KINEMAX PLUS MODULAR STABILIZER FEMORAL COMPONENT AND FEMORAL SPACERS

K963796 · Howmedica Corp. · JWH · Dec 9, 1996 · Orthopedic

Device Facts

Record IDK963796
Device NameKINEMAX PLUS MODULAR STABILIZER FEMORAL COMPONENT AND FEMORAL SPACERS
ApplicantHowmedica Corp.
Product CodeJWH · Orthopedic
Decision DateDec 9, 1996
DecisionSN
Submission TypeTraditional
Regulation21 CFR 888.3560
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Kinemax® Plus Modular Stabilizer Femoral Component and Femoral Spacers are intended to be used with Kinemax® and Kinemax® Plus tibial and patellar components as a total knee system. These components are intended to be used in total knee replacement procedures indicated due to inflammatory or non-inflammatory joint disease, failed previous prosthesis, or trauma. Specifically, this femoral component is intended to be used in situations where a stemmed component is desired to provide additional stability. This femoral component allows the surgeon to use (as an option) femoral spacers to augment bone loss on the distal or distal/posterior surfaces of the femur. These components are intended to be implanted using bone cement.

Device Story

Modular femoral component and spacers for total knee arthroplasty; used with Kinemax/Kinemax Plus tibial and patellar components. Provides stemmed stability; allows distal/posterior bone loss augmentation via spacers. Implanted via bone cement. Used in surgical settings by orthopedic surgeons. Output is physical structural support for knee joint; aids clinical decision-making by restoring joint function and stability in patients with degenerative or traumatic joint failure.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Modular femoral component and spacers; cemented fixation. Materials consistent with existing Kinemax/Duracon systems. Designed for use with Kinemax/Kinemax Plus tibial and patellar components.

Indications for Use

Indicated for patients requiring total knee replacement due to inflammatory or non-inflammatory joint disease, failed previous prosthesis, or trauma. Specifically for cases requiring stemmed components for additional stability or femoral spacers for bone loss augmentation.

Regulatory Classification

Identification

A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).

Predicate Devices

Submission Summary (Full Text)

{0} K963796 DEC 9 1996 510(k) Summary Device: Kinemax® Plus Modular Stabilizer Femoral Component and Femoral Spacers Common Name: Modular Femoral Component and Femoral Spacers Classification Name and Reference: 21 CFR 888.3560 Knee Joint patellofemorotibial semi-constrained cemented prosthesis Proposed Regulatory Class: Class II Device Product Code: JWH OR(87) The Kinemax® Plus Modular Stabilizer Femoral Component and Femoral Spacers are intended to be used with Kinemax® and Kinemax® Plus tibial and patellar components as a total knee system. These components are intended to be used in total knee replacement procedures indicated due to inflammatory or non-inflammatory joint disease, failed previous prosthesis, or trauma. Specifically, this femoral component is intended to be used in situations where a stemmed component is desired to provide additional stability. This femoral component allows the surgeon to use (as an option) femoral spacers to augment bone loss on the distal or distal/posterior surfaces of the femur. These components are intended to be implanted using bone cement. The Kinemax® Plus Modular Stabilizer Femoral Component and Femoral Spacers are substantially equivalent to several other legally marketed devices. Examples of these are listed below: 1. Kinemax® Plus Stabilizer Femoral Component (Howmedica - K910500) 2. Duracon® Stemmed Stabilizer Femoral Component (Howmedica - K932070) 3. Duracon® Femoral Spacers (Howmedica - K920034) For information contact: Margaret F. Crowe Manager, Regulatory Affairs Howmedica Inc. 359 Veterans Boulevard Rutherford, NJ 07070 Telephone: (201) 507-7431 Fax: (201) 507-6870
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...