K963954 · Caltag Laboratories, Inc. · GKZ · Nov 25, 1996 · Hematology
Device Facts
Record ID
K963954
Device Name
CD 19 R-PE, CD19 TRI-COLOR MONOCLONAL ANTIBODY
Applicant
Caltag Laboratories, Inc.
Product Code
GKZ · Hematology
Decision Date
Nov 25, 1996
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.5220
Device Class
Class 2
Indications for Use
CALTAG CD19 R-PE and CD19 TRI-COLOR are fluorochrome conjugated monoclonal antibody reagents that may be used to enumerate CD19+ lymphocytes in human peripheral blood by flow cytometric methods.
Device Story
Device consists of fluorochrome-conjugated (R-PE or TRI-COLOR) mouse monoclonal antibodies targeting human CD19 cell surface antigens. Used in clinical laboratories by trained personnel. Workflow: peripheral blood leukocytes incubated with antibody; unbound antibody washed away; red blood cells lysed; fixative added. Stained, fixed cells analyzed via flow cytometry. Output is percentage of CD19+ lymphocytes. Clinicians use this data for immunophenotyping and assessment of B-cell populations in patient samples. Benefits include standardized identification of B-cell subsets for diagnostic evaluation.
Clinical Evidence
Bench testing and clinical correlation study performed. Study included 155 healthy donors (age 16-72) to establish expected values and 20 abnormal donors for correlation. Correlation study (n=175) compared Caltag antibodies against Coulter predicates, yielding r² values between 0.96 and 0.98. Intra-lab and inter-lab reproducibility studies (n=30 per antibody) demonstrated low CVs across high, medium, and low ranges. Specificity data confirmed binding to lymphocyte regions with minimal cross-reactivity to monocytes, granulocytes, platelets, or RBCs.
Technological Characteristics
Monoclonal antibodies conjugated to R-PE or TRI-COLOR fluorochromes. Supplied as liquid in PBS. Sensing principle: immunofluorescence via flow cytometry. Analysis involves gating on lymphocyte populations. No specific ASTM material standards cited. Software/algorithm: standard flow cytometric data analysis (gating).
Indications for Use
Indicated for the enumeration of CD19+ B lymphocytes in human peripheral blood using flow cytometry. No specific age or gender contraindications are listed.
Regulatory Classification
Identification
An automated differential cell counter is a device used to identify one or more of the formed elements of the blood. The device may also have the capability to flag, count, or classify immature or abnormal hematopoietic cells of the blood, bone marrow, or other body fluids. These devices may combine an electronic particle counting method, optical method, or a flow cytometric method utilizing monoclonal CD (cluster designation) markers. The device includes accessory CD markers.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA document entitled “Class II Special Controls Guidance Document: Premarket Notifications for Automated Differential Cell Counters for Immature or Abnormal Blood Cells; Final Guidance for Industry and FDA.”
Predicate Devices
Coulter CD19 RD1 Monoclonal antibody
Coulter CD19 FITC monoclonal antibody
Submission Summary (Full Text)
{0}
CALTAG LABORATORIES
K963954
510(K) SUMMARY
SUMMARY OF SAFETY AND EFFECTIVENESS DATA
NOV 25 1996
CD19 R-PE, CD19 TRI-COLOR Mouse Monoclonal Antibodies
To Human Cell Surface Antigens by Flow Cytometry
## NAME AND LOCATION OF MANUFACTURER:
Caltag Laboratories, Inc.
1849 Old Bayshore Highway
Suite 200
Burlingame, CA 94010
(800) 874-4007
## NAME OF CONTACT PERSON:
Robert C. Johnson
Executive Vice President
Caltag Laboratories, Inc.
## DATE OF PREPARATION OF SUMMARY:
October 1, 1996
Caltag Laboratories Inc. • 1849 Bayshore Blvd., Suite #200 • Burlingame, CA 94010
Telephone (415) 652-0468 • (800) 874-4007 • FAX (415) 652-9030
{1}
-2-
## TRADE NAME OF THE DEVICE:
Caltag CD19 R-PE, CD19 TRI-COLOR Mouse Monoclonal Antibodies To Human Cell Surface Antigens by Flow Cytometry
## COMMON NAME:
Caltag CD19 R-PE, CD19 TRI-COLOR Monoclonal Antibody
## CLASSIFICATION NAME:
Automated Differential Cell Coulter (21 CFR 864.5220)
## LEGALLY MARKETED DEVICE (PREDICATE DEVICE) TO WHICH THE MANUFACTURER IS CLAIMING SUBSTANTIAL EQUIVALENCE:
Caltag CD19 R-PE Monoclonal Antibody to Human Cell Surface Antigens is substantially equivalent to the Coulter CD19 RD1 Monoclonal antibody for in-vitro diagnostic use.
