EYE FIXATION SPECULUM

K964289 · Eyefix, Inc. · HNC · Dec 17, 1996 · Ophthalmic

Device Facts

Record IDK964289
Device NameEYE FIXATION SPECULUM
ApplicantEyefix, Inc.
Product CodeHNC · Ophthalmic
Decision DateDec 17, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4350
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The Eye Fixation Speculum, by eyeFix™, Inc. is intended to reduce eye movement and stabilize the eye in any procedure in which the globe needs to be immobilized. The device is provided sterile and is disposable for single use only.

Device Story

Device consists of low-profile scleral vacuum ring attached to medical-grade plastic ophthalmic speculum via six-degrees-of-freedom resistance joint. Physician retracts eyelids, secures speculum, and places concentric elastomer ring on sclera. Vacuum applied via spring-loaded syringe secures ring to sclera; resistance joint holds eye steady against ocular muscle movement. Allows physician to adjust globe position within joint range. Luminescent plastic aids visualization in low-light conditions. Used in clinical settings by physicians to immobilize eye during procedures; benefits patient by preventing involuntary eye movement, improving surgical/procedural precision.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Materials: medical-grade plastic, elastomer ring. Principle: vacuum-assisted scleral fixation combined with mechanical resistance joint. Form factor: low-profile ring with attached speculum arm. Sterilization: provided sterile, single-use disposable.

Indications for Use

Indicated for patients requiring globe immobilization during ophthalmic procedures to reduce eye movement.

Regulatory Classification

Identification

A manual ophthalmic surgical instrument is a nonpowered, handheld device intended to aid or perform ophthalmic surgical procedures. This generic type of device includes the manual corneal burr, ophthalmic caliper, ophthalmic cannula, eyelid clamp, ophthalmic muscle clamp, iris retractor clip, orbital compressor, ophthalmic curette, cystotome, orbital depressor, lachrymal dilator, erisophake, expressor, ophthalmic forcep, ophthalmic hook, sphere introducer, ophthalmic knife, ophthalmic suturing needle, lachrymal probe, trabeculotomy probe, cornea-sclera punch, ophthalmic retractor, ophthalmic ring (Flieringa), lachrymal sac rongeur, ophthalmic scissors, enucleating snare, ophthalmic spatula, ophthalmic specula, ophthalmic spoon, ophthalmic spud, trabeculotome or ophthalmic manual trephine.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0} DEC 17 1996 K964289 510(k) Summary SUBMITTER: Submitted on behalf of: Company Name: eyeFix™, Inc. Address: Suite 750 Medical Arts Building 825 Nicollet Avenue Minneapolis, MN 55402 Phone: 612-338-4861 Fax: 612-333-8306 CONTACT PERSON: Neal Sher, M.D., F.A.C.S. OFFICIAL CORRESPONDENT: Richard E. Lippman, O.D., F.A.A.O. Firm Name: C.L. McIntosh & Associates, Inc. Address: 12300 Twinbrook Parkway, Suite 625 Rockville, MD 20852 Phone: 301-770-9590 Fax: 301-770-9584 DATE SUMMARY PREPARED: October 25, 1996 TRADE NAME: eyeFix™, Inc. Eye Fixation Speculum COMMON NAME: ophthalmic speculum SUBSTANTIALLY EQUIVALENT TO: The Eye Fixation Speculum, by eyeFix™, Inc. is substantially equivalent to two products currently legally marketed in the U.S. The Eye Fixation Speculum is substantially equivalent to the indication for use of the Lid-Set Eyelid Retractor manufactured by Universi, S.A. This device was determined by FDA to be exempt from 510(k) notification per K903322. The Eye Fixation Speculum is substantially equivalent to the stabilization method (technology) used in numerous manual ophthalmic instruments such as the Hessburg-Barron Vacuum Trephine as manufactured by Precision Instruments, Inc. This device was cleared by FDA in K864520. Additional products in the same family line were cleared under K843048 and K923540. DESCRIPTION of the DEVICE: The Eye Fixation Speculum, by eyeFix, Inc., consists of a low profile scleral vacuum ring attached to a medical quality plastic ophthalmic speculum using a six-degrees of freedom resistance joint. After retracting the eyelids, and securing the speculum, the vacuum ring is attached to the sclera by locating the medical quality concentric elastomer-ring to the cornea. Vacuum is applied 000017 {1} K964289 by way of the attached spring-loaded syringe. When vacuum is applied, the flexible elastomer portion of the ring sucks down on the scleral region of the eye securing the ring to the eye. The frictional forces within the resistance joint of the attaching arm of the speculum, together with the vacuum ring, hold the eye steady against the tendency of the eye muscles to the move the globe. Although this resistance is high enough to prevent the eye from moving, the force is not so high that the physician is prevented from conveniently moving the ring within the range of motion that the joint allows. Luminescent plastic on the upper portion of the ring helps in visualization of the eye under low examination lighting. INDICATIONS FOR USE: The Eye Fixation Speculum, by eyeFix™, Inc. is intended to reduce eye movement and stabilize the eye in any procedure in which the globe needs to be immobilized. The device is provided sterile and is disposable for single use only. 000018
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...