K964626 · Philips Medical Systems (Cleveland), Inc. · LNH · Feb 10, 1997 · Radiology
Device Facts
Record ID
K964626
Device Name
EDGE/VISTA SYSTEM ENHANCEMENT PACKAGE
Applicant
Philips Medical Systems (Cleveland), Inc.
Product Code
LNH · Radiology
Decision Date
Feb 10, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Picker International EDGE and VISTA systems are indicated for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
Device Story
The EDGE/VISTA System Enhancement Package is an upgrade for existing MRI systems. It incorporates water-cooled gradient subsystems, a 61 cm body coil, and increased SAR limits. The device uses magnetic resonance physics to acquire proton density, flow velocity, T1, and T2 signals. These inputs are processed by the system's computer subsystem to reconstruct images. Operated by trained technicians in clinical settings, the system provides images for physician interpretation to assist in diagnosis. The enhancement improves gradient performance and patient handling capacity. The output is displayed on a monitor for clinical review, aiding in the assessment of soft tissue structures.
Clinical Evidence
Bench testing only. No clinical data provided. Safety and effectiveness are supported by technical comparisons of gradient performance, RF absorption limits, acoustic noise levels, and system specifications against predicate devices.
Technological Characteristics
MRI system enhancement including water-cooled self-shielded gradient coils, 61 cm quadrature multi-conductor body coil, and increased SAR limits (3.2 W/kg). Connectivity via Ethernet (DICOM v3.0). Software operates on UNIX/X-Windows platform. Magnet enclosure is cylindrical fiberglass. Gradient performance levels: 16, 20, or 27 mT/m strength; 25, 40, or 72 mT/m/msec slew rates.
Indications for Use
Indicated for patients requiring MRI imaging of the head and whole body to visualize soft tissue structures for diagnostic purposes.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}
FEB 10 1997
# SUMMARY OF SAFETY AND EFFECTIVENESS
(As required by 21 CFR 807.92)
K964626
## 1. General Information
Classification: Class II
Magnetic Resonance (MR) Imaging System
Common/Usual Name: Magnetic Resonance Imaging Option
Proprietary Name: EDGE/VISTA System Enhancement Package
Establishment Registration: Picker International, Inc.
World Headquarters
595 Miner Road
Cleveland, Ohio 44143
FDA Owner Number: #1580240
FDA Registration Number: #1525965
Performance Standards: No applicable performance standards have been issued under section 514 of the Food, Drug and Cosmetic Act.
## 2. Intended Uses
The EDGE/VISTA System Enhancement Package does not change the existing indications for the EDGE/VISTA systems as defined below.
The Picker International EDGE and VISTA systems are indicated for use as a NMR device that produces images that: (1) correspond to the distribution of protons exhibiting NMR, (2) depend upon the NMR parameters (proton density, flow velocity, spin-lattice relaxation time (T1), and spin-spin relaxation time (T2)) and (3) display the soft tissue structure of the head and whole body. When interpreted by a trained physician, these images yield information that can be useful in the determination of a diagnosis.
## 3. Device Description
The EDGE/VISTA system enhancement package includes changes in the computer subsystem, the gradient subsystem, the RF subsystem, the magnet subsystem, the patient handling subsystem, and the magnet enclosure. The main features provided with this enhancement are a 61 cm body coil, three gradient performance levels, water cooled gradients, an increase in the SAR first controlled operating mode limit, and optional increased reconstruction rates.
PICKER INTERNATIONAL, INC.
(EDGE SE)
11/18/96
I - 1
{1}
# 4. Safety and Effectiveness
The EDGE/VISTA systems with the system enhancement package are similar in technological characteristics and intended use to EDGE/VISTA systems with standard or high performance gradients. The following chart has been created to demonstrate their substantial equivalence.
