K964444 · Rdh Medical · HSB · Aug 11, 1997 · Orthopedic
Device Facts
Record ID
K964444
Device Name
QUICKLOCK ORTHOPAEDIC TARGETING SYSTEM
Applicant
Rdh Medical
Product Code
HSB · Orthopedic
Decision Date
Aug 11, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Indications for Use
Locate incisions, guide bone drilling, and guide insertion of interlocking screws in the distal targeting procedure of intramedullary nailing.
Device Story
The QuickLock Orthopaedic Targeting System is an electromagnetic targeting device used during intramedullary nailing procedures to assist surgeons in locating distal interlocking screw holes. The system consists of a probe, a guide, and a display unit. The probe, inserted into the intramedullary nail, generates a low-energy, 25 KHz sinusoidal electromagnetic field. The guide, used by the surgeon to align the bone drill, detects this field. The system provides continuous visual feedback on a display, showing the alignment of the drill relative to the target hole. This replaces the traditional 'free-hand' technique, which relies on intermittent x-ray imaging and subjective visual perception. By providing continuous visual cues, the device allows surgeons to maintain a comfortable posture, reduces the need for repeated x-ray exposure, and improves drilling accuracy. Enhanced accuracy helps ensure correct hole size and shape, maximizing screw holding strength and reducing the risk of nail fatigue failure caused by drill nicks. The system is intended for use in an operating room setting by surgeons.
Clinical Evidence
Clinical trials conducted at the University of Arizona Health Sciences Center (1995-1997) compared the QuickLock system to predicate radiolucent drill systems. In 17 procedures using predicate devices, average targeting time was 16.1 minutes (range 10.4-30.0). In 22 procedures using the QuickLock system, average targeting time was 8.1 minutes (range 5.2-14.0), with a final configuration average of 6.6 minutes. The study demonstrated reduced procedure time and improved targeting efficiency.
Technological Characteristics
System utilizes a 25 KHz sinusoidal electromagnetic field generator. Probe is a sealed nickel-titanium alloy (Nitinol SE 508) tube. Guide is molded polyphenylsulfone (Radel R-5000). Drill bushings are molded polysulfone (UDEL P-1700 MG 11). Other components are surgical stainless steel. System is powered by 115 VRMS 60 Hz input, with internal +/-12 VDC and +5 VDC. Electrical safety certified to UL Standard 544, 2d Edition. Components are reusable and steam or Eto sterilizable.
Indications for Use
Indicated for use whenever distal interlocking screws are placed in intramedullary nails to which the QuickLock System has been adapted via accessory adapter components for mechanical attachment to proximal instrumentation and alignment during calibration.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
Howmedica (not specified)
Zimmer (not specified)
Richards (not specified)
Synthes (not specified)
Submission Summary (Full Text)
{0}
K964444
Page B-1
# Appendix B
## Summary of Safety and Effectiveness Information
AUG 11 1997
## Safety
### Predicate Devices
### QuickLock System
#### Radiation Exposure
Ionizing electromagnetic x-rays are generated by the C-arm image intensifier. Concern for minimizing the risk of x-ray exposure to surgeons and patients is a prominent factor in the free-hand technique currently employed by the predicate devices.
The low-energy, 25 KHz sinusoidal, nonionizing electromagnetic field generated by the probe coils is an nonsignificant radiation hazard* and produces negligible temperature rise of the probe surface.
* Opinion of Howard Bassen, Head, ElectroPhysics Branch, FDA Office of Science and Technology, telephone conversation July 29, 1994, in which Mr. Bassen confirmed that FDA does not impose standards for electromagnetic radiation unless the device causes heating of body tissue.
### Electric Shock
The C-arm image intensifier is an electronic device. It can contact operating room personnel, surgical apparatus in contact with the patient, or the patient directly, because during the distal targeting procedure it is close to the surgical field and is moved about by a technician to align the x-ray beam with the axis of the target hole. Operating room personnel and patients must, therefore, be protected from electric shock by C-arm compliance with safety standards for electronic medical devices.
The display, probe and guide come into contact with operating room personnel and patients, and are to be certified to UL Standard 544, 2d Edition. The display is in a sturdy, grounded, metal enclosure with 115 VRMS at 60 Hz input power, +/-12 VDC and +5 VDC internal power. Line power is limited by a 250V, 3AG UL 198.6, 2A, FAST type fuse. The AC to DC power supply and electroluminescent display are medical-grade components. Separation and insulation of internal display components prevent shock and fire hazardous conditions. Risk of shock by contact with the probe and guide is eliminated by low voltage levels, isolation, insulation and grounding. The probe coil voltage is below 12 VRMS with polyimide tube insulation breakdown above 1200 V. Probe coils are isolated from the display unit by transformer coupling, and the metal outer surface is grounded through the probe cable shield. Guide components reach voltages far less than 12 VRMS. The guide body is nonconductive, insulated by encapsulation of internal components, and is grounded through a wire in the guide cable.
{1}
Page B-2
# Predicate Devices
# QuickLock System
## Biocompatibility
Stainless steel guide wires or rods are routinely placed inside the medullary canal in reduction of fractures, reaming of the medullary canal, and guiding insertion of the nail. They are made of surgical stainless steel, are reusable, and steam or Eto sterilizable.
Probes are placed in nails after surgical insertion into the patient. They are sealed tubes of nickel-titanium alloy (Nitnol SE 508, Nitnol Devices and Componens, Inc.), a material used in the manufacture of vascular stents that are FDA approved for unrestricted implantation. The probe is sealed by laser welding a Nitnol disk into the end inserted into the nail, so there is no patient tissue or fluid contact with probe materials other than Nitnol. Probes are reusable and steam or Eto sterilizable. Probe cables have a silicone jacket and metal connectors that will withstand the same sterilization.
