← Product Code [ITI](/productcode/ITI) · K963808

# CHOICE POWERED WHEELCHAIR (K963808)

_American Medical Technology, Inc. · ITI · Feb 12, 1997 · Physical Medicine · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K963808

## Device Facts

- **Applicant:** American Medical Technology, Inc.
- **Product Code:** [ITI](/productcode/ITI.md)
- **Decision Date:** Feb 12, 1997
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The CHOICE powered wheelchair is intended to provide mobility for a person who, due to medical reasons, the debilitating nature of some illnesses, because of accident, injury and/or subsequent paralysis, is either missing or unable to use the lower extremities. Some, but not all of the specific indications for use of a power wheelchair are: paraplegia, quadriplegia, ALS, muscular dystrophy, muscular sclerosis, lower limb amputation and any one or more of numerous neurological or muscular diseases which render the limbs too weak or unstable for normal use.

## Device Story

Powered wheelchair providing mobility for individuals with lower extremity impairment. Fabricated steel frame; dual 12V gel cell batteries (24V DC system). Controller (Penny & Giles) manages drive motors and seating actuators; user interface via joystick or alternative controls (e.g., sip-n-puff). Features include power tilt-in-space, recline, elevating leg rests, and seat elevation. Used in various environments by patients with limited mobility. Healthcare providers use the device to restore independent mobility. Benefits include improved functional independence and pressure relief via seating adjustments.

## Clinical Evidence

No clinical data submitted. Bench testing only, conducted per ANSI/RESNA WC/14 standards (static/dynamic stability, braking, dimensions, obstacle climbing) and electromagnetic compatibility requirements.

## Technological Characteristics

Fabricated steel frame; 24V DC power (dual 12V gel batteries); Penny & Giles controller; electric actuators for seating (tilt, recline, lift). Dimensions: 34" length, 26-28" width. Compliant with ANSI/RESNA WC/14 standards. Electromagnetic compatibility tested to 40 V/m.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- EXCELSIOR SR powered wheelchair
- Chairman power wheelchair (Permobil, Inc.)
- Quickie P300 powered wheelchair ([K900330](/device/K900330.md))
- Action Storm powered wheelchair ([K940051](/device/K940051.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

AMERICAN MEDICAL TECHNOLOGIES, INC.
COMPASSION • DEDICATION • INNOVATION

# 510(k) Summary K963808

## (a) (1) Submitter's name, address

American Medical Technologies, Inc.
160 Lee Street
Canton, Georgia 30114

**Contact Person**
J.R. McLaughlin
Vice President, Operations
(770) 479-1695
FEB 13 1997

**Date of preparation of this summary:** 14 September 1996

## (2) Device trade or proprietary name:

CHOICE Powered Wheelchair
Device common or usual name or classification name

|  Product Nomenclature | Classification Number | Class | Panel  |
| --- | --- | --- | --- |
|  WHEELCHAIR, POWERED | 89 ITI | II | PHYSICAL MEDICINE  |

## (3) Substantial Equivalence

The CHOICE powered wheelchair is substantially equivalent in safety, efficacy, technology and intended use to its predicate model EXCELSIOR SR powered wheelchair currently marketed by American Medical Technologies as well as numerous other devices such as the Permobil, Inc. ‘Chairman’ power wheelchair, the Sunrise Medical ‘Quickie P300’ power wheelchair [K900330] and the Invacare ‘Action Storm’ power wheelchair [K940051].

## (4) Description of the new device

The CHOICE powered wheelchair is manufactured using fabricated steel for the power base and seat assemblies. All components determined to be a permanent part of the structural frame or seating are welded. Any parts which are required to be removable for service or adjustment to perform it’s function are attached with locking fasteners.

