ROCHE COBAS AMPLICOR ANALYZER

K964506 · Roche Molecular Systems, Inc. · JJF · May 21, 1997 · Clinical Chemistry

Device Facts

Record IDK964506
Device NameROCHE COBAS AMPLICOR ANALYZER
ApplicantRoche Molecular Systems, Inc.
Product CodeJJF · Clinical Chemistry
Decision DateMay 21, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.2170
Device ClassClass 1

Indications for Use

The COBAS AMPLICOR Analyzer is an in vitro diagnostic device intended for use in clinical laboratories for the automation of the AMPLICOR™ Polymerase Chain Reaction test procedures.

Device Story

COBAS AMPLICOR is an automated bench-top batch analyzer for PCR-based nucleic acid amplification and detection. Input: prepared clinical specimens in A-tubes. Operation: internal computer controls automated pipetting, thermal cycling, incubation, wash stations, and photometric detection. System uses probe-coated magnetic microparticles for amplicon capture; enzymatic color reaction for detection. Output: qualitative results based on absorbance cut-offs at 660 nm. Used in clinical laboratories by trained personnel. Benefits: automates manual molecular diagnostic steps; improves throughput and consistency; eliminates manual stop-reagent requirements.

Clinical Evidence

Clinical comparison study of 639 specimens (266 endocervical, 93 male urethral, 280 male urine) comparing COBAS AMPLICOR to manual AMPLICOR microwell assay. McNemar’s test P > 0.4 for all specimen types, indicating no statistical difference in results. Precision study (multi-operator, 3-day) evaluated reproducibility using spiked McCoy cell suspensions (0-50 IFU/PCR); total CVs ranged from 3.7% to 117.6% (highest at 0 IFU/PCR).

Technological Characteristics

Bench-top automated analyzer. Components: thermal cycler, pipetting station, incubation station, wash station, photometer. Sensing: fixed wavelength photometer (660 nm). Connectivity: internal computer for system control and diagnostics. Materials: polystyrene specimen cups, primer-coated magnetic microparticles. Software: internal control system for run management and result calculation.

Indications for Use

Indicated for use in clinical laboratories to automate PCR test procedures for the qualitative determination of Chlamydia trachomatis plasmid DNA in female endocervical, male urethral, and male urine specimens from symptomatic and asymptomatic patients.