Surgical stainless steel bone drills are supplied with nail manufacturer's instrumentation. They are reusable and steam or Eto sterilizable.
Drills are nonmagnetic, surgical stainless steel (SCF-23, Carpenter Steel Co.). They are reusable and steam or Eto sterilizable.
Drill bushings used in proximal targeting are surgical stainless steel. They are reusable and steam or Eto sterilizable.
Drill bushings are molded USP Class VI biocompatible polysulfone (UDEL P-1700 MG 11, Amoco Performance Products, Inc.). The bushings are reusable and steam or Eto sterilizable.
Drill or trocar holders of radioluscent plastic are used in the Howmedica and Zimmer devices. Radioluscent plastic components are used to hold the bone drill in the Richards and Synthes powered distal tageting devices. These are reusable and steam or Eto sterilizable.
Guides are molded USP Class III biocompatible polyphenylsulfone (Radel R-5000, Amoco Performance Products, Inc.). Guides are reusable and steam or Eto sterilizable. Guide cables have a silicone jacket and metal connectors that will withstand the same sterilization.
Screw drivers are surgical stainless steel, reusable and steam or Eto sterilizable.
Screw drivers are type 316 and 400 series stainless steel. They are reusable and steam or Eto sterilizable.
## Safety Standards Certification
Certification of compliance with UL Standard 544, 2d Edition will be obtained from an FDA approved testing laboratory before the product is commercially distributed.
{2}
Page B-3
# Effectiveness
| Predicate Devices | QuickLock System |
| --- | --- |
| Intermediate | Intermediate |
## Intended Use
Locate incisions, guide bone drilling, and guide insertion of interlocking screws in the distal targeting procedure of intramedullary nailing.
Same as for predicate devices
## Skill and Experience Required
Users must strive to hold the drill in alignment after short x-ray exposures that provide only intermittent visual control. This requires considerable skill and experience. Surgeons who infrequently perform the procedure fall out of practice and are markedly less effective.
Users have continuous visual control so the task is easier and less training, skill and experience are required. Surgeons learn the technique after a few minutes of training. Proficiency is retained without frequent performance of the procedure.
## Accuracy of Visual Cues
Adjustment of C-arm position to align the x-ray beam with target holes depends on visual perception of distal holes as true circles on the image intensifier screen. Subjective judgement limits reliable accuracy of this alignment. Operator skill strongly affects accuracy of drilling between visual snapshots.
The probe and guide are aligned mechanically with target holes during the calibration process, which electronically offsets errors for each procedure. Drilling accuracy is enhanced by the high visual sensitivity of dot and cross-hair sight adjustments and the continuous display of visual sighting information.
## Operator Comfort
The need to minimize exposure to x-ray radiation hampers the freehand technique used with the predicate devices. Surgeons work in awkward positions, reaching from the side to avoid the x-ray beam.
Because there is no need to stay out of an x-ray beam, surgeons can stand, hold the guide, and drill in a comfortable position.
## Quality of Drilled Holes
Holes started out of alignment must be elongated to pass through the nail and both cortices. This reduces the holding strength and endurance of bone screws and compromises the objective of nail interlocking.
Accurate alignment throughout the drilling process ensures correct hole size and shape, for maximum screw holding strength and endurance.
## Nicking of Intramedullary Nails
Drills often strike the edges of target holes, leaving nicks that can cause stress concentration in the nail material and lead to premature nail breakage due to fatigue fracture.
Improved accuracy reduces the occurrence of drill strikes and the resultant risk of induced nail fatigue failure.
{3}
Page B-4
# Predicate Devices
## QuickLock System
### Time Required for Distal Targeting*
In 17 procedures (10 femoral and 7 tibial) performed using Richards or Synthes radiolucent drills, the time required for distal targeting averaged 16.1 minutes with a range of 10.4 to 30.0 minutes.
In 22 procedures (13 femoral and 9 tibial) performed using the QuickLock system, the time required for distal targeting averaged 8.1 minutes with a range of 5.2 to 14.0 minutes. These data are scewed by changes in probe configuration during the trials; the average time over procedures performed with the final probe configuration was 6.6 minutes.
* Results of clinical trials performed at the University of Arizona Health Sciences Center, Tucson, Arizona, from September 1995 to present. These trials are conducted under IRB classification as a non-significant risk device, with approval by FDA, Center for Devices and Radiological Health IDE section's Mr. Ted Stevens, team leader of orthopedic devices, non-implantable. The principal investigator was John T. Ruth, M.D., Assistant Professor, Section of Orthopedic Surgery.
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
Roy B. Hollstien, Ph.D.
*RDH Medical
5775 Webster Road
Paso Robles, California 93446
AUG 11 1997
Re: K964444
Trade Name: QuickLock Orthopaedic Targeting System
Regulatory Class: II
Product Code: HSB
Dated: June 27, 1997
Received: June 30, 1997
Dear Dr. Hollstien:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}
Page 2 - Roy B. Hollstien, Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Enclosure
{6}
Page 1 of 1
510(k) Number (if known): K964444
Device Name: QuickLock Orthopaedic Targeting System
Indications For Use:
The QuickLock Orthopaedic Targeting System is indicated for use whenever distal interlocking screws are placed in intramedullary nails to which the QuickLock System has been adapted by making available accessory adapter components that enable mechanical attachment of the QuickLock Probe to the nail system proximal instrumentation and alignment of the QuickLock Guide during calibration.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐ (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.