The bipolar temporary myocardial pacing wire, Model 6495 is designed for temporary atrial and ventricular pacing and sensing, for a maximum of 7 days, during and after cardiac surgery. The Model 6495 is intended for single use only.
Device Story
Bipolar temporary myocardial pacing wire (Model 6495) used for temporary cardiac pacing and sensing; consists of distal/proximal electrodes, coaxial conductor, and fixation coil; includes atraumatic curved needle for placement and breakaway chest needle for chest wall exit. Used during/after cardiac surgery; operated by clinicians. Device provides electrical connection between external pacemaker and heart tissue; removed via gentle traction after use; no residual material remains in body. Benefits include temporary hemodynamic support post-surgery.
Clinical Evidence
Bench testing only. Integrity testing included visual verification, X-ray/dimensional verification, electrical verification, pull strength, flex life, connector compatibility, and break moment of thorax needle. All tests met specified requirements.
Indicated for temporary atrial and ventricular pacing and sensing in patients during and after cardiac surgery for a maximum of 7 days. Single use only.
Regulatory Classification
Identification
Temporary pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an external pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator. Permanent pacemaker electrode: A device consisting of flexible insulated electrical conductors with one end connected to an implantable pacemaker pulse generator and the other end applied to the heart. The device is used to transmit a pacing electrical stimulus from the pulse generator to the heart and/or to transmit the electrical signal of the heart to the pulse generator.
Predicate Devices
Temporary Myocardial Pacing Wire, Medtronic Model 6500 (K944957)
Bipolar temporary Myocardial Heart Wire, Oscor Model TME 65 C (K850622)
Submission Summary (Full Text)
{0}
K963898
MEDTRONIC CONFIDENTIAL
# G: 510(K) SUMMARY
## 1. SUBMITTER'S NAME, ADDRESS
Susan M. Noddin
Associate Product Regulation Manager
Medtronic, Inc.
7000 Central Avenue NE
Minneapolis, MN 55432
Phone: (612) 574-6191
Fax: (612) 574-6424
JUL 22 1997
| 2. TRADE NAME: | Bipolar temporary myocardial pacing wire, Model 6495 |
| --- | --- |
| Common Name: | Temporary Pacing Lead |
| Classification Name: | Temporary Pacemaker Electrode |
| Classification | This device has been classified by the Circulatory Systems Device Panel into Class II, (21 CFR 870.3680(a)). |
| 3. SUBSTANTIALLY EQUIVALENT DEVICE(S) | Temporary Myocardial Pacing Wire, Medtronic Model 6500, marketed via K944957
Bipolar temporary Myocardial Heart Wire, Oscor Model TME 65 C, marketed via K850622 |
## 4. DEVICE DESCRIPTION
The bipolar temporary myocardial pacing wire, Model 6495 consists of a distal electrode, proximal electrode, and a coaxial conductor which are crimped together. A blue polypropylene fiber proximally coiled (flattened coil) for fixation of the lead is attached to the distal electrode and terminated in an atraumatic curved needle. An atraumatic breakaway chest needle at the other end of the conductor wire permits running the lead through the chest wall. Proximal to the chest needle is a connector ring crimped to the conductor wire.
After removal of the pacing wire, which is performed by gentle traction, no part of the wire remains in the body.
## 5. INDICATIONS FOR USE
The bipolar temporary myocardial pacing wire, Model 6495 is designed for temporary atrial and ventricular pacing and sensing, for a maximum of 7 days, during and after cardiac surgery. The Model 6495 is intended for single use only.
## 6. TECHNOLOGICAL CHARACTERISTIC COMPARISONS
The bipolar temporary myocardial pacing wire, Model 6495 is substantially equivalent to the following products:
- Unipolar temporary myocardial pacing wire, Medtronic Model 6500, marketed via K944957
- Bipolar temporary myocardial heart wire, Oscor Medical Corporation Model TME 65 C, marketed via K850622
22
{1}
MEDTRONIC CONFIDENTIAL
The table below contains a comparison of the similarities and differences of the Model 6495 to the predicate devices to which it is substantially equivalent. Similarities between the Model 6495 and the comparison devices are noted.