Regulatory Classification

Identification

A micro chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. The distinguishing characteristic of the device is that it requires only micro volume samples obtainable from pediatric patients. This device is intended for use in conjunction with certain materials to measure a variety of analytes.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} Roche Molecular Systems a subsidiary of Hoffmann-La Roche K964506 MAY 21 1997 Roche Molecular Systems, Inc. 1080 U.S. Highway 202 Somerville, New Jersey 08876 # 510(k) Summary ## COBAS AMPLICOR™ Analyzer In accordance with the Safe Medical Devices Act of 1990, a 510(k) summary is provided as outlined in 21 CFR 807.92. ### I. Identification of 510(k) Sponsor: Roche Molecular Systems, Inc. 1080 U.S. Highway 202 Somerville, New Jersey 08876-1760 510(k) submission dated November, 1996 Contact: Alex Wesolowski Telephone: (908) 253-7540 Facsimile: (908) 253-7547 ### II. Device Name: Trade/Proprietary Name - COBAS AMPLICOR™ Analyzer Common or Usual Name - Automated batch analyzer for nucleic acid amplification and detection Classification Name - None applicable ### III. Identification of legally marketed device to which the 510(k) sponsor claims equivalence: The primary operational components of the COBAS AMPLICOR analyzer are the pipettor, thermal cycler, wash stations, and the photometric detection system. As independent components, each is categorized as Class I devices under 21 CFR 862 (Table 1). The COBAS AMPLICOR combines these operational functions into a single automated instrument that provides an interpretative qualitative result. The performance of the COBAS AMPLICOR is substantially equivalent to several currently marketed device used independently to perform the same functions. 1 120 {1} 1 121 TABLE 1 Predicate Devices for the COBAS AMPLICOR Analyzer | Predicate Device | Regulatory Class | Classification Number | Predicate Product Name | Predicate 510(k) Number | | --- | --- | --- | --- | --- | | Automated Pipettor | Class 1 | 862.2750 | Matrix Technologies Impact Pipettor | K940390 | | Thermal cycler (incubator) | Class 1 exempt | 862.2750 | Perkin-Elmer 9600 Thermal cycler | NA | | Incubator | Class 1 exempt | 862.2050 | Fisher Dry Heat Incubator Model 630D | NA | | Automated Microplate Washer | Class 1 exempt | 866.2500 | Bio-Tek Instruments Automated Microplate Washer Model Model ELP40 | NA | | Automated Microplate Reader | Class 1 | 862.2300 | Bio-Tek Instruments Automated EAI Microplate Reader Models Elx800/EL800 | K950104 | The COBAS AMPLICOR analyzer is designed for exclusive use with COBAS AMPLICOR reagents. The Roche AMPLICOR Chlamydia trachomatis Test (microwell plate) and the COBAS AMPLICOR Chlamydia trachomatis Test are both intended for the qualitative determination of Chlamydia trachomatis plasmid DNA in female endocervical, male urethral, and male urine specimens from symptomatic and asymptomatic patients. The Roche COBAS AMPLICOR Chlamydia trachomatis Test performed on the Roche COBAS AMPLICOR analyzer is substantially equivalent to the Roche AMPLICOR Chlamydia trachomatis Test (K922906/C, June 15,1993) performed on the Perkin-Elmer 9600 thermal cycler and the Bio-Tek EL800 (K950104) microtiter plate reader. ## IV. Description of the Device: The COBAS AMPLICOR is a flexible, automated bench top batch analyzer that automates the amplification and detection steps of the Polymerase Chain Reaction (PCR) process. The COBAS AMPLICOR combines the operations of automated sample handling, reagents deliver, thermal cycling, controlled temperature incubation, photometric detection and result reporting into a single automated analyzer. The instrument consists of five major sub-components: (1) a thermal cycler module; (2) an automated pipeting station; (3) an incubation station; (4) a wash station; and (5) a photometer. An internal computer controls and monitors the major components including system and run control, input/output, communication, results calculaton, and system diagnostic tests. Specimens are prepared off-line with AMPLICOR Specimen Preparation reagents before being transferred to amplification tubes (A-tubes) placed on the analyzer. Twelve A-tubes with caps are provided as a molded assembly called an A-ring. The COBAS AMPLICOR automatically performs test specific thermal cycling operations for PCR amplification. Amplified product is automatically transferred to polystyrene specimen cups (D cups) and primer coated magnetic beads are added. Following incubation, unbound material is removed by serial wash steps. Hybridized amplicon is measured by an enzymatically catalyzed color reaction. The intensity of the color is proportional to the amount of infectious organisms present in the specimen and qualitative results are determined based on a specified absorbance cut-off. The amplification and detection operations can be run in parallel or sequentially depending on the number of samples in the run and user preference. In sequential mode, the analyzer automatically proceeds to sample detection after amplification, and the pipettor transfers amplified products from the A-ring in the thermal cycling segments to the D-cups for hybridization, wash and detection. {2} In parallel operation, a second set of samples is amplified as the first set is detected. After amplification of the first set of specimens is complete, the system pauses for transfer of the first set of amplified A-rings to the ambient temperature chambers. A second set of A-rings are