← Product Code [ESD](/productcode/ESD) · K963915

# TELEX IN-THE-EAR CUSTOM HEARING AID WITH THRESHOLD COMPRESSION (K963915)

_Telex Communications, Inc. · ESD · Nov 8, 1996 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K963915

## Device Facts

- **Applicant:** Telex Communications, Inc.
- **Product Code:** [ESD](/productcode/ESD.md)
- **Decision Date:** Nov 8, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.3300
- **Device Class:** Class 1
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Therapeutic

## Indications for Use

to amplify and transmit sound to the ear.

## Device Story

In-the-ear (ITE) custom hearing aid; amplifies and transmits sound to ear. Input: ambient sound via microphone. Processing: amplifier with optional dynamic range compression (threshold compression) to adjust output levels. Output: amplified sound delivered to ear canal. Used in clinical or daily settings by patients with hearing loss. Healthcare provider adjusts settings via volume control, low/high cut trimmers, and threshold control. Benefits: improved auditory perception and communication for hearing-impaired users.

## Clinical Evidence

Bench testing only; compliance with ANSI S3.2-1987 standards for frequency response and technical specifications.

## Technological Characteristics

ITE form factor; standard hearing aid battery power (13, 312, 10A, or 5A); analog amplifier with dynamic range compression; manual controls (volume, trimmers, switches); complies with ANSI S3.2-1987.

## Regulatory Identification

An air-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing that conducts sound to the ear through the air. An air-conduction hearing aid is subject to the requirements in § 800.30 or § 801.422 of this chapter, as applicable. The air-conduction hearing aid generic type excludes the group hearing aid or group auditory trainer, master hearing aid, and the tinnitus masker, regulated under §§ 874.3320, 874.3330, and 874.3400, respectively.

## Predicate Devices

- Telex Model 28A ([K922066](/device/K922066.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

TELEX
Telex Communications, Inc.

Item C

NOV - 8 1996
K963915

9600 Aldrich Avenue South
Minneapolis, Minnesota 55420 USA
Telephone 612-884-4051
Fax 612-884-0043

510(K) SUMMARY

Trade Name: Telex In-the-Ear Custom Hearing Aid Compression®
Common Name: Hearing Aid, ITE
Classification Name: Air Conduction Hearing Aid, 77ESD
Equivalent to: Telex Model 28A K922066

Description:
Intended use: to amplify and transmit sound to the ear.
Features: Threshold Compression
Assembly: assembled from standard components that are widely used by other hearing aid manufacturers.
Technical characteristics: technical specifications comply with S3.2-1987 ANSI standards.
Fit: frequency response per S3.2-1987 ANSI standard as shown on specification filed with 510(k).
Controls: Volume control, Low cut Trimmer (optional), High cut Trimmer (optional), Threshold Control, Loudness Boost (optional), Set-screw Volume Control (optional), Low cut Switch (optional)
Power: Standard hearing aid battery (13, 312, 10A or 5A)

Comparison to predicate device: identical to the model 28A, except for different amplifier which incorporates dynamic range compression as an optional compensation strategy.

Submitted by:
Tom Scheller
Chief Engineer
Hearing Instruments Group
Telex Communications, Inc.
9600 Aldrich Ave S.
Minneapolis, MN 55420
(612) 884-4051 voice
(612) 884-0043 fax

Contact: Tom Scheller
Prepared: September 16, 1996

---

**Source:** [https://fda-staging.innolitics.com/device/K963915](https://fda-staging.innolitics.com/device/K963915)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).