Caltag CD19 TRI-COLOR Monoclonal Antibody to Human Cell Surface Antigens is substantially equivalent to the Coulter CD19 FITC and CD19 RD1 monoclonal antibodies for in-vitro diagnostic use.
## DESCRIPTION OF THE DEVICE:
The CALTAG CD19 R-PE and CD19 TRI-COLOR monoclonal antibodies bind to the surfaces of viable blood cells that express the CD19 antigen. To identify cells bearing the CD19 determinant, peripheral blood leukocytes are incubated with the monoclonal antibody, and washed to remove unbound antibody. Prior to removal of unbound antibody, lysis solution is added to lyse red blood cells. An appropriate fixative solution is added to lysed and washed cells. Stained and fixed cells are subsequently analyzed by flow cytometric methods.
## INTENDED USE OF THE DEVICE:
CALTAG CD19 R-PE and CD19 TRI-COLOR are fluorochrome conjugated monoclonal antibody reagents that may be used to enumerate CD19+ lymphocytes in human peripheral blood by flow cytometric methods.
{2}
-3-
SUMMARY OF THE TECHNICAL CHARACTERISTICS OF THE MANUFACTURER'S DEVICE COMPARED TO THE PREDICATE DEVICE:
## Comparisons of Caltag CD19 and Coulter CD19 Monoclonal Antibodies
| No. | Item | Caltag Antibodies | Coulter Antibodies | Comparison |
| --- | --- | --- | --- | --- |
| 1. | Intended Use | Flow Cytometry | Flow Cytometry
Immunofluorescence | Substantially equivalent |
| 2. | Specificity | CD19 | CD19 | Substantially equivalent |
| 3. | Target cell | B lymphocyte | B lymphocyte | Substantially equivalent |
| 4. | Chemical form | Monoclonal antibody | Monoclonal antibody | Substantially equivalent |
| 5. | Fluorochromes | R-PE,
TRI-COLOR | FITC,
RD1 | Substantially equivalent |
| 6. | Available forms
FITC
PE
TRI-COLOR | liquid, PBS
liquid, PBS
liquid, PBS | lyophilized
liquid, PBS
not available | Substantially equivalent |
| 7. | Sample prep. methods | whole blood | whole blood | Substantially equivalent |
| 8. | Expected values from this study (n=155)
R-PE
TRI-COLOR | 5-21%
4-24% | 4-21% (RD1)
3-23% (FITC) | Substantially equivalent |
**NON CLINICAL TESTS SUPPORTING A DETERMINATION OF SUBSTANTIAL EQUIVALENCE:**
### EXPECTED VALUE DATA
Blood samples were collected from a total of 155 apparently healthy normal donors in an age range of 16 to 72 with a mean age of 41. Samples were collected and analyzed in each of three independent laboratories. An approximately equal number of males and females were collected and analyzed in each laboratory.
The normal donor population included members of differing ethnic origins, including adult Caucasian, Black, Oriental and Hispanic.
{3}
-4-
Donors in geographically diverse areas of the United States, including the Western, Eastern and SouthCentral regions, participated in this study. Blood samples collected from each donor were stained with the CALTAG CD19 R-PE and CD19 TRI-COLOR monoclonal antibodies.
Summary of expected values for CALTAG CD19 monoclonal antibodies for all normal donors:
| procedure | mean % positive | S.D. ±2 S.D. | Range | n |
| --- | --- | --- | --- | --- |
| CD19 R-PE | 13.0 | 4.2 | 5-21 | 155 |
| CD19 TRI-COLOR | 13.9 | 5.0 | 4-24 | 155 |
## SPECIFICITY DATA
Blood samples were obtained from healthy normal donors of Caucasian, Black, Hispanic and Oriental ethnic origins. Samples of each donor were stained with CALTAG CD19 R-PE and CD19 TRI-COLOR monoclonal antibodies. Cells contained in the lymphocyte, monocyte and granulocyte regions were selected for analysis. Separate samples from the same donors were prepared for analysis of red blood cells and platelets and stained with each of the CALTAG monoclonal antibodies.
