K964590 · Chromatics Color Sciences Intl., Inc. · MQM · Jul 24, 1997 · Clinical Chemistry
Device Facts
Record ID
K964590
Device Name
COLORMATE III
Applicant
Chromatics Color Sciences Intl., Inc.
Product Code
MQM · Clinical Chemistry
Decision Date
Jul 24, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1113
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The Colormate III™ is a computer assisted non-invasive transcutaneous bilirubinometer which, through colorimetric technology, illuminates the skin of newborn babies and measures the yellow content of the skin color. The incremental changes of these color readings measured over time are compared to the newborn's baseline Colormate III™ readings. These data are then automatically processed to provide a numerical index of predicted bilirubin count in mg/dl that has been shown to correlate with total serum bilirubin concentration within a clinically useful range.
Device Story
Colormate III is a computer-assisted, non-invasive transcutaneous bilirubinometer; used in clinical settings by physicians to monitor newborns for hyperbilirubinemia. Device illuminates infant skin; measures yellow content via colorimetric technology. Input data (tristimulus color coordinates) are processed via internal algorithms to calculate a numerical index of estimated bilirubin (mg/dl). Output provides clinicians with a quantitative estimate of serum bilirubin concentration, replacing subjective visual assessment. Clinical benefit includes improved accuracy and precision in detecting hyperbilirubinemia, facilitating timely clinical management decisions.
Clinical Evidence
Clinical study of 1,317 newborns (diverse racial mix) compared Colormate III measurements to physician visual assessment and serum bilirubin blood tests. 171 newborns with 360 concurrent measurements were analyzed. Colormate III showed higher correlation with serum bilirubin (r=0.90) than visual assessment (r=0.67, p<0.05). Sensitivity for detecting bilirubin above 12.0 mg/dl was >89% for Colormate III vs 61% for visual assessment. Bias was <1.6 mg/dl in the 4.0-17.0 mg/dl range. Reproducibility studies showed coefficients of variation <3%.
Technological Characteristics
Non-invasive transcutaneous bilirubinometer using colorimetric technology. Measures skin reflectance in the visible spectrum (400-540 nm). Generates tristimulus color coordinates. Computer-assisted processing of color data to estimate serum bilirubin. Standalone device form factor.
Indications for Use
Indicated for use as an aid to physicians in monitoring newborn babies for the development of hyperbilirubinemia. Used for initial measurement and periodic monitoring of incremental changes in skin yellow content compared to baseline. Patients with results indicative of hyperbilirubinemia require re-evaluation by a physician.
Regulatory Classification
Identification
A bilirubin (total and unbound) in the neonate test system is a device intended to measure the levels of bilirubin (total and unbound) in the blood (serum) of newborn infants to aid in indicating the risk of bilirubin encephalopathy (kernicterus).
{0}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
2098 Gaither Road
Rockville MD 20850
Thomas M. Tsakeris
President
Devices & Diagnostics Consulting Group
16809 Briardale Road
Rockville, Maryland 20855
Re: K964590
Colormate™
Regulatory Class: II
Product Code: MQM
Dated: May 23, 1997
Received: May 23, 1997
Dear Mr. Tsakeris:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
JUL 24 1997
{1}
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven I. Gutman, M.D., M.B.A.
Director
Division of Clinical Laboratory Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}
APPENDIX F
(510 (K)) Summary
JUL 24 1997
K964590
This summary is provided in accordance with the Safe Medical Devices Act of 1990 (SMDA), ai U.S.C. §360c(i)(3) and FDA's implementing regulations, 21 C.F.R. 807.92.
## 1.0 Submitter of 510(k)
Chromatics Color Sciences International, Inc.
5 East 80th Street
New York, New York 10021-0109
Attention: Ms. Darby S. Macfarlane, CEO
Telephone: 212-717-6544
Facsimile: 212-717-6675
## 2.0 Name of Device
### 2.1 Trade/Proprietary Name
Colormate III™
### 2.2 Common/Usual Name
Transcutaneous Bilirubinometer (Colorimeter)
### 2.3 Classification Name
Bilirubin (total and unbound) in the neonate test system (21 CFR 862.1113)
### 3.0 Reason for Submitting the 510(k)
This 510(k) is being submitted to notify FDA of Chromatics Color Sciences International’s (CCSI) desire to commercially distribute for the first time the Colormate III.