SUBSTANTIAL EQUIVALENCE CHART
| ITEM | EDGE/VISTA with System Enhancement Package | Predicate Devices EDGE/VISTA (K931544) & High Performance Gradients Option (K954646) |
| --- | --- | --- |
| Computer Subsystem | Similar display/database computer, monitor and Ethernet communications systems. Ethernet communications link complies with DICOM v3.0 to allow image transfer to other MR, CT, Nuclear, workstation systems and cameras. Optional packages available for increased image reconstruction rates. | Display/Database computer and scan reconstruction hardware connected together by a dedicated Ethernet communications system. Allows for simultaneous scanning and image reconstruction/ manipulation. Single monitor for system operation and image reviewing. Second Ethernet interface for communication with systems on the external network. (See K931544) |
| Image Storage
Short Term:
Archival: | Same
Same | Magnetic Disk
Optical Disk
(See K931544) |
PICKER INTERNATIONAL, INC.
(EDGE SE)
11/13/96
I - 2
{2}
| ITEM | EDGE/VISTA with System Enhancement Package | Predicate Devices EDGE/VISTA (K931544) & High Performance Gradients Option (K954646) |
| --- | --- | --- |
| Gradient Subsystem
Gradient Coils: | Water-cooled self-shielded gradient system for all performance levels. | Standard EDGE- Air-cooled self-shielded gradient system. (see K931544)
EDGE with High Performance Gradients - Water-cooled self shielded gradient system (see K954646) |
| Max. Gradient Strength | 16, 20 or 27 mT/m | 16 or 27 mT/m |
| Max. Slew Rate | 25, 40, 72 mT/m/msec | 25 or 72 mT/m |
| Gradient Amps: | 150 A RMS, 300 A peak, 400 V
or
270 A RMS, 440 A peak, 600 V | 135 A RMS, 220 A peak, 350 V
or
270 A RMS, 440 A peak, 600 V |
| RF Subsystem
DTR Spectrometer: | Same | Single Channel with fast sampling ADC and additional digital filtering. Three additional receive channels are available as an option. |
| RF Amplifiers: | Single frequency RF amplifiers with increased output power to give equivalent B_{1} field with larger body coil. | Single frequency RF amplifiers. (See K931544) |
| Body Coil | 61 cm quadrature multi-conductor transmit/receive coil. | 58 cm quadrature multi-conductor transmit/receive coil. (See K931544) |
| Head Coil | Same | Quadrature multi-conductor receive only. (See K931544) |
PICKER INTERNATIONAL, INC.
(EDGE SE)
11/13/96
I - 3
{3}
| ITEM | EDGE/VISTA with System Enhancement Package | Predicate Devices EDGE/VISTA (K931544) & High Performance Gradients Option (K954646) |
| --- | --- | --- |
| Receive Only Coil Connection | Same | All receive only coils plug into single couch RF connector. |
| Transmit/Receive Box | Same | Four receive channels in system. (See K931544) |
| Magnet Subsystem | Same | Active Shield Magnets offered at 1.5 T and 1.0 T. (See K931544) |
| Patient Handling | Computer controlled patient transport system with increased weight capacity. New method of patient illumination used to increase patient comfort. | Computer controlled patient transport system. (See K931544) |
| Magnet Facade | Cylindrical fiberglass enclosure. Hinge on front facade removed. | Cylindrical fiberglass enclosure. Hinged front facade. (See K931544) |
| Power Distribution Subsystem | Same | Isolation transformer, transient suppression circuitry, and power distribution center all contained in a single cabinet. (See K931544) |
| Operating Software | Same | UNIX - X Windows based operating software. Graphical User Interface - windows and multi-tasking capability provided.
SCAN, VIEW, FILM and UTILITIES operations all accessed from single console.
Able to switch between on-going tasks. (See K931544) |
PICKER INTERNATIONAL, INC.
(EDGE SE)
11/13/96
I - 4
{4}
| ITEM | EDGE/VISTA with System Enhancement Package | Predicate Devices EDGE/VISTA (K931544) & High Performance Gradients Option (K954646) |
| --- | --- | --- |
| Operational Features | Same | SCAN capabilities include: Pilot positioning on three different reference images. Preloaded anatomical protocol categories.
