← Product Code [DZE](/productcode/DZE) · K964330

# BIOHORIZONS DENTAL IMPLANT SYSTEM (NAME SUBJECT TO CHANGE) (K964330)

_BioHorizons Implant Systems, Inc. · DZE · Jan 17, 1997 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K964330

## Device Facts

- **Applicant:** BioHorizons Implant Systems, Inc.
- **Product Code:** [DZE](/productcode/DZE.md)
- **Decision Date:** Jan 17, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3640
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

The BioHorizons Dental Implant System may be used in the mandible and maxilla for use as an artificial root structure for single tooth replacement or as abutments for fixed bridgework and denture retention.

## Device Story

System comprises endosseous screw-form implants, surgical components, and prosthetic components. Implants designed for specific bone density classifications (D1-D4) in mandible and maxilla to optimize strain distribution and promote bone growth. Available in 4mm and 5mm diameters; various lengths. Materials include titanium alloy with optional titanium plasma-spray (TPS) or hydroxyapatite (HA) coatings. Used by dental clinicians for tooth replacement or prosthetic support. Device provides artificial root structure; clinical benefit includes functional loading support and interface maintenance.

## Clinical Evidence

Bench testing only. Evaluation consisted of mechanical testing of the implant and bioactive coating mechanical tests.

## Technological Characteristics

Titanium alloy (ASTM F 136). Screw-form endosseous implants. Coatings: Uncoated, titanium plasma spray (TPS), or hydroxyapatite (HA). Diameters: 4mm, 5mm. Lengths: 9mm-15mm. Design optimized for bone density classifications D1-D4.

## Regulatory Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

## Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

## Predicate Devices

- BioHorizons Dental Implant System ([K960026](/device/K960026.md))
- Steri-Oss® Hex-Lock (HL) Threaded Titanium Implants

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

JAN 17 1997
K964330

BioHorizons Implant Systems, Inc.
510(k) Notification
October 24, 1996

# 510(k) Summary of Safety and Effectiveness

## Proprietary Name
BioHorizons Dental Implant System (name subject to change)

## Common Name
Uncoated, titanium plasma spray, and hydroxyapatite screw-form implants, surgical components, and prosthetic components

## Classification Name
Endosseous implants, surgical components, and prosthetic attachments

## Classification
Class III

## Official Contact
R. Steven Boggan, M.S., M.BA.
President and Chief Operating Officer
BioHorizons Implant Systems, Inc.
2129 Montgomery Highway
Birmingham, AL 35209
(205) 871-1345
FAX (205) 870-0304

## Device Description
The BioHorizons Dental Implant System is a comprehensive system containing implants, surgical components, and prosthetic components. The implants are specifically designed to optimize strain distribution to contiguous bone under functional loading in order to promote strain-induced bone growth and interface maintenance over the long term. This improvement in biomechanical performance is achieved by optimizing implant designs specifically for each bone density classification (D1, D2, D3, D4) in the mandible and maxilla.

Four implant designs, corresponding to each bone density classification, are available in four and five mm diameters. Each implant design, manufactured from titanium alloy

16

{1}

BioHorizons Implant Systems, Inc.
510(k) Notification
October 24, 1996

conforming to ASTM F 136, is available in two lengths and may feature a titanium plasma-spray (TPS) or hydroxyapatite (HA) coating. The following table provides a comprehensive summary of implant diameter, length, and coating.

|  Diameter (mm) | Design | Lengths (mm) | Coating  |
| --- | --- | --- | --- |
|  φ4.00 | D1 | 10, 12 | Uncoated  |
|   | D2 | 11, 13 | Uncoated  |
|   | D3 | 12, 14 | TPS  |
|   | D4 | 13, 15 | HA  |
|  φ5.00 | D1 | 9, 11 | Uncoated  |
|   | D2 | 10, 12 | Uncoated  |
|   | D3 | 11, 13 | TPS  |
|   | D4 | 12, 14 | HA  |

## Product Evaluation

Evaluation of the BioHorizons Dental Implant System consisted mechanical testing of the implant and bioactive coating mechanical tests. These analyses indicate that the BioHorizons Dental Implant System should be safe and effective when used as intended.

## Indications

The BioHorizons Dental Implant System may be used in the mandible and maxilla for use as an artificial root structure for single tooth replacement or as abutments for fixed bridgework and denture retention.

## Substantial Equivalence Information

The BioHorizons Dental Implant System is substantially equivalent in all features which could affect safety or effectiveness to the BioHorizons Dental Implant System (K960026) and the Steri-Oss® Hex-Lock (HL) Threaded Titanium Implants.

17

---

**Source:** [https://fda-staging.innolitics.com/device/K964330](https://fda-staging.innolitics.com/device/K964330)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).