### CD19 R-PE
| Ethnic
Origin | Lymph. | Percent of Stained Cells | | | |
| --- | --- | --- | --- | --- | --- |
| | | Mono. | Gran. | Plt. | RBC |
| Caucasian | 18.0 | 0.6 | 0.9 | 0.5 | 0.5 |
| Caucasian | 13.3 | 1.1 | 0.8 | 0.3 | 0.7 |
| Hispanic | 12.2 | 0.7 | 0.8 | 0.4 | 1.0 |
| Oriental | 11.2 | 1.6 | 1.3 | 0.4 | 0.5 |
| Black | 14.6 | 0.0 | 0.5 | 0.6 | 0.9 |
| Mean | 13.9 | 0.8 | 0.9 | 0.4 | 0.7 |
| ±1 S.D. | 2.6 | 0.6 | 0.3 | 0.1 | 0.2 |
### CD19 TRI-COLOR
| Ethnic
Origin | Lymph. | Percent of Stained Cells | | | |
| --- | --- | --- | --- | --- | --- |
| | | Mono. | Gran. | Plt. | RBC |
| Caucasian | 18.3 | 0.0 | 1.0 | 0.3 | 0.4 |
| Caucasian | 12.1 | 0.2 | 1.1 | 0.4 | 0.1 |
| Hispanic | 11.6 | 0.2 | 0.3 | 0.2 | 0.7 |
| Oriental | 11.0 | 1.0 | 0.9 | 0.3 | 0.2 |
| Black | 12.0 | 0.7 | 0.9 | 0.4 | 0.2 |
| Mean | 13.0 | 0.4 | 0.8 | 0.3 | 0.3 |
| ±1 S.D. | 3.0 | 0.4 | 0.3 | 0.1 | 0.2 |
Specific and/or nonspecific antibody Fc binding to monocytes in a patient sample can be excluded by proper gating on lymphocytes on the flow cytometer.
{4}
-5-
# REPRODUCIBILITY DATA (INTRA-LAB)
Intra-lab reproducibility for the CALTAG CD19 R-PE and CD19 TRI-COLOR conjugated monoclonal antibodies was determined by performing 10 replicated determinations for each antibody in each of three ranges; high, medium and low. Thus, a total of 30 determinations were performed for each form of CD19. In this manner, reproducibility was demonstrated throughout the entire measuring range.
The 10 determinations for each range were performed by the staining, processing and analysis of 10 separate samples. Lymphocytes were selected for the analysis of percent cells stained in each of the three ranges.
To perform this study, anticoagulated blood was obtained from an abnormal donor expressing a high percentage of CD19+ cells. Mid range and low range samples were obtained by adding known CD19- cells in appropriate ratios, while maintaining approximately the same total cell concentrations for the three ranges.
The study was performed in each of three independent laboratories, in the manner that each laboratory obtained, stained and analyzed separate blood samples. The following data are representative:
| procedure | Level | mean % positive | S.D. | % CV | n |
| --- | --- | --- | --- | --- | --- |
| CD19 R-PE | high | 66.6 | 0.4 | 0.5 | 10 |
| | mid | 46.7 | 0.8 | 1.6 | 10 |
| | low | 14.9 | 0.6 | 4.3 | 10 |
| procedure | Level | mean % positive | S.D. | % CV | n |
| CD19 | high | 65.4 | 0.7 | 1.0 | 10 |
| TRI-COLOR | mid | 46.1 | 0.6 | 1.3 | 10 |
| | low | 15.5 | 0.4 | 2.7 | 10 |
# REPRODUCIBILITY, (INTER-LAB)
Inter-lab reproducibility for the CALTAG CD19 R-PE and CD19 TRI-COLOR conjugated monoclonal antibodies was determined by performing 10 replicated determinations for each antibody in each of three ranges; high, medium and low. Thus, a total of 30 determinations were performed for each form of CD19. In this manner, reproducibility was demonstrated throughout the entire measuring range.
The 10 determinations for each range were performed by the staining, processing and analysis of 10 separate samples. Lymphocytes were selected for the analysis of percent cells stained in each of the three ranges.
