K964017 · Avecor Cardiovascular, Inc. · DTQ · Aug 5, 1997 · Cardiovascular
Device Facts
Record ID
K964017
Device Name
AFFINITY BLOOD PUMP SYSTEM
Applicant
Avecor Cardiovascular, Inc.
Product Code
DTQ · Cardiovascular
Decision Date
Aug 5, 1997
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.4220
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Affinity™ Blood Pump System is intended to be used in an extracorporeal perfusion circuit to pump blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
Device Story
Affinity Blood Pump System is an extracorporeal perfusion device used during cardiopulmonary bypass (CPB) procedures. System components include Model 2000 Affinity Pump Console, rotor, holder, pump chamber, and connecting tubing. Operates on peristaltic principle: polyurethane pump chamber fills passively via reservoir fluid height; pump rotor rollers exert peristaltic action against chamber to generate forward blood flow. Console controls variable speed motor and monitors flow rate, RPM, pressure, timed events, and battery status. Used in clinical settings by perfusionists/clinicians to maintain circulation during surgery. Output provides controlled blood flow to support patient hemodynamics. Benefits include reliable extracorporeal circulation during routine bypass procedures.
Clinical Evidence
No clinical data. Evidence consists of in vitro bench testing, including operating characteristics, cellular damage (hemolysis, platelet and white cell response), material biocompatibility, sterility, pyrogenicity, and sterilant residuals. Software validation and verification were also performed.
Technological Characteristics
Peristaltic pump system. Materials: Polyurethane pump chamber, polycarbonate connectors, PVC protective caps. Console features: variable speed motor control, monitoring for flow, RPM, pressure, and battery. Power: 100-120/220-240 Vac. Standards: IEC 601, UL 544. CE marked.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass procedures requiring extracorporeal blood circulation for up to 6 hours.
Regulatory Classification
Identification
A cardiopulmonary bypass heart-lung machine console is a device that consists of a control panel and the electrical power and control circuitry for a heart-lung machine. The console is designed to interface with the basic units used in a gas exchange system, including the pumps, oxygenator, and heat exchanger.
Predicate Devices
Stockert Instrumente (Sorin) Pump Console Model No. 10-10-00
Submission Summary (Full Text)
{0}
K 904017
# SECTION II
## AVECOR CARDIOVASCULAR INC.
### AFFINITY BLOOD PUMP SYSTEM
### SUPPORTING SUMMARY FOR 510 (k) NOTIFICATION
1. Date Prepared: September 30, 1996
2. Submitter: AVECOR Cardiovascular Inc.
13010 County Road 6
Plymouth, MN 55441
3. Contact: Dennis E. Steger
Director Regulatory Affairs/
Quality Assurance
(800) 328-3320
4. Device Name: Cardiopulmonary Bypass Roller-Type Pumps, Consoles and Bypass Pump Tubing have been classified by the Cardiovascular Device Classification Panel as Class II devices, as proposed in Section 870.4220/4370/4380/4390, Title 21 of the Code of Federal Regulations
5. Device Description & Comparison to Predicate Device:
To establish the substantial equivalency of the AVECOR Affinity Pump System to the Stockert Instrumente (Sorin) Pump design, specifications, materials, indications for use, and performance characteristics were compared.
## DESIGN
The AVECOR Affinity Blood Pump System is intended for use in an extracorporeal perfusion circuit to pump blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.
The AVECOR Cardiovascular Affinity Pump System consists of the Model 2000 Affinity Pump Console, Affinity Pump Rotor, Affinity Pump Holder, Affinity Pump Chamber, 5/8" Connecting Tube and an AVECOR venous reservoir with 5/8" outlet.
The AVECOR Affinity Blood Pump Console and components shall conform with all the applicable requirements of IEC 601, UL 544 and will display the European Community CE marking.
The Affinity Pump Console, Model 2000 provides a system for controlling a variable speed pump motor and additional systems for monitoring flow rate, pump speed (RPM), pressure, timed events and battery status.
{1}
The AVECOR Affinity Pump has been designed to operate on the principles of a peristaltic pump for extracorporeal use. The pump consists of a polyurethane chamber incorporating polycarbonate inlet and outlet connectors, which fills passively due to the filling pressure exhibited by the fluid height in a reservoir. The peristaltic action of the rollers against the filled pump chamber results in forward flow. Flow rate is affected by the combination of reservoir volume, RPM's of the pump rotor and the total circuit resistance.
The Stockert Instrumente (Sorin) Pump Console Model No. 10-10-00, provides a system for controlling a variable speed pump motor and additional systems for monitoring flow rate, pump speed (RPM), pressure and timed events.