VIEW capabilities include: Multi planar reconstruction and curvilinear reformatting.
FILM capabilities include: ability to set film formats and load print queue directly from Display/Database computer. (See K931544) |
| Standard Imaging Sequences | Same | 2DFT: Field Echo, Spin Echo, Multiple Echo, Inversion Recovery and FAST.
3DFT: FAST |
| Acquisition and Reconstruction Techniques | Same | Main features include: Multi-angle oblique, presaturation, phase conjugate symmetry, TrueRes and TrueSlice. (See K931544) |
PICKER INTERNATIONAL, INC.
(EDGE SE)
11/13/96
I - 5
{5}
| ITEM | EDGE/VISTA with System Enhancement Package | Predicate Devices EDGE/VISTA (K931544) & High Performance Gradients Option (K954646) |
| --- | --- | --- |
| Optional Receive Only Coils | All coils for the EDGE/VISTA are compatible with the system enhancement package.
New coils/accessories described in this submission:
Quadrature Lower Extremity Coil
Coil Combiner | Large Joint Coil
Small Joint Coil
Volume Neck Coil
Quad Spine Coil and Positioner
General Purpose Flex Coil
Bilateral TMJ Coil (Linear and Phased Array versions)
Bilateral Breast Coil
C/T/L Phased Array
Pelvic Phased Array
Head-Neck Vascular Phased Array
Flexible Body Coil (Quadrature and Phased Array versions)
Quadrature Wrist Coil
Shoulder Phased Array
(See K931544, K932693, K941996, K944469, K952530, K962117, K963356) |
PICKER INTERNATIONAL, INC.
(EDGE SE)
11/13/96
I - 6
{6}
| ITEM | EDGE/VISTA with System Enhancement Package | Predicate Devices EDGE/VISTA (K931544) & High Performance Gradients Option (K954646) |
| --- | --- | --- |
| Software Options | Software options are the same except for the following additions:
Variable Fast Spin Echo: 2 and 32 echo sequences have been added. | Angiography
Cardiac Imaging
Variable Fast Spin Echo (4, 8, 16, 48 and 64 echo sequences with minimum TE’s of 16 to 32 msec)
Gradient-Recalled and Spin-Echo (GRaSE) Technique
Echo Planar Imaging
(See K931544, K945397, K945828, K954646) |
| Time Varying Magnetic Field | All gradient performance levels:
Normal Operating Mode: dB/dt ≤ 40 T/s
First Controlled Operating Mode: 40 T/s < dB/dt ≤ 60 T/s | Standard EDGE: dB/dt ≤ 20 T/s for all sequences.
(see K931544)
EDGE with high performance gradients:
Normal Operating Mode: dB/dt ≤ 40 T/s
First Controlled Operating Mode: 40 T/s < dB/dt ≤ 60 T/s
(see K954646) |
PICKER INTERNATIONAL, INC.
(EDGE SE)
11/13/96
I - 7
{7}
| ITEM | EDGE/VISTA with System Enhancement Package | Predicate Devices EDGE/VISTA (K931544) & High Performance Gradients Option (K954646) |
| --- | --- | --- |
| Radiofrequency Absorption | Normal Operating Mode: Limited to a maximum level of 1.2 W/kg.
First Level Controlled Operating Mode: Limited to a maximum value of 3.2 W/kg.
Change in method for acquiring the RF scale factor. | Normal Operating Mode: Limited to a maximum level of 1.2 W/kg.
First Level Controlled Operating Mode: Limited to a maximum value of 2.4 W/kg. (See K931544) |
| Acoustic Noise
Typical
Worst Case | 80.6 dBA (average)
93.3 dB (peak)
115.1 dBA (average)
123.8 dB (peak) | 90.7 dBA (average)
100.0 dB (peak)
118.7 dBA (average)
125.6 dB (peak)
(See K931544) |
PICKER INTERNATIONAL, INC.
(EDGE SE)
11/13/96
I - 8
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.