The study was performed in each of three laboratories. All laboratories stained and analyzed blood samples from the same blood donors. Lysed unstained samples containing cells in the appropriate ranges were prepared by one of the participating laboratories for staining and analysis in each of the laboratories. The following data were obtained:
{5}
-6-
| SITE 1 | | | | | |
| --- | --- | --- | --- | --- | --- |
| procedure | Level | mean % positive | S.D. | % CV | n |
| CD19 R-PE | high | 84.7 | 4.4 | 5.2 | 10 |
| | mid | 70.4 | 1.2 | 1.8 | 10 |
| | low | 42.7 | 3.2 | 7.6 | 10 |
| procedure | Level | mean % positive | S.D. | % CV | n |
| CD19 | high | 86.7 | 1.0 | 1.2 | 10 |
| TRI-COLOR | mid | 70.9 | 1.3 | 1.9 | 10 |
| | low | 42.7 | 0.9 | 2.1 | 10 |
| SITE 2 | | | | | |
| procedure | Level | mean % positive | S.D. | % CV | n |
| CD19 R-PE | high | 83.9 | 1.5 | 1.7 | 10 |
| | mid | 71.1 | 2.8 | 3.9 | 10 |
| | low | 42.5 | 1.9 | 4.6 | 10 |
| procedure | Level | mean % positive | S.D. | % CV | n |
| CD19 | high | 85.3 | 2.2 | 2.6 | 11 |
| TRI-COLOR | mid | 65.7 | 3.1 | 4.8 | 9 |
| | low | 32.6 | 2.0 | 6.0 | 10 |
| SITE 3 | | | | | |
| procedure | Level | mean % positive | S.D. | % CV | n |
| CD19 R-PE | high | 87.5 | 0.6 | 0.7 | 10 |
| | mid | 69.9 | 0.8 | 1.1 | 10 |
| | low | 38.8 | 1.6 | 4.2 | 10 |
| procedure | Level | mean % positive | S.D. | % CV | n |
| CD19 | high | 85.3 | 0.9 | 1.0 | 10 |
| TRI-COLOR | mid | 66.1 | 1.0 | 1.5 | 10 |
| | low | 30.7 | 2.4 | 7.9 | 10 |
{6}
-7-
# CLINICAL TESTS SUPPORTING A DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
## CORRELATION DATA
The correlation study was performed on 175 donors, including 155 normal and 20 abnormal donors.
Comparison of the CALTAG CD19 R-PE conjugated monoclonal antibody with the Coulter CD19 RD1 conjugated monoclonal antibody:
| procedure | mean % positive | r² value | slope | Y intercept | n |
| --- | --- | --- | --- | --- | --- |
| CD19 R-PE | 16.4 | 97.7 | 0.92 | 1.30 | 175 |
| CD19 RD1 | 16.2 | | | | |
CD19 R-PE
Linear regression $y = 1.30 + 0.92x$
Comparison of the CALTAG CD19 R-PE conjugated monoclonal antibody with the Coulter CD19 FITC conjugated monoclonal antibody:
| procedure | mean % positive | r² value | slope | Y intercept | n |
| --- | --- | --- | --- | --- | --- |
| CD19 R-PE | 16.4 | 96.3 | 0.93 | 0.69 | 175 |
| CD19 FITC | 16.7 | | | | |
CD19 R-PE
Linear regression $y = 0.69 + 0.93x$
Comparison of the CALTAG CD19 TRI-COLOR conjugated monoclonal antibody with the Coulter CD19 RD1 conjugated monoclonal antibody:
| procedure | mean % positive | r² value | slope | Y intercept | n |
| --- | --- | --- | --- | --- | --- |
| CD19 TRI-COLOR | 17.1 | 96.7 | 0.92 | 2.14 | 175 |
| CD19 RD1 | 16.2 | | | | |
CD19 TRI-COLOR
Linear regression $y = 2.14 + 0.92x$
Comparison of the CALTAG CD19 TRI-COLOR conjugated monoclonal antibody with the Coulter CD19 FITC conjugated monoclonal antibody:
| procedure | mean % positive | r² value | slope | Y intercept | n |
| --- | --- | --- | --- | --- | --- |
| CD19 TRI-COLOR | 17.1 | 97.4 | 0.94 | 1.36 | 175 |
| CD19 FITC | 16.7 | | | | |
CD19 TRI-COLOR
Linear regression $y = 1.36 + 0.94x$
{7}
-8-
Comparison of the CALTAG CD19 TRI-COLOR conjugated monoclonal antibody with the CALTAG CD19 R-PE conjugated monoclonal antibody:
