DSBAS OPTION THE MAMMOMAT 3000 MAMMOGRAPHY X-RAY SYSTEM

K964069 · Siemens Medical Solutions USA, Inc. · IZH · May 8, 1997 · Radiology

Device Facts

Record IDK964069
Device NameDSBAS OPTION THE MAMMOMAT 3000 MAMMOGRAPHY X-RAY SYSTEM
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeIZH · Radiology
Decision DateMay 8, 1997
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1710
Device ClassClass 2

Indications for Use

The Siemens Digital Stereotactic Biopsy Attachment System (DSBAS) marketed as OPDIMA (Optimized Digital Mammography) is a digital small field x-ray option for the commercially available Mammomat 3000 mammographic x-ray system. The Mammomat 3000 with DSBAS option is intended for use in small field mammographic x-ray imaging. Such small field imaging is used during stereotactic biopsy and spot localization.

Device Story

OPDIMA is a digital small field x-ray attachment for the Mammomat 3000 mammographic system. It utilizes a CCD detector to capture X-ray images during stereotactic biopsy and spot localization. The system includes a biopsy unit, controller, and workstation for image processing and display. Operators are trained healthcare professionals. The workstation provides image manipulation tools including zoom (2x, 4x, 8x), magnifying glass, contrast enhancement, image inversion, histogram display, electronic rulers, and selectable image filters. These tools assist clinicians in visualizing small field areas, facilitating accurate biopsy targeting and localization, thereby potentially improving diagnostic and procedural precision.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and bench testing of system components to ensure compliance with 21 CFR Subchapter J performance standards.

Technological Characteristics

System comprises Mammomat 3000 X-ray stand/generator, CCD detector (49x85 mm imaging area), biopsy unit, and workstation. Pixel size: 24 µm (spot) or 48 µm (stereo). Matrix: 2048x3584 (spot) or 1024x1792 (stereo); 12-bit depth. Resolution: ≥10 lp/mm (stereo), ≥11 lp/mm (spot). Workstation: 32-bit, 140 MHz processor, 64 MB RAM, 2 GB HDD. Monitor: 1024x1280 resolution, 8-bit depth. Complies with 21 CFR Subchapter J.

Indications for Use

Indicated for small field mammographic x-ray imaging, specifically for stereotactic biopsy and spot localization procedures in patients requiring mammographic examination.

Regulatory Classification

Identification

A mammographic x-ray system is a device intended to be used to produce radiographs of the breast. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0} K964069 # SIEMENS ## 510 (k) SUMMARY ### I. GENERAL INFORMATION: #### Establishment: - **Address:** Siemens Medical Systems, Inc. 186 Wood Avenue South Iselin, NJ. 08830 MAY - 8 1997 - **Registration Number:** 2240869 - **Contact Person:** Kathleen Rutherford Manager, Regulatory Submissions Telephone: (908) 321-4779 TELEFAX: (908) 321-4841 #### Date of Summary Preparation: ##### Device Name: - **Trade Name:** OPDIMA Mammographic X-ray System Mammomat 3000 with DSBAS (Digital Stereotactic Attachment System) - **Common Name:** Mammographic X-ray System - **Classification Name:** Mammographic X-ray System - **Classification:** Class II **Performance Standards:** 21 CFR, Subchapter J All system components to which the above standard applies are certified to conform with 21 CFR subchapter J ### II. SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING THE SUBSTANTIAL EQUIVALENCE DETERMINATION. #### Device Description: The fully functional OPDIMA mammographic x-ray system configuration consists of the commercially available Mammomat 3000 mammographic x-ray system configured with the DSBAS (Digital Stereotactic Biopsy Attachment System) option. The Mammomat 3000 with DSBAS option is a Digital Stereotactic Biopsy Mammography System intended for use in small field digital mammographic examinations. The Mammomat 3000 with the DSBAS option consists of the Mammomat 3000 stand and generator configured with a camera assembly, i.e., CCD detector, biopsy unit and a workstation for the display and processing of images, #### Siemens Medical Systems, Inc. Imaging Systems Group 186 Wood Avenue South Iselin, NJ 08830 Tel: (908) 321-4500 Fax: (908) 494-2250 {1} # SIEMENS modem, MO drive, trolley, Biopsy controller and DSBAS software (Application Software and Biopsy Controller software). The user can select three modes of operation, Spot, biopsy or normal (non-digital) mode. The system is designed with key image processing functions, Zoomfactor, Magnifying Glass, Contrast Enhancement, Image invert, Histogram, Electronic ruler and image filter. Zoomfactor and Magnifying glass provide 2x, 4x and 8x magnification. For image filtering, five image filters are selectable. The Histogram feature displays the population for each grey scale value. ## Intended Use: The Siemens Digital Stereotactic Biopsy Attachment System (DSBAS) marketed as OPDIMA (Optimized Digital Mammography) is a digital small field x-ray option for the commercially available Mammomat 3000 mammographic x-ray system. The Mammomat 3000 with DSBAS option is intended for use in small field mammographic x-ray imaging. Such small field imaging is used during stereotactic biopsy and spot localization. ## Technological Characteristics: The DSBAS mammographic X-ray system consists of the Mammomat 3000 configured with the DSBAS (Digital Stereotactic Biopsy Attachment System) Option. The major system components consist of the Mammomat 3000 X-ray stand and generator, CCD detector, biopsy unit, Biopsy controller, DSBAS system software and workstation. The CCD detector has an imaging area of 49 x 85 mm. The pixel sizes vary according to the mode of operation; 24 µm in spot mode and 48 µm stereo mode. The pixel matrices are 2048 x 3584 pixel in spot mode and 1024 x 1792 pixel in stereo mode with a pixel depth of 12 bits. The image resolution is ≥ 10 lp/mm in stereo mode and ≥ 11 lp/mm in spot mode. The DSBAS workstation is a 32 bit, 140 Mhz processor with 64 MB RAM, 32 kB and 2GB hard disk. The monitor has a resolution of 1024 x 1280 with 8-bit pixel depth. The system is equipped 2GB hard disc for archiving of approximately 500 images and optional disc archiving of 2.6 GB (250 images in spot mode and 1000 images in biopsy mode) per additional disc. ## General Safety and Effectiveness Concerns: Instructions for use are included within the device labeling and the information provided will enable the trained healthcare professional to operate the device in a safe and efficacious manner. Furthermore the operators are health care professionals familiar with and responsible for the X-ray examinations to be performed. To minimize electrical, mechanical and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. {2} # SIEMENS Substantial Equivalence: In the opinion of Siemens Medical Systems, Inc., Mammomat 3000 with the DSBAS option is of comparable type and substantially equivalent to the following currently marketed Small Field Digital Mammographic X-ray Systems: | Predicate Device | Manufacturer | | --- | --- | | Mammovision | Fischer Imaging, Inc. | | DSM | Lorad, Inc. | ![img-0.jpeg](img-0.jpeg) Kathleen Rutherford Manager, Regulatory Submissions 10/8/96 Date
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...