The CHOICE is supplied with two, 12 volt gel cell batteries, either Group 22, Group 24 or Group 27 with a #891 style post, wired in series to provide 24 volts DC [Vortex Batteries, Doraville, GA]. The controller of the device is completely manufactured, tested and certified by Penny &amp; Giles, Dorset, UK, which also contains the software programming and power output to operate the chair’s drive motors, as well as interface to various optional features. A joystick, ‘sip-n-puff’, or alternative control mechanism is used to direct the function of the controller. Two drive motors are located on the power base and attached to a gearbox which provides the appropriate speed control and torque for the chair through a direct drive to the drive wheels.

Electrically controlled actuators are used to provide the various movement features of the seating system, including: reduced-shear recline, power elevating leg rests, ‘tilt-in-space’ and seat elevation. The chair is recharged overnight using a dual-mode, fully automatic battery charger [Lester Electrical, Lincoln, NE Model Number 24EL8, 120V/60Hz] certified for safety and fire resistance in compliance with current ANSI/RESNA standards.

AMERICAN MEDICAL TECHNOLOGIES, INC.
Tippens Building • 160 Lee Street at North Street • Canton, Georgia 30114 • Phone: (770) 479-1695 • Fax: (770) 479-1705

{1}

510(k) Summary
CHOICE Powered Wheelchair

## (5) Intended use of the device.

The CHOICE powered wheelchair is intended to provide mobility for a person who, due to medical reasons, the debilitating nature of some illnesses, because of accident, injury and/or subsequent paralysis, is either missing or unable to use the lower extremities. Some, but not all of the specific indications for use of a power wheelchair are: paraplegia, quadriplegia, ALS, muscular dystrophy, muscular sclerosis, lower limb amputation and any one or more of numerous neurological or muscular diseases which render the limbs too weak or unstable for normal use.

## (6) Technological characteristics of the device.

### Sizes, configurations and functions of each device component

The CHOICE is available with either a 16- or 18-inch wide base in combination with seats widths from 14- to 20-inches and seat depths from 14- to 24-inches to accommodate most users.

The CHOICE is available as a basic, powered wheelchair with a fixed seat, or with any or all of the following seating and positioning options:

- seat elevator providing 8 to 10 inches of vertical lift
- reduced-shear recliner, with or without power elevating leg rests
- ‘Tilt-in-Space’, 0 to 55 degrees

Both tilt and recline are used to provide pressure relief and, can be used either in combination or independently as necessary.

The elevator feature provides many vocational and functional benefits associated with position. While not providing any physically therapeutic benefit, the practical value tends to make this feature very desirable for those persons with enough use of arms and hands.

### Specifications

Weight: 199 lb. w/o batteries, 270 to 300 lb. with batteries
Length: 34 inches, plus footrest
Height: 19.5 inches, floor to seat
Overall Width: customized, from 26 to 28 inches
Seat Width: customized, from 14 to 22 inches
Front Wheels: 14” OD x 3” wide
Rear Wheels: 8” OD x 2” wide

Page 2

{2}

510(k) Summary
CHOICE Powered Wheelchair

(b) (1) Summary of nonclinical tests submitted with the premarket notification for the device.

The CHOICE was tested in accordance with the ANSI/RESNA Standard, Wheelchairs - Testing of Power and Control Systems for Electric Wheelchair. WC/14, December 1991, parts 00, 01, 02, 03, 05 and 10:

WC 01: Static Stability Test Results

|  Test Number | 7.1 |   | 7.1 |   |   | 7.2 | 7.3  |   |
| --- | --- | --- | --- | --- | --- | --- | --- | --- |
|  Chair position | Downhill |   | Uphill |   |   | Uphill | Sideways  |   |
|  Wheel locks | locked |   | locked |   |   | unlocked | locked  |   |
|  Action | Slide | Tip | Slide | Tip |   | Tip | Slide | Tip  |
|  Block Size | 0" | 1.6" | 0" | 0" | 1.6" | 3.9" | 0" | 1.6"  |
|  seat position fully down | no slide | 17.5° | 18° |  | 43° | 49° | 22° | 35°  |
|  seat position fully up | no slide | 12° | 18° |  | 43° | 49° | 19° | 26°  |

