DELTA N95 RESPIRATOR

K964722 · Racal Filter Technologies, Inc. · FXX · Feb 7, 1997 · General, Plastic Surgery

Device Facts

Record IDK964722
Device NameDELTA N95 RESPIRATOR
ApplicantRacal Filter Technologies, Inc.
Product CodeFXX · General, Plastic Surgery
Decision DateFeb 7, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4040
Device ClassClass 2

Indications for Use

The Delta N95 Respirator and Surgical Mask will be used for the protection of the health care practitioner from blood and body fluids and particulates and to meet the NIOSH / CDC TB Prevention Guidelines.

Device Story

Disposable, one-piece, lightweight half-mask respirator and surgical mask. Used by healthcare practitioners to provide respiratory protection against particulates (aerosols free of oil) and fluid exposure. Features multi-layer filtering media, wire nose clip, foam nose piece, and elastic headbands to ensure face-to-mask seal. Available in four sizes (extra small, small, medium, large). Provides 95% filtration efficiency. NIOSH approved (TC-84A-0003).

Clinical Evidence

Bench testing only. Device meets 42 CFR Part 84 performance standards for N95-class respiratory protective devices and CDC guidelines for preventing Mycobacterium tuberculosis transmission. Filtration efficiency is 95% against particulate aerosols free of oil.

Technological Characteristics

N95-class particulate respirator. Materials: inner scrim layer, netting layer, meltblown protection layer, outer scrim layer, elastic headband. Features: wire nose clip, foam nose piece. Performance: 95% filtration efficiency against non-oil aerosols. NIOSH approved (TC-84A-0003).

Indications for Use

Indicated for health care practitioners requiring respiratory protection from blood, body fluids, and particulates, and to meet NIOSH/CDC TB prevention guidelines.

Regulatory Classification

Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible. (ii) Analysis and nonclinical testing must: (A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and (B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device. (iii) NIOSH approved under its regulation. (2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0} FILE No. 708 01/29 '97 16:06 ID:RACAL HEALTH & SAFETY 301 695 3017 PAGE 5 RACAL HEALTH & SAFETY INC. PREMARKET SUBMISSION DELTA N95 RESPIRATOR AND SURGICAL MASK 510(K) K964722 510(k) SUMMARY - K964722 DELTA N95 RESPIRATOR & SURGICAL MASK FEB - 7 1997 January 29, 1997 This summary is provided in accordance with the Safe Medical Devices Act of 1990 (SMDA). The information provided in the 510(k), Premarket Notification, was in accordance with 21 CFR 807.87. 1. Submitter of 510(k): Company Name: Racal Health & Safety Inc. 7305 Executive Way Frederick, Maryland U.S.A. 21704-8368 Contact Person (U.S.A.): Mr. J. Randall Hoggle Telephone: (301)695-8200 Fax: (301)695-4413 Contact Person (Canada): John Campbell Telephone: (613)345-0111 Fax: (613)345-2639 2. Name of Device Trade/Proprietary Name: DELTA N95 RESPIRATOR AND SURGICAL MASK Common/Usual Name: Masks, Surgical Classification Name: 21 CFR 878.4040 "Surgical Apparel", Class II 3. Legally Marketed Predicate Devices DELTA N95 RESPIRATOR AND SURGICAL MASK is substantially equivalent to many legally marketed Surgical masks, including those listed in Table 1 below. Page 1 of 3 CONFIDENTIAL {1} FILE No. 708 01/29 '97 16:07 ID:RACAL HEALTH & SAFETY 301 695 3017 PAGE 6 RACAL HEALTH & SAFETY INC. DELTA N95 RESPIRATOR AND SURGICAL MASK PREMARKET SUBMISSION 510(K) K964722 # 510(k) SUMMARY ## TABLE 1 ### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS ### PREDICATE DEVICES AND DELTA N95 RESPIRATOR AND SURGICAL MASK | Manufacturers | Racal Health & Safety | 3M | Louis M Gerson | | --- | --- | --- | --- | | Device | DELTA N95 Respirators | Model 1860 Health Care Respirators | Isolair APR Type N95 Respirators | | 510(k) Number | K964722 | K955382 | K960778 | | Standards Met | Performance Code of Federal Regulations Title 42 Part 84 for N95-class of Respiratory Protective Devices Meets CDC guidelines for preventing transmission of Mycobacterium tuberculosis | Performance Code of Federal Regulations Title 42,Part 84 for N95-class of Respiratory Protective Devices Meets CDC guidelines for preventing transmission of Mycobacterium tuberculosis | Performance Code of Federal Regulations Title 42, Part 84 for N95 - class of Respiratory Protective Devices Meets CDC guidelines for preventing transmission of Mycobacterium tuberculosis | | NIOSH Approval Number | TC- 84A-003 | TC - 84A- 006 | TC-84A-160 | | Device Description | • N95 Class particulate Respirator • Multi-Layer Filtering media • Fluid resistant barrier • Wire nose clip • Foam nose piece inside mask • Elastic Headband • Mint colored | • N95 Class particulate Respirator • Multi-Layer Filtering media • Fluid resistant barrier • Wire nose clip • Foam nose piece inside mask • Elastic Headband • Blue colored | • N95 Class particulate Respirator • Multi-Layer Filtering media • Fluid resistant barrier • Wire nose clip • Foam nose piece inside mask • Elastic Headband • Blue-green colored | | Intended Use | Respiratory Protection of Health Care workers | Respiratory Protection of Health Care of workers | Respiratory Protection of Health Care workers | ## 4. Device Description DELTA N95 RESPIRATOR AND SURGICAL MASK is indicated for respiratory protective device. This disposable Surgical mask is a one-piece, light weight half mask with filter material forming the mask. The component includes an inner scrim layer, netting material forming the mask. The component includes an inner scrim layer, netting layer, meltblown protection layer, outer scrim layer and elastic band. DELTA Page 2 of 3 CONFIDENTIAL {2} FILE No. 708 01/29 '97 16:07 ID:RACAL HEALTH & SAFETY 301 695 3017 PAGE 7 RACAL HEALTH & SAFETY INC. PREMARKET SUBMISSION DELTA N95 RESPIRATOR AND SURGICAL MASK 510(K) K964722 N95 RESPIRATOR AND SURGICAL MASK is NIOSH approved (Approval #TC-84A-0003) and has a filtration efficiency level of 95% and is effective against particulate aerosols free of oil. DELTA N95 RESPIRATOR AND SURGICAL MASK is available in four sizes, extra small, small, medium and large. The nose clip and elastic headband ensures effective face to mask seal. Based on the information provided, DELTA N95 RESPIRATOR AND SURGICAL MASK is substantially equivalent to legally marketed predicate devices. ## 5. Indications For Use The Delta N95 Respirator and Surgical Mask will be used for the protection of the health care practitioner from blood and body fluids and particulates and to meet the NIOSH / CDC TB Prevention Guidelines. Page 3 of 3 CONFIDENTIAL
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...