FORCE ARGON II ARGON ENHANCED ELECTROSURGICAL SYSTEM

K964636 · Valleylab, Inc. · GEI · Mar 11, 1997 · General, Plastic Surgery

Device Facts

Record IDK964636
Device NameFORCE ARGON II ARGON ENHANCED ELECTROSURGICAL SYSTEM
ApplicantValleylab, Inc.
Product CodeGEI · General, Plastic Surgery
Decision DateMar 11, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Force Argon™ II Argon Enhanced Electrosurgical System is intended for use in both open, laparoscopic, and thoracoscopic surgical procedures (general, neurosurgical, gynecologic) where monopolar electrosurgery (cutting and coagulation) is normally used. The Force Argon™ II Argon Enhanced Electrosurgical System provides a controlled flow of argon to electrosurgical handsets during cutting and coagulation. When the handset is activated in the gas enhanced modes, an argon gas plasma is created between the electrode and the tissue.

Device Story

System delivers controlled flow of inert argon gas to electrosurgical handsets; creates argon gas plasma between electrode and tissue during cutting/coagulation. Used in open, laparoscopic, and thoracoscopic procedures. Features two flow modes: Low Flow (0.5-4.0 SLPM) for laparoscopy; Standard Flow (0.5-12 SLPM) for open surgery. Includes integrated peritoneal pressure monitoring for laparoscopic use; provides user-settable pressure limits with audible alarms to prevent over-pressurization. Operated by surgeons in OR settings. Unit designed for tabletop or cart mounting; compatible with argon tanks (14-42 cubic feet). Enhances electrosurgical performance by providing gas-shrouded cutting and coagulation.

Clinical Evidence

Bench testing only. Preclinical testing performed in a swine model and simulated laparoscopic model. Verified system functionality, argon delivery, and peritoneal pressure monitoring against specifications.

Technological Characteristics

System provides controlled argon gas flow for electrosurgical handsets. Materials for peritoneal pressure tubing meet USP Class V biocompatibility standards. Complies with IEC 601-1, IEC 601-2-2, EN 60601-1-2, IEC 801-3, and ANSI/AAMI HF18. Tabletop/cart-mounted unit. Software-controlled gas flow and pressure monitoring.

Indications for Use

Indicated for patients undergoing open, laparoscopic, or thoracoscopic surgical procedures (general, neurosurgical, gynecologic) requiring monopolar electrosurgical cutting and coagulation.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0} K96+636 MAR 11 1997 # VALLEYLAB FORCE ARGON™ II ARGON ENHANCED ELECTROSURGICAL SYSTEM SUMMARY OF SAFETY AND EFFECTIVENESS ## I. Product Description: The Force Argon™ II Argon Enhanced Electrosurgical System is designed to provide a controlled flow of argon (an inert, non-reactive gas) to custom designed electrosurgical handsets powered by Valleylab electrosurgical generators. The handsets allows the surgeon to perform argon enhanced electrosurgery in open and laparoscopic procedures. Depending on the handpiece, the system can function in the following argon enhanced modes: - Argon Shrouded Cut - Argon gas is delivered through the handset while the attached electrosurgical generator is activated in the cut mode. - Argon Enhanced Coag - Argon gas is delivered through the handset while the generator is activated in the coag mode. The Force Argon™ II Argon Enhanced Electrosurgical System has two gas flow modes; one for laparoscopic procedures (Low Flow Mode, 0.5 - 4.0 SLPM) and one for open procedures (Standard Flow Mode, 0.5-12 SLPM). In addition, the system monitors the peritoneal pressure during laparoscopic procedures. The physician can set the desired pressure on the Force Argon™ II Argon Enhanced Electrosurgical System. When the peritoneal pressure approaches a predefined percentage of the set pressure, an alarm is activated to alert the user. A second alarm sounds when the actual peritoneal pressure reaches the set point. The physician can then vent the peritoneal cavity to relieve the pressure. The Force Argon™ Argon Enhanced Electrosurgical System is designed to allow the physician to continue to use argon enhanced electrosurgery when the peritoneal pressure set limit is reached. The Force Argon™ II unit is designed to fit on a table top or a cart. Tank holders will be offered for use with the Valleylab Gas Delivery System carts which will accommodate argon tank sizes ranging from 14 to 42 cubic feet. ## II. Intended Use The Force Argon™ II Argon Enhanced Electrosurgical System is intended for use in both open, laparoscopic, and thoracoscopic surgical procedures (general, neurosurgical, gynecologic) where monopolar electrosurgery (cutting and {1} coagulation) is normally used. The Force Argon™ II Argon Enhanced Electrosurgical System provides a controlled flow of argon to electrosurgical handsets during cutting and coagulation. When the handset is activated in the gas enhanced modes, an argon gas plasma is created between the electrode and the tissue. ## III. Safety and Performance Data Safety and performance of the Force Argon™ II Argon Enhanced Electrosurgical System were verified through preclinical testing, hardware validation and testing, and software validation. ### A. Hardware Validation The Force Argon™ II Argon Enhanced Electrosurgical System has been tested in accordance with applicable sections of the following standards: - IEC 601-1 (1988), Medical Electrical Equipment Part 1: General Requirements for Safety - IEC 601-2-2 (1991), Medical Electrical Equipment Part 2: Particular Requirements for the Safety of High Frequency Surgical Equipment - EN 60601-1-2 (1993) Medical Electrical Equipment, Part 1 General Requirements for Safety, 2, Collateral Standard: Electromagnetic Compatibility- Requirements and Tests - IEC 801-3 (1984), Radiated Electromagnetic Field Requirements - ANSI/AAMI HF18 (1993), Electrosurgical Devices ### B. Software Validation The software for the Force Argon™ II unit has been designed and is being validated in accordance with accepted industry standards. ### C. Biocompatibility Testing The biological safety of the peritoneal pressure tubing has been assured through the selection of materials which demonstrate appropriate levels of biocompatibility. The materials have been tested in accordance with USP Class V. ### D. Preclinical and Laboratory Bench Testing Preclinical and laboratory bench testing were performed on the Force Argon™ II Argon Enhanced Electrosurgical System in conjunction with a Valleylab electrosurgical generator and associated accessories using laparoscopic procedures in a swine model and a simulated laparoscopic model, respectively. This testing {2} showed the Force Argon™ II Argon Enhanced Electrosurgical System functioned according to specification and monitored peritoneal pressure as designed. ## IV. Summary of Substantial Equivalence The Force Argon™ II Argon Enhanced Electrosurgical System is substantially equivalent in function and intended use to the following legally marketed devices: - Valleylab Force GSU® Argon Enhanced Electrosurgical System. - Valleylab Force GSU® System with Laparoscopic option. - Valleylab Force GSU® - SLPM System. The peritoneal pressure monitoring function of the Force Argon™ II Argon Enhanced Electrosurgical System is comparable to similar systems in currently marketed insufflators.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...