| procedure | mean % positive | r² value | slope | Y intercept | n |
| --- | --- | --- | --- | --- | --- |
| CD19 TRI-COLOR | 17.1 | 97.6 | 0.99 | -0.56 | 175 |
| CD19 R-PE | 16.4 | | | | |
CD19 TRI-COLOR
Linear regression $y = -0.56 + 0.99x$
# BIBLIOGRAPHY
1. Hsu S., Cossman J., Jaffe E.; Lymphocyte subsets in normal human lymphoid tissues. Am. J. Clin. Path. 80:21-30, 1983.
2. Morimoto C., Letvin N.L., Distaso J.A., et al: The cellular basis for the induction of antigen-specific T8-suppressor cells. Eur. J. Immunol. 16:198-204, 1986.
3. Dorken B., Moller P., Pezzutto A., Schwartz-Albiez R., Moldenhauer G., B-cell antigens:CD19, in Fourth International Workshop and Conference On Human Leucocyte Differentiation Antigens, pp 34-36, Vienna, 1989.
4. Pirruccello S.J., Johnson D.R., Reagents for flow cytometry: Monoclonal antibodies and hematopoietic cell antigens, in Flow Cytometry And Clinical Diagnosis, pp 56-78, pub. American Society of Clinical Pathologists, 1994.
5. Shields J.G., Rigley K.P., Callard R.E., Regulation of human B cell proliferation by the CD19 cell-surface glycoprotein, in Fourth International Workshop and Conference On Human Leucocyte Differentiation Antigens, pp 40-43, Vienna, 1989.
6. Nadler L.M., in Leukocyte Typing II, Volume 2, B cell/Leukemia Panel Workshop:Summary and Comments, Chapter 1:2-20, Springer Verlag, New York, 1986.
7. Wedgwood R.J., X-linked agammaglobulinemia, in CRC Handbook Series In Clinical Laboratory Science, CRC Press, West Palm Beach Florida, pp 41-50, 1978.
8. Pearl E.R., Vogler L.B., Okos A.J. et al, B lymphocyte precursors in bone marrow. An analysis of normal individuals and patients with antibody-deficiency states, J. Immunol. 120:1169-175, 1978.
9. Spickett G.P., Webster A.D., Farrant J., Cellular abnormalities in common variable immunodeficiency, Immunodefic. Rev. 2:199-219, 1990.
10. Small T.N., Keever C., Collins N. et al, Characterization of B cells in severe combined immunodeficiency disease, Hum. Immunol. 25:181-193, 1991.
11. Goldstein R., Izaguirre L.C., Douglas J. et al, Systemic lupus erythematosus and common variable panhypogammaglobulinemia. Theoretical and practical considerations, Fed. Proc. 29:1606-1611, 1985.
12. Loken M.L., Brosnan N.N., Back B.A., Ault K.A., Establishing optimal lymphocyte gates for immunophenotyping for flow cytometry. Cytometry 11:453-459, 1990.
13. Guidelines for the Performance of CD4+ T-Cell Determinations in Persons with Human Immunodeficiency Virus Infection, Morbidity And Mortality Weekly Report (MMWR), Volume 41/No. RR-8, May 8, 1992.
14. Nicholson J.K.A., Green T.A. and Collaborative Laboratories, Selection of anticoagulants for lymphocyte immunophenotyping. J. Immunol. Methods 165:31-35, 1993.
15. Tennant J.R., Evaluation of the Trypan Blue technique for determination of cell viability, Transplantation 2:685-694, 1964.
16. Koepke J.A., Landay A.L.: Precision and Accuracy of Absolute Lymphocyte Counts, Clin. Immunol. and Immunopath. 52:19-27, 1989.
17. Brown M.C., Hoffman R.A., Kirchanski S., Controls for flow cytometers in hematology and cellular immunology, Ann. N.Y. Acad. Sci. 468:93-103, 1986.
18. Durrand R.E., Calibration of flow cytometer detector systems, Cytometry 2:192-193, 1981.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.