LONG MEDIUM PALMAZ-SCHATZ BALLOON EXPANDABLE STENT
Applicant
Cordis Corp.
Product Code
FGE · Gastroenterology, Urology
Decision Date
Jun 20, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The stents are intended for palliation of malignant neoplasms in the biliary tree.
Device Story
Long Medium PALMAZ-SCHATZ™ Balloon-Expandable Stent is a stainless steel, multi-articulated, slotted tube designed for biliary duct support. Device is provided unmounted and manually crimped onto a qualified balloon catheter using a polyethylene crimping tube and a reusable metallic crimping tool. Upon deployment, balloon inflation expands the stent into a diamond-pattern open lattice, which conforms to the inner luminal surface of the bile duct to maintain patency. Used by clinicians in a medical setting to palliate malignant neoplasms in the biliary tree. Stent provides structural support to prevent collapse of the duct, thereby improving biliary drainage and patient comfort.
Clinical Evidence
Bench testing only. Performance parameters evaluated included inflation time, balloon burst pressure, crossing profile, distension, deflation time, stent retention, recoil, radial force, percent open area, MRI compatibility, and fatigue/corrosion testing. Results demonstrated that the device meets all internal performance criteria and maintains structural integrity under clinically relevant conditions.
Technological Characteristics
Surgical grade stainless steel, multi-articulated slotted tube design. Balloon-expandable. Nominal lengths: 42mm, 56mm, 78mm; expansion range: 6-10mm. Open mesh design with 80-90% open area. MRI compatible. Delivered via balloon catheter. Crimping tool (CRT 80) used for manual mounting.
Indications for Use
Indicated for palliation of malignant neoplasms in the biliary tree.
Regulatory Classification
Identification
A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.
Special Controls
*Classification.* Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Predicate Devices
PALMAZ™ Balloon-Expandable Stent
PALMAZ-SCHATZ™ Balloon-Expandable Stent
Schneider's Wallstent® Biliary Endoprosthesis
Submission Summary (Full Text)
{0}
JUN 20 1997
K964688
pg 1 of 4
510(k) Summary
Submitter's name:
Cordis Corporation, a Johnson & Johnson Company
40 Technology Drive
Warren, New Jersey 07059
908-755-8300
908-412-3915 (fax)
Contact person:
Joan Martin
Manager, Regulatory Affairs
(908) 412-7250
Date prepared:
November 20, 1996
Trade name:
Long Medium PALMAZ-SCHATZ™
Balloon-Expandable Stent
Common name:
Biliary Stent
Classification name:
§876.5010 Biliary Catheter and Accessories
Predicate devices:
PALMAZ™ and PALMAZ-SCHATZ™ Balloon-Expandable Stents
Schneider's Wallstent® Biliary Endoprosthesis
Device description:
The Long Medium PALMAZ-SCHATZ™ Balloon-Expandable Stent is a stainless steel, multi-articulated, slotted tube with nominal lengths of 42mm (PS424), 56mm (PS564), and 78mm (PS784) and an expansion range of 6-10mm. The Stents are intended for palliation of malignant neoplasms in the biliary tree. The Stents are provided unmounted to be crimped onto commercially available Cordis Opta 5™ Catheters which have been qualified by Cordis for stent delivery. The Stent is expanded via inflation of the balloon catheter and deployed. Upon expansion, the Stent forms a strong open lattice in a diamond pattern. The expanded Stent conforms to the inner luminal surface of the bile duct. The ability of the Stent to palliate malignant neoplasms is based upon the ability of the metallic structure of the Stent to plastically deform when the balloon is inflated and maintain its expanded state.
Accessory Devices - The following accessory devices have been slightly modified to accommodate the longer Stent lengths. Shorter versions of these devices are
{1}
K964688
Pg 2 of 4
# 510(k) Summary (p.2)
marketed for use with our current line of biliary stents, as well as with our iliac stent.
The **Crimping Tube** is constructed of polyethylene. It is supplied sterile, for one use only and is provided with the unmounted Stent. The Crimping Tube protects the Stent/balloon assembly during the manual crimping process. Crimping tube modifications consist of a longer length to accommodate the longer stents and a slightly decreased wall thickness.
