← Product Code [JIX](/productcode/JIX) · K963543

# SIGMA DIAGNOSTICS CX -3 CALIBRATION STANDARD, LEVEL 2 (K963543)

_Sigma Diagnostics, Inc. · JIX · Oct 30, 1996 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K963543

## Device Facts

- **Applicant:** Sigma Diagnostics, Inc.
- **Product Code:** [JIX](/productcode/JIX.md)
- **Decision Date:** Oct 30, 1996
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1150
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

Sigma Diagnostics CX®3 Calibration Standard Level 2 is used to calibrate the Beckman SYNCHRON CX®3 System. This reagent is used for the quantitative determination of BUN/urea, calcium, and total carbon dioxide in serum or plasma; of creatinine, potassium, and sodium in serum, plasma, or urine; and of chloride and glucose in serum, plasm, urine, or cerebrospinal fluid. Calibration Standard Level 2 should be used with CX®3 Calibration Standard Level 2.

## Device Story

Calibration standard for Beckman SYNCHRON CX®3 clinical chemistry analyzer; contains specific concentrations of chloride, potassium, sodium, and total CO2; used in conjunction with Level 1 standard to establish calibration curves for quantitative analyte measurement; utilized in clinical laboratory settings by trained technicians; output enables accurate patient sample quantification; supports clinical decision-making regarding metabolic and electrolyte status.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Liquid reagent calibration standard; contains chloride (65 mmol/L), potassium (10.0 mmol/L), sodium (100 mmol/L), and total CO2 (30 mmol/L); non-reactive components included; designed for use with Beckman SYNCHRON CX®3 system.

## Regulatory Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

## Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

## Predicate Devices

- Beckman Calibration Standard 2 (Part No. 443365)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

k963543

510(k) NOTIFICATION

Sigma Diagnostics
545 South Ewing Avenue
St. Louis, MO 63103

CX®-3 Calibration Standard, Level 2
Procedure No. C7433
August 30, 1996

OCT 30 1996

# ATTACHMENT 1

## Summary of Safety and Effectiveness

{1}

k963543

510(K) NOTIFICATION

Sigma Diagnostics
545 South Ewing Avenue
St. Louis, MO 63103

CX®3 Calibration Standard, Level 2
Procedure Number C7433
August 31, 1996

# SUMMARY OF SAFETY AND EFFECTIVENESS

Sigma Diagnostics CX®3 Calibration Standard Level 2 is used to calibrate the Beckman SYNCHRON CX®3 System. This reagent is used for the quantitative determination of BUN/urea, calcium, and total carbon dioxide in serum or plasma; of creatinine, potassium, and sodium in serum, plasma, or urine; and of chloride and glucose in serum, plasm, urine, or cerebrospinal fluid. Calibration Standard Level 2 should be used with CX®3 Calibration Standard Level 1.

CX®3 Calibration Standard Level 2 contains the following analytes and other non-reactive components:

|  ANALYTE | SI UNITS  |
| --- | --- |
|  Chloride | 65 mmol/L  |
|  Potassium | 10.0 mmol/L  |
|  Sodium | 100 mmol/L  |
|  Total CO₂ | 30 mmol/L  |

The safety and effectiveness of Sigma Diagnostics CX®3 Calibration Standard, Level 2, Procedure Number C7433, are demonstrated by its substantial equivalency to Beckman Calibration Standard 2, Part No. 443365.

---

**Source:** [https://fda-staging.innolitics.com/device/K963543](https://fda-staging.innolitics.com/device/K963543)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
