The Sarns Aortic Arch Cannulae with Duraflo Treatment, Flexible Arterial Cannulae with Duraflo Treatment, High Flow Aortic Arch Cannulae with Duraflo Treatment, Soft Arc Cannula with Duraflo Treatment, D4 Cannulae with Duraflo Treatment, and the Soft Flow Cannulae with Duraflo Treatment are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. The Sarns Flexible Aortic Arch Cannula with Duraflo Treatment is indicated for use in perfusion of the descending thoracic aorta during cardiopulmonary bypass surgery.
Device Story
Sarns Arterial Cannulae are sterile, single-use devices used during cardiopulmonary bypass surgery to perfuse the ascending or descending thoracic aorta. The device consists of a cannula tip, tube, and connector, available in various sizes (3.8 mm to 8.0 mm OD), styles (straight/angled, soft/rigid, metal/plastic), and construction (wire-reinforced or non-wire-reinforced). Some models include an optional suture flange or Luer port for air venting. The subject devices incorporate Duraflo Treatment, which deposits a heparin layer on internal and external surfaces to improve hemocompatibility. The device is operated by clinical staff in a surgical setting. It functions as a conduit for blood flow during bypass, with the heparin coating intended to reduce thrombogenicity.
Clinical Evidence
No clinical data. Bench testing only.
Technological Characteristics
Sterile, single-use arterial cannulae. Materials include metal or plastic tips, wire-reinforced or non-wire-reinforced tubing. Features heparin-based Duraflo treatment applied to internal and external surfaces. Dimensions range from 3.8 mm to 8.0 mm outer diameter. No dimensional changes compared to predicate devices.
Indications for Use
Indicated for perfusion of the ascending aorta during cardiopulmonary bypass surgery; Sarns Flexible Aortic Arch Cannula also indicated for perfusion of the descending thoracic aorta during cardiopulmonary bypass surgery.
Regulatory Classification
Identification
A cardiopulmonary bypass vascular catheter, cannula, or tubing is a device used in cardiopulmonary surgery to cannulate the vessels, perfuse the coronary arteries, and to interconnect the catheters and cannulas with an oxygenator. The device includes accessory bypass equipment.
{0}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUN 18 1997
Mr. Jason Smith
Regulatory Affairs Coordinator
3M Health Care
Cardiovascular Systems
Sarns and CDI Products
6200 Jackson Road
Ann Arbor, Michigan 48103-9300
Re: K964373
Sarns Arterial Cannulae with Duraflo® Treatment
Regulatory Class: II (Two)
Product Code: 74 DWF
Dated: April 9, 1997
Received: April 10, 1997
Dear Mr. Smith:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{1}
Page 2 - Mr. Jason Smith
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "dsmo@fdadr.cdrh.fda.gov."
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{2}
510(k) Number (if known): K 964373/5001
Device Name: Sarns Arterial Cannulae with Duraflo Treatment
## Indications For Use:
The Sarns Aortic Arch Cannulae with Duraflo Treatment, Flexible Arterial Cannulae with Duraflo Treatment, High Flow Aortic Arch Cannulae with Duraflo Treatment, Soft Arc Cannula with Duraflo Treatment, D4 Cannulae with Duraflo Treatment, and the Soft Flow Cannulae with Duraflo Treatment are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery.
The Sarns Flexible Aortic Arch Cannula with Duraflo Treatment is indicated for use in perfusion of the descending thoracic aorta during cardiopulmonary bypass surgery.
( PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED )
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K 964373
Description Use ☑
Order 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
{3}
JUN 18 1997
510(k) SUMMARY
k964373
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is _______________.
Submitter's Name: 3M Health Care
Submitter's Address: 6200 Jackson Road, Ann Arbor, Michigan 48103
Contact Person: Jason Smith
Phone Number: (313) 663-4145
FAX Number: (313) 663-5062
Summary Date:
## Device Trade Names:
Sarns Aortic Arch Cannulae with Duraflo® Treatment, Sarns Flexible Aortic Arch Cannula with Duraflo® Treatment, Sarns Flexible Arterial Cannulae with Duraflo® Treatment, Sarns High Flow Aortic Arch Cannulae with Duraflo® Treatment, Sarns Soft-Arc Cannula with Duraflo® Treatment, Sarns D4 Cannulae with Duraflo® Treatment, Sarns Soft Flow Cannulae with Duraflo® Treatment
## Device Classification Name:
"Cardiopulmonary Bypass Vascular Catheter, Cannula, or Tubing" (21 CFR 870.4210)
## Predicate Devices:
The Sarns Arterial Cannulae with Duraflo® Treatment are substantially equivalent to the existing Sarns Arterial Cannulae [510(k) clearance numbers listed in the table below].
| Cannula Type | 510(k) Number | Clearance Date |
| --- | --- | --- |
| Soft Flow | K934127 | 2/22/94 |
| D4 | K874896 | 2/2/88 |
| Soft-Arc | K834134 | 4/17/84 |
| Flexible Arterial | K771499 | 8/16/77 |
| High Flow Aortic Arch | K770429 | 3/16/77 |
| Aortic Arch, Flexible Aortic Arch | Preamendment | |
## Device Description:
The only difference between the Sarns Arterial Cannulae and the Sarns Arterial Cannulae with Duraflo® Treatment is the addition of the Duraflo® Treatment. As a result of this process, a layer of heparin is deposited on the inside and outside of the cannula tip, tube, and connector. Like their predicate devices, the Sarns Arterial Cannulae with Duraflo® Treatment are sterile, single-use medical devices. They are available in sizes ranging from 3.8 mm to 8.0 mm outer diameter tip sizes, with tips being straight or angled, soft or rigid, metal or plastic. Several styles of the Sarns Arterial Cannulae are available in both wire-reinforced and non-wire-reinforced tube styles. Other features include an
Page 14 of 210
{4}
optional suture flange on some models as well as an optional Luer port for the venting of air.
## Indications for Use:
The Sarns Aortic Arch, Flexible Aortic Arch, High Flow Aortic Arch, Flexible Arterial, D4, and Soft Flow Cannulae with Duraflo® Treatment are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery. The Sarns Flexible Aortic Arch Cannula with Duraflo® Treatment is indicated for use in perfusion of the descending thoracic aorta during cardiopulmonary bypass surgery.
## Technological Characteristics:
The only difference between the Sarns Arterial Cannulae and the Sarns Arterial Cannulae with Duraflo® Treatment is the addition of the Duraflo® Treatment. As a result of this process, a layer of heparin is deposited on the inside and outside of the cannula tip, tube, and connector. There will be no dimensional changes to the cannulae due to the addition of the Duraflo® Treatment
## Nonclinical Performance:
The performance characteristics of selected models (these models were chosen as being representative of the technological attributes of all models of Sarns Arterial Cannulae) of the Sarns Arterial Cannulae with Duraflo® Treatment were exhaustively tested and compared with the performance characteristics of the currently marketed Sarns Arterial Cannulae. All new and existing performance characteristics of the Sarns Arterial Cannulae with Duraflo® Treatment have been validated.
## Clinical Performance:
Clinical testing was not performed on these devices.
## Conclusions from Nonclinical Tests:
The Sarns Arterial Cannulae with Duraflo® Treatment perform as intended according to their performance specifications. The Sarns Arterial Cannulae with Duraflo® Treatment are substantially equivalent to their predicate devices.
Page 15 of 210
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.