The **Crimping Tool** (CRT 80) is a hand-held, reusable, metallic instrument which is used to manually crimp the unmounted Stents onto the recommended balloon catheter. The Crimping Tool is provided nonsterile and must be sterilized prior to use by standard autoclave procedures. The tool delivers a radial compressive force to the Stent, thereby securing the Stent to the balloon catheter. The Crimping Tool die has been modified to accommodate the additional stent lengths. Our currently marketed Crimping Tool (CRT40) is used to crimp the 42mm Long Medium Stents.
---
**Intended use statement:**
The stents are intended for palliation of malignant neoplasms in the biliary tree.
---
**Labeled indication:**
The Long Medium PALMAZ-SCHATZ™ Balloon Expandable Stent is indicated for palliation of malignant neoplasms in the biliary tree.
---
**Technological characteristics:**
The Long Medium PALMAZ-SCHATZ™ Balloon-Expandable Stents are substantially equivalent in design, function and intended use to the currently marketed PALMAZ™ and PALMAZ-SCHATZ™ Biliary Stents and to the Schneider Wallstent® Biliary Endoprosthesis.
The subject and predicate devices are constructed of surgical grade stainless steel which is a known biocompatible implant material. The subject and predicate devices all have an open mesh design which is
{2}
K964688
pg 3 of 4
# 510(k) Summary (p.3)
intended for palliation of malignant neoplasms in the biliary tree. Both the currently marketed and the modified Cordis stents are constructed from stainless steel tubing. The Wallstent® stents are constructed of braided wire.
The subject device and the currently marketed Cordis biliary stents are balloon-expandable. The Wallstent® stents are self-expanding. All of the devices utilize a catheter to deliver the stent. The range of lengths and diameters of the subject and predicate devices are comparable and are intended to meet clinical need.
Diameter: 6-10mm for the subject devices, 4-12mm for the currently marketed Cordis Stents and predetermined 8-10mm for the Wallstent®.
Lengths: 42mm, 56mm and 78mm for the subject device, 10-39mm for the currently marketed Cordis Stents and 15mm to 150mm for the Wallstent® device.
## Nonclinical performance:
The differences between the subject and predicate devices are minor and are not relevant to the ability of the subject devices to palliate malignant neoplasms in the biliary tree. The descriptive characteristics are precise enough to demonstrate equivalence. However, qualification testing was performed to confirm the acceptable performance and properties of the subject devices. The following parameters were studied:
Inflation time: Samples were fully expanded to the nominal diameter within the allotted time period.
Balloon burst: Mounted stents did not modify the catheters’s labeled maximum burst pressure.
Crossing profile: A 7F sheath is required for 6mm diameter balloons with mounted stents and an 8F sheath is required for 8mm and 10mm diameter balloons with mounted stents.
Distension: All stents reached their nominal diameter within the recommended pressure.
Deflation time/deflatability: All balloons with mounted stents deflated within the allotted time period. The
{3}
K964688
Pg 4 of 4
# 510(k) Summary (p.4)
presence of the stent did not cause any interference with balloon deflatability.
**Stent retention:** All samples met or exceeded our internal criteria for stent retention.
**Recoil and Length:** The stent remains within its nominal diameter after recoil. Lengths were measured to determine stent length at each expanded diameter (for labeling purposes).
**Radial Force:** The stent will not collapse within the clinically relevant range.
**Percent Open Area:** All stent sizes have an open area of 80-90% in their expanded diameter range.
**MRI Compatibility:** The subject devices were found to be MRI compatible.
**Fatigue/Corrosion:** No cracks or other metallurgical defects resulted from fatigue testing.
TOTAL P.06
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
JUN 20 1997
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Ms. Joan Martin
Manager, Regulatory Affairs
Cordis Corporation,
a Johnson & Johnson Company
40 Technology Drive
Warren, New Jersey 07059
Re: K964688
Long Medium PALMAZ-SCHATZ™ Balloon
Expandable Stent
Dated: March 31, 1997
Received: April 1, 1997
Regulatory class: II
21 CFR §876.5010/Product code: 78 FGE
Dear Ms. Martin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
510(k) Number (if known): K964688
Device Name: Long Medium PALMAZ-SCHATZ™ Balloon Expandable Stent
Indications For Use:
The Long Medium PALMAZ-SCHATZ™ Balloon Expandable Stents are indicated for palliation of malignant neoplasms in the billiary tree.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964688
Prescription Use ☑
(Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.