← Product Code [LJT](/productcode/LJT) · K964117

# NON-PVC FLUID PATH PORT ACCESS INFUSION SET WITH SLIT SEPTUM INJECTION SITE (K964117)

_Kawasumi Laboratories Co., Ltd. · LJT · Mar 14, 1997 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K964117

## Device Facts

- **Applicant:** Kawasumi Laboratories Co., Ltd.
- **Product Code:** [LJT](/productcode/LJT.md)
- **Decision Date:** Mar 14, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5965
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

THE NON-PVC PORT ACCESS INFUSION SET IS ROUTINELY USED TO ACCESS IMPLANTED MEDICATION PORTS FOR THE DELIVERY OF MEDICATIONS.

## Device Story

Sterile, single-use winged access set; consists of needle, wings, and non-PVC tubing. Used by clinicians in hospital settings to access implanted medication ports. Device punctures port septum to establish fluid path for medication delivery. Coextruded tubing prevents fluid contact with PVC. Benefits include non-reactive fluid path for safe medication administration.

## Clinical Evidence

Bench testing only. Biocompatibility testing performed on body fluid-contacting materials. Device meets USP XXII Bacterial Endotoxins Test (LAL) requirements.

## Technological Characteristics

Non-PVC coextruded tubing; winged needle assembly; sterile, single-use. Materials are non-reactive. No electronic or software components.

## Regulatory Identification

A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.

## Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.

## Predicate Devices

- Kawasumi Laboratories Port Access Infusion Set
- Baxter Health Care Interlink Slit Septum Injection Site

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

K 96 4/17

# MAR 14 1997 510(K) SUMMARY

## 1. SUBMITTER

KAWASUMI LABORATORIES, INC.
3-28-15 MINAMI-OH
SHINAGAWA-KU, TOKYO 140 JAPAN
PHONE: 81-3-376-1151
FAX: 81-3-376-3235
CONTACT: MR. SHOZO MORIYAMA

U.S. AGENT:
KAWASUMI LABORATORIES AMERICA, INC
5100 W. KENNEDY BLVD., SUITE 455
TAMPA, FL 33609
PHONE: (813) 282-3733
FAX: (813) 282-3544
CONTACT: MR. JACK PAVLO

## 2. NAME OF DEVICE:

KAWASUMI LABORATORIES NON-PVC FLUID PATH PORT ACCESS INFUSION SET WITH SLIT SEPTUM INJECTION SITE

COMMON NAME: WINGED ACCESS SET
CLASSIFICATION: UNCLASSIFIED

## 3. PREDICATE DEVICE:

KAWASUMI LABORATORIES PORT ACCESS INFUSION SET AND BAXTER HEALTH CARE'S "INTERLINK" SLIT SEPTUM INJECTION SITE.

## 4. DESCRIPTION OF THE DEVICE:

A NON-PVC PORT ACCESS INFUSION SET IS A STERILE, SINGLE USE DEVICE WITH A NEEDLE, WING, NON-PVC I.D. TUBING AND NON-PVC COMPONENTS USED FOR ACCESSING IMPLANTED MEDICATION PORTS.

BASIC CONCEPT:
A DEVICE USED FOR ACCESSING AN IMPLANTED MEDICATION PORT BY PUNCTURING THE SEPTUM OF THE MEDICATION PORT USED FOR THE DELIVERY OF MEDICATION. FLUID ADMINISTRATION THROUGH THE NON-PVC FLUID PATH PORT ACCESS INFUSION SET ARE THOSE GENERALLY USED IN HOSPITALS AND FOR CHEMOTHERAPY.

SIGNIFICANT PERFORMANCE CHARACTERISTICS: THERE ARE NO NEW PERFORMANCE CHARACTERISTICS OF THIS DEVICE COMPARED TO SUBSTANTIALLY EQUIVALENT DEVICES MARKETED FOR SALE IN INTERSTATE COMMERCE. BOTH DELIVER FLUID TO THE VASCULAR SYSTEM THROUGH A NON-REACTIVE MATERIAL.

## 5. INTENDED USE:

THE NON-PVC PORT ACCESS INFUSION SET IS ROUTINELY USED TO ACCESS IMPLANTED MEDICATION PORTS FOR THE DELIVERY OF MEDICATIONS.

## 6. TECHNOLOGICAL CHARACTERISTICS:

THERE ARE NO TECHNOLOGICAL CHARACTERISTICS OF THIS DEVICE TO SUBSTANTIALLY EQUIVALENT DEVICES INCLUDING KAWASUMI AND BAXTER SETS BEING MARKETED FOR SALE IN INTERSTATE COMMERCE. THIS DEVICE USES A COEXTRUDED TUBE WHICH DOES NOT EXPOSE THE FLUID DELIVERED INTRAVASCULARY TO PVC.

9

{1}

7. PERFORMANCE DATA: KAWASUMI LABORATORIES HAS CONDUCTED BIOCOMPATIBILITY TESTS ON THE BODY FLUID CONTACTING MATERIAL PORTIONS OF THE DEVICE, AND KL BELIEVES THE BIOCOMPATIBILITY DATA SHOW THE DEVICE IS SUITABLE FOR ITS INTENDED USE.

THE DEVICE MEETS THE LIMULUS AMEBOCYTE LYSATE ("LAL") TEST DESCRIBED AT "(85) BACTERIAL ENDOTOXINS TEST", USP XXII, PP 1493-1495.

8. CONCLUSIONS: THE DEVICE MEETS ALL BIOCOMPATIBILITY AND PYROGENICITY TEST REQUIREMENTS. THEREFORE, IT IS AS SAFE AS THE PREDICATE DEVICE AND PERFORMS AS WELL AS THE PREDICATE DEVICE.

---

**Source:** [https://fda-staging.innolitics.com/device/K964117](https://fda-staging.innolitics.com/device/K964117)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).