IMX CA 15-3

K964407 · Abbott Laboratories · MOI · Nov 10, 1997 · Immunology

Device Facts

Record IDK964407
Device NameIMX CA 15-3
ApplicantAbbott Laboratories
Product CodeMOI · Immunology
Decision DateNov 10, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.6010
Device ClassClass 2

Indications for Use

The IMx® CA 15-3™ assay is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative measurement of CA 15-3 assay values in human serum and plasma (EDTA) to aid in the management of Stage II and III breast cancer patients. Serial testing for patient CA 15-3 assay values should be used in conjunction with other clinical methods for monitoring breast cancer.

Device Story

IMx CA 15-3 is a microparticle enzyme immunoassay (MEIA) used for quantitative measurement of CA 15-3 antigen in human serum and EDTA plasma. Operated on the IMx System in clinical laboratory settings by trained personnel. The device processes patient samples to provide quantitative concentration values (U/mL). These results are used by physicians as an aid in the management of Stage II and III breast cancer patients, specifically for serial monitoring of disease status. The output assists clinicians in tracking disease progression or response to therapy when used alongside other clinical diagnostic methods.

Clinical Evidence

Clinical validation included 561 specimens for correlation analysis (r=0.910). ROC analysis on 160 healthy, 30 benign, and 228 malignant breast cancer patients showed AUC of 0.71 (IMx) vs 0.70 (predicate). Sensitivity/specificity at reference values were 29.8%/97.4% (IMx) vs 30.3%/98.4% (predicate). Concordance across 629 subjects was 98.4%. Serial tracking of 24 patients showed comparable trends. Additional study of 79 Stage II/III patients (359 specimens) showed 91% concordance; sensitivity at relapse was 54% (IMx) vs 62% (predicate).

Technological Characteristics

Microparticle Enzyme Immunoassay (MEIA) technology. Analyte: CA 15-3. Sample type: human serum and EDTA plasma. Dynamic range: 0-250 U/mL. Sensitivity: 0.2 U/mL. System: IMx automated analyzer. Connectivity: Standalone clinical laboratory instrument.

Indications for Use

Indicated for management of Stage II and III breast cancer patients. Used for quantitative measurement of CA 15-3 in human serum and EDTA plasma. Intended for serial testing in conjunction with other clinical monitoring methods.

Regulatory Classification

Identification

A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.

Special Controls

*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} K 964407 510(k) SUMMARY IMx® CA 15-3™ NOV 0 1997 # SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING A SUBSTANTIALLY EQUIVALENT DETERMINATION The following information as presented in the Premarket Notification [510(k)] for IMx CA 15-3 constitutes data supporting a substantially equivalent determination. IMx CA 15-3 is a microparticle enzyme immunoassay for the quantitative measurement of CA 15-3 assay values in human serum and EDTA plasma on the IMx System. IMx CA 15-3 employs Abbott Calibrators and Controls. Substantial equivalence has been demonstrated between the Abbott IMx CA 15-3 assay and the BIOMIRA Diagnostics Inc. TRUQUANT® BR™ RIA assay. Both assays are intended to be used as an aid in the management of stage II and stage III breast cancer patients. A linear regression analysis between these two assays, using 561 specimens with IMx CA 15-3 assay values ranging from 4.0 to 246.7 U/mL, yielded a correlation coefficient of 0.910, slope of 0.69, and y-intercept of 3.1 U/mL. The dynamic range of IMx CA 15-3 is 0 - 250 U/mL with a sensitivity of 0.2 U/mL. The dynamic range of TRUQUANT BR RIA is 0 - 200 U/mL with a sensitivity of 7.0 U/mL. Receiver Operating Characteristic (ROC) analyses on 160 apparently healthy females plus 30 benign breast patients vs. 228 malignant breast patients gave substantially equivalent areas under the curve of 0.71 for IMx CA 15-3 and 0.70 for TRUQUANT BR RIA. At the claimed reference values for the assays (31.3 U/mL for IMx CA 15-3 and 37.7 U/mL for TRUQUANT BR RIA), similar sensitivities of 29.8% and 30.3% and specificities of 97.4% and 98.4% were obtained for IMx CA 15-3 and TRUQUANT BR RIA, respectively. Based on the claimed reference values for the two assays, concordance was 98.8%, 100%, 93.4%, and 98.4% for 160 apparently healthy females, 30 benign breast patients, 228 malignant breast patients, and 629 total subjects, respectively. Serial tracking data on 24 malignant breast patients showed comparable trending results for both assays. Seventy nine Stage II and Stage III breast cancer patients were evaluated in a blinded study using the AxSYM CA 15-3 assay and the TRUQUANT BR RIA. 359 specimens from 77 of the evaluable patients gave a concordance(agreement) between the two assays of 91% at their respective reference values. At the claimed reference values for the assays (31.1 U/mL for AxSYM CA 15-3 and 37.7 U/mL for TRUQUANT BR RIA), when using values obtained within 6 months of relapse, similar sensitivities of 54% (95% CI=25-81) and 62% (95% CI=32-86) and specificities of 94% (95% CI=85-99) and 91% (95% CI=80-97) were obtained for AxSYM CA 15-3 and TRUQUANT BR RIA, respectively. The IMx CA 15-3 assay was compared to the AxSYM CA 15-3 assay on 2337 specimens with IMx CA 15-3 assay values ranging from 3.0 to 250.0 U/mL, yielded a correlation coefficient of 0.989, slope of 1.04, and y-intercept of -0.30 U/mL. Page 1 of 2 {1} In conclusion, these data demonstrate that the Abbott IMx CA 15-3 assay is as safe and effective as, and is substantially equivalent to the BIOMIRA Diagnostics Inc. TRUQUANT BR RIA assay. Prepared and Submitted November 1, 1996 (edited September 15, 1997) by: Joy C. Sonsalla 200 Abbott Park Road Abbott Laboratories Abbott Park, IL 60064 Page 2 of 2 {2} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Joy C. Sonsalla Regulatory Affairs Section Leader ADD Regulatory Affairs Abbott Laboratories D49C AP31 200 Abbott Park Road Abbott Park, Illinois 60064-3537 Re: K964407 Trade Name: Abbott IMz® CA 15-3™ Regulatory Class: II Tier III Product Code: MOI Dated: September 16, 1997 Received: September 17, 1997 Dear Ms. Sonsalla: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations. NOV 10 15 {3} Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html" Sincerely yours, Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4} 510(k) Number (if known): K#964407 Device Name: IMx CA 15-3 Indications For Use: The IMx® CA 15-3™ assay is a Microparticle Enzyme Immunoassay (MEIA) for the quantitative measurement of CA 15-3 assay values in human serum and plasma (EDTA) to aid in the management of Stage II and III breast cancer patients. Serial testing for patient CA 15-3 assay values should be used in conjunction with other clinical methods for monitoring breast cancer. ![img-0.jpeg](img-0.jpeg) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use ☑ (Per 21 CFR 801.109) OR Over-The-Counter Use ☐ (Optional Format 1-2-96) IMx CA 15-3 510(k) Response to FDA Letter dated 1/13/97 mdn1531wp
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...