then added to the thermal cycler for amplification, and the automated pipettor transfers amplified product from each of the first set of A-tubes in the ambient temperature segments to D-cups for hybridization, wash and detection. ## V. Statement of Intended Use: The COBAS AMPLICOR is an *in vitro* diagnostic device intended for use in the amplification and detection steps of the Polymerase Chain Reaction (PCR) process. ## VI. Summary of the technological characteristics of the new device in comparison to those of the predicate: The COBAS AMPLICOR Analyzer is a flexible, automated, bench top analyzer that automates the amplification and detection steps of the Polymerase Chain Reaction (PCR) process. The principles of operation of the COBAS AMPLICOR are substantially equivalent to manual in-vitro diagnostic PCR testing using general laboratory equipment. The COBAS AMPLICOR Analyzer consists of five major sub-components: (1) a thermal cycler module; (2) an automated pipeting station; (3) an incubation station; (4) a wash station; and (5) a photometer. An internal computer controls and monitors the major components including system and run control, input/output, communication, results calculaton, and system diagnostic tests. The COBAS AMPLICOR Analyzer operating as a system performing automatically various steps such as pipetting, heating and measuring color intensity is substantially equivalent to several currently marketed devices used independently to perform the same functions. The COBAS AMPLICOR is designed to be used in conjunction with AMPLICOR Specimen Preparation Kits (for off-line specimen preparation), AMPLICOR Amplification Kits and COBAS AMPLICOR Detection Kits. The AMPLICOR Test for *Chlamydia trachomatis* used with the Perkin-Elmer 9600 thermal cycler and Bio-Tek plate reader, and the COBAS AMPLICOR *Chlamydia trachomatis* are both intended for the qualitative determination of *Chlamydia trachomatis* from urogenital swab specimens in specimen transport medium and male urine. The COBAS AMPLICOR *Chlamydia trachomatis* Test is substantially equivalent to the AMPLICOR *Chlamydia trachomatis* Test (K922906/C, June 15,1993) performed on the Perkin-Elmer 9600 thermal cycler and the Bio-Tek EEL800 (K950104) microtiter plate reader. ## Similarities and Differences to Comparable Commercial Products The specimen preparation, amplification and detection methods used in the COBAS AMPLICOR system are similar to that previously described for the microwell plate AMPLICOR Test. The notable similarities and differences are as follows: ## Similarities 1. The COBAS AMPLICOR Analyzer provides an automated method for performing the same basic procedures used in manual molecular diagnostic testing. These include, sample handling, reagent preparation, sample incubation, thermal cycling washing and photometric measurement 2. The procedures used specimen preparation, amplification and detection portions of the COBAS AMPLICOR procedure utilize identical reagent formulations as the AMPLICOR microplate assay. Identical reagent formulations include specimen diluent, master mix, free nucleotides, nucleotide primers and probes, TMB chromagen, peroxide substrate, and wash buffers. 3. The COBAS AMPLICOR analyzer performs thermal cycling to tolerances that meet or exceed operational characteristics of conventional general laboratory equipment (e.g. the Perkin Elmer 9600). {3} 4. The COBAS AMPLICOR analyzer measures absorbance using a fixed wavelength photometer with comparable performance to other conventional general laboratory equipment and to automated clinical instrumentation such as the Roche COBAS MIRA. ## Differences 1. The COBAS AMPLICOR system automates all sample handling, washing and pipetting steps for test amplification and detection compared to manual requirements with the AMPLICOR microplate assay. 2. PCR amplified products are captured by hybridization to probe-coated paramagnetic microparticles on the COBAS AMPLICOR compared to probe-coated microwell plates in the AMPLICOR test format. 3. The COBAS AMPLICOR method does not require a stop reagent following the substrate incubation as does the AMPLICOR because the automated system precisely times each portion of the hybridization and detection procedures. 4. The optical density of the detection system is measured at 660 nm with the COBAS AMPLICOR system compared to 450 nm in the AMPLICOR system. The stop reagent used in the AMPLICOR microplate assay modifies the oxidation state and thus, the absorbance maximum, of the chromophore. ## VI. A brief discussion of the nonclinical and clinical performance data ### Non-Clinical Performance #### Precision A multi-operator study was performed to determine the reproducibility of the COBAS AMPLICOR *Chlamydia trachomatis* Test. The study design was based upon the study suggested in the NCCLS document EP5-T2. Three independent operators tested a series of samples once a day for three days. A test run consisted of the amplification and detection of the following samples in triplicate: COBAS AMPLICOR *Chlamydia trachomatis* Test CT (+) and CT (-) kit controls and specimens that were prepared by spiking AMPLICOR Specimen Transport Medium with infected McCoy cell suspensions to obtain levels of 0, 1, 10 and 50 C. *trachomatis* IFU/PCR. Calculations were performed using the Analysis of Variance model suggested in the NCCLS document to derive estimates of between-day, between-operator, within-operator, and total variance for the Test at each sample concentration. The results