The Stockert Instrumente (Sorin) Pump is designed to operate on the principles of a peristaltic pump for extracorporeal use. The pump consists of a polyvinyl chloride tubing segment with polycarbonate connectors at each end. Flow rate is affected by the RPM's, of the pump rotor, total circuit resistance and the size of the tubing segment used.
## SPECIFICATIONS & MATERIALS
A detailed comparison of specifications and materials can be found in Table 1. Specifications for the Stockert (Sorin) console are taken from Stockert's labeling.
| | Affinity Console | Stockert Console |
| --- | --- | --- |
| Input Power | • 100 or 110-120 Vac, 50-60 Hz,
• 220-240 Vac, 50-60 Hz | • 100 or 110-120 Vac, 50-60 Hz
• 220-240 Vac, 50-60 Hz |
| Internal Batteries | Yes | No (External Battery Pack Optional) |
| Case Dimensions | 13" H x 7" W x 17" D | 12" H x 7" W x 19" D |
| Weight | 50 lb. | 48 lb. |
{2}
# Pump Chamber Comparison of Materials
| | Affinity | Stockert (Sorin) |
| --- | --- | --- |
| Chamber | Polyurethane | Polyvinyl Chloride |
| Connectors | Polycarbonate | Polycarbonate |
| Protective Caps | PVC | PVC |
## INDICATIONS FOR USE
The Affinity Blood Pump System and the Stockert Instrumente (Sorin) Pump System are indicated for use in an extracorporeal perfusion circuit to pump blood during cardiopulmonary bypass procedures.
## PERFORMANCE CHARACTERISTICS
The following tests were conducted by AVECOR to establish the safety and effectiveness of the Affinity Blood Pump System.
| Test Description | Location |
| --- | --- |
| In Vitro Operating Characteristics | Section III |
| In Vitro Cellular Damage | Section III |
| Material Biocompatibility | Section IV |
| Sterility | Section V |
| Pyrogenicity | Section V |
| Sterilant Residuals | Section V |
| Software Validation & Verification | Appendix E |
Protocols describing the test methods utilized in this application may be found in the appropriate Appendix.
The results indicate that the Affinity Blood Pump System is substantially equivalent in performance to the Stockert Instrumente (Sorin) Pump System.
## BLOOD TRAUMA SUMMARY STATEMENT
Blood trauma testing shows that in all aspects evaluated (hemolysis, platelet and white cell response) the Affinity Blood Pump System and the Stockert Instrumente (Sorin) Pump System are equivalent.
{3}
# CONCLUSION
WE HAVE CONSIDERED DESIGN, MATERIALS, INDICATIONS FOR USE, PERFORMANCE CHARACTERISTICS, ALONG WITH A DIRECT COMPARISON OF TEST RESULTS BETWEEN THE AFFINITY AND STOCKERT PUMP SYSTEMS IN OUR APPLICATION. IT IS BASED UPON THE REVIEW OF ALL OF THESE PARAMETERS THAT AVECOR CARDIOVASCULAR INC. DETERMINES THAT THE AFFINITY BLOOD PUMP SYSTEM IS SUBSTANTIALLY EQUIVALENT TO THE CURRENTLY MARKETED STOCKERT INSTRUMENTS (SORIN) BLOOD PUMP SYSTEM.
{4}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
Mr. Dennis E. Steger
Director Regulatory Affairs/
Quality Assurance
Avecor Cardiovascular, Inc.
7611 Northland Drive
Minneapolis, Minnesota 55428
AUG - 5 1997
Re: K964017
AFFINITY™ Blood Pump System
Regulatory Class: II (Two)
Product Code: 74 DTQ
Dated: May 9, 1997
Received: May 12, 1997
Dear Mr. Steger:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}
Page 2 - Mr. Dennis E. Steger
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html."
Sincerely yours,
Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{6}
870.4370 - DWB II - Roller-type CPB Blood Pump
510(k) Number (if known): K964017
Device Name: AFFINITY Blood Pump System
Indications For Use:
The Affinity™ Blood Pump System is intended to be used in an extracorporeal perfusion circuit to pump blood during routine cardiopulmonary bypass procedures up to 6 hours in duration.

Dennis E. Steger
Director Regulatory Affairs/
Quality Assurance
AVECOR Cardiovascular Inc.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of Cardiovascular, Respiratory,
and Neurological Devices
510(k) Number K964017
Prescription Use ☑ (Per 21 CFR 801.109)
OR
Over-The-Counter Use ☐
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.