patient is being exposed to radiation. The diodes produce a current directly proportional to the ionizing radiation to which they are exposed. The diodes are more rugged and more sensitive than ionization chambers and are waterproof. They must be connected to an appropriate display device. The purpose of the diode is to measure ionizing radiation independently at various points of interest. The responsible operator should verify the characteristics of each specific detector if the exact parameter is critical to any given clinical application.
Device Story
VeriDose Diodes are solid-state silicon radiation detectors used for relative measurement of ionizing radiation in therapeutic oncology. The device consists of a waterproof, color-coded molded plastic hemisphere containing buildup material and a radiation-sensitive diode connected to a coaxial cable with a BNC connector. During radiation therapy, the diode is attached to the patient's skin; it generates an electrical current directly proportional to the incident ionizing radiation. The device is operated by qualified clinical personnel who connect the diode to an external display device to read the output. The measurements provide independent verification of radiation dose at specific points of interest, assisting in quality assurance and clinical decision-making regarding patient treatment delivery. The device is designed to be more rugged and sensitive than traditional ionization chambers.
Clinical Evidence
Bench testing only. Performance specifications provided include linearity (better than 0.3% over 2-1000 cGy/min), angular dependence (less than 5% from 0-75 degrees, less than 10% from 75-90 degrees), and temperature sensitivity (less than 0.3% per degree C).
Technological Characteristics
Silicon diode radiation detector; waterproof, non-hygroscopic molded plastic housing; hemispherical buildup material (H2O equivalent or Al equivalent); BNC coaxial connector; 2.0 cGy/min to 1000 cGy/min range; 2 mm² effective detection area; 1.0 nC/R nominal output; 2-meter cable length; passive sensing (current generation proportional to radiation).
Indications for Use
Indicated for dose verification and quality assurance for patients undergoing radiation therapy.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
ISORAD Photon Detectors that are part of the Dual Dosimetry System, Model number 37-720 (K940660/S1)
Submission Summary (Full Text)
{0}
Exhibit "C"
K96.4785
# Premarket Notification [510(k)] Summary as required by section 807.92(c)
**Date Summary was prepared:**
November 21, 1996
**Submitter’s Name:**
JUN 13 1997
Victoreen, Inc.
6000 Cochran Road
Cleveland, Ohio 44139-3395
**Contact Person:**
Linda S. Morin
Director of Regulatory Affairs
and Quality Assurance
Phone: 216-248-9300
Fax: 216-248-9301
**Device Name:**
Veridose Diodes
**Classification Name:**
Medical Charged-Particle Radiation Therapy System
**Predicate Device:**
ISORAD Photon Detectors that are part of the Dual Dosimetry System, Model number 37-720, 510(k) number K940660/S1
**Intended Use:**
(Function): The VeriDose solid state detectors are used to make relative measurements of radiation in therapeutic oncology. Silicon diode detectors have been in use for many years for the measurement of both photons and electrons used in radiation. The VeriDose is intended for the detection of ionizing radiation, it should only be used by persons who have the proper interpretation of its readings and the appropriate safety procedures to be followed in the presence of radiation. The diodes are taped or otherwise attached to a
{1}
patient's skin during a therapeutic oncology session where the patient is being exposed to radiation. The diodes produce a current directly proportional to the ionizing radiation to which they are exposed. The diodes are more rugged and more sensitive than ionization chambers and are waterproof. They must be connected to an appropriate display device. The purpose of the diode is to measure ionizing radiation independently at various points of interest. The responsible operator should verify the characteristics of each specific detector if the exact parameter is critical to any given clinical application.
## Product Description:
The product consists of a waterproof color coded molded plastic hemisphere containing the appropriate buildup material and a radiation sensitive diode attached to a coaxial cable. A BNC connector is supplied for connection to a readout device.
## Model no. table
| Model # | Description | Color Code |
| --- | --- | --- |
| 30-471 | VeriDose Diode, 1-4 MV Photon, (positive polarity) | Blue |
| 30-471-8000 | VeriDose Diode, 1-4 MV Photon, (negative polarity) | Blue |
| 30-472 | VeriDose Diode, 6-12 MV Photon, (positive polarity) | Yellow |
| 30-472-8000 | VeriDose Diode, 6-12 MV Photon, (negative polarity) | Yellow |
| 30-473 | VeriDose Diode, 12-18 MV Photon, (positive polarity) | Red |
| 30-473-8000 | VeriDose Diode, 12-18 MV Photon, (negative polarity) | Red |
| 30-474 | VeriDose Diode, 18-25 MV Photon, (positive polarity) | Green |
| 30-474-8000 | VeriDose Diode, 18-25 MV Photon, (negative polarity) | Green |
| 30-475 | VeriDose Diode, 6-25 MeV Electron, (positive polarity) | Silver |
| 30-475-8000 | VeriDose Diode, 6-25 MeV Electron, (negative polarity) | Silver |
| 30-476 | VeriDose Diode, Energy Compensated Photon, (positive polarity) | Black |
| 30-476-8000 | VeriDose Diode, Energy Compensated, (negative polarity) | Black |
## GENERAL SPECIFICATIONS:
1. Range: 2.0 cGy / minute to 1000 cGy / minute.
2. Effective Detection Area: 2 mm²
3. Output, all models: 1.0 nC/R nominal
4. Cable length: 2 meters
{2}
| Range | Particle | Optimal Build-up | mg/cm² | Same as Isorad | |
| --- | --- | --- | --- | --- | --- |
| | | | | Build-up | Color Dot |
| 1-4 MV | Photon | 5 mm H₂O equiv. | 826 | Yes | Blue |
| 6-12 MV | Photon | 15 mm H₂O equiv. | 1500 | Yes | Yellow |
| 12-18 MV | Photon | 28 mm H₂O equiv. | 2800 | No | Red |
| 18-25 MV | Photon | 36 mm H₂O equiv. | 3600 | No | Green |
| 6-25 MeV | Electron | 1 mm Al equiv. | 275 | Yes | Silver |
| 100-662 keV | Photons | Energy Compensated | | No | Black |
6. The sensitive area buildup material is hemispherical in shape.
7. The angular dependence is less than 5% over the 2 pi hemisphere from perpendicular incidence to 75 degrees and less than 10% from 75 to 90 degrees.
8. Linearity is better than 0.3% over the range of 2 cGy / min to 1000 cGy/min provided that dose-per-pulse remains constant.
9. There are both positive and negative polarity versions of each.
10. The temperature sensitivity is less than 0.3% per degree C.
11. The detectors are waterproof and are washable in cold disinfectant. They are molded in plastic that is non-hygroscopic.
12. The device has a BNC coaxial connector.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN 13 1997
Linda S. Morin
Corporate Director of Regulatory
Affairs and Quality Assurance
Victoreen, inc.
6000 Cochran Road
Cleveland, Ohio 44139-3395
Re: K964785
VeriDose Diodes Medical Charged-Particle
Radiation Therapy System
Dated: April 18, 1997
Received: April 25, 1995
Regulatory class: II
21 CFR 892.5050/Procode: 90 IYE
Dear Ms. Morin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591 for Radiology devices, or 594-4613 for Ear, Nose and Throat devices. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Director, Division of Reproductive,
Abdominal, Ear, Nose and Throat,
and Radiological Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{4}
510(k) Number: K964785
Device Name: Veridose Diodes
Indications for Use:
These diodes are used to provide dose verification and quality assurance for patients undergoing radiation therapy.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seymon
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964785
Prescription Use ☑ or Over-The-Counter Use ☐
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.