## 4.0 Device Description
The Colormate III™ (Colormate III or CM III) is a computer assisted noninvasive transcutaneous bilirubinometer which, through colorimetric technology, illuminates the skin of newborn babies and measures the yellow content of the skin color. The incremental changes of these color readings measured over time are compared to the newborn’s baseline readings. These data are then automatically processed to provide a numerical index of estimated bilirubin count in mg/dl that has been shown to correlate with total serum bilirubin within a clinically useful
{3}
range. The Colormate III, like other colorimeters, generates a set of tristimulus color coordinates for each measurement. It then performs a series of algorithms to produce final measurement results.
## 5.0 Indications for Use & Intended Use
### 5.1 Indications For Use
The Colormate III is to be used as an aid to the physician in monitoring the status of newborn babies for the development of hyperbilirubinemia. Following the physician's examination within the first hours after birth, newborn babies are initially measured and periodically monitored by the Colormate III for incremental changes in the yellow content of the skin color as compared to the baseline Colormate III measurements. Babies with Colormate III test results indicative of hyperbilirubinemia are to be re-evaluated by the attending physician for appropriate patient management.
### 5.2 Intended Use
The Colormate III is a computer assisted non-invasive transcutaneous bilirubinometer which, through colorimetric technology, illuminates the skin of newborn babies and measures the yellow content of the skin color. The incremental changes of these color readings measured over time are compared to the newborn's baseline Colormate III readings. These data are then automatically processed to provide a numerical index of predicted bilirubin count in mg/dl that has been shown to correlate with total serum bilirubin concentration within a clinically useful range.
### 6.0 Substantial Equivalence
The Colormate III is "substantially equivalent" (as defined in 21 U.S.C. § 360c(i)) in principle, performance, and safety and effectiveness to at least two legally marketed predicate devices. The first predicate device, the Unistat Bilirubinometer manufactured by Leica, Inc. (K922770), is a bilirubinometer which operates by measuring the absorption of visible light through a sample of unreacted serum. The absorption at several wavelengths determines the level of bilirubin present in the sample. The second device, the Ingram Icterometer, commercially distributed preenactment (marketed since 1950's) by Cascade Health Care Products, Salem, Oregon, is a noninvasive aid used as a color reference by the physician during his/her visual assessment of the newborn's skin for jaundice or yellowing. The yellow color of the subcutaneous tissue is matched to a corresponding "yellow hue" stripe for a reference score.
{4}
The Colormate III and predicate devices can be used as aids in monitoring the development of hyperbilirubinemia in the newborn.
Further, the Colormate III has comparable technological characteristics to those of the predicate devices. The underlying principles of the Colormate III and the predicate devices are the same in that the Colormate III and the predicate devices base their measurements on the unique absorbance characteristics of bilirubin in the visible spectrum. That is, the level of bilirubin (whether intravascular or extravascular) is easily detected by its characteristic absorption of visible light. Bilirubin strongly absorbs light in the wavelengths between approximately 400 to 540 nm. These correspond to the green, blue and violet portions of the visible spectrum. Therefore, an increase in bilirubin is seen as a direct absorption of these colors in transmitted or reflected light or as a relative increase in the complementary color of yellow. This latter effect is seen in the yellowing (jaundice) of the infant's skin during hyperbilirubinemia.
The Colormate III, like the predicate Unistat Bilirubinometer, directly measures the bilirubin concentration via its absorption of visible light. The Colormate III measures the interstitial subcutaneous concentration and the Unistat measures the intravascular concentration.
The Colormate III and the noninvasive predicate device, as well as the visual assessment by the physician, operate on the principle that increasing levels of serum bilirubin during hyperbilirubinemia results in corresponding increases in the level of interstitial bilirubin. The painted stripes of precisely graded yellow hues in the Ingram Icterometer provide the physician with a visual aid to gauge the depth of jaundice in the newborns and estimate the serum bilirubin levels. The Colormate III expands on this concept. The Colormate III replaces the human eye with a colorimeter having more accuracy and precision and provides the physician with a numerical index of estimated serum bilirubin in mg/dl.
## 7.0 Performance Data
The performance testing provided demonstrates that the Colormate III is statistically significantly more accurate and precise than visual assessment alone and can aid the physician in monitoring hyperbilirubinemia in the newborn. Further, these data demonstrate that the Colormate III interstitial readings are highly correlated (r = 0.90, p<0.01) with serum or intravascular bilirubin readings within a clinically useful range.