WC 02: Dynamic Stability Test Results

|  Test Number | 7.1 | 7.2 | 7.3  |
| --- | --- | --- | --- |
|  description | Up ramp with max.acceleration | Braking down 5 deg. [lift, tip, slide, stable] | Tuming [lift, tip or stable]  |
|  seat position fully down | 22° to lift unit is front-wheel-drive | stable. unit is equipped with anti-tip devices | stable. unit is equipped with anti-tip devices  |
|  seat position fully up | 19° to lift unit is front-wheel-drive | stable. unit is equipped with anti-tip devices | stable. unit is equipped with anti-tip devices  |

WC 03: Stopping Distances from Maximum Speed
Maximum Effort (Max) and Automatic (Auto) Braking

|  Test Number |   | 7.2.1a |   | 7.2.1b |   | 7.2.1c  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Maximum Speed (mph) |   | Forward, Horizontal Plane (feet) |   | Backward, Horiz. Plane (feet) |   | Forward, down 5' Incline (feet)  |   |
|  Forward | Reverse | Max | Auto | Max | Auto | Max | Auto  |
|  5.2 | 1.3 | 3.3 | 3.8 | 2.3 | 2.8 | 3.3 | 3.8  |

WC 05: Overall Dimensions

|  5.1.1 |   | 5.1.2 |   | 5.1.3 |   | 5.1.4  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Overall length with leg support and footrest |   | Overall length without leg support and footrest |   | Overall width |   | Overall height with backrest in upright position  |   |
|  inches | mm | inches | mm | inches | mm | inches | mm  |
|  45 | 1143 | 31.5 | 800 | 26 | 660 | 41 | 1041  |

{3}

510(k) Summary
CHOICE Powered Wheelchair

WC 05: Overall Dimensions (turning space)

|  7.1 |   | 7.2  |   |
| --- | --- | --- | --- |
|  Turning radius |   | Turnaround width between limiting walls  |   |
|  inches | mm | inches | mm  |
|  24 | 610 | 44 | 1118  |

WC 10: Obstacle Climbing Test Results

|  7.1 |   | 7.2 |   | 7.3 |   | 7.4  |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|  Forward no run-up |   | Backward no run-up |   | Forward 20 inch run-up |   | Backward 20 inch run-up  |   |
|  inches | mm | inches | mm | inches | mm | inches | mm  |
|  2.25 | 57.15 | 2.25 | 57 | 3 | 76 | 2.25 | 57  |

Additionally, the CHOICE was tested in accordance with the ANSI/RESNA “Proposal, Addition to ANSI/RESNA WC/14 - Electromagnetic Compatibility Requirements for Powered Wheelchairs and Motorized Scooters”, Ver. 1.5; 1/11/94 to conclude that the CHOICE is fully compliant with the EMC specification requirements with a test level of 40 V/m.

The CHOICE was also tested for compliance to the specification, Electromagnetic Compatibility for Industrial Process Measurement and Control Equipment, also related to the above ANSI/RESNA Proposal to determine that the CHOICE meets or exceeds all requirements of the test specification.

(b) (2) Summary of clinical tests submitted with the premarket notification for the device.

No clinical tests were submitted with the pre-market notification for the CHOICE Powered Wheelchair.

(b) (3) Conclusions drawn from the clinical and nonclinical trials.

Analysis of the comparison of design, function and features of the CHOICE Powered Wheelchair to other devices currently legally marketed for the same intended use, together with the results of testing conducted to assess the CHOICE’s compliance with existing ANSI/RESNA standards for powered wheelchairs demonstrates the CHOICE to be substantially equivalent to these predicate devices in terms of safety, efficacy, intended use and technology.

Page 4

---

**Source:** [https://fda-staging.innolitics.com/device/K963808](https://fda-staging.innolitics.com/device/K963808)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).