CONVERTORS DRAPE MATERIAL

K964125 · Allegiance Healthcare Corp. · KKX · Dec 23, 1996 · General, Plastic Surgery

Device Facts

Record IDK964125
Device NameCONVERTORS DRAPE MATERIAL
ApplicantAllegiance Healthcare Corp.
Product CodeKKX · General, Plastic Surgery
Decision DateDec 23, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4370
Device ClassClass 2

Indications for Use

The Convertors® Drapes and Leggings are devices intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination.

Device Story

Convertors® Drapes and Leggings function as sterile barriers; composed of nonwoven polyester fibers adhesively laminated to 1.25 mil embossed polyethylene film. Used in clinical/surgical settings to isolate incision sites from contaminants. Protects patients by preventing microbial transfer; aids clinicians in maintaining sterile fields. No active components; passive barrier technology.

Clinical Evidence

Bench testing only. Biocompatibility evaluated per ISO 10993-1 (cytotoxicity, sensitization, irritation/intracutaneous reactivity). Physical performance testing included hydrostatic head, impact penetration, synthetic blood penetration (ASTM 1670-1995), Phi X174 viral penetration (ASTM 1671-1995), Trapezoid Tear, Grab Tensile, Mullen Burst, Basis Weight, and flammability (Class 2).

Technological Characteristics

Nonwoven adhesively bonded polyester fibers laminated to 1.25 mil embossed blended polyethylene film. Passive barrier device. Flammability Class 2. Tested per ASTM 1670-1995 and ASTM 1671-1995.

Indications for Use

Indicated for use as a protective patient covering to isolate surgical incision sites from microbial and other contamination during surgical procedures.

Regulatory Classification

Identification

A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.

Special Controls

*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.

Predicate Devices

Submission Summary (Full Text)

{0} DEC 20 '96 08:57:47 3-4- 3-4- 3-4- 3-4- 3-4- 3-4- 3-4- 3-4- 3-4- 3-4- 3-4- 3-4- 3-4- 3-4- 3-4- 3-4- 3-4 DEC 23 1996 Allegiance Allegiance Healthcare Corporation 1500 Waukegan Road McGaw Park, Illinois 60085-6787 847.473.1500 FAX: 847.785.2460 K964125 ## XII. **SMDA REQUIREMENTS** ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS ### Convertors® Drapes and Leggings **Manufacturer:** Allegiance Healthcare Corporation Convertors® Operations One Butterfield Trail El Paso, Texas 79906 **Regulatory Affairs Contact:** Sharon Robbins Allegiance Healthcare Corporation 1500 Waukegan Road MPK McGaw Park, IL 60085 **Telephone:** (847) 785-3311 **Date Summary Prepared:** September, 1996 **Common Name:** Convertors® Drapes and Leggings **Classification:** Class II per 21CFR § 878.4370 **Predicate Device:** Convertors® Drape Materials **Description:** These drapes and leggings will be composed of a nonwoven fabric adhesively laminated to a synthetic polymer film. The nonwoven contains adhesively bonded polyester fibers. The film is a 1.25 mil finely embossed blended polyethylene construction. 23 {1} DEC 20 '96 08:50AM 344-345.2460 P.4/4 # Allegiance Allegiance Healthcare Corporation 1500 Waukegan Road McGaw Park, Illinois 60085-6787 847.473.1500 FAX: 847.785.2460 ## XII. **SMDA REQUIREMENTS** (continued) ### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS **Convertors® Drapes and Leggings** **Intended Use:** The Convertors® Drapes and Leggings are devices intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. **Substantial Equivalence:** The Convertors® Drapes and Leggings are substantially equivalent to the Convertors® legging drape materials in that: - the intended use is the same - the performance attributes are similar **Summary of testing:** All materials used in the fabrication of this Convertors® drapes and leggings were evaluated through biological qualification safety tests as outlined in in ISO 10993 Part-1 “Biological Evaluation of Medical Devices”. The biocompatibility tests performed were cytotoxicity, sensitization, and irritation/intracutaneous reactivity. These materials also met the requirements of hydrostatic head, impact penetration, synthetic blood penetration (ASTM 1670-1995), Phi X174 (ASTM 1671-1995) Trapazoid Tear, Grab Tensile, Mullen Burst, Basis Weight, and flammability, Class 2. These materials have met the requirements of the identified tests and were found to be acceptable for the intended use. 24
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...