The bipolar temporary myocardial pacing wire, Model 6495 contains technologies comparable to Medtronic, Inc. temporary myocardial pacing lead, Model 6500. The proximal end of the Model 6495 ends in a straight chest needle, similar to the Model 6500. The conductor wire material is stainless steel, identical to that of the Model 6500 and the Oscar Model TME 65 C. The insulation materials for the Model 6495 are identical to the Model 6500 and the Model TME 65 C. The distal end is identical to the Model 6500, and similar to the Model TME 65 C. The Model 6495 utilizes the double pouch package configuration, identical to that utilized by the Model 6500. The Model 6495 uses the same 100% EtO sterilization process as the Model 6500. The Model 6495 is a bipolar heartwire, as is the Model TME 65 C myocardial heartwire.
| Manufacturer | Medtronic | Oscor Medical | Medtronic |
| --- | --- | --- | --- |
| Model No. | 6495 | TME 65 C | 6500 |
| 510(k) Number | | K850622 | K944957 |
| Intended Use | Temporary Myocardial Pacing and Sensing | Same | Same |
| | Bipolar | Bipolar | Unipolar |
| Electrode | Two discrete electrodes | Ring electrode and wire | Tip |
| Electrode and Conductor Material | Stainless Steel | Same | Same |
| Fixation Mode | “Pigtail” Coil | Zigzag and suturing | “Pigtail” Coil |
| Outer Insulation Material | Polyethylene | Same | Same |
| Insulated Wire Diameter (nominal) | 0.7mm | 0.45 mm | 0.7mm |
| Proximal Needle Shape and Breakaway Feature Size (nominal) | Straight
Breakaway
L = 88mm
D = 1.0mm | Slightly curved
L= 80mm
D = 1.12 mm | Straight
Breakaway
L = 90 mm
D = 1.0 mm |
| Distal Needle Shape Size (nominal) | 3/8
Curved
L = 32mm
D = 0.38mm | 1/2
Curved
L = 16mm
D = 0.52 mm | 3/8
Curved
L = 25mm
D = 0.66mm |
| Sterile Package Configuration | Blister/Pouch | Pouch/Pouch | Blister/Pouch |
| Sterile Packaging Materials | spunbounded olefin-PE/Polyester | not known | spunbounded olefin-PE/Polyester |
| Sterilization & Aeration Process | 100% EtO | not known | 100% EtO |
{2}
MEDTRONIC CONFIDENTIAL
## 7. SUMMARY OF STUDIES
Medtronic, Inc. performed device integrity testing to support the bipolar temporary myocardial pacing wire, Model 6495 is substantially equivalent to the predicate devices.
Device integrity testing included:
- Visual verification
- X-ray & Dimensional verification
- Electrical verification
- Pull strength verification
- Flex life verification
- Connector compatibility
- Break moment of thorax needle
All device integrity test results for the bipolar temporary myocardial pacing wire, Model 6495 met specified requirements.
## 8. CONCLUSION (STATEMENT OF EQUIVALENCE)
Through the data and information provided in this submission, numerous similarities support a substantial equivalence determination, and, therefore, clearance of the 510(k) notification for the bipolar temporary myocardial pacing wire, Model 6495.
{3}
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration
9200 Corporate Boulevard
Rockville MD 20856
Ms. Susan Noddin
Product Regulation Manager
Medtronic, Inc.
7000 Central Avenue, N.E.
Minneapolis, Minnesota 55432-3576
JUL 22 1997
Re: K963898
Model 6495 Bipolar Temporary Pacing Lead
Regulatory Class: III (three)
Product Code: LDF
Dated: April 24, 1997
Received: April 25, 1997
Dear Ms. Noddin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for *in vitro* diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,

Thomas J. Callahan, Ph.D.
Director
Division of Cardiovascular, Respiratory, and Neurological Devices
Office of Device Evaluation
Center for Devices and Radiological Health
Enclosure
{5}
Medtronic Confidential
K963898
# Indications For Use
## INDICATIONS FOR USE
The Model 6495 bipolar temporary myocardial pacing lead is designed for temporary atrial and ventricular pacing and sensing, for a maximum of 7 days, during and after cardiac surgery. The Model 6495 bipolar temporary myocardial pacing lead is intended for SINGLE USE ONLY.
*(Division Sign-Off)*
Division of Cardiovascular, Respiratory and Neurological Devices
510(k) Number **K963898**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.