K964617 · GE Medical Systems · IYO · Oct 3, 1997 · Radiology
Device Facts
Record ID
K964617
Device Name
GE LOGIQ 700 DIAGNOSTIC ULTRASOUND SYSTEM
Applicant
GE Medical Systems
Product Code
IYO · Radiology
Decision Date
Oct 3, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1560
Device Class
Class 2
Attributes
Pediatric
Indications for Use
LOGIQ 700 with 3D is a general purpose ultrasound imaging system intended for use by, or under the direction of, a qualified physician in the evaluation of soft tissue and vascular disease of the head, neck chest, abdomen pelvis, male and female reproductive organs, limbs and pregnant uterus of the adult, pediatric or neonatal patient.
Device Story
GE LOGIQ 700 3D is a mobile diagnostic ultrasound console (128-channel) with integrated 3D cine capabilities. System utilizes various transducers to acquire ultrasound signals from soft tissue and vascular structures. Signals are processed to generate B-mode, M-mode, and Doppler (Color, Power, PWD) images displayed on a color monitor. Operated by qualified physicians or under their direction in clinical settings. Output provides real-time or stored diagnostic images for evaluation of disease in head, neck, chest, abdomen, pelvis, reproductive organs, limbs, and pregnant uterus. Benefits include non-invasive visualization of internal anatomy to support clinical decision-making.
Clinical Evidence
Bench testing only. Device evaluated for acoustic output, biocompatibility, and thermal, electrical, and mechanical safety. Conformance to applicable medical device safety standards verified through independent evaluation.
Technological Characteristics
Mobile console (70x120x120 cm) with 128-channel capability. Includes adjustable keyboard, A/N display, and color video monitor. Supports various transducers (e.g., 326s, 618E, 348C, LA39, 547L, 618C, 227s, M3C, 739L, 548C, 546L, M12L). Energy source: electrical. Connectivity: integrated image storage/hard-copy devices. Sterilization: not specified.
Indications for Use
Indicated for diagnostic ultrasound imaging or Doppler analysis (B, M, PWD, Color, Power Doppler) of adult, pediatric, and neonatal patients. Applications include fetal, abdominal, cardiac, small organ (breast, testes, thyroid), and peripheral vessel evaluation. Contraindicated for in vitro fertilization and percutaneous umbilical blood sampling.
Regulatory Classification
Identification
An ultrasonic pulsed echo imaging system is a device intended to project a pulsed sound beam into body tissue to determine the depth or location of the tissue interfaces and to measure the duration of an acoustic pulse from the transmitter to the tissue interface and back to the receiver. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A biopsy needle guide kit intended for use with an ultrasonic pulsed echo imaging system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
GE Medical Systems LOGIQ 700 diagnostic ultrasound system (K930768)
Submission Summary (Full Text)
{0}
OCT 3 1997
GE LOGIQ 700 3D Premarket Notification - October 25, 1996
K964617
510(k) Summary of Safety and Effectiveness: GE LOGIQ 700 3D Cine
This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
**Identification of Submitter:**
GE Medical Systems
PO Box 414
Milwaukee, WI 53201
**Contact:**
D. Duersteler, Safety and Regulatory Programs
414-647-4385
**Date Prepared:**
October 25, 1996
**Product Identification:** GE LOGIQ 700 Diagnostic Ultrasound System with 3D Cine.
**Marketed Devices:** The GE LOGIQ 700 Diagnostic Ultrasound System with 3D is of a comparable type and substantially equivalent to the GE Medical Systems LOGIQ 700 diagnostic ultrasound system, 510(k) Number K930768, currently in commercial distribution:
**Device Description:** The LOGIQ 700 diagnostic ultrasound system with 3D is the same in all other aspects as the currently marketed LOGIQ 700, consisting of a mobile console approximately 70 cm wide, 120 cm deep and 120 cm high that provides full 128 channel capability, and assorted transducers. The user interface is an adjustable height keyboard, small A/N display panel and a color video display monitor. Optional image storage or hard-copy devices are integrated into the design.
**Indications for Use:** LOGIQ 700 with 3D is a general purpose ultrasound imaging system intended for use by, or under the direction of, a qualified physician in the evaluation of soft tissue and vascular disease of the head, neck chest, abdomen pelvis, male and female reproductive organs, limbs and pregnant uterus of the adult, pediatric or neonatal patient.
**Comparison with Predicate Device:** The GE LOGIQ 700 with 3D is comparable in key safety and effectiveness features, uses the same design, construction, and materials, and has the same intended uses and operating modes as the predicate device.
**Summary of Studies:** The device has been evaluated for acoustic output, biocompatibility, and thermal, electrical and mechanical safety, and has been found to conform with applicable medical device safety standards.
**Conclusion:** Intended uses and other key features are consistent with traditional clinical practice, FDA guidelines, and established methods of patient examination. The design and development process of the manufacturer conforms with GMP standards. The product is designed to conform with applicable medical device safety standards and compliance is verified through independent evaluation. Diagnostic ultrasound has accumulated a long history of safe and effective performance. Therefore, it is the opinion of GE Medical Systems that the GE LOGIQ 700 with 3D Cine is substantially equivalent with respect to safety and effectiveness to devices currently cleared for market.
