DELIVERY SYSTEM(SDS 1400 XL)/DELIVERY SYSTEM CUSPIDOR(SDS1450,SDS1650 XL)

K964619 · Sds Dental, Inc. · EIA · Jan 22, 1997 · Dental

Device Facts

Record IDK964619
Device NameDELIVERY SYSTEM(SDS 1400 XL)/DELIVERY SYSTEM CUSPIDOR(SDS1450,SDS1650 XL)
ApplicantSds Dental, Inc.
Product CodeEIA · Dental
Decision DateJan 22, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6640
Device ClassClass 1

Indications for Use

The 1400 XL Deluxe Dental Delivery System is designed to conveniently position the dentist's dynamic instruments for use during dental procedures.

Device Story

Chair-mounted dental delivery system; mounts to 2-inch post. Provides pneumatic power and water coolant to three dynamic instruments (e.g., handpieces) and one air/water syringe. Operation: automatic instrument selection via pneumatic valve in hanger assembly; activation via foot-controlled pneumatic valve. Features: pneumatic brake for arm positioning; master on/off control; manual adjustments for air/water coolant volume; water line purge toggle. Utilities supplied via polyurethane tubing in flexible umbilical connected to floor-mounted utility box with filter/regulator assemblies. Used by dentists in clinical settings to facilitate instrument access during procedures.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Materials: PVC, polyurethane tubing, stainless steel, painted steel, anodized/plated alloys, ABS plastics. Energy: Pneumatic (80 PSI air, 40 PSI water). Form factor: Chair-mounted, three-section arm system (rigid, flex, control mount). Connectivity: Hard-plumbed to floor-mounted utility box. Sterilization: Autoclavable syringe tips. Control: Mechanical pneumatic valves and toggles.

Indications for Use

Indicated for use by dental professionals to position and supply power to dynamic dental instruments, such as handpieces and air/water syringes, during dental procedures.

Regulatory Classification

Identification

A dental operative unit and accessories is an AC-powered device that is intended to supply power to and serve as a base for other dental devices, such as a dental handpiece, a dental operating light, an air or water syringe unit, and oral cavity evacuator, a suction operative unit, and other dental devices and accessories. The device may be attached to a dental chair.

Reference Devices

Submission Summary (Full Text)

{0} SDS SUMMIT DENTAL SYSTEMS K964619 510 W. 21st Avenue, Suite 141 • Ft. Lauderdale, Florida USA 33309 • (954) 733-4302 • (800) 275-3368 • FAX (954) 733-4303 510(K) Summary JAN 22 1997 Contact Person: Cesar Coral President Date Prepared: September 24, 1996 Trade Name: SDS Model 1400 XL Deluxe Dental Delivery System Common Name: Chair Mounted Dental Unit The 1400 XL Deluxe Dental Delivery System is designed to conveniently position the dentist's dynamic instruments for use during dental procedures. The arm system the control head is affixed to consists of three pieces. The first section is a rigid arm that mounts on a two inch outside diameter post. Attached to the end of the rigid arm section is a flex arm section that allows for vertical positioning. The flex arm section contains a pneumatic brake that helps provide positive positioning. The third piece is a control mount that allows flexibility in the horizontal positions. The delivery system provides connections for three pneumatically operated instruments. The common instrument used is the dental handpiece. The unit provides automatic instrument selection by means of a three way pneumatic valve located in each instrument hanger assembly. The automatic selection takes place when a static air pressure signal is removed from the top of a diaphragm located in the handpiece control block. The actual activation of the instrument is initiated by pressing on a foot control containing a three way pneumatic valve. The delivery system also provides an air and water syringe for use as the dental procedure may require. The air and water syringe has an autoclavable syringe tip. The control also houses a master on/off control, air volume adjustments for the pneumatic instruments, water coolant volume adjustments, flex arm brake toggle, and an air coolant volume adjustment. There is also a toggle valve located on the control that allows for water purging of the water coolant lines that supply the pneumatic instruments with water coolant. The air and water utilities are supplied through polyurethane tubing housed in a flexible umbilical. The components that make up this delivery system in many instances are generic to the dental industry and are made of the same materials utilized by other manufacturers that provide an equivalent product. These materials include poly vinyl compounds (PVC), polyurethane tubing, stainless steel, painted steel, anodized and plated alloy compounds, and ABS plastics. These manufacturers would include, but not be limited to, Forest Medical Products, Hillsboro, OR, Beaverstate Dental Inc., Tualatin, OR, Ampco Dental Equipment Company, Santa Maria, CA, and Marus Dental International, Bend, OR. {1} The delivery system is designed as a dental chair mounted unit and mounts on a two inch steel post. The two inch post is accepted as a dental industry standard. The polyurethane air and water supply lines are connected in a floor mounted metal utility box. The utilities should be provided by contractors, and/or plumbers, that bring the utilities to the required space in compliance with all local codes. The air and water lines have both manual shut off valves and pneumatically controlled automatic shut valves. The pneumatic shut off valves are controlled by the master on/off toggle located on the control section of the delivery system. Also provided in the utility box for both the air and water are filter/regulator assemblies. Recommended air pressure is 80 PSI. Recommended water pressure is 40 PSI.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...