← Product Code [DFH](/productcode/DFH) · K964260

# IMMAGE IMMUNOCHEMISTRY SYSTEM KAPPA LIGHT CHAIN (KAP) REAGENT IMMAGE IMMUNOCHEMISTRY SYSTEM LAMDA LIGHT CHAIN (LAM) REAG (K964260)

_Beckman Instruments, Inc. · DFH · Feb 10, 1997 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K964260

## Device Facts

- **Applicant:** Beckman Instruments, Inc.
- **Product Code:** [DFH](/productcode/DFH.md)
- **Decision Date:** Feb 10, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5550
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

The IMMAGE Immunochemistry System Kappa (KAP) Reagent, when used in conjunction with Beckman IMMAGE™ Immunochemistry Systems and Beckman Calibrator 1, is intended for the quantitative determination of kappa light chains (free and bound) by rate nephelometry. The IMMAGE Immunochemistry System Lambda (LAM) Reagent, when used in conjunction with Beckman IMMAGE™ Immunochemistry Systems and Beckman Calibrator 1, is intended for the quantitative determination of lambda light chains (free and bound) by rate nephelometry.

## Device Story

IMMAGE Immunochemistry System reagents utilize rate nephelometry to measure kappa and lambda light chain concentrations in human serum and urine. System uses goat-derived antibodies to bind light chains; antigen excess testing solution is integrated into the reagent cartridge to automate detection of high-concentration samples. Operated by laboratory personnel on the IMMAGE Immunochemistry System. Output is a quantitative concentration value used by clinicians to assess light chain levels. Benefits include automated processing and improved workflow compared to manual antigen excess testing.

## Clinical Evidence

Bench testing only. Method comparison studies performed against predicate Beckman reagents across serum (n=220 for KAP, n=236 for LAM) and urine (n=103 for KAP, n=40 for LAM) samples showed high correlation (r=0.954 to 0.991). Within-run imprecision studies (n=30 to 80 per level) demonstrated %C.V. values ranging from 1.2% to 3.0%. Stability testing confirmed 24-month shelf-life and 14-day open container/calibration stability.

## Technological Characteristics

Rate nephelometry; goat-derived antibodies; integrated antigen excess testing solution; reagent cartridge form factor. Quantitative determination of free and bound light chains.

## Regulatory Identification

An immunoglobulin (light chain specific) immunological test system is a device that consists of the reagents used to measure by immunochemical techniques both kappa and lambda types of light chain portions of immunoglobulin molecules in serum, other body fluids, and tissues. In some disease states, an excess of light chains are produced by the antibody-forming cells. These free light chains, unassociated with gamma globulin molecules, can be found in a patient's body fluids and tissues. Measurement of the various amounts of the different types of light chains aids in the diagnosis of multiple myeloma (cancer of antibody-forming cells), lymphocytic neoplasms (cancer of lymphoid tissue), Waldenstrom's macroglobulinemia (increased production of large immunoglobulins), and connective tissue diseases such as rheumatoid arthritis or systemic lupus erythematosus.

## Predicate Devices

- Beckman Kappa Light Chain (KAP) Reagent (K884276A, [K902484](/device/K902484.md))
- Beckman Lambda Light Chain (LAM) Reagent (K884597A, [K902484](/device/K902484.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

200 South Kraemer Boulevard, Box 8000, Brea, CA 92822-8000 • (714) 993-5321

# BECKMAN

Summary of Safety &amp; Effectiveness

IMMAGE™ Immunochemistry System Kappa (KAP) and Lambda (LAM) Light Chain Reagents

## 1.0 Submitted By:

**K94260**

Annette Hellie
Regulatory Specialist, Product Submissions
Beckman Instruments, Inc.
200 S. Kraemer Blvd., W-337
Brea, California 92822-8000
Telephone: (714) 993-8767
FAX: (714) 961-4457

FEB 10 1997

## 2.0 Date Submitted:

24 October 1996

## 3.0 Device Name(s):

### 3.1 Proprietary Names

IMMAGE™ Immunochemistry System Kappa Light Chain (KAP) Reagent
IMMAGE™ Immunochemistry System Lambda Light Chain (LAM) Reagent

### 3.2 Classification Name

Immunoglobulin (light chain specific) immunological test system (21 CFR § 866.5550)

## 4.0 Predicate Device(s):

|  IMMAGE System Reagent | Predicate | Manufacturer | Docket Number  |
| --- | --- | --- | --- |
|  IMMAGE System Kappa (KAP) Reagent | Beckman Kappa Light Chain (KAP) Reagent | Beckman Instruments, Inc. | K884276A
K902484  |
|  IMMAGE System Lambda (LAM) Reagent | Beckman Lambda Light Chain (LAM) Reagent | Beckman Instruments, Inc. | K884597A
K902484  |

Beckman Instruments, Inc.

bex: 910-592-1260 • telex: 06-78413

{1}

Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Kappa (KAP) and Lambda (LAM) Light Chain Reagents
Summary of Safety &amp; Effectiveness
2

# 5.0 Description:

The IMMAGE Immunochemistry System KAP and LAM Reagents in conjunction with Beckman Calibrator 1, are intended for use in the quantitative determination of kappa and lambda light chain concentrations respectively in human serum and urine samples on Beckman's IMMAGE Immunochemistry System.