A series of reproducibility studies were conducted using the Colormate III. Measurements were made using (A) a standard
{5}
calibration tile, (B) adult subjects and (C) newborn babies with varying degrees of jaundice. The coefficients of variation (CV) for the tristimulus color coordinates calculated by the Colormate III and used in its numerical index are below 3%.
Prior to undertaking a pivotal clinical trial to evaluate the Colormate III's performance, two phases of studies were conducted to validate the clinical approach and test criteria and determine calibration algorithms for subsequent validation.
A clinical study was conducted to compare (1) the performance of the Colormate III using colorimetric measurement technology to estimate serum bilirubin concentration (BRC) from the yellow content of the color of the skin and (2) the ability of the physician to visually assess the same yellow color of the skin and make a clinical estimate of the BRC. Neonates enrolled in the study underwent the usual course of physical examination (including a visual assessment of the yellowness of the skin for bilirubinemia), monitoring and care. Colormate III skin color measurements, employing a number of anatomical sites on an infant, were made concurrently with serum bilirubin blood tests and physicians' visual estimates.
The trial involved a sequence of 1317 newborns. The study included a mix of the following races: 487 (37%) Caucasians, 298 (23%) Blacks, 427 (32%) Hispanics, 82 (6%) Orientals and 20 (2%) others. One Thousand Thirty-seven (1037) babies were normal (i.e., did not develop hyperbilirubinemia) and, therefore, did not undergo blood serum tests. One Hundred Nine (109) babies did not have initial measurements made within acceptable time frames. The remaining 171 premature and full term newborns with 360 concurrent Colormate III measurements, clinical visual estimates and blood serum tests demonstrated that the Colormate III was better correlated (p<0.05) with serum bilirubin concentration (r = .90) than was the clinicians' visual assessment (r = 67).
The estimation of bias between the Colormate III test and comparative serum did not exceed 1.6 mg/dl in the clinically useful range of 4.0 to 17.0 mg/dl (less than 1 mg/dl from 7 to 17 mg/dl). In addition, the correlation of Colormate III with serum readings was relatively unaffected by race (Caucasian r = .91, Black r = .88, Hispanic r = .89, Oriental r = .91).
Additionally, the relative sensitivities of the visual measurement and the Colormate III results to serum values were computed and compared using a contingency table approach to data presentation and a cutoff of 12.0 mg/dl of bilirubin as an acceptable level of clinical concern. The clinicians' visual assessment agreed above the cutoff of 12.0 with serum only 61% of the time while the Colormate III agreement (sensitivity of the Colormate III) was over 89% in the same patients. The sensitivity of the Colormate III™ was statistically
4
{6}
significantly better than that for the visual assessment, in their correlation to serum values above and below the accepted level of clinical concern of 12.0 mg/dl, p<.05. Further, exactly the same conclusions were reached when higher or lower levels were examined (e.g., 8.0 or 14.0). Sensitivity calculations again demonstrated significantly higher sensitivity for the Colormate III than that for the visual assessment.
In conclusion, the Colormate III performance was superior to visual examination in estimating the true level of serum bilirubin, in addition to an excellent corrrelation to serum bilirubin throughout a clinically useful range.
{7}
# Indications For Use
Device Name: Colormate III™
## Indications For Use:
The Colormate III™ is to be used as an aid to the physician in monitoring the status of newborn babies for the development of hyperbilirubinemia. Following the physician's examination within the first hours of birth, newborn babies are initially measured and periodically monitored by the Colormate III™ for incremental changes in the yellow content of the skin color as compared to the baseline Colormate III™ measurements. Babies with Colormate III™ test results indicative of hyperbilirubinemia are to be re-evaluated by the attending physician for appropriate patient management.
## Intended Use:
The Colormate III™ is a computer assisted non-invasive transcutaneous bilirubinometer which, through colorimetric technology, illuminates the skin of newborn babies and measures the yellow content of the skin color. The incremental changes of these color readings measured over time are compared to the newborn's baseline Colormate III™ readings. These data are then automatically processed to provide a numerical index of predicted bilirubin count in mg/dl that has been shown to correlate with total serum bilirubin concentration within a clinically useful range.
* A rationale for clinical utility is provided on the following page.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use: ☐ OR Over-The-Counter: ☐
(Per 21 CFR 801.109)
(Optional Format 1-2-96)

Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.