The ERA 300 is intended for use in the following conditions: - Temporary External Pacing Function. Provides temporary stimulation under DDD, DDI, VDD, DOO, VVI, VOO, AAI, or AOO modalities during implantable pacemaker procedures or physician evaluations. - Lead Test Function. Determines in situ lead characteristics of impedance, capture threshold, P/R wave amplitude and P/R wave slew rate. Determines the in vivo retrograde conduction time. - Pacemaker Test Function. Tests and analyzes the in vitro operation of external or implantable pulse generators. Determines the parameters: modality, pulse amplitude and width, sensitivity, refractory period, A/V delay, and rate/interval.
Device Story
ERA 300 is a portable, non-implantable dual-chamber pacing system analyzer. Used by physicians during pacemaker implantation or invasive troubleshooting. Inputs include electrical signals from pacing leads and pulse generators. Device functions as an external programmable pulse generator and diagnostic tester. It measures lead impedance, capture threshold, P/R wave amplitude, slew rate, and retrograde conduction time. It also analyzes pulse generator parameters like modality, amplitude, width, sensitivity, refractory period, A/V delay, and rate/interval. Output is displayed for the clinician to evaluate pacing system performance and verify pulse generator function. Benefits include ensuring proper lead placement and verifying correct operation of implantable or external pulse generators during clinical procedures.
Clinical Evidence
No clinical data. Qualification testing performed, including system checks, electromagnetic compatibility, circuit design, firmware, battery, and test cable performance.
Technological Characteristics
Portable dual-chamber pacing system analyzer. Includes main unit, battery recharger (ACD 300), and pacemaker test cable (EK-4-E). Features firmware-based control for pacing stimulation and signal analysis. Operates as an external programmable pulse generator. Testing included electromagnetic compatibility and circuit design verification.
Indications for Use
Indicated for patients undergoing invasive pacing system procedures, including implantable pacemaker procedures or physician evaluations, requiring temporary external pacing, lead threshold determination, or pulse generator function testing.
Regulatory Classification
Identification
An external pacemaker pulse generator (EPPG) is a prescription device that has a power supply and electronic circuits that produce a periodic electrical pulse to stimulate the heart. This device, which is used outside the body, is used as a temporary substitute for the heart's intrinsic pacing system until a permanent pacemaker can be implanted, or to control irregular heartbeats in patients following cardiac surgery or a myocardial infarction. The device may have adjustments for impulse strength, duration, R-wave sensitivity, and other pacing variables.
Special Controls
*Classification.* Class II (special controls). The special controls for this device are:(1) Appropriate analysis/testing must validate electromagnetic compatibility (EMC) within a hospital environment.
(2) Electrical bench testing must demonstrate device safety during intended use. This must include testing with the specific power source (
*i.e.,* battery power, AC mains connections, or both).(3) Non-clinical performance testing data must demonstrate the performance characteristics of the device. Testing must include the following:
(i) Testing must demonstrate the accuracy of monitoring functions, alarms, measurement features, therapeutic features, and all adjustable or programmable parameters as identified in labeling;
(ii) Mechanical bench testing of material strength must demonstrate that the device and connection cables will withstand forces or conditions encountered during use;
(iii) Simulated use analysis/testing must demonstrate adequate user interface for adjustable parameters, performance of alarms, display screens, interface with external devices (
*e.g.* data storage, printing), and indicator(s) functionality under intended use conditions; and(iv) Methods and instructions for cleaning the pulse generator and connection cables must be validated.
(4) Appropriate software verification, validation, and hazard analysis must be performed.
(5) Labeling must include the following:
(i) The labeling must clearly state that these devices are intended for use in a hospital environment and under the supervision of a clinician trained in their use;
(ii) Connector terminals should be clearly, unambiguously marked on the outside of the EPPG device. The markings should identify positive (+) and negative (−) polarities. Dual chamber devices should clearly identify atrial and ventricular terminals;
(iii) The labeling must list all pacing modes available in the device;
(iv) Labeling must include a detailed description of any special capabilities (
*e.g.,* overdrive pacing or automatic mode switching); and(v) Appropriate electromagnetic compatibility information must be included.