# 6.0 Intended Use:

The IMMAGE Immunochemistry System Kappa (KAP) Reagent, when used in conjunction with Beckman IMMAGE™ Immunochemistry Systems and Beckman Calibrator 1, is intended for the quantitative determination of kappa light chains (free and bound) by rate nephelometry.

The IMMAGE Immunochemistry System Lambda (LAM) Reagent, when used in conjunction with Beckman IMMAGE™ Immunochemistry Systems and Beckman Calibrator 1, is intended for the quantitative determination of lambda light chains (free and bound) by rate nephelometry.

# 7.0 Comparison to Predicate(s):

The following table shows similarities and differences between the predicates identified in Section 4.0 of this summary.

|  Reagent | Aspect/Characteristic | Comments  |
| --- | --- | --- |
|  SIMILARITIES  |   |   |
|  IMMAGE System KAP and LAM Reagents | Initial Analytic Range | Same as Beckman Kappa and Lambda Light Chain reagents  |
|   |  Nephelometric methodology  |   |
|   |  Antibody source (goat)  |   |
|  DIFFERENCES  |   |   |
|  IMMAGE System KAP and LAM Reagents | Antigen excess testing solution | IMMAGE KAP & LAM have antigen excess testing solution included in the reagent cartridge, while the Beckman Kappa and Lambda require off-line preparation of the solution.  |
|   | Antibody concentration | IMMAGE KAP and LAM have a higher antibody concentration than the Beckman Kappa and Lambda reagents  |

file: id510k.sse

{2}

Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Kappa (KAP) and Lambda (LAM) Light Chain Reagents
Summary of Safety &amp; Effectiveness

## 8.0 Summary of Performance Data:

The data in the Premarket Notification on safety and effectiveness supports a finding of substantial equivalence to chemistry test systems already in commercial distribution. Equivalence is demonstrated through method comparison, stability, and imprecision experiments that relate results obtained from the Beckman Reagents to the IMMAGE System Reagents.

### Method Comparison Study Results
IMMAGE Kappa (KAP) and Lambda (LAM) Reagents

|  Analyte | Sample Type | Slope | Intercept | r | n | Predictor Method  |
| --- | --- | --- | --- | --- | --- | --- |
|  IMMAGE KAP Reagent | serum | 1.037 | -9.16 | 0.987 | 220 | Beckman KAP Reagent  |
|   |  urine | 0.970 | 1.36 | 0.980 | 103  |   |
|  IMMAGE LAM Reagent | serum | 1.020 | -23.5 | 0.991 | 236 | Beckman LAM Reagent  |
|   |  urine | 1.009 | -0.08 | 0.954 | 40  |   |

### Stability Study Results

|  Reagent | Product Claim  |
| --- | --- |
|  IMMAGE KAP & LAM | 24 month shelf-life
14 day open container stability
14 day calibration stability  |

file: k510k.sse

{3}

Beckman Instruments, Inc., Section 510(k) Notification
IMMAGE™ Immunochemistry System Kappa (KAP) and Lambda (LAM) Light Chain Reagents
Summary of Safety &amp; Effectiveness
4

## Estimated Within-Run Imprecision

|  Sample | Mean (μg/mL) | S.D. (μg/mL) | %C.V. | N  |
| --- | --- | --- | --- | --- |
|  Kappa Light Chain (KAP) Serum  |   |   |   |   |
|  Level 1 | 540 | 15.6 | 2.9 | 80  |
|  Level 2 | 1113 | 24.1 | 2.2 | 80  |
|  Level 3 | 2377 | 61.2 | 2.6 | 80  |
|  Sample | Mean (μg/mL) | S.D. (μg/mL) | %C.V. | N  |
| --- | --- | --- | --- | --- |
|  Kappa Light Chain (KAP) Urine  |   |   |   |   |
|  Level 1 | 2.33 | 0.059 | 2.5 | 30  |
|  Level 2 | 9.34 | 0.236 | 2.5 | 30  |
|  Level 3 | 19.2 | 0.25 | 1.3 | 30  |
|  Sample | Mean (μg/mL) | S.D. (μg/mL) | %C.V. | N  |
| --- | --- | --- | --- | --- |
|  Lambda Light Chain (LAM) Serum  |   |   |   |   |
|  Level 1 | 247 | 5.1 | 2.1 | 80  |
|  Level 2 | 718 | 14.0 | 1.9 | 80  |
|  Level 3 | 1233 | 37.2 | 3.0 | 80  |
|  Sample | Mean (μg/mL) | S.D. (μg/mL) | %C.V. | N  |
| --- | --- | --- | --- | --- |
|  Lambda Light Chain (LAM) Urine  |   |   |   |   |
|  Level 1 | 8.69 | 0.204 | 2.4 | 30  |
|  Level 2 | 17.1 | 0.20 | 1.2 | 30  |
|  Level 3 | 30.0 | 0.89 | 3.0 | 30  |

This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.

file: id510k.sse

---

**Source:** [https://fda-staging.innolitics.com/device/K964260](https://fda-staging.innolitics.com/device/K964260)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).