of this study are presented in Table 2. {4} 1 124 # TABLE 2 COBAS AMPLICOR Chlamydia trachomatis Test Precision Analysis of Variance Test Results | | SAMPLE | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | Chlamydia trachomatis Spiked STM (IFU/PCR) | | | | Kit Controls | | | | 0 | 1 | 10 | 50 | Negative | Positive | | Total Replicates | 27 | 27 | 27 | 27 | 27 | 27 | | Mean Absorbance | 0.004 | 3.856 | 3.685 | 3.639 | 0.004 | 3.831 | | Minimum | 0.000 | 3.646 | 3.220 | 3.220 | 0.000 | 3.279 | | Maximum | 0.017 | 4.000 | 4.000 | 4.000 | 0.011 | 4.000 | | Between Day Variance | 0.0000 | 0.0000 | 0.0040 | 0.0037 | 0.0000 | 0.0022 | | Standard Deviation | 0.0013 | 0.0000 | 0.0632 | 0.0610 | 0.0000 | 0.0468 | | CV (%) | 32.0 | 0.0 | 1.7 | 1.7 | 0.0 | 1.2 | | Between Operator Variance | 0.0000 | 0.0145 | 0.0271 | 0.0230 | 0.0000 | 0.0587 | | Standard Deviation | 0.0029 | 0.1206 | 0.1647 | 0.1517 | 0.0027 | 0.2422 | | CV (%) | 72.4 | 3.1 | 4.5 | 4.2 | 69.7 | 6.3 | | Within Operator Variance | 0.0000 | 0.0059 | 0.0162 | 0.0125 | 0.0000 | 0.0109 | | Standard Deviation | 0.0035 | 0.0767 | 0.1274 | 0.1118 | 0.0026 | 0.1045 | | CV (%) | 87.0 | 2.0 | 3.5 | 3.1 | 67.4 | 2.7 | | Total Variance | 0.0000 | 0.0204 | 0.0474 | 0.0392 | 0.0000 | 0.0718 | | Standard Deviation | 0.0047 | 0.1429 | 0.2176 | 0.1980 | 0.0037 | 0.2679 | | CV (%) | 117.6 | 3.7 | 5.9 | 5.4 | 97.0 | 7.0 | ## Clinical Performance The clinical data reported here were obtained using the AMPLICOR Chlamydia trachomatis Test. The AMPLICOR microwell assay and the COBAS AMPLICOR Chlamydia trachomatis Test procedures are highly conserved and differ only in the method of detection. The COBAS AMPLICOR utilizes a magnetic microparticle based solid support compared to microwell plates and the automated analyzer enables the elimination of a endpoint hydrolysis Stop reagent that changes the absorbance maximum of the chromophore. The COBAS AMPLICOR and AMPLICOR Tests use the exact same specimen preparation, detection and amplification reagents, including identical amplification primer (CP24 and CP27) and detection probe (CP35) sequences. Therefore, the equivalent clinical performance is expected from the microwell and automated COBAS Tests. ## Clinical Results - COBAS AMPLICOR To ensure that equivalent clinical performance of the COBAS AMPLICOR Chlamydia trachomatis Test and the AMPLICOR Chlamydia Trachomatis Test are obtained, a comparison of the two test procedures was performed in which 639 clinical specimens were tested. The specimens included 266 endocervical swabs, 93 male urethral swabs, and 280 male urine specimens. Table 3 shows the comparative clinical performance obtained when these specimens were tested by the two test procedures. The results for each specimen type were evaluated for statistical difference in comparative clinical performance using McNemar’s Test. For each of the three specimen types tested, the McNemar’s Test found a value of $P > 0.4$. This finding provides strong evidence that there are no differences in the positive and negative results obtained when testing these specimens by either the COBAS AMPLICOR Chlamydia Trachomatis Test or the AMPLICOR Chlamydia Trachomatis Test. {5} 1 125 Table 3 Comparative Clinical Performance of the COBAS AMPLICOR Chlamydia Trachomatis Test versus the AMPLICOR Chlamydia Trachomatis Test | COBAS AMPLICOR | Pos Neg | Cervical | | Male Urethral | | Male Urine | | | --- | --- | --- | --- | --- | --- | --- | --- | | | | Pos | Neg | Pos | Neg | Pos | Neg | | | | 34 | 1* | 16 | 1† | 63 | 4‡ | | | | 1** | 230 | 0 | 76 | 5‡‡ | 208 | * Sample was culture positive ** Sample was culture negative/MOMP negative † Sample was culture negative/MOMP negative ‡ Two samples were culture positive/MOMP positive; one was culture negative/MOMP positive ‡‡ Two samples were culture positive/MOMP positive; three samples were culture negative/MOMP positive VIII. Conclusions The COBAS AMPLICOR is an automated bench-top analyzer for the in vitro diagnostic analysis of clinical laboratory specimens using PCR based nucleotide amplification methods. The instrument gives substantially equivalent analytical performance to other general laboratory equipment for thermal cycling and photometric detection of colormetric clinical assays. Based on the clinical performance data, substantially equivalent results were obtained for the qualitative COBAS AMPLICOR Chlamydia trachomatis Test on the COBAS AMPLICOR and the Roche AMPLICOR Chlamydia trachomatis Test (microwell plate) performed using the Perkin-Elmer 9600 thermal cycler and a conventional photometric microwell plate reader. {6} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 MAY 21 1997 Alex Wesolowski Director, Regulatory and Clinical Affairs Roche Molecular Systems, Inc. 1080 U.S. Highway 202 Branchburg, New Jersey 08876-1760 Re: K964506 COBAS AMPLICOR™ Analyzer Regulatory Class: I Product Code: JJF Dated: March 24, 1997 Received: March 25, 1997 Dear Mr. Wesolowski: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {7} Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {8} Page 1 of 1 510(k) Number (if known) ___________ Device Name: COBAS AMPLICOR™ Analyzer Indications for Use: The COBAS AMPLICOR Analyzer is an *in vitro* diagnostic device intended for use in clinical laboratories for the automation of the AMPLICOR™ Polymerase Chain Reaction test procedures. (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use (Optional Format 1-2-96) ![img-0.jpeg](img-0.jpeg)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...