← Product Code [OCY](/productcode/OCY) · K964780

# BARD DIRECTOR GUIDEWIRE (000560) (K964780)

_C.R. Bard, Inc. · OCY · Feb 26, 1997 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K964780

## Device Facts

- **Applicant:** C.R. Bard, Inc.
- **Product Code:** [OCY](/productcode/OCY.md)
- **Decision Date:** Feb 26, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

For use with devices having a .035" guidewire compatible lumen. The Director guidewire is used to guide endoscopic accessories into the GI tract and for cannulation of the pancreatobiliary ducts.

## Device Story

Bard Director Guidewire is a medical device used to facilitate the placement of endoscopic accessories within the GI tract and to assist in the cannulation of pancreatobiliary ducts. The device is operated by a physician during endoscopic procedures, typically performed under endoscopic and/or fluoroscopic visualization. The guidewire is advanced through a duodenoscope to the target anatomical location. Its design features a multi-segment core construction intended to provide kink resistance, torque control, and radiopacity, aiding the clinician in navigating complex ductal anatomy. By providing a stable track, the device enables the successful delivery of therapeutic or diagnostic accessories, potentially improving procedural efficiency and patient outcomes during GI interventions.

## Clinical Evidence

Bench testing only. Performance was evaluated through comparative bench testing for stiffness, torque, and radiopacity against the Microvasive Geenen Endotorque and Wilson-Cook Tracer wire guides. Biocompatibility testing was performed to confirm the safety of the materials used in the device construction.

## Technological Characteristics

Guidewire with three-segment core: proximal segment (Bisphenol A epoxy with fibers, polyethylene jacket), mid-section (PTFE-coated stainless steel), and distal section (nitinol tapered core, polyester weave, polyurethane, hydrophilic coating). Radiopaque. Designed for .035" lumen compatibility.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Microvasive® Geenen Endotorque™ Guidewire ([K942677](/device/K942677.md)/A)
- Wilson-Cook® Tracer™ Wire Guide ([K910497](/device/K910497.md))
- Davol® Laparoscopic Suction Irrigation Probe ([K941334](/device/K941334.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K964780/S1
p192
BARD

Bard Interventional Products Division
C.R. Bard, Inc.
129 Concord Road
P.O. Box 7031
Billerica, MA 01821-7031
508-663-8989
FEB 26 1997
BAARD

## 6.0 510(k) SUMMARY FOR THE BARD®DIRECTOR™ GUIDEWIRE

As required under Section 12, part (a)(i)(3A) of the Safe Medical Device Act of 1990, an adequate summary of any information respecting safety and effectiveness follows.

## 6.1 General Information

- Name and address of submitter:

Bard Interventional Products Division, C.R. Bard, Inc.
129 Concord Road, Building #3
Billerica, MA 01821-7031

- Contact:

Beth A. Rochette, R.A.C.          Phone: (508) 663-8989
Regulatory Affairs Manager   Fax: (508) 670-9827

- Date of Summary:

November 15, 1996

- Name of Device:

Trade Name: Bard® Director™ Guidewire
Common/Usual Name: Guidewire
Classification Name: Endoscope and accessory

- Predicate Device(s):

Microvasive® Geenen Endotorque™ Guidewire
Wilson-Cook® Tracer™ Wire Guide
Davol® Laparoscopic Suction Irrigation Probe

- Description and Intended Use of Device:

For use with devices having a .035" guidewire compatible lumen. The Director guidewire is used to guide endoscopic accessories into the GI tract and for cannulation of the pancreatobiliary ducts.

039

{1}

K964780
P292

## 6.2 Summary of Similarities and Differences

The Bard® Director™ Guidewire is substantially equivalent to the currently marketed Microvasive® Geenen Endotorque™ Guidewire (#K942677/A), the Wilson-Cook® Tracer™ Wire Guide (#K910497), and the Davol® Laparoscopic Suction Irrigation Probe (#K941334).

The indication statements are different, however, the Director guidewire has the same intended use as the Geenen Endotorque guidewire and the Tracer wire guide. The Director guidewire is used to guide endoscopic accessories into the GI tract and for cannulation of the pancreatobiliary ducts. The general design and functionality of the Director guidewire is similar to these devices. All three of the guidewires are advanced through a duodenoscope (side viewing endoscope) under endoscopic and/or fluoroscopic visualization into the GI tract to the desired location. The major differences between the guidewires are:

The Director guidewire is constructed with three core segments that are bonded together. The proximal core is a Bisphenol A epoxy with fibers, covered with a polyethylene jacket, the mid section of the guidewire is a PTFE coated stainless steel, and the distal section is a nitinol tapered core covered with a polyester weave, polyurethane, and proprietary hydrophilic coating. Both the Geenen Endotorque guidewire and the Tracer wire guides have a single nitinol distal tapered core with a teflon jacket and a distal tip coating for lubricity.

Although the materials and construction are different; these materials are used in currently marketed devices. The Bisphenol A epoxy used in the proposed Director guidewire is the same material utilized with the Davol Laparoscopic Suction Irrigation Probe tip. The Davol Laparoscopic Suction Irrigation Probe is a nonconductive fiberglass probe intended for suction and irrigation of the operative site during electrosurgical laparoscopic procedures. PTFE and stainless steel are standard materials used in the manufacturing of guidewires. The jacket material covering the distal tapered nitinol core is different from the Geenen Endotorque and the Tracer wire guide, however, biocompatibility testing has confirmed the materials are safe for contacting mucosa and tissue.

The distal tapered core with the polyurethane coating provides a kink resistant, radiopaque, lubricious guidewire. Stiffness, torque, and radiopacity bench testing has confirmed the device is substantially equivalent to the Microvasive Geenen Endotorque™ guidewire and the Wilson-Cook Tracer™ wire guide.

040

---

**Source:** [https://fda-staging.innolitics.com/device/K964780](https://fda-staging.innolitics.com/device/K964780)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).