{1}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
David Duersteler
Safety/Regulatory Project Engineer
GE Medical Systems
PO Box 414
Milwaukee, WI 53201
Re: K964617
Dated: August 25, 1997
Received: August 27, 1997
Regulatory Class: II
21 CFR 892.1550/Procode: 90 IYO
Dear Mr. Duersteler:
We have reviewed your section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug and Cosmetic Act. You may, therefore, market the device, subject to the general controls provisions Act (Act). The general controls provisions of the Act include requirements for registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use with the LOGIQ 700 with 3D View, as described in your premarket notification:
| Transducer Model Number | | | |
| --- | --- | --- | --- |
| 326s | 618E | 348C | LA39 |
| 547L | 618C | 227s | M3C |
| 739L | 548C | 546L | M12L |
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}
Page 2 - Mr. David Duersteler
Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
Please be advised that the determination above is based on the fact that no medical devices have been demonstrated to be safe and effective for *in vitro* fertilization or percutaneous umbilical blood sampling, nor have any devices been marketed for these uses in interstate commerce prior to May 28, 1976, or reclassified into class I (General Controls) or class II (Special Controls). FDA considers devices specifically intended for *in vitro* fertilization and percutaneous umbilical blood sampling to be investigational, and subject to the provision of the investigational device exemptions (IDE) regulations, 21 CFR, Part 812. Therefore, your product labeling must be consistent with FDA’s position on this use.
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, “Misbranding by reference to premarket notification” (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597 or at its Internet address “http://www.fda.gov/cdrh/dsmamain.html”.
If you have any questions regarding the content of this letter, please contact Maureen Butler at (301) 594-1212.
Sincerely yours,

Lillian Yin, Ph.D.
Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure(s)
{3}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 326s Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | X | X | X | | X | X | | X | |
| Abdominal | | X | X | X | | X | X | | X | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | X | X | X | | X | X | | X | |
| Cardic Pediatric | | X | X | X | | X | X | | X | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)

(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
{4}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 547L Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | X | X | X | | X | X | | X | |
| Abdominal | | | | | | | | | | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) | | X | X | X | | X | X | | X | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardic Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | X | X | X | | X | X | | X | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
{5}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 739L Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | X | X | X | | X | X | | X | |
| Abdominal | | | | | | | | | | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | X | X | X | | X | X | | X | |
| Small Organ (Specify) | | X | X | X | | X | X | | X | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | X | X | X | | X | X | | X | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David G. Seppman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
{6}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 618E Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | X | X | X | | X | X | | X | |
| Abdominal | | | | | | | | | | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | X | X | X | | X | X | | X | |
| Trans-vaginal | | X | X | X | | X | X | | X | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Sigman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
{7}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 618C Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | | | | | | | | | |
| Abdominal | | | | | | | | | | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | X | X | X | | X | X | | X | |
| Small Organ (Specify) | | | | | | | | | | |
| Neonatal Cephalic | | X | X | X | | X | X | | X | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | X | X | X | | X | X | | X | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David G. Segman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
{8}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 548C Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | X | X | X | | X | X | | X | |
| Abdominal | | X | X | X | | X | X | | X | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) | | X | X | X | | X | X | | X | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardic Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Seyman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
{9}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 348C Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | X | X | X | | X | X | | X | |
| Abdominal | | X | X | X | | X | X | | X | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) | | X | X | X | | X | X | | X | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardic Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Seyman
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
{10}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 227s Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | X | X | X | | X | X | | X | |
| Abdominal | | X | X | X | | X | X | | X | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) | | X | X | X | | X | X | | X | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardic Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH. Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
{11}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 546L Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | X | X | X | | X | X | | X | |
| Abdominal | | | | | | | | | | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) | | X | X | X | | X | X | | X | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardic Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | X | X | X | | X | X | | X | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David G. Segerson
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
{12}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 LA39 Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | X | X | X | | X | X | | X | |
| Abdominal | | | | | | | | | | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | X | X | X | | X | X | | X | |
| Small Organ (Specify) | | X | X | X | | X | X | | X | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardiac Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | X | X | X | | X | X | | X | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David A. Seymour
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
{13}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 M3C Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | X | X | X | | X | X | | X | |
| Abdominal | | X | X | X | | X | X | | X | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) | | X | X | X | | X | X | | X | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardic Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | | | | | | | | | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David C. Sypson
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
{14}
Ultrasound Device Indications Statement
Page 1 of 1
510(k) Number (if known): K964617
Device Name: LOGIQ 700 M12L Transducer
Fill out one form for each ultrasound system or transducer
Indications for use: Diagnostic ultrasound imaging or Doppler analysis of the human body as follows:
Mode of Operation
| Clinical Application | A | B | M | PWD | CWD | Color Doppler | Power (Amplitude) Doppler | Color Velocity Imaging | Combined (Specify) | Other (Specify) |
| --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- |
| Opthalmic | | | | | | | | | | |
| Fetal and other | | X | X | X | | X | X | | X | |
| Abdominal | | | | | | | | | | |
| Musculoskeletal | | | | | | | | | | |
| Intra-operative | | | | | | | | | | |
| Neurosurgical | | | | | | | | | | |
| Pediatric | | | | | | | | | | |
| Small Organ (Specify) | | X | X | X | | X | X | | X | |
| Neonatal Cephalic | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | |
| Cardiac Adult | | | | | | | | | | |
| Cardic Pediatric | | | | | | | | | | |
| Trans-esophageal | | | | | | | | | | |
| Trans-rectal | | | | | | | | | | |
| Trans-vaginal | | | | | | | | | | |
| Intra-luminal | | | | | | | | | | |
| Trans-urethral | | | | | | | | | | |
| Peripheral vessel | | X | X | X | | X | X | | X | |
| Laparoscopic | | | | | | | | | | |
Other Indications or Modes: Small organ includes breast, testes, thyroid. Fetal and Other includes urological. Combined modes are B/M, B/Color, B/PWD, B/Color/PWD.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109)
David C. Sigmon
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices
510(k) Number K964617
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.