Predicate Devices
SeaMED Model 3300 Dual-Chamber Pacing System Analyzer (K883930)
Medtronic Model 5311B A-V Pacing System Analyzer (K884331)
Submission Summary (Full Text)
{0}
October 18, 1996
510(k)
ERA 300
K964190
July 10, 1997
# A. 510(k) Summary of Safety and Effectiveness
## 1. General Information
| Device Name | |
| --- | --- |
| Proprietary Name | ERA 300 Dual Chamber Pacing System Analyzer |
| Classification Name | External Pacemaker Pulse Generator
Pacemaker Electrode Function Tester
Pacemaker Generator Function Analyzer |
| Manufacturer | BIOTRONIK GmbH & Co.
Woermannkehre 1
D-12359 Berlin
Germany |
| Manufacturer's Registration Number | 7010992 |
| Applicant's Name & Address | BIOTRONIK, Inc.
6024 Jean Road
Lake Oswego, OR 97035 |
| Establishment Registration Number | 1028232 |
| Performance Standards | No applicable performance standards have been promulgated for these devices. |
## 2. Device Description
The ERA 300 is a portable, dual-chamber pacing system analyzer (non-implantable) designed to test the electrical performance of the pulse generator and the pacing lead system at the time of pacemaker implantation and during invasive pacemaker troubleshooting or evaluation procedures. It can also operate as an external programmable pulse generator.
## 3. Substantial Equivalence
The ERA 300 is substantially equivalent to the FDA-cleared SeaMED Model 3300 Dual-Chamber Pacing System Analyzer, approved 8/19/86 (K883930), and the Medtronic Model 5311B A-V Pacing System Analyzer, approved (K884331). Data to support this statement are provided in the premarket notification.
## 4. Intended Use
The ERA 300 is intended for use in the following conditions:
- Temporary External Pacing Function. Provides temporary stimulation under DDD, DDI, VDD, DOO, VVI, VOO, AAI, or AOO
2
w
{1}
October 18, 1996
510(k)
ERA 300
modalities during implantable pacemaker procedures or physician evaluations.
- Lead Test Function. Determines in situ lead characteristics of impedance, capture threshold, P/R wave amplitude and P/R wave slew rate. Determines the in vivo retrograde conduction time.
- Pacemaker Test Function. Tests and analyzes the in vitro operation of external or implantable pulse generators. Determines the parameters: modality, pulse amplitude and width, sensitivity, refractory period, A/V delay, and rate/interval.
## 5. Qualification Testing
Qualification testing for the ERA 300 is provided in this submission for the tests described in Table III, pages 14–33. Table III is organized into six main categories: ERA 300 Main Unit (System Checks); ERA 300 Main Unit (Electromagnetic Compatibility and Circuit Design); ERA 300 Main Unit (Firmware); ERA 300 Battery; Battery Recharger (ACD 300); and Pacemaker Test Cable (EK-4-E). Internal standards and specifications were used when relevant international standards were not available.
## 6. Labeling
Proposed labeling for the ERA 300 is included in the premarket notification. The product labeling includes instructions for use adequate to assure safe and effective operation of the device.
{2}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20850
JUL 10 1997
Mr. Taras Tatarko, M.S.
Biotronik, Inc.
6024 Jean Road
Lake Oswego, Oregon 97035-5369
Re: K964190
ERA 300 Dual Chamber Pacing System Analyzer
Regulatory Class: III (three)
Product Code: 74 DTE
Dated: May 20, 1997
Received: May 21, 1997
Dear Mr. Tatarko:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}
Page 2 - Mr. Taras Tatarko, M.S.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular,
Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices
and Radiological Health
Enclosure
{4}
# INDICATIONS FOR USE
The ERA 300 is indicated for use during invasive pacing system procedures in the following activities:
- **TEMPORARY EXTERNAL PACING**
To provide temporary stimulation under DDD, DDI, DOO, VVI, VDD, VOO, AAI, or AOO modalities during implantable pacemaker procedures or physician evaluations.
- **LEAD THRESHOLD DETERMINATION**
To determine in situ lead characteristics of impedance, capture threshold, P/R wave amplitude and P/R wave slew rate. To determine the in vivo retrograde conduction time.
- **PACEMAKER FUNCTION TEST**
To test and analyze the in vitro operation of external or implantable pulse generators. To determine the following parameters: pulse amplitude and width, sensitivity, refractory period, A/V delay, and rate/interval.


Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.